K Number
K090034

Validate with FDA (Live)

Date Cleared
2009-06-02

(147 days)

Product Code
Regulation Number
868.1840
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ndd Medical Technologies Easy on-PC Spirometer is intended for prescriptions use only to conduct diagnostic spirometry testing of adults and pediatric patients over 4 years old, in general practice and specialty physician, industrial and hospital settings.

Device Description

Easy on-PC is a PC based diagnostic spirometer and consists of a PC software and the Easy on-PC sensor (Spiroson-AS). The Easy on-PC software is installed on a Windows based PC or laptop computer to which the Easy on-PC sensor is connected.

In order to conduct simple diagnostic spirometry testing, the Easy on-PC sensor is used in combination with a commercially available disposable breathing tube with integrated mouthpiece (spirette).

The sensor is an ultrasound flow sensor that measures the transit-time to determine flow velocity, volume and molar mass of the gas. The collected data is transferred to the PC for pulmonary function evaluation and data management. The results of the testing are stored in a database and reports can be displayed or printed.

AI/ML Overview

Here's an analysis of the acceptance criteria and study detailed in the provided document:

The document focuses on the Easy on-PC Spirometer, and the primary performance evaluation mentioned is against standards for diagnostic spirometry.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriterionReported Device Performance
Adherence to American Thoracic Society (ATS) recommendations for accuracy and precision in diagnostic spirometry tests."Dynamic wave form testing confirmed that the Easy on-PC Spirometer meets recommendations published by the American Thoracic Society (ATS) for accuracy and precision for the intended diagnostic spirometry tests."
Conformance with IEC 60601-1 requirements for electrical safety."The device was tested to demonstrate conformance with IEC 60601-1 and IEC 60601-1-2 requirements for electrical safety."
Conformance with IEC 60601-1-2 requirements for electrical safety (specifically electromagnetic compatibility)."The device was tested to demonstrate conformance with IEC 60601-1 and IEC 60601-1-2 requirements for electrical safety."
Biocompatibility of materials used."The materials used meet the requirements for biocompatibility in accordance with ISO 10993."
Software verification and validation to meet specified criteria."Software verification and validation revealed that the Easy on-PC software meets the specified criteria."
Substantial equivalence to the predicate device (Brentwood/Midmark Diagnostics IQmark Digital Spirometer) in terms of technological characteristics, specifically an equivalent level of accuracy for flow measurement."The device has the same technological characteristics as the predicate devices except for the use of ultrasonic technology for flow measurement. Testing was conducted that demonstrates this method of flow measurement is as accurate as the methods used in predicate devices and therefore substantially equivalent to the predicate device." (The ATS compliance listed above implicitly supports this, as the predicate would also need to meet such standards.)

2. Sample Size for the Test Set and Data Provenance

The document mentions "Dynamic wave form testing" to confirm ATS recommendations. This typically involves using a calibrated syringe or similar device that generates known flow-volume curves. The document does not specify the sample size (e.g., number of waveforms, number of tests, specific range of volumes/flows tested) or the data provenance in terms of patient data (e.g., country of origin, retrospective/prospective). The testing described is a technical performance validation rather than a clinical study with patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. For technical performance testing against ATS standards, the "ground truth" would be established by the precise and calibrated nature of the waveform generation equipment, not by clinical experts making diagnoses.

4. Adjudication Method for the Test Set

This information is not mentioned in the document, as the testing described is primarily technical performance validation against a standard, not a subjective clinical assessment requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

A MRMC comparative effectiveness study was not performed, nor is it applicable to this device. This device is a diagnostic spirometer, which directly measures physiological parameters (flow, volume). It does not involve human "readers" interpreting output in the way an AI-assisted diagnostic imaging system would. The "assistance" it provides is the measurement itself and presentation of results, not interpretive aid to a human reader in the context of an MRMC study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The primary performance evaluation described is "standalone" in the sense that the device itself (the sensor and software) takes measurements and processes them to produce results in accordance with ATS standards. There's no mention of a human-in-the-loop performance evaluation in the context of improving diagnostic accuracy; rather, the device is the diagnostic tool.

7. The Type of Ground Truth Used

The ground truth used for the primary performance claim (meeting ATS recommendations) was established by standardized, simulated waveforms generated by "dynamic wave form testing." This is a form of technical or engineering ground truth based on established physical and physiological standards for spirometry. It is not expert consensus, pathology, or outcomes data.

8. The Sample Size for the Training Set

The document does not describe any "training set" as this device is a measurement instrument based on physical principles (ultrasound transit-time) and does not rely on machine learning or AI models that require data training.

9. How the Ground Truth for the Training Set Was Established

As there is no mention of a training set, the establishment of ground truth for such a set is not applicable to this device.

{0}------------------------------------------------

K090034

510(k) Summary 5

JUN - 2 2009

In accordance with 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitter Information:

Submitter name:ndd Medizintechnik AGTechnoparkstrasse 18005 Zurich, Switzerland
United statescontact person:Mr. Jerry MasielloOperations Managerndd Medical Technologies, Inc
Phone:(877) 904.0090 x11
Fax:(978) 824.3853
Mailto:jmasiello@nddmed.com
Date prepared:December 05, 2008

Device name:

Proprietary name:Easy on-PC Spirometer
Common name:Spirometer
Classification name:Diagnostic spirometer
Product code:BZG

Predicate device:

Substantial equivalence is claimed to the Brentwood/Midmark Diagnostics IQmark Digital Spirometer, cleared for commercial distribution per K002499.

Device Description:

Easy on-PC is a PC based diagnostic spirometer and consists of a PC software and the Easy on-PC sensor (Spiroson-AS). The Easy on-PC software is installed on a Windows based PC or laptop computer to which the Easy on-PC sensor is connected.

In order to conduct simple diagnostic spirometry testing, the Easy on-PC sensor is used in combination with a commercially available disposable breathing tube with integrated mouthpiece (spirette).

The sensor is an ultrasound flow sensor that measures the transit-time to determine flow velocity, volume and molar mass of the gas. The collected data is transferred to the PC for pulmonary function evaluation and data management. The results of the testing are stored in a database and reports can be displayed or printed.

{1}------------------------------------------------

Intended Use:

The Easy on-PC Spirometer is intended for prescriptions use only to conduct diagnostic spirometry testing of adults and pediatric patients over 4 years old, in general practice and specialty physician, industrial and hospital settings.

Comparison of technological characteristics:

The device has the same technological characteristics as the predicate devices except for the use of ultrasonic technology for flow measurement. Testing was conducted that demonstrates this method of flow measurement is as accurate as the methods used in predicate devices and therefore substantially equivalent to the predicate device.

Summary of testing:

Dynamic wave form testing confirmed that the Easy on-PC Spirometer meets recommendations published by the American Thoracic Society (ATS) for accuracy and precision for the intended diagnostic spirometry tests.

The device was tested to demonstrate conformance with IEC 60601-1 and IEC 60601-1-2 requirements for electrical safety.

The materials used meet the requirements for biocompatibility in accordance with ISO 10993.

Software verification and validation revealed that the Easy on-PC software meets the specified criteria.

Conclusion:

Based on the above, ndd Medical Technologies concluded that the Easy on-PC Spirometer is substantially equivalent to the legally marketed predicate device and is safe and effective for its intended use.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus-like symbol in the center, consisting of three abstract shapes resembling human figures. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the symbol. The text is in all caps and is evenly spaced around the circle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

ndd Medizintechnik AG C/O Mr. Jerry Masiello Operations Manager ndd Medical Technologies, Incorporated Two Dundee Park Andover, Massachusetts 01810

JUN - 28 2009

Re: K090034

Trade/Device Name: Easy on-PC Spirometer Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: II Product Code: BZG Dated: May 27, 2009 Received: May 28, 2009

Dear Mr. Masiello:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Masiello

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/cdrh/comp/ for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Susan Russer

Susan Runner, D.D.S., MA Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if known):

Device Name: Easy on-PC Spirometer

Indications for Use:

The ndd Medical Technologies Easy on-PC Spirometer is intended for prescriptions use only to conduct diagnostic spirometry testing of adults and pediatric patients over 4 years old, in general practice and specialty physician, industrial and hospital settings.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

L. Schulten

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K090034

§ 868.1840 Diagnostic spirometer.

(a)
Identification. A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.(b)
Classification. Class II (performance standards).