Intended Use

MIRS ™ is used to create treatment plans for patients for whom radiation therapy has been prescribed. The system will calculate and display three dimensional radiation dose distributions within the patient for a given treatment plan set up. MIRS can be used for different treatment modalities of radiation therapy, such as conventional therapy, Conformal Radiation Therapy, Intensity Modulated Radiation Therapy, and Radiosurgery with Conic Collimators.

Device Description

MIRS ™ is used to create treatment plans for patients for whom external beam radiation therapy has been prescribed. MIRS is used to create three dimensional radiation dose distributions within the patient for a given treatment plan set up. MIRS (Module Integrated Radiotherapy System) is a modular patient treatment planning system for different treatment modalities of radiation therapy, from three dimensional radiotherapy to Conformal Radiation Therapy, Intensity Modulated Radiation Therapy and Conventional radiotherapy with Conic Collimators. MIRS supports any stereotactic frame. MIRS has automatic multimodality image registration, with and without fiducials. Besides 3D RMI and CT images studies with any orientation (axial, coronal or sagital), MIRS supports registration of functional RMI, PET and angiography.

AI/ML Overview

The provided 510(k) summary for the MIRS™ (Module Integrated Radiotherapy System) device does not contain specific acceptance criteria or an explicit study describing formal performance testing against such criteria. The document claims substantial equivalence to predicate devices based on "Technological Characteristics Comparison" and "Performance Data" sections, but these sections only state that items were found to be substantially equivalent without detailing the actual comparisons or performance metrics.

Therefore, the requested information regarding acceptance criteria, reported device performance, sample sizes, ground truth establishment, expert involvement, adjudication methods, and MRMC studies cannot be extracted from the provided text.

Based on the available information:

  1. Table of Acceptance Criteria and Reported Device Performance: This information is not provided in the document. The document states, "All of the items contained in the tables of section 10 have been found to be substantially equivalent," but does not present these tables or their content.

  2. Sample sized used for the test set and the data provenance: This information is not provided.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not provided.

  4. Adjudication method: This information is not provided.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This information is not provided. The device described is a treatment planning system, not an AI-assisted diagnostic tool that would typically involve human readers for interpretation.

  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: The document does not describe specific standalone performance studies. Its claims are centered around substantial equivalence to predicate devices, implying that its performance is comparable, but details of such a comparison are absent.

  7. The type of ground truth used: This information is not provided.

  8. The sample size for the training set: This information is not provided.

  9. How the ground truth for the training set was established: This information is not provided.

Summary of what is available:

  • Device Name: MIRS™ Module Integrated Radiotherapy System
  • Intended Use: To create treatment plans for patients for whom radiation therapy has been prescribed, calculating and displaying three-dimensional radiation dose distributions. It supports various modalities (conventional, conformal, IMRT, radiosurgery).
  • Comparison Method: Substantial equivalence claimed against multiple predicate devices based on technological characteristics and performance data, but specific details of the comparison are not given in the provided text.
  • Submission Type: Traditional Premarket Notification (510(K)).
  • Submission Date: December 8th, 2008
  • Clearance Date: May 4, 2009

{0}------------------------------------------------

Ke90682

V. Arredondo 2684-2ª Piso ♦ 1426 Buenos Aires ♦ Argentina Tel: (54) 114896-0989

510(K) SUMMARY

MAY 4 2009

MIRS ™

Submitter:Nuclemed S.A.V. Arredondo 2684 - 2ª Piso1426 Buenos Aires - ArgentinaPhone: (54) 11 4896-0989
Contact Person:Raymond J Kelly IVArazy Group56B Shadowbrook, Dr Hudson, NH 0305Phone: (603) 595-0095

Fax: (603) 595-0095

E-mail: ray@arazygroup.com

December 8th, 2008

Device Names:

Date:

MIRS ™ Module Integrated Radiotherapy System Trade Name: Module Integrated Radiotherapy System Common Name: System, Planning, Radiation Therapy Treatment Classification Name:

Legally Marketed Device to Which Substantial Equivalence is Claimed:

The MIRS ™ unit is substantially equivalent (SE) to Rad Calc (K010464) cleared in 2001, CDMS (K082606) cleared in 2008, PLATO SRS (K010784) cleared in 2001, FOCUS SRS (K973936 + K002147) cleared in 1998, IMSure (K031975) cleared in 2000, ERGO++ (K080601) cleared in 2008, ADAC Pinnacle3 (K951581) cleared in 1996, Pinnacle3 (K993923) cleared in 2000, and the FOCUS RTP (K915691) cleared in 1995. .'

Page #:
5.1

Traditional Premarket Notification for MIRS, Nuclemed S.A.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for NUCLEMED. The logo consists of a stylized letter N above the word "NUCLEMED". The letter N is black and white and appears to be made up of several smaller squares. The word "NUCLEMED" is written in a simple, sans-serif font.

Nuclem V. Arredondo 2684-2ª Piso ♦ 1426 Buenos Aires ♦ Argentina Tel: (54) 114896-0989

510(K) SUMMARY (CONTINUED)

MIRS ™

Device Description:

MIRS ™ is used to create treatment plans for patients for whom external beam radiation therapy has been MIRS ** Is used to create trouncell pails to pay three dimensional radiation dose distributions within the patient for a given treatment plan set up. MIRS (Module Integrated Radiotherapy System) is a modular patient for a given reatment planning system for different treatment modalities of radiation therapy, from three univelsional radiotherapy to Conformal Radiation Therapy, Intensity Modulated Radiation Therapy and Conventional radioneraly with Conic Collimators. MIRS supports any stereotactic frame. MIRS has automatic multimodality image registration, with and without fiducials. Besides 3D RMI and CT images studies with any orientation (axial, coronal or sagital), MIRS supports registration of functional RMI, PET and angiography.

Intended Use:

MIRS ™ is used to create treatment plans for patients for whom radiation therapy has been prescribed. The system will calculate and display three dimensional radiation dose distributions within the patient for a given treatment plan set up. MIRS can be used for different treatment modalities of radiation therapy, as conventional therapy. Conformal Radiation Therapy, Intensity Modulated Radiation Therapy, and Radiosurgery with Conic Collimators.

Technological Characteristics Comparison to Predicate Device:

The proposed device was compared to the predicate devices as described in section 9 of this submission. All of the items contained in the tables of section 9 have been found to be substantially equivalent.

Performance Data:

The proposed device's performance was compared to the predicate device's performance as described in section 10 of this submission. All of the items contained in the tables of section 10 have been found to be substantially equivalent.

Traditional Premarket Notification for MIRS, Nuclemed S.A.

Page #: 5.2

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol. The logo is in black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

MAY 4 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Nuclemed S.A. % Mr. Raymond Kelly Senior Consultant Arazy Group, Medes Argentina 56B Shadowbrook Dr. HUDSON NH 03051

Re: K090182 -

Trade/Device Name: MIRSTM Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: January 15, 2009 Received: February 3, 2009

Dear Mr. Kelly:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse, events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxx (Gastroenterology/Renal/Urology) (240) 276-0115 (240) 276-0115 (Obstetrics/Gynecology) 21 CFR 884.xxx (Radiology) (240) 276-0120 21 CFR 892.xxx Other (240) 276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Lois McMurtry

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

7

Nuclemed S.A. V. Arredondo 2684-2ª Piso ♦ 1426 Buenos Aires ♦ Argentina Tel: (54) 114896-0989

INDICATIONS FOR USE

2090182 510(k) Number (if known):

MIRS™M Device Name:

Indications For Use:

MIRS ™ is used to create treatment plans for whom radiation therapy has been prescribed.
The batient the count of the last direct signed redistion dose distributions within t MIRS ™ is used to create treatment plans to patients to whoirs within the patient for The system will calculate and display international treatment modelities of radiation therapy, as
a given treatment plan set up. MIRS during interest the Modulation Therany, a given treatment plan set up. MIRS can be used for unterests and the same of the appy, and Radiosurgery with Conic Collimators.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

Traditional Premarket Notification for MIRS, Nuclemed S.A.

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.