K Number
K090211

Validate with FDA (Live)

Date Cleared
2009-07-01

(154 days)

Product Code
Regulation Number
888.3027
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Intended for the reduction and fixation of fractures in cancellous bone in the spine for kyphoplasty (for use with Cardinal Health Radiopaque Bone Cement).

Device Description

The Inflatable Bone Tamp (IBT) was designed for use in Balloon kyphoplasty. The balloon serves to create a cavity in the vertebral body, thereby reducing the fracture and preventing cement leakage, while still allowing for cement interdigitation. The balloon catheter is the functional part of the device that creates a cavity and reduces the fracture. The balloon catheter provides a conduit through which the physician can inflate the balloon at the distal end of the catheter. The wire stiffener provides stiffness to the balloon catheter to facilitate insertion through the access cannula

AI/ML Overview

This document, K090211, describes an Inflatable Bone Tamp (IBT) submitted for 510(k) clearance by Cardinal Health. The submission focuses on demonstrating substantial equivalence to predicate devices, rather than establishing acceptance criteria and conducting a study to prove a specific level of performance for the device. Here's a breakdown of the requested information based on the provided text:

Key Takeaway: The provided document is a 510(k) summary, which aims to demonstrate "substantial equivalence" of a new device to existing predicate devices. It does not outline specific acceptance criteria for a new performance benchmark for the device, nor does it detail a study that proves the device meets such criteria as one might find for a novel device with new claims. Instead, it relies on demonstrating that the new device shares similar technological characteristics and performs comparably to already cleared devices.

1. A table of acceptance criteria and the reported device performance

Based on the provided text, specific acceptance criteria in the sense of quantitative performance thresholds (e.g., "Must achieve XYZ accuracy" or "Tensile strength of X MPa") are not explicitly stated. The "acceptance criteria" for this 510(k) submission are implicitly tied to demonstrating "substantial equivalence" to the predicate devices. The performance reported is that the device is substantially equivalent to its predicates.

Acceptance Criteria (Implicit)Reported Device Performance
Materials meet biological safety standards (ISO 10993-1).All materials used were evaluated through biological qualification safety tests as outlined in ISO 10993 Part-1 "Biological Evaluation of Medical Devices" and found acceptable.
Materials meet industry-recognized test methods.Materials also were tested in accordance with industry recognized test methods and were found to be acceptable for the intended use.
Functional characteristics are substantially equivalent to predicate devices.Performance testing demonstrated that the proposed device is substantially equivalent to the currently marketed predicate devices with regard to functional characteristics.
Design, materials, and manufacturing are same or similar to predicate devices.The proposed device and the predicate devices are composed of the same or similar design, materials, and manufacturing characteristics.

2. Sample sized used for the test set and the data provenance

The document mentions "Performance testing" and "biological qualification safety tests" but does not specify the sample sizes used for these tests or the data provenance (e.g., country of origin, retrospective/prospective). This information is typically detailed in the full 510(k) submission, not necessarily in the public summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This type of information is not applicable or not mentioned in this 510(k) summary. The context is not one of diagnostic accuracy or expert consensus on findings, but rather physical and biological performance of a medical device. "Ground truth" in this context would likely refer to established engineering and material science standards for device components, not expert interpretation of medical data.

4. Adjudication method for the test set

Not applicable or not mentioned. Adjudication methods are typically used in studies involving human interpretation or subjective assessments. The tests described (material safety, functional characteristics) are objective engineering and biological tests that wouldn't require an adjudication method in the clinical sense.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an Inflatable Bone Tamp, a physical medical device used in surgery, not an AI or imaging diagnostic tool. Therefore, an MRMC study related to human reader improvement with AI assistance is entirely outside the scope of this device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. As stated above, this is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical testing described:

  • Material safety: The "ground truth" would be established by validated ISO 10993 standards and other industry-recognized test methods for biocompatibility and material properties.
  • Functional characteristics: The "ground truth" would be the established performance characteristics of the predicate devices as measured against engineering specifications and validated test methods.

8. The sample size for the training set

Not applicable. This is not a study involving machine learning or algorithms that require a training set. The term "training set" is not relevant to the evaluation of this physical medical device.

9. How the ground truth for the training set was established

Not applicable. As there is no training set for this type of device evaluation, there is no ground truth to establish for it.

{0}------------------------------------------------

K090211

CardinalHealt McGaw Park, Illinois 60085-6787 AX: 847.785.2506

JUL - 1 2009

SMDA REQUIREMENTS

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Inflatable Bone Tamp

Sponsor:

Cardinal Health 1430 Waukegan Road MPKB McGaw Park, IL 60085

Regulatory Affairs: Contact

Telephone:

Date Summary Prepared:

Common Name:

Regulation Description:

Device Class and Regulation Number:

Predicate Devices:

Description:

Sharon Nichols

(847) 578-6610

January 2009

Inflatable Bone Tamp

Primary - Arthroscope Secondary - Cement, bone, vertebroplasty

Class II per 21CFR §888.1100, Procode HRX: Class II per 21CFR §888.3027, Procode NDN:

Kyphx Inflatable Bone Tamp, K041454, K981251 Radiopaque Bone Cement, K043518

The Inflatable Bone Tamp (IBT) was designed for use in Balloon kyphoplasty. The balloon serves to create a cavity in the vertebral body, thereby reducing the fracture and preventing cement leakage,

Image /page/0/Picture/23 description: The image shows the number 000035. The number is written in a simple, sans-serif font. The digits are evenly spaced and the number is horizontally oriented.

{1}------------------------------------------------

Intended Use:

Summary of Technological Characteristics:

Summary of testing:

Non-Clinical Testing:

while still allowing for cement interdigitation. The balloon catheter is the functional part of the device that creates a cavity and reduces the fracture. The balloon catheter provides a conduit through which the physician can inflate the balloon at the distal end of the catheter. The wire stiffener provides stiffness to the balloon catheter to facilitate insertion through the access cannula

Intended for the reduction and fixation of fractures in cancellous bone in the spine for kyphoplasty (for use with Cardinal Health Radiopaque Bone Cement).

The proposed device and the predicate devices are composed of the same or similar design, materials and manufacturing characteristics.

All materials used in the fabrication of the Inflatable Bone Tamp were evaluated through biological qualification safety tests as outlined in ISO 10993 Part-1 "Biological Evaluation of Medical Devices". These materials also were tested in accordance with industry recognized test methods and were found to be acceptable for the intended use.

Performance testing demonstrated that the proposed device is substantially equivalent to the currently marketed predicate devices with regard to functional characteristics.

000036

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

Public Health Service

JUL -- 1 2009.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Cardinal Health % Ms. Sharon Nichols 1430 Waukegan Road MPKB-3B McGaw Park, Illinois 60085

Re: K090211

Trade/Device Name: Inflatable Bone Tamp Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: II Product Code: NDN, HRX Dated: May 29, 2009 Received: June 3, 2009

Dear Ms. Nichols:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Sharon Nichols

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/cdrh/comp/ for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Barbarafneum

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and . Radiological Health

Enclosure

{4}------------------------------------------------

CardinalHealt

Cardinal Health 1430 Waukegan Road McGaw Park, Illinois 60085-6787 847.578.6610 FAX: 847.785.2506

Indication for Use

510(k) Number (if known):

Device Name:

Indications For Use:

unknown at this time K0902 ||

Inflatable Bone Tamp

Intended for the reduction and fixation of fractures in cancellous bone in the spine for kyphoplasty (for use with Cardinal Health Radiopaque Bone Cement).

Prescription Use X

Over-The Counter Use _

(Per 21 CFR 801 109)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE)

or

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K090211

Image /page/4/Picture/17 description: The image shows a sequence of numbers, specifically "000030". The numbers are printed in a bold, sans-serif font. The image appears to be a close-up of the numbers, with a slightly grainy texture.

N/A