K Number
K090481

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-03-19

(23 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DEMASQ Imaging Software Device™ is a software device that receives medical images and data from various imaqing sources (including but not limited to CT. MR. US, RF units, computed and direct radiographic devices, and secondary capture devices, (scanners, imaging gateways or imaging sources). Images and data can be stored. communicated, processed and displayed within the system or across computer networks at distributed locations.

Only pre-processed DICOM for presentation images can be interpreted for primary image diagnosis in mammography.

Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Typical users of this system are trained professionals, e.g. physicians, radiologists, nurses, medical technicians, and assistants.

Device Description

DEMASQ Imaging Software Device™ (DISD) is a user network and or a web-based PACS DICOM viewing software running on an IBM Compatible PC with a Windows 98 SE, Windows NT 4.0 (with SP 5 or later), XP or Vista operating system.

DISD is adapted for, storing, processing, routing and report generating. DISD supports the DICOM standard. To assist interpretation a user-friendly interface will allow the enduser to color, crop, zoom, rotate, measure and save the image. Images can be saved to a desktop or network as a standard file format.

DISD allows users to take full advantage of the radiographic images from digital X-ray and various modalities in order to obtain invaluable mission critical diagnostic data and images.

AI/ML Overview

This 510(k) submission for the DEMASQ Imaging Software Device™ is a declaration of substantial equivalence to predicate devices, rather than a submission for a novel device requiring extensive clinical studies to demonstrate performance against acceptance criteria. Therefore, the information provided focuses on demonstrating similar technological characteristics and intended use to existing, FDA-cleared devices, rather than establishing new performance metrics through a dedicated study.

Based on the provided text, a detailed study proving the device meets specific acceptance criteria, as one would expect for a new diagnostic algorithm or AI product, is not present or described.

Here's an analysis based on the structure requested, noting the absence of certain information:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria: Not explicitly stated as quantifiable performance metrics (e.g., sensitivity, specificity, accuracy) for a diagnostic task. The "acceptance criteria" for this type of device (a PACS/DICOM viewer) are implicitly tied to its ability to perform its stated functions effectively and be substantially equivalent to predicate devices. These functions include:

  • Receiving medical images and data from various imaging sources.
  • Storing, communicating, processing, and displaying images and data.
  • Supporting DICOM standard.
  • Providing user-friendly interface for color, crop, zoom, rotate, measure, and save images.
  • Allowing full advantage of radiographic images from digital X-ray and various modalities.

Reported Device Performance:
The submission asserts that the DEMASQ Imaging Software Device™ (DISD) has essentially the same technological characteristics, general function, application, and intended use as its predicate devices (INFINITT G3 PACS and XEBRA DICOM IMAGE BROWSER). This claim of substantial equivalence serves as the primary "performance" assertion for this type of device.

Feature/Criterion (Implicit)DEMASQ Imaging Software Device™ Performance (Claimed)
Technological Equivalence to Predicates"Essentially the same technological characteristics" as INFINITT G3 PACS (K052545) and XEBRA DICOM IMAGE BROWSER (K071894).
Functional Equivalence"Essentially the same... general function, application, and intended use" as predicates. Includes: advanced reviewing (contrast, magnification), parameter evaluation (pixel values, measurements), remote diagnoses, Ethernet/TCP/IP/DICOM 3.0 data acquisition, Windows UI, internal DICOM 3.0 handling, on-line help.
Image Handling & ModalitiesReceives medical images and data from CT, MR, US, RF units, computed and direct radiographic devices, and secondary capture devices. Stores, communicates, processes, and displays these within the system or across computer networks.
DICOM ComplianceSupports DICOM standard. Internal handling of images in DICOM 3.0 format. DICOM 3.0 radiology image file standards for data acquisition.
Image Manipulation ToolsUser-friendly interface allows color, crop, zoom, rotate, measure, and save images.
Operating System CompatibilityRuns on IBM Compatible PC with Windows 98 SE, Windows NT 4.0 (SP 5+), XP, or Vista.
Maneuver to avoid lossy compressed mammography review*Only pre-processed DICOM for presentation images can be interpreted for primary image diagnosis in mammography. Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

*These are limitations/warnings on use, not performance metrics.


2. Sample Size Used for the Test Set and Data Provenance

The provided document describes a 510(k) submission for a Picture Archiving Communications System (PACS) software, which is a medical image viewing and management device. It aims to demonstrate substantial equivalence to existing predicate devices, not to prove the statistical performance of a novel diagnostic algorithm.

Therefore, there is no mention of a specific "test set" of medical images used to evaluate the device's diagnostic performance in the way one would test an AI algorithm. The evaluation focuses on functional comparisons to predicate devices and adherence to standards.


3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Since there is no "test set" for diagnostic performance, there is no mention of experts or ground truth establishment related to diagnostic accuracy for this device in the provided text.

The "experts" involved are implied to be "trained professionals, e.g. physicians, radiologists, nurses, medical technicians, and assistants" as typical users who interpret images and information displayed by the system, providing "ample opportunity for competent human intervention." These are users, not experts establishing ground truth for the device's internal performance.


4. Adjudication Method for the Test Set

Given the absence of a test set for diagnostic performance, there is no adjudication method described in the submission.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This submission is for a PACS viewing software, not an AI-assisted diagnostic tool. Therefore, no MRMC comparative effectiveness study was done and no effect size for human reader improvement with AI assistance is reported in this document. The device is a "viewer," and human interpretation is still paramount.


6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is a PACS viewing software, not a standalone diagnostic algorithm. Therefore, no standalone performance study was done for an algorithm without human-in-the-loop performance. The document explicitly states: "A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed."


7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

Since there is no diagnostic performance study, no type of ground truth is mentioned as being used to evaluate the device's diagnostic capabilities.


8. The Sample Size for the Training Set

This submission is for a PACS viewing software, not a machine learning algorithm requiring a training set. Therefore, there is no training set mentioned in the provided document.


9. How the Ground Truth for the Training Set Was Established

As there is no training set for an AI algorithm, there is no mention of how ground truth for a training set was established.

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

Date Prepared: November 20, 2008

Submitter's Information: 21 CFR 807.92(a)(1) Mr. Stuart Gall, CEO DEMASQ Limited 8th Floor Eastgate House 35-43 Newport Road Cardiff, Wales CF240AB United Kingdom +44 114 275555 Tel: +44 292 0491275 Fax: Email: stuartgall(@demasq.com

Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)

Trade Name:DEMASQ Imaging Software Device™
Common Name:Picture Archiving Communications System
Classification Name:system, image processing, radiological
Product code:LLZ
Device Classification:892.2050

Predicate Device: 21 CFR 807. 92(a)(3)

510(k) NumberK052545K071894
Device Classification Namesystem, image processing,radiologicalsystem, image processing,radiological
Regulation Number892.2050892.2050
Device NameINFINITT G3 PACSXEBRA DICOM IMAGEBROWSER
ApplicantINFINITT CO., LTD.HX TECHNOLOGIES, INC.
Product CodeLLZLLZ
Decision Date11/08/200508/16/2007

Device Description: 21 CFR 807 92(a)(4)

DEMASQ Imaging Software Device™ (DISD) is a user network and or a web-based PACS DICOM viewing software running on an IBM Compatible PC with a Windows 98 SE, Windows NT 4.0 (with SP 5 or later), XP or Vista operating system.

DISD is adapted for, storing, processing, routing and report generating. DISD supports the DICOM standard. To assist interpretation a user-friendly interface will allow the enduser to color, crop, zoom, rotate, measure and save the image. Images can be saved to a desktop or network as a standard file format.

DISD allows users to take full advantage of the radiographic images from digital X-ray and various modalities in order to obtain invaluable mission critical diagnostic data and images.

l

{1}------------------------------------------------

510(k) Summary of Safety and Effectiveness

Indications for Use: 21 CFR 807 92(a)(5)

DEMASQ Imaging Software Device™ is a software device that receives medical images and data from various imaging sources (including but not limited to CT, MR, US, RF units, computed and direct radiographic devices, and secondary capture devices, (scanners, imaging gateways or imaging sources). Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations.

Only pre-processed DICOM for presentation images can be interpreted for primary image diagnosis in mammography.

Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Typical users of this system are trained professionals, e.g. physicians, radiologists, nurses, medical technicians, and assistants.

Technological Characteristics: 21 CFR 807 92(a)(6)

DEMASQ Imaging Software Device™ is a software device that does not contact the patient, nor does it control any life sustaining devices.

A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.

DEMASQ Imaging Software Device ™ and its predicates have essentially the same technological characteristics.

The predicate devices and the new device are software PACS for use in medical image acquisition, management, manipulation, and distribution. All systems have been developed to replace traditional film handling in radiology.

The new device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use.

The predicate device and the new device include the following similar features and functionality:

  • Advanced reviewing features as contrast adjustments, magnification and more. .
  • Evaluation of general parameters such as pixel values and measurements of . distances and angles in the images.
  • Possibility of remote diagnoses through transmission via public or private . telecommunications network (including the Internet) of an examination.
  • Image data acquisition based on Ethernet communication standard with TCP/IP file . transfer protocol and DICOM 3.0 radiology image file standards
  • A user interface using Microsoft Windows operating systems. .
  • Internal handling of images in DICOM 3.0 format. .
  • On-line help system t

Conclusion: 21 CFR 807 92(b)(1)

The 510(k) Pre-Market Notification for DEMASQ Imaging Software Device™ contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device. The subject device has been designed and will be delivered in accordance with the voluntary standards.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three lines forming its body and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

MAR 1 9 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEMASQ Limited % Mr. Jeff D. Rongero Senior Project Engineer Underwriters Laboratories, Inc. 12 Laboratory Drive Research Triangle Park, NC 27709

Re: K090481

Trade/Device Name: DEMASQ Imaging Software Device™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 20, 2009 Received: February 24, 2009

Dear Mr. Rongero:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxx(Gastroenterology/Renal/Urology)(240) 276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)(240) 276-0115
21 CFR 892.xxx(Radiology)(240) 276-0120
Other(240) 276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division-of-Surveillance-Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppor/index.html.

Sincerely yours,

Laura M. Morris

anine M. Morris Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K09048

Device Name: DEMASQ Imaging Software Device™

Indications for Use:

DEMASQ Imaging Software Device™ is a software device that receives medical images and data from various imaqing sources (including but not limited to CT. MR. US, RF units, computed and direct radiographic devices, and secondary capture devices, (scanners, imaging gateways or imaging sources). Images and data can be stored. communicated, processed and displayed within the system or across computer networks at distributed locations.

Only pre-processed DICOM for presentation images can be interpreted for primary image diagnosis in mammography.

Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Typical users of this system are trained professionals, e.g. physicians, radiologists, nurses, medical technicians, and assistants.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED).

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off Office of In Vitro Øiagnostic Device Evaluation and Safety

510(k)

Hedy Lamarr

(Division Sign-Off) Division of Reproductive. Abdominal Radiological Devices 510(k) Number

Page 1 of 1 Updated June 4, 2007

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).