K Number
K090101

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-02-10

(26 days)

Product Code
Regulation Number
876.5540
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EQUISTREAM™ Iong-term hemodialysis catheter is indicated for use in attaining short-term or long-term vascular access for hemodialysis, hemoperfusion, and apheresis therapy. Access is attained via the internal jugular vein, external jugular vein, subclavian vein or femoral vein. Catheters greater than 40cm are intended for femoral vein insertion.

Device Description

Not Found

AI/ML Overview

The provided text describes a 510(k) premarket notification for the EQUISTREAM™ Long-Term Hemodialysis Catheter Family. However, it does not include information about acceptance criteria or a study proving the device meets specific performance metrics in the way that an AI/ML device submission would.

The document primarily focuses on demonstrating substantial equivalence to an existing predicate device (HemoSplit® Long-Term Hemodialysis Catheter, K030020). The concept of substantial equivalence in 510(k) means that the new device is as safe and effective as a legally marketed predicate device. This is typically achieved by showing similar technological characteristics, intended use, and safety/performance testing without raising new questions of safety or effectiveness.

Therefore, I cannot provide the requested information as it pertains to an AI/ML device's performance study with acceptance criteria, sample sizes, ground truth establishment, or expert reviews. The provided text is for a traditional medical device (catheter) regulated under the 510(k) pathway, which primarily relies on comparison to a predicate device rather than de novo performance studies with acceptance criteria in the context of AI/ML.

Here's why the requested information cannot be extracted from the provided text:

  • Acceptance Criteria & Reported Device Performance: These are not explicitly stated for the EQUISTREAM™ catheter in a quantifiable manner (e.g., sensitivity, specificity, accuracy). The document states "the subject EQUISTREAM™ catheters met the minimum requirements that are considered adequate for its intended use" and "Based on the indications for use, technological characteristics, and safety and performance testing, the subject EQUISTREAM™ catheters met the minimum requirements that are considered adequate for its intended use and is substantially equivalent...". This implies that the device performed comparably to its predicate in relevant tests, but the specific metrics are not detailed.
  • Sample Size (Test Set) & Data Provenance: Not applicable in this context. The "test set" for a traditional device like this would involve bench testing for material properties, flow rates, durability, etc., not a clinical data set for algorithm performance.
  • Number of Experts & Qualifications: Not applicable. Ground truth establishment with experts is a concept for AI/ML device validation.
  • Adjudication Method: Not applicable.
  • Multi-Reader Multi-Case (MRMC) Study: Not applicable, as this is for evaluating human performance with and without AI assistance.
  • Standalone Performance: While the catheter's standalone performance was assessed (e.g., flow rates, material compatibility), these are not presented with quantitative "acceptance criteria" and results in the provided summary.
  • Type of Ground Truth: Not applicable in the AI/ML sense. The "ground truth" for a catheter would be its physical and mechanical properties confirmed by engineering and material tests, and its clinical performance monitored post-market.
  • Sample Size (Training Set): Not applicable, as this is not an AI/ML device.
  • How Ground Truth for Training Set was Established: Not applicable.

In summary, the provided document is a 510(k) summary for a physical medical device (catheter) demonstrating substantial equivalence to a predicate, not an AI/ML device submission with performance metrics against acceptance criteria.

{0}------------------------------------------------

K090101 Page 1 of 2

Bard Access Systems EQUISTREAM™ Hemodialysis Catheter Special 510(k) Premarket Notification

Section 5 - 510(k) Summary

Page 17 of 77

510(k) Summary 21 CFR 807.92(a)

FEB 1 0 2009

EQUISTREAM™ Long-Term Hemodialysis Catheter Family Prepared January 14, 2009

Bard Access Systems, Inc. (BAS) [Subsidiary of C.R. Bard, Inc.] Submitter of 510(k) Premarket Notification: Salt Lake City, Utah 84116 Phone: (801) 522-5651 Fax: (801) 522-5425 General Provisions Contact Person: Jessica Agnello Requlatory Affairs Specialist EQUISTREAM™ Device Trade Name: Device Generic Name: Long-Term Hemodialysis Catheters HemoSplit® Trade Name: Common/Usual Name: Long-Term Hemodialysis Catheters Classification Name: 78 MSD-Catheter, Hemodialysis, Implanted 21 CFR §876.5540(b)(1), Class III + CFR Reference: Predicate Classification Panel: Gastroenterology and Urology Device Premarket Notification: See below Predicate Device Name HemoSplit® Long-Term K030020 June 16, 2003 Hemodialysis Catheter 21 CFR §876.5540(b)(1), Class III Classification 78 MSD-Catheter, Hemodialysis, Implanted Performance Performance standards have not been established by FDA under section 514 Standards of the Federal Food, Drug and Cosmetic Act. EQUISTREAM™ catheters are recommended for use in attaining short-term Intended Use or long-term vascular access for hemodialysis, hemoperfusion, and apheresis treatments. The EQUISTREAM™ Iong-term hemodialysis catheter is indicated for use in attaining short-term or long-term vascular access for hemodialysis, hemoperfusion, and apheresis therapy. Access is attained via the internal Indications for juqular vein, external jugular vein, subclavian vein or femoral vein. Catheters Use greater than 40cm are intended for femoral vein insertion. Technological similarities between the subject EQUISTREAM™ catheters Technological and the predicate device remain identical. There are no new questions raised regarding safety or efficacy of the subject EQUISTREAM™ catheters. Characteristics

{1}------------------------------------------------

K090101 Page 2 of 2

Bard Access Systems
EQUISTREAM™ Hemodialysis Catheter
Special 510(k) Premarket Notification

Section 5 - 510(k) Summary

Page 18 of 77

Summary of Substantial Equivalence

Based on the indications for use, technological characteristics, and safety and performance testing, the subject EQUISTREAM™ catheters met the minimum requirements that are considered adequate for its intended use and is substantially equivalent in design, materials, sterilization, principles of operation and indications for use to current commercially available catheters/cited predicates.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it, and three diagonal lines above it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the symbol.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 0 2009

Ms. Jessica Agnello Regulatory Affairs Specialist C. R. Bard, Inc. Bard Access Systems 605 North 5600 West SALT LAKE CITY UT 84116

Re: K090101

Trade/Device Name: EQUISTREAM™ Long-Term Hemodialysis Catheter Family Regulation Number: 21 CFR 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: MSD Dated: January 14, 2009 Received: January 15, 2009

Dear Ms. Agnello:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Agnello

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification'(21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (240) 276-3150. or at its Internet address

http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Heilsteen
Janine M. Morris

Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K090101

Section 4 - Indications for Use Statement

Page 15 of 77

Bard Access Systems EQUISTREAM™ Hemodialysis Catheter
Special 510(k) Premarket Notification

510(k) Number (if known): ¸ K ( 90 l 0 l

Device Name:

EQUISTREAM™ Long-Term Hemodialysis Catheter Family

Indications for Use:

The EQUISTREAM™ Iong-term hemodialysis catheter is indicated for use in attaining short-term or long-term vascular access for hemodialysis, hemoperfusion, and apheresis therapy. Access is attained via the internal jugular vein, external jugular vein, subclavian vein or femoral vein. Catheters greater than 40cm are intended for femoral vein insertion.

Prescription Use (Part 21 CFR §801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR §801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Huls

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

§ 876.5540 Blood access device and accessories.

(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.