K Number
K090408

Validate with FDA (Live)

Date Cleared
2009-03-20

(30 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DSS™ Stabilization System - Rigid is intended as a single-level system for noncervical pedicle fixation from the T4 to S1 vertebrae in skeletally mature patients to help provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion. The DSS™ Stabilization System - Rigid Coupler is intended to be used with autograft and/or allograft.

Device Description

This 510(k) adds additional screw sizes to the DSS™ Stabilization System. The DSSTM Stabilization System is comprised of a variety of pedicle screws sizes, and couplers that act as longitudinal spacers that are uniquely fitted for each individual case. The pedicle screws and couplers are manufactured from medical grade titanium alloy (Ti6Al4V).

AI/ML Overview

Here's an analysis of the provided text regarding the Paradigm Spine DSS™ Stabilization System and its 510(k) submission (K090408), focusing on the acceptance criteria and the study (or lack thereof) proving the device meets them:

No clinical performance study or specific acceptance criteria are described in the provided 510(k) summary documents.

The provided documents are a 510(k) summary and the FDA's clearance letter. For a Class III device like a pedicle screw spinal system, the FDA typically determines substantial equivalence based on a comparison to a legally marketed predicate device. This often involves demonstrating equivalent technological characteristics and performance through non-clinical (e.g., mechanical) testing, rather than extensive clinical efficacy trials, especially for devices introducing minor modifications (like additional screw sizes as stated here).

Therefore, many of the requested categories related to clinical study design and performance metrics cannot be filled from the given text.


1. Table of Acceptance Criteria and Reported Device Performance

Based on the provided documents, specific acceptance criteria for clinical performance and reported device performance are NOT detailed.

The approval is based on "substantial equivalence" to a predicate device (K080241). The performance stated is that:

Acceptance Criteria (Implied)Reported Device Performance
Mechanical Soundness (compared to predicate devices)"Testing performed indicate that the DSS™ is as mechanically sound as predicate devices."
Indications for Use (matching predicate device)"DSS™ Stabilization System was shown to be substantially equivalent to previously cleared devices (K080241) and has the same indications for use..."
Design, Function, and Materials (matching predicate device)"...has the same indications for use, design, function, and materials used."

Note: "Mechanical Soundness" implies that the device met certain mechanical testing standards (e.g., fatigue, static strength), but the specific criteria (e.g., force levels, cycles) are not provided in this summary.


2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size for Test Set: Not applicable/not provided for clinical performance. The "test set" would refer to mechanical testing samples, which are not detailed here.
  • Data Provenance: Not applicable/not provided for clinical performance. The mechanical testing data would likely be from the manufacturer's internal labs.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • Not applicable, as no clinical test set requiring expert ground truth establishment is described.

4. Adjudication Method for the Test Set

  • Not applicable, as no clinical test set requiring adjudication is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, an MRMC comparative effectiveness study was not done or described in these documents. This is a mechanical orthopedic device, not an imaging AI diagnostic device, so MRMC studies are generally not relevant for 510(k) clearance of this type of product.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not applicable. This is a medical implant device, not an algorithm.

7. The Type of Ground Truth Used

  • Not applicable for clinical performance. For mechanical testing, the "ground truth" would be the engineering specifications and performance benchmarks derived from the predicate device and established standards for pedicle screw spinal systems.

8. The Sample Size for the Training Set

  • Not applicable. This is a medical implant device, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. This is a medical implant device, not an AI/ML algorithm.

Summary of 510(k) Clearance for K090408:

The K090408 submission was a Special 510(k), specifically to add additional screw sizes to an already cleared DSS™ Stabilization System (predicate K080241). The primary basis for clearance was a demonstration of substantial equivalence to the predicate device in terms of:

  • Mechanical Soundness: Testing indicated the new screw sizes were "as mechanically sound as predicate devices."
  • Indications for Use: The system maintained the same indications as the predicate.
  • Design, Function, and Materials: These were also consistent with the predicate.

The FDA's review concluded that there was a reasonable likelihood that the device could be used for spinal stabilization without fusion, which could cause harm. Therefore, a specific Warning was mandated in the labeling: "The safety and effectiveness of this device has not been established for the intended use of spinal stabilization without fusion. This device is only intended to be used when fusion with autogenous bone graft is being performed at all instrumented levels." This indicates a careful consideration of the intended use and potential off-label use, even in the absence of a de novo clinical trial for this specific 510(k).

{0}------------------------------------------------

Paradigm Spine DSS™ Special 510(k)

K090408 (pg. 10f1)

5. 510(k) Summary

Contact:Mr. Adam HerderMusculoskeletal Clinical & Regulatory Advisers, LLC1331 H Street NW, 12th FloorWashington, DC 20005202.552.5800
Device Trade Name:DSSTM Stabilization SystemMAR 20 2009
Manufacturer:Paradigm Spine, LLC505 Park Ave. 14th FloorNew York, NY 10022
Classification:21 CFR §888.3070, Pedicle screw spinal system
Class:III
Product Code:NKB, MNH, MNI

Indications For Use:

The DSS™ Stabilization System - Rigid is intended as a single-level system for noncervical pedicle fixation from the T4 to S1 vertebrae in skeletally mature patients to help provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion. The DSS™ Stabilization System - Rigid Coupler is intended to be used with autograft and/or allograft.

Device Description:

This 510(k) adds additional screw sizes to the DSS™ Stabilization System. The DSSTM Stabilization System is comprised of a variety of pedicle screws sizes, and couplers that act as longitudinal spacers that are uniquely fitted for each individual case. The pedicle screws and couplers are manufactured from medical grade titanium alloy (Ti6Al4V).

Predicate Device(s):

DSS™ Stabilization System was shown to be substantially equivalent to previously cleared devices (K080241) and has the same indications for use, design, function, and materials used.

Performance Standards:

Testing performed indicate that the DSS™ is as mechanically sound as predicate devices.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three horizontal lines representing the branches of government. The logo is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Public Health Service

MAR 2 0 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Paradigm Spine, LLC % Musculoskeletal Clinical Regulatory Advisors, LLC Mr. Adam Herder 1331 H Street Northwest, 12th Floor Washington, DC 20005

Re: K090408

Trade/Device Name: DSS Stabilization System-Rigid Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: III Product Code: NKB, MNH, MNI Dated: February 17, 2009 Received: February 18, 2009

Dear Mr. Herder:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of device's labeling:

The safety and effectiveness of this device has not been established for the intended use of spinal stabilization without fusion. This device is only intended to be used when fusion with autogenous bone graft is being performed at all instrumented levels.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

{2}------------------------------------------------

Page 2 - Mr. Adam Herder

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Christy Foreman for

Donna-Bea Tillman, Ph.D., M.P.A Director -- - - -Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Paradigm Spine DSSTM Special 510(k)

4. Indications for Use

510(k) Number (if known): KO90408

Device Name: DSS™ Stabilization System

The DSS™ Stabilization System - Rigid is intended as a single-level system for noncervical pedicle fixation from the T4 to S1 vertebrae in skeletally mature patients to help provide immobilization and stabilization of spinal segments as an adjunct to fusion. for the following indications: degenerative disc. disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); turnor; pseudarthrosis; and failed previous fusion. The DSS™ Stabilization System - Rigid Coupler is intended to be used with autograft and/or allograft.

Prescription Use V . (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Pet D. Rumm mo dep 10,2

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number K090408

Page 1 of

Page 23 of 36