K Number
K090483

Validate with FDA (Live)

Date Cleared
2009-03-20

(23 days)

Product Code
Regulation Number
870.2340
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

Intended Use: For monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility.

Device Description

The names of the subject devices are the Philips SureSigns VM Series Patient Monitors, the SureSigns VS2 Vital Signs Monitor, and the SureSigns VS3 Vital Signs Monitor. The common usual name is multi-parameter patient monitor. The modification is to replace the currently cleared CAS NBP module with the picoNIBP(non invasive blood pressure) OEM Module (which was cleared in K051366) to the VS2 Vital Signs Monitor, the VS3 Vital Signs Monitor, and the VM Series (VM 4/6/8) Patient Monitors. In addition, this submission includes a change in plastics to Valox and the addition of new SP0- sensors and the disclosure of IPX1 rating for ingress protection to vertical water drops for all subject devices.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study information for the Philips SureSigns VM Series Patient Monitors, SureSigns VS2 Vital Signs Monitor, and SureSigns VS3 Vital Signs Monitor:

Based on the provided K090483 510(k) summary, much of the requested information regarding specific acceptance criteria, detailed device performance metrics, and study methodologies for the modified device is not explicitly stated. The submission focuses on demonstrating substantial equivalence to a predicate device (K052707) rather than providing a detailed de novo performance study with specific criteria.

The key takeaway is that the manufacturer states that "Pass/Fail criteria were based on the specifications cleared for the predicate device" and that testing showed "substantial equivalence." This implies that the acceptance criteria for the modified device are implicitly aligned with those established for the previously cleared predicate device.


Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria (from document)Reported Device Performance (from document)
General Performance"Pass/Fail criteria were based on the specifications cleared for the predicate device.""Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the Philips SureSigns VM Series Patient Monitors, SureSigns VS2 Vital Signs Monitors and SureSigns VS3 Vital Signs Monitors meet all claims and support a determination of substantial equivalence."
Technological EquivalenceSame fundamental technological characteristics as the legally marketed predicate devices. Same design as the predicate device.Modified devices have the same fundamental technological characteristics and use the same design as the predicate device (M3046B Compact Configurable Portable Patient Monitor, K052707). The modification is to replace the NBP module with picoNIBP module (cleared in K051366), change in plastics, and new SpO2 sensors.
Intended UseSame intended use as the legally marketed predicate device."The modified devices have the same intended use as the legally marketed predicate device." For monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments and transport within a healthcare facility.
SafetySafety testing from hazard analysis. Implicitly, to be as safe as the predicate device."Safety testing from hazard analysis" was performed. Results supported substantial equivalence.
ReliabilityReliability characteristics of the modified device with respect to the predicate. Implicitly, to be as reliable as the predicate device."Reliability characteristics of the modified device with respect to the predicate" were established. Results supported substantial equivalence.
FunctionalityFunctionality characteristics of the modified device with respect to the predicate. Implicitly, to be as functional as the predicate device."Functionality characteristics of the modified device with respect to the predicate" were established. Results supported substantial equivalence.

Study Information Table

FeatureDetails from K090483 Document
Test Set Sample SizeNot explicitly stated. The document refers to "system level tests, performance tests, and safety testing," but does not provide specific sample sizes for these tests.
Test Set Data ProvenanceNot explicitly stated. The testing appears to have been internal verification and validation activities conducted by Philips Medical Systems in Andover, MA, United States. It's likely prospective testing in a lab setting, given the nature of device modifications.
Number of Experts for Ground Truth (Test Set)Not applicable/Not stated. The document describes engineering and performance testing against predicate device specifications, not a clinical study involving experts to establish a "ground truth" in the traditional sense of diagnostic accuracy.
Qualifications of Experts (Test Set)Not applicable/Not stated.
Adjudication Method (Test Set)Not applicable/Not stated.
Multi-Reader Multi-Case (MRMC) Study?No. This submission describes modifications to patient monitors, not a diagnostic imaging or AI-assisted interpretation device that would typically undergo a MRMC study.
Effect Size of Human Readers (with/without AI assist)Not applicable. This is not an AI-assisted diagnostic device.
Standalone Performance (Algorithm Only)?Not applicable. This is a medical device (patient monitor) with integrated hardware and software components for physiological parameter monitoring. The "performance" described is the overall device performance, not a standalone algorithm.
Type of Ground Truth UsedThe "ground truth" in this context refers to the established specifications and performance metrics of the predicate device (K052707) and the pre-cleared picoNIBP module (K051366), against which the modified device's performance was measured. It's based on engineering specifications and validated performance, not clinical outcomes or pathology from a patient cohort for diagnostic purposes.
Training Set Sample SizeNot applicable/Not stated. This device is not described as utilizing machine learning or AI that would require a "training set" in the common sense of the term for algorithm development. The development process would involve traditional engineering design, simulation, and testing.
How Ground Truth for Training Set was EstablishedNot applicable/Not stated. As above, this doesn't appear to be an AI/ML device requiring a training set with established ground truth. The "ground truth" for its development would be engineering principles, physiological models, and regulatory standards for vital signs monitoring.

{0}------------------------------------------------

K090483
p1/2

MAR 2 0 2009

510K Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807-92(c).

    1. The submitter of this pre-market notification is:
      Larry Milana Philips Medical Systems 3000 Minuteman Road Andover, MA 01810 United States

Tel: 978-659-7860 Fax: 978-685-5624 Email: larry.milana@philips.com

This summary was prepared on January 27, 2009.

  • The names of the subject devices are the Philips SureSigns VM Series Patient Monitors, the 2. SureSigns VS2 Vital Signs Monitor, and the SureSigns VS3 Vital Signs Monitor.
    1. The trade names of the devices are the SureSigns VM Series Patient Monitors, the SureSigns VS2 Vital Signs Monitor, and the SureSigns VS3 Vital Signs Monitor.
  • The common usual name is multi-parameter patient monitor. 4.
    1. The classification of the SureSigns VS2 Vital Signs Monitor, the SureSigns VS3 Vital Signs Monitor, and the SureSigns VM Series Patient Monitors are not changed by the addition of the picoNIBP (non - invasive blood pressure) OEM Module.
  • The modified devices are substantially equivalent to previously cleared Philips device, ୍. M3046B Compact Configurable Portable Patient Monitor marketed pursuant to K052707.
    1. The modification is to replace the currently cleared CAS NBP module with the picoNIBP(non invasive blood pressure) OEM Module (which was cleared in K051366) to the VS2 Vital Signs Monitor, the VS3 Vital Signs Monitor, and the VM Series (VM 4/6/8) Patient Monitors.

In addition, this submission includes a change in plastics to Valox and the addition of new SP0- sensors and the disclosure of IPX1 rating for ingress protection to vertical water drops for all subject devices.

{1}------------------------------------------------

  • The modified devices have the same intended use as the legally marketed predicate device. 8. The Philips SureSigns VM Series Patient Monitors, SureSigns VS2 Vital Signs Monitors and SureSigns VS3 Vital Signs Monitors are for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. For monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare environments. Additionally, the monitors may be used in transport situations within a healthcare facility.
  • The modified devices have the same fundamental technological characteristics as the legally တ် marketed predicate devices. The subject devices use the same design as the predicate device.
    1. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the Philips SureSigns VM Series Patient Monitors, SureSigns VS2 Vital Signs Monitors and SureSigns VS3 Vital Signs Monitors meet all claims and support a determination of substantial equivalence.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its head facing left and its wings spread. The eagle is composed of three thick, curved lines. Surrounding the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular pattern.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 9 0 2009

Philips Medical System c/o Mr. Larry Milana Regulatory Affairs Specialist 3000 Minuteman Road Andover, MA 01810

Re: K090483

Suresigns VS2 monitor, Suresigns VS3 vital signs monitor, Suresigns VM series patient monitors

Regulation Number: 21 CFR 870.1025

Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm)

Regulatory Class: Class II (Two)

Product Code: MXH, (DPS, DXN, DQA, CCK and FFL)

Dated: February 24, 2009

Received: February 25, 2009

Dear Mr. Milana:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Larry Milana

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html,

Sincerely yours,

1 Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Eyaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510 (k) Number (if known):

Device Name: SureSigns VS2 (reference numbers: 863079, 863080, 863081, 863082) SureSigns VS3 (reference numbers: 863069, 863070, 863071, 863072, 863073, 863074) SureSigns VM Series Patient Monitors (VM1/3/4/6/8) (reference numbers: 863063, 863064, 863065, 863066, 863068, 863077,863078)

Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

Intended Use: For monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility.

Prescription Use:YESAND/OR over-the-counter Use:NO
(Part 21 CFFR 801 Subpart D)(21 CFR 807 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Signature

(Division Sign-Off)
Division of Cardiovascular Devices
510(k) NumberK090483

February 24, 2009

Page 23 of 71

§ 870.2340 Electrocardiograph.

(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).