K Number
K090902

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-07-22

(112 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SPIDER™ (Single Port Instrument Delivery Extended Reach) is intended to establish a path of entry for laparoscopic instruments for use during minimally invasive abdominal laparoscopic surgery.

Device Description

The SPIDER™ is a single port, single incision device to facilitate multi-instrument access during laparoscopic surgical procedures. Following a Hasson cut down incision, the multi-channel cannula is inserted through a small abdominal incision. The channels are deployed allowing laparoscopic instruments to pass through each channel into the abdomen to perform laparoscopic surgery. Two channels known as IDTs (Instrument Delivery Tubes) are positioned east to west and include extended lumens to facilitate manipulation of flexible surgical instruments, enabling control of the instruments over extended distances. These 2 IDTs are flexible and allow for x, y, and z motion for a multidirectional approach of the surgical field, mimicking the approach of standard laparoscopic surgery. Two (2) rigid channels, north to south, can accommodate an endoscope or a rigid surgical instrument. The device includes ports for insufflation or smoke evacuation. Pneumoperitoneum is maintained in the abdomen during the surgical procedure. The single use SPIDER™ device is provided pre-sterilized and is a single use device. The SPIDER™ device also includes a Support Arm accessory. The support arm is provided to mount and stabilize the device.

AI/ML Overview

The provided text describes a medical device called the SPIDER™ Single Port Surgical Device (K090902), not a system for generating unique identifiers (UIDs). Therefore, the information regarding acceptance criteria, study details, sample sizes, expert involvement, and ground truth for a UID generation system is not available in the provided document.

However, I can extract the information related to the SPIDER™ medical device:

1. Acceptance Criteria and Reported Device Performance (Summary based on the provided text):

Acceptance Criteria CategoryReported Device Performance
BiocompatibilityVerification and validation studies were conducted to demonstrate.
SterilityVerification and validation studies were conducted to demonstrate.
FunctionalityVerification and validation studies were conducted to demonstrate.
Equivalency to Standard Multi-Incision Laparoscopic SurgeryPre-clinical studies were conducted to demonstrate.
Intended Use ValidationPre-clinical studies were conducted to validate.
Support Arm Cleaning and AutoclavingTested and validated.
Support Arm Retention StrengthTested and found adequate.
Support Arm Mechanical Strength (simulated use)Tested and found adequate.
Support Arm Functionality (repeated use)Tested and found adequate.
Overall Safety and EffectivenessConcluded to be as safe and effective as predicate devices, performs similarly, and does not raise new safety or effectiveness issues.
Substantial EquivalenceDetermined by the FDA to be substantially equivalent to legally marketed predicate devices.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the specific sample sizes used for the verification and validation studies or pre-clinical tests. It refers generally to "studies" and "tests." It also does not specify the country of origin of the data or whether the studies were retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

This information is not provided in the document. The text indicates that "pre-clinical studies were conducted," but details on expert involvement or ground truth establishment are absent.

4. Adjudication Method for the Test Set:

This information is not provided in the document.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No mention of an MRMC comparative effectiveness study, human readers, or AI assistance is made in the provided text, as the device is a surgical instrument, not an AI diagnostic tool.

6. Standalone (Algorithm Only) Performance:

This is not applicable as the device is a physical surgical instrument, not an algorithm or software requiring standalone performance testing in that context.

7. Type of Ground Truth Used:

The document mentions "pre-clinical studies were conducted to demonstrate equivalency to the medical effect achieved with standard multiple incision laparoscopic surgery and to validated the intended use of the device." This suggests the ground truth was likely established by comparing the device's performance (e.g., surgical outcomes, ease of use, ability to perform tasks) against established standards of care or predicate devices in a pre-clinical setting. Specific types of ground truth (e.g., pathology, outcomes data with specific metrics) are not detailed.

8. Sample Size for the Training Set:

This information is not provided in the document. The mention of "pre-clinical studies" suggests testing, but not a distinct "training set" in the context of machine learning, which is not applicable here.

9. How the Ground Truth for the Training Set Was Established:

This information is not provided and is not applicable, as the device is a physical surgical tool and not an AI/ML algorithm requiring a training set with established ground truth.

In summary, the provided document is a 510(k) summary for a medical device (surgical instrument) and, as such, focuses on demonstrating substantial equivalence to predicate devices through various verification and validation tests rather than specific AI-related study design elements.

{0}------------------------------------------------

Ko90902

P1 of 3

510(k) Summary - K090902

This summary of safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: June 26, 2009

JUL 2 2 2009

    1. Submitter:
      ":

Submitted by:

TransEnterix. Inc. 3908 Patriot Drive, Suite 170 Durham, NC 27703

TEL: 919-541-9977 FAX: 919-541-9978

Contact:

Tammy Carrea Director, Regulatory Affairs and Quality Assurance

    1. Device:
      SPIDER TM Single Port Surgical Device Propriety Name Common Name Laparoscopic instrument and accessory Classification Name: Endoscope and accessories Classification: Class II 21 CFR 876.1500 Product Code: GCJ

3. Predicate Device:

Covidien, SILS Port - K082619 Endopath III Bladeless Trocar - K032676

4. Description:

The SPIDER™ is a single port, single incision device to facilitate multi-instrument access during laparoscopic surgical procedures.

Following a Hasson cut down incision, the multi-channel cannula is inserted through a small abdominal incision. The channels are deployed allowing laparoscopic instruments to pass through each channel into the abdomen to perform laparoscopic surgery.

Two channels known as IDTs (Instrument Delivery Tubes) are positioned east to west and include extended lumens to facilitate manipulation of flexible surgical instruments, enabling control of the instruments over extended distances. These 2 IDTs are flexible and allow for x, y, and z motion for a multidirectional approach

{1}------------------------------------------------

Ko 90902

Pz. B 3

of the surgical field, mimicking the approach of standard laparoscopic surgery. Two (2) rigid channels, north to south, can accommodate an endoscope or a rigid surgical instrument. The device includes ports for insufflation or smoke evacuation. Pneumoperitoneum is maintained in the abdomen during the surgical procedure. The single use SPIDER™ device is provided pre-sterilized and is a single use device.

5. Indications for Use:

្នូ រ

The SPIDER™ (Single Port Instrument Delivery Extended Reach) is intended to establish a path of entry for laparoscopic instruments for use during minimally invasive abdominal laparoscopic surgery.

6. Comparison of Technological Characteristics with Predicate:

The SPIDER™ Device has the same intended use and function of other currently marketed single port access devices and single port trocars in that it is intended to establish a path of entry for laparoscopic instruments in minimally invasive abdominal procedures. Like the single port predicates it achieves this function via a single port and single incision. The device has the same main port and channel dimensions and similarly employs a similar number of rigid and flexible ports like the predicates. Like the predicate devices it includes valves to maintain pneumoperitoneum and insufflation ports and stopcocks to achieve insufflation and smoke evacuation. Similar to the predicate devices it is provided pre-sterilized, is disposable, and is a single use device, the same as other predicate devices.

It incorporates the same design features as the predicates to facilitate minimally invasive surgical procedures. It has multiple channels to facilitate introduction of numerous surgical instruments. It may be inserted using a Hasson technique under direct visualization. It has IDTs to allow placement. control and manipulation of rigid and flexible surgical instruments.

The IDTs, while longer than the channels of the predicate devices, are not longer than the effective length of surgical instruments and are presented and positioned within the surgical space in a similar manner. The IDTs are able to achieve triangulation in a manner similar to that of standard multi-port laparoscopic equipment.

The SPIDER™ device also includes a Support Arm accessory. The support arm is provided to mount and stabilize the device. The support arm has been tested for mechanical stability, strength, and repeated use and found to be adequate to stabilize and maintain the SPIDER™ device in a fixed position during surgical procedures.

{2}------------------------------------------------

Ko-90902

Image /page/2/Picture/1 description: The image shows the handwritten text "P3 83". The text is written in a cursive style with a black ink. The number 3 is written twice, once after the letter P and once after the number 8. There is a line under the number 83.

7. Performance Data:

Verification and validation studies were conducted for the SPIDER™ to demonstrate biocompatibility, sterility, and functionality. Pre-clinical studies were conducted to demonstrate equivalency to the medical effect achieved with standard multiple incision laparoscopic surgery and to validated the intended use of the device.

The Support Arm was tested to validate cleaning and autoclaving cycles. The Support Arm was also tested for retention strength, mechanical strength under simulated use, and functionality following repeated use.

Any technological differences between the SPIDER™ device and the predicates have been mitigated via verification and validation testing. Thus the SPIDER™ device does not introduce any new issues of safety or effectiveness compared to other similar single port access devices currently marketed.

8. Conclusion:

The conclusion drawn from the test data is that the SPIDER™ device is as safe and effective as the predicate devices, performs similarly to other legally marketed predicate devices for laparoscopic surgery, and does not raise any new issues of safety or effectiveness.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular pattern around the eagle. The logo is black and white.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 222009

TransEnterix, Inc. % Ms. Tammy B. Carrea Director Regulatory Affairs and Quality Assurance 3908 Patriot Drive, Suite 170 Durham, North Carolina 27703

Re: K090902

Trade/Device Name: SPIDER™ (Single Port Instrument Delivery Extended Reach) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: June 27, 2009 Received: June 30, 2009

Dear Carrea:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Ms. Tammy B. Carrea

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

for D.O.K

Mark N. Melkerson Director

Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

090902

510(k) No.

If known

KO90902

Indications For Use Statement

SPIDER™ (Single Port Instrument Delivery Extended Reach) Device Name:

Indications For Use:

The SPIDER™ (Single Port Instrument Delivery Extended to establish a path of entry for laparoscopic instruments for use during minimally invasive abdominal laparoscopic surgery.

Prescription Use

1. Introduction

This document describes the design and implementation of a system for generating and managing unique identifiers (UIDs) at scale. UIDs are essential for identifying and tracking entities in distributed systems, databases, and various applications. The system aims to provide a reliable, efficient, and scalable solution for generating UIDs while ensuring uniqueness and minimizing collisions.

2. Requirements

The UID generation system must meet the following requirements:

  • Uniqueness: The generated UIDs must be globally unique to avoid conflicts and ensure data integrity.
  • Scalability: The system should be able to handle a large number of UID generation requests concurrently without performance degradation.
  • Low Latency: UID generation should be fast and efficient to minimize the impact on application performance.
  • Fault Tolerance: The system should be resilient to failures and continue to generate UIDs even in the presence of hardware or software issues.
  • Sortability: The generated UIDs should be sortable by time to facilitate efficient querying and analysis of data.
  • Customization: The system should allow customization of UID generation parameters to meet specific application requirements.

3. Design

The UID generation system is designed as a distributed service that consists of multiple UID generation nodes. Each node is responsible for generating UIDs within a specific range. The system uses a combination of techniques to ensure uniqueness, scalability, and fault tolerance.

3.1. UID Structure

The generated UIDs are 64-bit integers that consist of the following components:

  • Timestamp: A 42-bit timestamp that represents the number of milliseconds since a custom epoch. This component ensures sortability and provides a rough estimate of the UID generation time.
  • Worker ID: A 10-bit worker ID that identifies the UID generation node. This component ensures uniqueness across multiple nodes.
  • Sequence Number: A 12-bit sequence number that is incremented for each UID generated within the same millisecond on the same node. This component ensures uniqueness within a node.

$$UID = Timestamp | WorkerID | SequenceNumber$$

3.2. UID Generation Algorithm

The UID generation algorithm is as follows:

  1. Get the current timestamp in milliseconds.
  2. If the timestamp is the same as the previous timestamp, increment the sequence number. If the sequence number reaches its maximum value (4095), wait until the next millisecond.
  3. If the timestamp is greater than the previous timestamp, reset the sequence number to 0.
  4. Combine the timestamp, worker ID, and sequence number to form the UID.
  5. Return the UID.

3.3. Worker ID Allocation

Each UID generation node is assigned a unique worker ID. The worker IDs are allocated by a central coordination service, such as ZooKeeper or etcd. The coordination service ensures that each node receives a unique worker ID and prevents conflicts.

3.4. Fault Tolerance

The UID generation system is designed to be fault-tolerant. If a UID generation node fails, the coordination service automatically reassigns its worker ID to another node. The system also supports replication of UID generation nodes to provide redundancy and ensure high availability.

4. Implementation

The UID generation system is implemented in Java using the Spring Framework. The system uses ZooKeeper for worker ID allocation and coordination. The system is deployed on a cluster of virtual machines using Docker and Kubernetes.

4.1. Dependencies

The following dependencies are required to build and run the UID generation system:

  • Java 8 or later
  • Maven 3 or later
  • ZooKeeper 3.4 or later
  • Spring Framework 5 or later
  • Docker 18.09 or later
  • Kubernetes 1.13 or later

4.2. Configuration

The UID generation system can be configured using a properties file. The following properties can be configured:

  • worker.id: The worker ID of the UID generation node.
  • epoch: The custom epoch to use for the timestamp.
  • zookeeper.connect: The ZooKeeper connection string.

4.3. API

The UID generation system provides a simple API for generating UIDs. The API consists of a single endpoint that returns a new UID.

GET /uid

json { "uid": 1234567890 }

5. Testing

The UID generation system is thoroughly tested to ensure uniqueness, scalability, and fault tolerance. The following tests are performed:

  • Uniqueness Test: This test generates a large number of UIDs and verifies that they are all unique.
  • Scalability Test: This test measures the throughput and latency of the system under high load.
  • Fault Tolerance Test: This test simulates failures of UID generation nodes and verifies that the system continues to generate UIDs without errors.

6. Conclusion

The UID generation system provides a reliable, efficient, and scalable solution for generating UIDs. The system is designed to meet the requirements of distributed systems, databases, and various applications. The system is implemented in Java using the Spring Framework and deployed on a cluster of virtual machines using Docker and Kubernetes.

×

AND/OR

Over-the-Counter Use

(Per 21 CFR 801 Subpart D)

(Per 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NECESSSARY)

Nike R.P. Alence for mxm
(Division Sign-Off) Division of CDRH, Office of Device Evaluation (ODE)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K090902

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.