K Number
K090996

Validate with FDA (Live)

Device Name
PONTO PRO
Manufacturer
Date Cleared
2009-07-29

(112 days)

Product Code
Regulation Number
874.3302
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ponto Pro is intended for the following patients and indications;

  • Patients with conductive or mixed hearing losses, who can still benefit from amplification of the sound. The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1, 2 and 3 kHz).
  • Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 4 kHz, or less than 15 dB at individual frequencies.
  • Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an AC CROS. The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should then be better than 20 dB HL (measured at 0.5. 1, 2 and 3 kHz).

The placement of a bone anchored implant is contraindicated for patients below the age of 5.

The Ponto Pro sound processor is intended to be connected to the Oticon Medical bone anchored implant system or to the Baha® Abutment snap coupling from Cochlear BAS. In addition, the Oticon Medical abutments can be used for connection of the Baha® sound processors with snap coupling from Cochlear BAS (Baha Divino®, Baha IntensoTM, Baha Cordelle II, Baha BP100),

Device Description

The Ponto Pro is a bone conduction-type hearing aid. The Ponto Processor is connected to an implant with a skin penetrating abutment which as been surgically anchored in the bone behind the ear. Vibrations generated by the sound processor are transmitted directly through the skull bone to the cochlea as bone conduction sound. The Ponto Pro sound processor has a coupling so that it can be easily connected to and disconnected from the abutment by the user. The Ponto Pro can alternatively be connected to head band accessories, to function as a conventional bone conductor. Using a computer based fitting system the Ponto Pro can be adjusted to the patient's individual hearing requirements. The sound processor is intended to work with the Oticon Medical bone anchored implant system or the Baha® Abutment snap coupling from Cochlear BAS (ref no. 90410, 90305, 90434, 90480). In addition, the Oticon Medical abutments can be used for connection of the Baha® sound processors with snap coupling from Cochlear BAS (Baha Divino®, Baha Intenso"" Baha Cordelle II, Baha BP100; ref no. 90500, 90510, 90511, 90502, 90512, 90503, 90513, 90730, 90731, 90732, 90733, HBC 400-0, HCB 401-0, HCB 402-0, 91301, 91302, 91303, 91304, 91305).

AI/ML Overview

The provided document describes the Ponto Pro bone anchored sound processor, which falls under the classification of a Class II medical device (Hearing Aid, Bone Conduction, Product Code: LXB). The device is intended for improving hearing in patients with conductive and mixed hearing losses, bilateral fitting, and single-sided deafness.

The submission is a 510(k) summary, which typically demonstrates substantial equivalence to a predicate device rather than presenting full clinical study data for new acceptance criteria. Therefore, the information provided focuses on the comparison to predicate devices and functional testing, not a detailed study with specific acceptance criteria and detailed device performance outcomes in a traditional clinical trial sense.

Here's an analysis based on the provided text, addressing your points where information is available:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in the formal sense found in a clinical trial report or a performance goal. Instead, the submission relies on demonstrating substantial equivalence to predicate devices. The "reported device performance" is essentially that it functions equivalently to these established devices.

CharacteristicAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (Ponto Pro)
DesignBone conduction sound processor connected to surgically placed implantBone conduction sound processor connected to surgically placed implant
Intended UseImprove hearing for patients with conductive/mixed hearing loss, bilateral fitting, single-sided deafnessSame as predicate; specific audiometric thresholds are given for intended use
MaterialSound processor coupling: PEEK; Implant: TitaniumSound processor coupling: PEEK; Implant: Titanium
Power RequirementZinc-air batteryZinc-air battery
Max Gain33dB33dB
Frequency Response125 Hz - 8 kHz125 Hz - 8 kHz
Sound ProcessingDigitalDigital
Patient FittingIndividual adjustment via computer-based fitting systemIndividual adjustment via computer-based fitting system
AccessoriesVarious accessories for direct audio input (DAI)Various accessories for direct audio input (DAI)
Safety & EfficacyNo new concerns compared to predicate devicesFunctional testing and software validation confirm equivalence, raising no new safety/efficacy concerns.

2. Sample Size Used for the Test Set and Data Provenance

The document explicitly states: "Testing of the Ponto Pro include functional testing and software validation. These tests verify that the Ponto Pro is functionally equivalent to the OBC, why we have come to the conclusion that further testing will not raise new issues of safety and efficacy."

This indicates that the "test set" for performance evaluation was likely functional testing and software validation of the device itself, rather than a clinical study on a patient population.

  • Sample Size: Not applicable in the context of a patient-based test set, as no patient study is detailed for proving functional equivalence in this section. The "testing" refers to internal device verification.
  • Data Provenance: Not applicable for a patient-based test set. The functional and software testing would have been conducted by the manufacturer (Oticon Medical AB in Sweden).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not provided. Since the submission relies on functional and software testing to demonstrate equivalence, and not on a clinical ground truth evaluation by experts for a patient test set, this detail is not relevant to the provided text.

4. Adjudication Method for the Test Set

Not applicable. There's no mention of a human-reviewed "test set" for which adjudication would be required. The "testing" described is functional and software-based.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, a multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned or indicated. The submission focuses on substantial equivalence based on technological characteristics and functional testing.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

The device is a hearing aid, which by its nature requires human interaction (the patient wearing it and an audiologist fitting it). The concept of "standalone algorithm only" performance doesn't directly apply in the same way it would for an AI diagnostic imaging algorithm. The "functional testing and software validation" could be considered a form of standalone testing of the device's inherent functionality.

7. The Type of Ground Truth Used

For the purpose of this 510(k) submission, the "ground truth" for demonstrating safety and effectiveness relied on:

  • Predicate Device Performance: The established safety and efficacy of the legally marketed predicate devices (OBC, Baha Divino, and air conduction hearing aids with digital sound processing). The Ponto Pro's performance is deemed acceptable because it is substantially equivalent to these devices.
  • Functional Testing and Software Validation: Internal verification that the Ponto Pro performs according to its design specifications, ensuring it functions as intended and safely, similar to its predicates.

No external "expert consensus, pathology, or outcomes data" is described as being used to establish a ground truth for a novel performance claim beyond equivalence.

8. The Sample Size for the Training Set

Not applicable. As this is a medical device (a hearing aid), not an AI algorithm trained on a dataset of patient cases, there is no "training set" in the context of machine learning. The device's design and software are developed through engineering processes, and its functionality is validated through testing and comparison to predicate devices, not by training on a data set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" in the context of this device. The "ground truth" for the device's development and validation is based on established audiological principles, engineering standards, and the performance of predicate devices.

{0}------------------------------------------------

K090196

SECTION 2. SUMMARY AND CERTIFICATION

510(K) SUMMARY ব

JUL 2 9 2002

Summary of Safety and Effectiveness

In accordance with 21 CFR 807.92, the following information constitutes the Oticon Medical summary for the Ponto Pro bone anchored sound processor

SUBMITTER'S NAME:Oticon Medical AB
ADDRESS:Ekonomiv. 2SE-436 33 AskimSweden
CONTACT PERSON:Karolin Isberg
TELEPHONE NUMBER:+46 31 748 6153
FAX NUMBER:+46 31 687 756
E-MAIL:kai@oticonmedical.se
DATE OF SUBMISSION:April 6, 2009

1. Identification of device

Proprietary Name: Ponto Pro Common Name: Hearing Aid, Bone Conduction Classification Status: Class II per regulations 21 CFR § 874.3300 Product Codes: LXB

2. Equivalent devices

Oticon Medical believes that the Ponto Pro, regarding intended use, function and procedure, is substantially equivalent to the OBC cleared in 510(k) K082108. has substantially equivalent fitting technology as air conduction hearing aids with digital sound processing (exempt from 510(k)) and has substantially equivalent accessories for direct audio input as the Baha® Divino cleared in 510(k) K042017.

3. Description of the Device

The Ponto Pro is a bone conduction-type hearing aid. The Ponto Processor is connected to an implant with a skin penetrating abutment which as been surgically anchored in the bone behind the ear. Vibrations generated by the sound processor are transmitted directly through the skull bone to the cochlea as bone conduction sound. The Ponto Pro sound processor has a coupling so that it can be easily connected to and disconnected from the abutment by the user. The Ponto Pro can alternatively be connected to head band accessories, to function as a conventional bone conductor. Using a computer based fitting system the Ponto Pro can be adjusted to the patient's individual hearing requirements. The sound processor is intended to work with the Oticon Medical bone anchored implant system or the Baha® Abutment snap coupling from Cochlear BAS (ref no. 90410, 90305, 90434, 90480). In addition, the Oticon Medical abutments can be used for connection of the Baha® sound processors with snap coupling from Cochlear BAS (Baha Divino®, Baha Intenso"" Baha Cordelle II, Baha BP100; ref no. 90500, 90510, 90511, 90502, 90512, 90503, 90513, 90730, 90731, 90732, 90733, HBC 400-0, HCB 401-0, HCB 402-0, 91301, 91302, 91303, 91304, 91305).

{1}------------------------------------------------

4. Intended use

The Ponto Pro sound processor is intended for improvement of hearing for patients with conductive and mixed hearing losses, bilateral fitting and single sided deafness.

ડ. Technological characteristics, comparison to predicate device

CharacteristicsPonto ProOBCBAHA DivinoAir conductionHearing Aidswith digitalsound processingS/Eq
DesignBone conductionsound processorconnected to animplant which hasbeen surgicallyplaced in the bonebehind the earBone conductionsound processorconnected to animplant which hasbeen surgicallyplaced in the bonebehind the earBone conductionsound processorconnected to animplant which hasbeen surgicallyplaced in the bonebehind the earAir conductionsound processorplaced behind theear or in the earcanal.Yes
Intended useImprovement ofhearing forpatients withconductive andmixed hearinglosses, bilateralfitting and singlesided deafnessImprovement ofhearing forpatients withconductive andmixed hearinglosses, bilateralfitting and singlesided deafnessImprovement ofhearing forpatients withconductive andmixed hearinglosses, bilateralfitting and singlesided deafnessImprovement ofhearingYes
MaterialSound processorcoupling: PEEKImplant: TitaniumSound processorcoupling: PEEKImplant: TitaniumSound processorcoupling: PEEKImplant: TitaniumMultipleYes
PowerrequirementZinc-air batteryZinc-air batteryZinc-air batteryZinc-air batteryYes
Max gain33dB33dB33dBN/AYes
Frequencyresponse125 Hz-8 kHz125 Hz - 8 kHz125 Hz - 8 kHzN/AYes
Sound processingDigitalDigitalDigitalDigitalYes
Patient fittingIndividualadjustment topatient audiogramand needs by acomputer basedfitting system usedby the audiologistIndividualadjustment topatient audiogramand needs by acomputer basedfitting system atthe manufactureron request fromthe audiologist.Individualadjustment topatient audiogramand needs bypotentiometersadjusted by theaudiologist using ascrewdriver.Individualadjustment topatient audiogramand needs by acomputer basedfitting system usedby the audiologistYes
AccessoriesVariousaccessories fordirect audio input(DAI)Variousaccessories fordirect audio input(DAI)Variousaccessories fordirect audio input(DAI)N/AYes
ManufacturerOticon MedicalABOticon MedicalABEntific MedicalSystems ABN/A
K-numberNo number yetK082108K042017Exempt from510(k)

Comnarison table

6. Discussion of testing

Testing of the Ponto Pro include functional testing and software validation. These tests verify that the Ponto Pro is functionally equivalent to the OBC, why we have come to the conclusion that further testing will not raise new issues of safety and efficacy.

7. Conclusion

Based on the comparison to the predicate devices, the Oticon Medical Ponto Pro, is substantially equivalent to devices already on the market, both cleared by and exempt from the 510(k) process, and presents no new concerns about safety and effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Oticon Medical AB C/O Karolin Isberg Quality and Regulatory Affairs Manager Ekomomiv. 2 SE-436 33 Askim Sweden

JUL 2 9 2009

Re: K090996

Trade/Device Name: Ponto Pro Bone Anchored Sound Processor with Accessories Regulation Number: 21 CFR 874,3300 Regulation Name: Hearing Aid, Bone Conduction Regulatory Class: Il Product Code: LXB Dated: July 10, 2009 Received: July 14, 2009

Dear Ms. Isberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address. http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Ra. An. no/p

Malvina B. Eydelman, M.I Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Kojo 996

Indications for Use

510(k) Number (if known): K090996

Device Name: Ponto Pro

Indications For Use:

The Ponto Pro is intended for the following patients and indications;

  • Patients with conductive or mixed hearing losses, who can still benefit from amplification of the sound. The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1, 2 and 3 kHz).
  • I Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 4 kHz, or less than 15 dB at individual frequencies.
  • l Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an AC CROS. The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should then be better than 20 dB HL (measured at 0.5. 1, 2 and 3 kHz).

The placement of a bone anchored implant is contraindicated for patients below the age of 5.

The Ponto Pro sound processor is intended to be connected to the Oticon Medical bone anchored implant system or to the Baha® Abutment snap coupling from Cochlear BAS. In addition, the Oticon Medical abutments can be used for connection of the Baha® sound processors with snap coupling from Cochlear BAS (Baha Divino®, Baha IntensoTM, Baha Cordelle II, Baha BP100),

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Jams K. Kane ph.D

Page 1 of 1

(Division Sign-Off) vision of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K090996

§ 874.3302 Bone-conduction hearing aid.

(a)
Identification. A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.(b)
Classification. Class II.