K Number
K090214

Validate with FDA (Live)

Date Cleared
2009-03-13

(43 days)

Product Code
Regulation Number
892.1560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  • Abdomen, B-Mode, per Indications for Use Ultrasound Form .
  • The BladderScan® BVM 9500 is intended to project ultrasound energy through . the lower abdomen to obtain an image of the bladder and measure urinary bladder volume and mass non-invasively.
  • Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Abdominal, Other (Bladder)
Device Description

The BladderScan® BVM 9500 is a B-mode ultrasonic instrument, portable and battery operated, intended for the noninvasive measurement of urinary bladder volume and bladder mass. The BladderScan® BVM 9500 calculates the bladder volume and mass using patented Vmode® technology. Vmode ultrasound is easy to use and comfortable for the patient. When you release the scan button, within seconds, the Vmode technology measures ultrasonic reflections on multiple planes inside the body and produces a 3-dimensional image. Based on this image, the BladderScan® calculates and displays the bladder volume and bladder mass. Volume measurements made with Vmode ultrasound are more accurate than those from conventional 2-dimensional ultrasound, as they are based on a more complex, 3-D image of the bladder. Bladder volume and mass, mode (uterus detection or standard BladderScan), directional aiming with real-time feedback, battery status, and usage rate indicators are all displayed on the BladderScan® BVM 9500 LCD display. The instrument contains a thermal printer that will allow the user to print exams with a single button click.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Verathon Inc. BladderScan® BVM 9500 Ultrasound System, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the comparison to previously cleared devices. The study compares the performance of the BVM 9500 against known volumes and wall thicknesses of phantoms.

Performance MetricAcceptance Criteria (Implied)Reported Device Performance (Implied)
Bladder Volume AccuracySubstantially equivalent to Verathon BladderScan® BVI 9400 (K071217)Met, via comparison to phantoms
Bladder Mass AccuracyEquivalent to Verathon BladderScan® BVM 6500 (K030763)Met, via comparison to phantoms
Bladder Wall Thickness Measurement AccuracyImproved due to higher resolution (7.3MHz signal)Assumed improved and accurate, demonstrated via phantoms
Acoustic OutputRemains within Preamendment limitsMet

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated, but the study used "tissue equivalent phantoms with known dimensions." The text implies multiple phantoms were used to demonstrate accuracy for both volume and mass.
  • Data Provenance: The testing was conducted using synthetic phantoms manufactured by CIRS. This is a controlled, laboratory-based environment, not real-world patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Not applicable.
  • Qualifications of Experts: Not applicable.
  • Method of Ground Truth Establishment: The ground truth was established by the known dimensions of the tissue-equivalent phantoms provided by CIRS. These phantoms are designed to have quantifiable and precise characteristics (known volume and wall thickness).

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. The "ground truth" was the known physical properties of the phantoms, not subjective expert assessment.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

  • MRMC Study: No, an MRMC comparative effectiveness study was not conducted or reported in the provided text. This study focused on the standalone performance of the device against phantoms.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

  • Standalone Study: Yes, the described study is a standalone performance evaluation. The BladderScan® BVM 9500's ability to measure bladder volume and mass was assessed directly against the known values of the phantoms, without involving human readers or human-in-the-loop performance.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The ground truth used was phantom-based known dimensions. Specifically, the phantoms had "known volume and wall thickness."

8. The Sample Size for the Training Set

  • Sample Size: The document does not provide information about a separate "training set" or its size. The device calculates volume and mass using "patented Vmode® technology" and algorithms, but the development and training process for these algorithms are not detailed in this section. The focus is on the validation of the finalized device.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not discussed in the provided text. The document describes the validation of the final product against phantoms, not the internal development or training of its underlying algorithms.

{0}------------------------------------------------

K090214/
17

510(K) SUMMARY (As required by 21 CFR 807.92(a))

MAR 1 3 2009

Submitter Information
Company:Verathon Inc.
20001 North Creek Parkway
Bothell, WA 98011
Phone:425-867-1348 ext.5642
Fax:425-883-2896
Email:krisn@verathon.com
Contact:Kristian Nielsen
Regulatory Affairs Specialist
Date:January 29, 2009
Device Information
Trade/Proprietary Name:Verathon Inc. BladderScan® BVM 9500 Ultrasound System
Common Name:Diagnostic Ultrasound System with Accessories
Classification Name(s):Regulatory Class: II
Review Category: Tier II
Classification Panel: Radiology
Ultrasonic Pulsed Echo Imaging System
FR Classification 892.1560
Product Code 90-IYO
Diagnostic Ultrasound Transducer
FR Classification 892.1570
Product Code 90-ITX
Predicate Device(s):Verathon BladderScan® BVI 9400 Ultrasound System (K071217)
And
Verathon BladderScan® BVM 6500 Ultrasound System (K030763)
Device Description:The BladderScan® BVM 9500 is a B-mode ultrasonic

: : : : :

{1}------------------------------------------------

instrument, portable and battery operated, intended for the noninvasive measurement of urinary bladder volume and bladder mass. The BladderScan® BVM 9500 calculates the bladder volume and mass using patented Vmode® technology. Vmode ultrasound is easy to use and comfortable for the patient. When you release the scan button, within seconds, the Vmode technology measures ultrasonic reflections on multiple planes inside the body and produces a 3-dimensional image. Based on this image, the BladderScan® calculates and displays the bladder volume and bladder mass. Volume measurements made with Vmode ultrasound are more accurate than those from conventional 2-dimensional ultrasound, as they are based on a more complex, 3-D image of the bladder.

Bladder volume and mass, mode (uterus detection or standard BladderScan), directional aiming with real-time feedback, battery status, and usage rate indicators are all displayed on the BladderScan® BVM 9500 LCD display. The instrument contains a thermal printer that will allow the user to print exams with a single button click.

(Same as previously cleared Verathon BVM 6500 Ultrasound System): The BladderScan® BVM 9500 is intended to project ultrasound energy through the lower abdomen to obtain an image of the bladder and measure urinary bladder volume and mass non-invasively.

The Verathon Inc. BladderScan® BVM 9500 Ultrasound System retains the same bladder volume measurement features of the Verathon inc. BladderScan® BVI 9400 System (K071217), but also has the Bladder mass measurement function and intended use as the previously cleared Verathon Inc. BladderScan® BVM 6500 (K030763).

The BVM 9500 is identical to the BVI 9400, with regards to hardware and calibration method however, the BVM 9500 transducer has been changed to a trifrequency transducer for better bladder detection and measurement.

Intended Use:

Comparison of Required Technological Characteristics:

{2}------------------------------------------------

The Verathon Inc. BladderScan® BVM 9500's center section of the transducer is generating higher-frequency ultrasound at 7.3MIIz. This special design is for bladder wall thickness measurement, which can yield higher resolution information and improves the thickness measurement accuracy. The ultrasonic power transmitted from the system is not user adjustable.

The portable Verathon Inc. BladderScan® BVM 9500 Ultrasound System is applied to the patient's abdomen along with Sontac Ultrasound Coupling Gel to obtain bladder volume and bladder mass. The BVM9500 Ultrasound System transducer collects two group of information: Group 1 includes cross-sectional Bmode images from 3MHz signal and the harmonic ratio from 1.74MHz: Group 2 includes the RF signal from higher frequency signal at 7.3MHz. Then the BVM9500 Ultrasound System utilizes the first group information to automatically calculate the urine volume via volumetric integration. At the same time, the 3D bladder surface area is also calculated based on the segmentations from every plane. Bladder wall thickness is measured automatically based on the7.3MHz RF signal in the following step. The higher resolution guarantees the accuracy of the thickness measurement of the thin bladder wall. Lastly, the 3D bladder surface area and the bladder wall thickness are computed.

In order to demonstrate the BladderScan® accuracy claimed in 0270-0425 -xx-55, a third party vendor was contracted to build tissue equivalent phantoms with known dimensions. The supplier of the tissue equivalent phantoms is CIRS and is known for supplying medical imaging phantoms to the medical marketplace.

The tissue equivalent phantom is essentially a balloon which gets filled with urine mimicking material and then the filled balloon is surrounded with tissue mimicking material. This balloon is a known volume and wall thickness and can be compared to the BladderScan® measurements from the BVM 9500.

{3}------------------------------------------------

Bladder volume and mass accuracy have been demonstrated by comparing the bladder mass and bladder volume phantom measurements to the measurements of the Verathon Inc. BladderScan® BVM 9500. The same procedures are used to prove the measurement accuracy of the previously cleared Verathon Inc.BVM 6500 (K030763) during production.

The calibration system for the Verathon Inc. BladderScan® BVM 9500 Ultrasound System is exactly the same as the Verathon Inc. BladderScan® BVI 9400 Ultrasound System (K071217).

Design and testing of the Verathon Inc. BladderScan® BVM 9500 Ultrasound System indicates that the BVM 9500 is substantially equivalent to the Verathon Inc. BladderScan® BVI 9400 Ultrasound System (K071217) for bladder volume measurements and equivalent to the Verathon Inc. BVM 6500 Ultrasound System (K030763) for bladder mass measurements.

All acoustic output measurements for the BladderScan® BVM 9500 Ultrasound System remains within Preamendment limits.

Summary and Conclusion of Non Clinical and Clinical Testing

{4}------------------------------------------------

Image /page/4/Picture/16 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized symbol that resembles an abstract human figure or a bird-like shape. The logo is presented in black and white.

MAR 17 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Kristian Neilsen Regulatory Affairs Specialist Verathon, Inc. Corporate Headquarters 20001 North Creek Parkway BOTHELL WA 98011

Re: K090214

Trade/Device Name: Verathon Inc. BladderScan® BVM 9500 Ultrasound System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO and ITX Dated: January 29, 2009 Received: January 29, 2009

Dear Mr. Neilsen:

This letter corrects our substantially equivalent letter of March 13, 2009.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include . requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Verathon Inc. BladderScan® BVM 9500 Ultrasound System, as described in your premarket notification:

Transducer Model Number

3.0 / 1.74 Second/ 7.3 MHz Third Harmonic Transducer

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device

{5}------------------------------------------------

can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

If you have any questions regarding the content of this letter, please contact Lauren Hefner at (240) 276-3666.

Sincerely yours,

Hulbert Leeman

ine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K090214

BladderScan® BVM 9500 Ultrasound System Device Name:________

Indications For Use:

  • Abdomen, B-Mode, per Indications for Use Ultrasound Form .
  • The BladderScan® BVM 9500 is intended to project ultrasound energy through . the lower abdomen to obtain an image of the bladder and measure urinary bladder volume and mass non-invasively.

Contraindications:

  • The BladderScan® BVM 9500 is contraindicated for fetal use and for use on . pregnant patients.
    Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Helus flemen

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

Page 1 of 1

{7}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM System: BladderScan® BVM 9500 Ultrasound System 3.0 / 1.74 Second/ 7.3MHz Third Harmonic Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track Ionly)Specific(Tracks I & III)BMPWDCWDColorDopplerCombined(Spec.)Other(Spec.)
OphthalmicOphthalmic
FetalImaging& OtherFetal
AbdominalP
Intra-operative(Abdominal organs andvascular)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (breast,thyroid, testicles)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel.(Convent.)
Musculo-skel.(Superfic.)
Intra-luminal
Other (spec.)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (card.)
Other (Abdominal AortaMeasurement)
Peripheral vessel
Other (Bladder)P

N= new indication; P= previously cleared by FDA (K071217 & K030763)

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.