K Number
K083746

Validate with FDA (Live)

Device Name
BIOSIGN FLU A+B
Date Cleared
2010-11-10

(693 days)

Product Code
Regulation Number
866.3328
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BioSign® Flu A+B test is an in vitro rapid qualitative test that detects influenza type A and type B antigens directly from nasal swab, nasopharyngeal swab, and nasopharyngcal aspirate/wash specimens obtained from patient with signs and symptoms of respiratory infection. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. A negative test result is presumptive and it is recommended these results be confirmed by viral culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions. The test is intended for professional and laboratory use.

Device Description

BioSign® Flu A+B is an immuno-chromatographic test for the rapid, qualitative detection of influenza A and B. The test device has two test lines, thereby allowing the separate identification of type A and/or type B viral antigens from the same specimen.

In the test procedure, a specimen is collected and placed into the Extraction Well of the test device containing extraction solution for one minute, during which time antigen is extracted from disrupted virus particles. The test device is then raised, tapped and laid back down onto a level surface to allow the solution in the Extraction Well to migrate through the pads containing lyophilized detector antibodies conjugated to gold dye and then through the test membrane. If influenza antigens are present in the specimen, they will react with anti-influenza antibody coupled to gold dye particles, migrate through the membrane as antigen-antibody-dye complexes, bind to the immobilized anti-influenza antibody on the membrane, and generate a colored line in the Test line position (A and/or B). The rest of the sample and unbound/bound dye complexes continue to migrate to the Control line position, where antibody to the antiinfluenza antibody is immobilized, and anti-influenza antibody-unbound/bound dye complexes form the Control line (internal procedural control).

AI/ML Overview

Here's an analysis of the provided 510(k) summary, extracting the requested information:

The device under review is the BioSign® Flu A+B / Status™ Flu A & B, an immuno-chromatographic test for the rapid, qualitative detection of influenza A and B antigens.


1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are implicitly set by the reported performance (sensitivity and specificity) when compared against viral culture as the reference method. The document presents performance for different sample types (nasopharyngeal aspirate, nasopharyngeal swab, nasal swab).

Table of Acceptance Criteria and Reported Device Performance:

Sample Type & AntigenAcceptance Criteria (Implicit from Study Results)Reported Device Performance (Sensitivity)Reported Device Performance (Specificity)
Nasopharyngeal AspirateHigh sensitivity and specificity required for clinical utility.
Flu A (Sensitivity)-95.3% (95% CI: 92.1-98.5%)-
Flu A (Specificity)--85.7% (95% CI: 83.3-88.1%)
Flu B (Sensitivity)-91.6% (95% CI: 83.6-99.6%)-
Flu B (Specificity)--97.5% (95% CI: 96.5-98.5%)
Nasopharyngeal SampleHigh sensitivity and specificity required for clinical utility.
Flu A (Sensitivity)-89.6% (95% CI: 84.0-95.2%)-
Flu A (Specificity)--77.0% (95% CI: 74.2-79.8%)
Flu B (Sensitivity)-86.8% (95% CI: 81.4-92.2%)-
Flu B (Specificity)--92.9% (95% CI: 91.2-94.6%)
Nasal Swab SampleHigh sensitivity and specificity required for clinical utility.
Flu A (Sensitivity)-91.7% (95% CI: 78.2-97.1%)-
Flu A (Specificity)--75.2% (95% CI: 70.2-79.6%)
Flu B (Sensitivity)-82.4% (95% CI: 59.0-93.8%)-
Flu B (Specificity)--88.3% (95% CI: 84.4-91.3%)

Note: The document does not explicitly state numerical "acceptance criteria" thresholds, but rather reports the observed performance and implies these values were sufficient for regulatory clearance. The wide confidence intervals in some cases suggest that the performance might vary.


2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set:
    • Total Patients: 862
    • Nasopharyngeal Aspirate: 253 cases
    • Nasopharyngeal Sample: 251 cases
    • Nasal Swab Sample: 358 cases
      (Note: The sum of individual sample types is 862, matching the total number of patients, indicating these might be overlapping or distinct subsets of samples from the same patient pool, or simply different sample types evaluated from the overall patient population.)
    • Archived Samples: 80 samples for both Flu A and Flu B verification (Columbia NY Presbyterian Hospital).
  • Data Provenance:
    • Country of Origin: USA (samples collected at 5 sites in the USA).
    • Retrospective or Prospective: Prospective clinical study, with additional testing on "Archived Samples" which would be considered retrospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The document does not specify the number of experts used to establish the ground truth or their qualifications. The ground truth method was viral culture, and it's common for laboratory professionals to perform and interpret viral cultures, but no details on their expertise are provided.


4. Adjudication Method for the Test Set

The primary reference method for initial comparison was viral culture. For samples that produced discrepant results between the BioSign® Flu A+B test and viral culture, a third, more definitive method was used for resolution: proFLU plus by Prodesse (real-time RT-PCR, or PCR). This constitutes an adjudication method, where ambiguous cases are resolved by a higher-tier test. The document refers to the results from the PCR as "further analyzed" or used to identify cases that were "positive by both BioSign and PCR" or "negative by both BioSign and PCR".


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This study design is typically used for image-based diagnostic aids where human readers interpret results with and without AI assistance to assess the impact of AI on reader performance. The BioSign® Flu A+B is a rapid diagnostic test, not an AI-powered diagnostic imaging device.


6. Standalone (Algorithm Only) Performance Study

Yes, a standalone performance study was done for the device. The reported sensitivity and specificity values directly reflect the performance of the BioSign® Flu A+B device itself, without any human interpretation adjustments beyond standard operation and reading of a rapid diagnostic test. The entire clinical study section, including the performance tables, represents the standalone performance of the device against the ground truth.


7. Type of Ground Truth Used

The type of ground truth used was primarily viral culture, which is a laboratory-based method for identifying and growing viruses. For discrepant results, real-time RT-PCR (proFLU plus by Prodesse) was used as a confirmatory or adjudicating ground truth.


8. Sample Size for the Training Set

The document does not explicitly mention a "training set" or its sample size. This type of device (rapid diagnostic test) is typically developed and optimized during its R&D phase using various sample sets, but the clinical study described is primarily for validation/testing of the finalized device.


9. How the Ground Truth for the Training Set Was Established

As no explicit training set is described, the method for establishing ground truth for a training set (if one existed in the R&D phase) is not provided. The ground truth for the clinical test set was established by viral culture (and PCR for discrepants), as described in point 7.

{0}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR Part 807.92.

The assigned 510(k) number is: K083746

    1. Date of Summary: August 17, 2010
    1. Submitted by: Princeton BioMeditech Corporation 4242 U.S. Hwy 1, Monmouth Jct., NJ 08852 Phone: 732-274-1000 Fax: 732-274-1010

Contact Person: Jemo Kang, Ph.D., j.kang@pbmc.com Alternate Contact Person: Kyung-ah Kim, Ph.D., kakim@pbmc.com

    1. Device Name: Trade Names: BioSign® Flu A+B Status™ Flu A & B Common or Usual Name: Influenza test Classification Name: Influenza virus serological reagents (21CFR 866.3330)
    1. Identification of legally marketed devices to which claims of equivalence are made: K053146: QuickVue Influenza A+B Test by Quidel Corp.
    1. Device Description:

BioSign® Flu A+B is an immuno-chromatographic test for the rapid, qualitative detection of influenza A and B. The test device has two test lines, thereby allowing the separate identification of type A and/or type B viral antigens from the same specimen.

In the test procedure, a specimen is collected and placed into the Extraction Well of the test device containing extraction solution for one minute, during which time antigen is extracted from disrupted virus particles. The test device is then raised, tapped and laid back down onto a level surface to allow the solution in the Extraction Well to migrate through the pads containing lyophilized detector antibodies conjugated to gold dye and then through the test membrane. If influenza antigens are present in the specimen, they will react with anti-influenza antibody coupled to gold dye particles, migrate through the membrane as antigen-antibody-dye complexes, bind to the immobilized anti-influenza antibody on the membrane, and generate a colored line in the Test line position (A and/or B). The rest of the sample and unbound/bound dye complexes continue to migrate to the Control line position, where antibody to the antiinfluenza antibody is immobilized, and anti-influenza antibody-unbound/bound dye complexes form the Control line (internal procedural control).

    1. Intended Use:
      The BioSign® Flu A+B test is an in vitro rapid qualitative test that detects influenza type A and type B antigens directly from nasal swab, nasopharyngeal swab, and nasopharyngcal aspirate/wash specimens obtained from patient with signs and symptoms of respiratory infection. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. A negative test result is presumptive and it is recommended these results be confirmed by viral

{1}------------------------------------------------

culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions. The test is intended for professional and laboratory use.

7. Technological Characteristics

Both BioSign® Flu A+B and the predicate, QuickVue® Influenza A+B tests are in virro rapid qualitative tests that detect influenza type A and type B antigens directly from nasal swab, nasopharyngeal swab, and nasopharyngeal aspirate/wash specimens. The scientific principle of both tests is a solid phase chromatographic immunoassay. Both tests are lateral flow rapid assays that employ specific antibodies immobilized onto solid phases to capture and visualize influenza nucleoprotein antigens.

8. Performance Summary

Clinical Study

A prospective clinical study was conducted from January 2007 to March 2008 and during March and April 2009 to determine the performance of BioSign® Flu A+B for aspirate, nasopharyngeal swab, and nasal swab specimens. The samples were collected at 5 sites in the USA from patients who visited physicians' offices and clinics with signs and symptoms of respiratory infection during the study period. All collected samples were tested with BioSign® Flu A+B, and were cultured to confirm the results of BioSign® Flu A+B. The total number of patients tested was 862, of which 30% were 5 and younger, 38% were 6-21 years old, and the rest were older than 21. Forty eight (48) percent were male and 52% were female.

The combined data from all sites of the prospective study are presented in the tables below.

The samples that produced discrepant results between BioSign® Flu A+B and viral culture were further analyzed with proFLU plus by Prodesse (real time RT-PCR, PCR hereafter). These results are presented in the footnote below each table.

Reference (Virus Culture)ResultsReference (Virus Culture)Results
BioSignFlu A+ BFlu APositiveFlu ANegativeTotalPerformanceBioSignFlu A+ BFlu BPositiveFlu BNegativeTotalPerformance
Flu APositive4130*71Sensitivity:95.3%95% CI: 92.1-98.5%Flu BPositive116*17Sensitivity:91.6%95%CI: 83.6-99.6%
Flu ANegative2**180182Specificity:85.7%95% CI: 83.3-88.1%Flu BNegative1**235236Specificity:97.5%95% CI:96.5-98.5%
Total43210253Total12241253

Nasopharyngeal Aspirate Sample ... .

  • Of 30 discrepant results, 22 were positive by both BioSign and PCR ; . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ** Of 2 discrepant results, 1 was negative by both BioSign and PCR

** The discrepant sample was positive by PCR

{2}------------------------------------------------

Nasopharyngeal Sample

Reference (Virus Culture) Results
BioSignFlu A+ BFlu A PositiveFlu A NegativeTotalPerformance
Flu APositive2651*77Sensitivity:89.6%95% CI: 84.0-95.2%
Flu ANegative3**171174Specificity:77.0%95% CI: 74.2-79.8%
Total29222251
Reference (Virus Culture) Results
BioSignFlu A+ BFlu BPositiveFlu BNegativeTotalPerformance
Flu BPositive3315*48Sensitivity:86.8%95% CI: 81.4-92.2%
Flu BNegative5**198203Specificity:92.9%95% CI: 91.2-94.6%
Total38213251

*Of 51 discrepant results, 41 were positive by both BioSign and PCR ** Of 3 discrepant results, 1 was negative by both BioSign and PCR

*Of the 15 discrepant results, 8 were positive by both BioSign and PCR

** Of the 5 discrepant results, 2 were negative by both BioSign and PCR

Reference (Virus Culture)Reference (Virus Culture)Results
ResultsBioSignFlu A+ BFlu BPositiveFlu BNegativeTotalPerformance
BioSignFlu A+ BFlu APositiveFlu ANegativeTotalPerformanceFlu BPositive1440*54Sensitivity:82.4%95%CI: 59.0–93.8%
Flu BNegative3**301304Specificity:88.3%95% CI: 84.4–91.3%
Flu APositive3380*113Sensitivity:91.7%95% CI: 78.2-97.1%Total17341358
Flu ANegative3**242245Specificity:75.2%95% CI: 70.2-79.6%
Total36322358

Nasal Swab Sample

.

*Of 80 discrepant results, 65 were positive by both BioSign and PCR ** Of 3 discrepant results, all 3 were positive by PCR

*Of 40 discrepant results, 18 were positive by both BioSign and PCR

** Of 3 discrepant results, 1 was negative by both BioSign and PCR

{3}------------------------------------------------

As further verification of the PCR test results shown from the samples with discrepant results between BioSign and viral culture, available archived remnant samples from the clinical studies with concordant results were also tested by PCR. The specificity for both Flu A and Flu B was 100%, while the sensitivity for Flu A was 90% and the sensitivity for Flu B was 91.7%.

Archived Sample Test

Aspirate Sample

Eighty (80) frozen archived samples originally obtained from influenza positive patients visiting Columbia NY Presbyterian Hospital and confirmed as positive for either influenza A or Influenza B by viral culture were tested with BioSign Flu A+B.

The tables below present test results with archived samples.

Reference (Virus Culture) Results
BioSignFlu A+ BFlu APositiveFlu ANegativeTotalAgreement
Flu APositive50050100%
Flu ANegative03030100%
Total503080
Reference (Virus Culture) Results
BioSignFlu A+ BFlu BPositiveFlu BNegativeTotalAgreement
• Flu BPositive30030100%
Flu BNegative05050100%
Total305080

Swab Sample ¨¨¨ ¨ ¨ ¨¨:

TReference (Virus Culture) ResultsTotalAgreement
BioSignFlu A+ BFlu APositiveFlu ANegative
Flu APositive50050100%
Flu ANegative03030100%
Total503080
Reference (Virus Culture) Results
BioSignFlu A+ BFlu BPositiveFlu BNegativeTotalAgreement
Flu BPositive30030100%
Flu BNegative05050100%
Total305080

Reproducibility Study

The reproducibility study for the BioSign® Flu A+B test was conducted at two Physicians' Offices and one laboratory using a panel of 180 coded specimens for each site. Testing was performed by two people for five days at each site following the same test protocol as would be used for fresh patient sample. The panel contained high negative, low positive and moderate positive specimens. Each specimen level was tested at each site in replicates of 15 over a period of five days.

{4}------------------------------------------------

The results obtained at each site agreed 100% with the expected results. No differences were observed within run (15 replicates), between runs (three different days), or between sites (three POL sites and one lab).

Analytical Sensitivity

Limit of Detection (LOD)

The LODs were determined for each of the two strains selected from the influenza type A and type B strains. The sensitivity level of each selected viral strain was tested 60 times to confirm the sensitivity level as LOD level, which gives 95% detection rate.

All four viral strains tested were detected 96.7% of the time in 60 replicates at the level listed in the table below.

Influenza TypeViral StrainTCID50/mL#Positive/#Total% Positive
AA/PR/8/34(H1N1)$1.05 \times 10^2$58/6096.7
AA/Victoria/3/75(H3N2)$9.95 \times 10^1$58/6096.7
BB/Taiwan/2/62$1.58 \times 10^3$58/6096.7
BB/Maryland/1/59$1.99 \times 10^1$58/6096.7

. . . . .

Analytical Inclusivity

The analytical inclusivity was established for a total of 21 influenza strains: 11 strains of influenza A type and 10 strains of influenza B type. The results are shown in the table below

InfluenzaTypeViral StrainTCID50/mLInfluenzaTypeViral StrainTCID50/mL
AA/PR/8/34(H1N1)$1.05 × 10^2$BB/Lcc/40$5.00 × 10^0$
AA/FM/1/47(H1N1)$1.73 × 10^1$BB/Allen/45$1.58 × 10^0$
AA/NWS/33(H1N1)$4.10 × 10^3$BB/GL/1739/54$9.95 × 10^2$
AA/Hong Kong/8/68(H3N2)$8.5 × 10^2$BB/Taiwan/2/62$1.58 × 10^3$
AA/Denver/1/57(H1N1)$7.20 × 10^0$BB/Maryland/1/59$1.99 × 10^1$
AA/Aichi/2/68(H3N2)$9.95 × 10^0$BB/Mass/3/66$5.00 × 10^1$
AA/Port Chalmers/1/73$1.99 × 10^2$BB/R22 Barbara$1.6 × 10^{-1}$
AA/Victoria/3/75(H3N2)$9.95 × 10^1$BB/R75$2.94 × 10^3$
AA/New Jersey/8/76(H1N1)$9.95 × 10^1$BB/Russia/69$3.16 × 10^3$
AA/WS/33(H1N1)$5.00 × 10^1$BB/Hong Kong/5/72$2.88 × 10^1$
AA/Swine/1976/31$1.58 × 10^2$
A2009 H1N1 Clinical Isolate*(Swine Origin Influenza A)$1.00 × 10^3$

ー ::! ・ ・ ・ i \ .: ' ' \ '', ' '', ' ! ! ・( ・;

{5}------------------------------------------------

A2009 H1N1 Clinical Isolate*(Swine Origin Influenza A)$1.00 \u00d7 10^3$
AA/CA/07/2009(H1N1)$6.15 \u00d7 10^3$
AA/CA/08/2009(H1N1)$9.31 \u00d7 10^3$
AA/NY/18/2009(H1N1)$2.5 \u00d7 10^3$
AA/Mexico/4108/2009(H1N1)$8.51 \u00d7 10^3$
AA/CA/07/2009 NYC, X-179A(H1N1)$1.08 \u00d7 10^3$
AA/Virginia/ATCC2/2009(H1N1)$2.32 \u00d7 10^3$
AA/Virginia/ATCC3/2009(H1N1)$5.00 \u00d7 10^4$

· Clinical Isolate cultured and tittered. Culture confirmed positive for 2009 H1N1 Influenza A strain using proFLU Influenza A Subtyping

The performance of BioSign® Flu A+B was evaluated with nasal and nasopharyngeal swab samples obtained from patients infected with the 2009 HIN1 influenza virus consisting of sixty six (66) frozen clinical Nasal and Nasopharyngeal samples that had previously tested positive for 2009 HIN by the cleared CDC RT-PCR test. The BioSign® Flu A+B test detected 71% (47/66) of the CDC RT-PCR test positive specimens. The detection rate was 91% with the higher tittered specimens and 38% with the lower tittered specimens.

NOTE:

The performance characteristics of the test with cultured avian influenza A subtype H5N1 virus, or with specimens from human infected with H5N1 or other avian influenza viruses has not been established. ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...

Analytical Specificity

Cross-reactivity

The potential cross-reactivity of the non-influenza respiratory pathogens and other microorganisms with which the majority of the population may be infected was tested on the BioSign® Flu A+B test at medically relevant levels, 108 cfu/mL for bacteria and 10 pfu/mL for non-flu viruses. None of the organisms or viruses listed in the table below gave a positive result with BioSign® Flu A+B at the tested concentration.

Viruses Tested
Adenovirus*Measles**
Human coronavirus**Human metapneumovirus**
Cytomegalovirus**Mumps virus**
Enterovirus**Respiratory syncytial virus; Type B*
Epstein Barr Virus**Rhinovirus; Type 1A**
Human parainfluenza; Type 1, 2 and 3*

** In the study the virus was confirmed using commercially available PCR (not approved by FDA).

  • In the study the virus was confirmed using FDA approved immuno-fluorecence assay.

{6}------------------------------------------------

Bacteria Tested
Bordetella pertussisNeisseria sp.
Chlamydia pneumoniaePseudomonas aeruginosa
Corynebacterium sp.Staphylococcus aureus: Protein A Producer
Escherichia coliStaphylococcus epidermidis
Hemophilus influenzaeStreptococcus pneumoniae
Lactobacillus sp.Streptococcus pyogenes
Legionella sppStreptococcus salivarius
Moraxella catarrhalis
Mycobacterium tuberculosis avirulent
Mycoplasma pneumoniae
Neisseria meningitides

Interference

The interference study was conducted using medically relevant concentrations of the potentially interfering substances listed below with two strains each of influenza type A and type B to assess the potential interference of the substances on the performance of the BioSign® Flu A+B test.

The test was conducted by spiking each substance into samples containing the lowest detectable virus level of influenza Type A or Type B for the positive interference testing and into samples without influenza virus for the negative interference testing. Each substance had no inhibitory effect on the BioSign® Flu A+B test at the concentration listed in the table below.

Substances TestedConcentration Tested
Mucin1 mg/ml
Whole Blood1 %
Phenylephrine10 mg/mL
Oxymetazoline10 mg/mL
Sodium Chloride with preservative20%
Beclomethasone1 mg/mL
Dexamethasone1 mg/mL
Flunisolide1 mg/mL
Triamcinolone1 mg/mL
Budesonide1 mg/mL
Mometasone1 mg/mL
Fluticasone0.5 mg/mL
Luffa opperculata, sulfur1%
Galphimia glauca1%
Histaminum hydrochloricum1%
Live intranasal influenza virus vaccine1%
Benzocaine1 mg/mL
Menthol1 mg/mL
Zanamivir1 mg/mL
Mupirocin1 mg/mL
Tobramycin1 mg/mL

ﺮ ﺩﺭ ۱۰۰۰

7

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The bird is facing to the right. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Building 66 - Room 5645 Silver Spring. MD 20993-0002

Dr. Kyung-ah Kim Associate Director Princeton BioMeditech Corporation 4242 U.S Route 1 Monmouth Junction, NJ 08852-1905

NOV 1 0 2010

Re: K083746

Trade/Device Name: BioSign Flu A+B Test Regulation Number: 21 CFR 866.3330 Regulation Name: Influenza Virus Serological Reagents Regulatory Class: Class I Product Code: GNX Dated: November 8, 2010 Received: November 9, 2010

Dear Dr. Kyung-ah Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{8}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jauzaotgn

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

{9}------------------------------------------------

Indications for Use

510(k) Number (if known): K083746 NOV 1 0 2010

Device Name: BioSign® Flu A+B, Status Flu A & B

Indications For Use:

The BioSign® Flu A+B test is an in vitro rapid qualitative test that detects influenza type A and type B antigens directly from nasal swab, nasopharyngeal swab, nasopharyngeal aspirate/wash specimens of patients with signs and symptoms of respiratory infection. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. A negative test result is presumptive and it is recommended these results be confirmed by viral culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions. The test is intended for professional and laboratory use.

・・・・・・・・・・・・・

Prescription Use V (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C) .

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) a consideraria por use

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off

Office of In Vitro Diagnostic Device
Evaluation and Safety

Page 1 of 1

510(k) K083746

§ 866.3328 Influenza virus antigen detection test system.

(a)
Identification. An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria:
(i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method:
(A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method:
(A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies.
(3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria:
(i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains.
(ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or
(B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
(4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.