K Number
K083644

Validate with FDA (Live)

Date Cleared
2009-06-19

(192 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Tactile Guidance System - Hip is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. The Tactile Guidance System - Hip is indicated for use in surgical hip procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include: Acetabular reaming during total hip arthoplasty (THA)

Device Description

The TGS - Hip is a stereotaxic instrument that includes an optical detector, computer, dedicated instrumentation, operating software, tools and accessories, and a robotic arm. The TGS – Hip is enabled for use in conjunction with a 30 party drill system (e.g., Stryker drill system) in order to support the surgeon's preparation of the acetabulum during total hip arthroplasty. The TGS – Hip uses patient CT data to assist the physician with presurgical planning and interpretive navigation. The TGS - Hip robotic arm serves as an "intelligent" tool holder or tool guide used by a surgeon for stereotactic guidance during minimally invasive orthopedic surgical procedures. The TGS - Hip robotic arm, an electromechanical arm, is passively constrained by software-defined spatial boundaries implemented through the use of the robotic arm and is designed to support a surgeon's preparation of an anatomical site for an orthopedic implant with standard surgical tools such as burrs and reamers.

AI/ML Overview

The provided text describes the acceptance criteria and the study for the MAKO Surgical Corp.'s Tactile Guidance System - Hip (TGS - Hip).

1. Table of Acceptance Criteria and Reported Device Performance:

The document states that the performance data for the TGS-Hip satisfied all required acceptance criteria. However, the specific quantitative acceptance criteria (e.g., specific accuracy thresholds for reaming, registration accuracy) are not detailed within the provided text. The description of the reported device performance is qualitative.

Acceptance CriteriaReported Device Performance
(Specific quantitative criteria not provided in the document, but implicitly met for):Setup, registration, and overall accuracy and functionality of the system in supporting acetabular reaming during THA were evaluated and found satisfactory.
Validation of system's intended usePostoperative x-rays/CT scans of cadaveric material were obtained and evaluated, and the results supported the system's intended use.

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: The document mentions "sawbone models" and "cadaveric material." It does not specify the exact number of either for the testing.
  • Data Provenance: The testing was conducted in a "laboratory." The provenance of the sawbone models is not specified. The cadaveric material is a prospective acquisition for the lab testing. No country of origin for the cadaveric data is given, but it can be inferred as originating from a laboratory setting. No retrospective human patient data was used.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

  • Number of Experts: The document does not specify the number of experts.
  • Qualifications of Experts: The document does not specify the qualifications of the experts who evaluated the results from the cadaveric testing (e.g., who evaluated the postoperative x-rays/CT scans). It only implies that an evaluation was performed.

4. Adjudication Method for the Test Set:

  • The document does not specify an adjudication method. It notes that postoperative x-rays/CT scans were "obtained and evaluated," but the process for consensus or resolution of discrepancies, if any, is not described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

  • No, an MRMC comparative effectiveness study was not done. The study described focuses on the device's performance in a laboratory setting on sawbones and cadavers, not on comparing human reader performance with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Yes, in spirit, a standalone performance was essentially done for certain aspects. The described testing focused on the system's ability to support the surgeon and operate accurately. While a surgeon uses the robotic arm, the "accuracy and functionality of the system" and the validation of its "intended use" through post-operative imaging evaluate the device's technical performance. The robotic arm is passively constrained by software-defined spatial boundaries, indicating the algorithm's direct role in guiding the tool.

7. The Type of Ground Truth Used:

  • For the sawbone models, the ground truth would likely be design specifications or engineering measurements for the desired reaming and registration.
  • For the cadaveric material, the ground truth was established by postoperative x-rays/CT scans, which provide anatomical verification of the reaming and implant placement (though implant placement is not explicitly mentioned for validation, only "acetabular reaming"). This is a form of imaging-based anatomical truth.

8. The Sample Size for the Training Set:

  • The document does not provide information on the sample size for a training set. The described studies are verification and validation tests, not machine learning model training. The system uses "patient CT data to assist the physician with presurgical planning," but this is patient-specific input for operation, not a training dataset for an AI model in the modern sense.

9. How the Ground Truth for the Training Set was Established:

  • As no training set is described for a machine learning model, this information is not applicable based on the provided text. The device uses patient CT data for "presurgical planning and interpretive navigation," which means the CT itself serves as the anatomical ground truth for planning each individual surgery.

{0}------------------------------------------------

083644

JUN 19 2009

fort lauderdale_florida 33317 fəx 954.927 0446 tel 954,927 2044 2555 davie road

EXHIBIT 9 510(K) SUMMARY

Submitter: Address: Phone number: Fax number: Contact Person: Date Prepared: Device Trade Name: Regulation Number: Regulation Name: Regulatory Class: Product Code:

MAKO Surgical Corp. 2555 Davie Road, Fort Lauderdale, FL, 33317 954-927-2044 x 605 954-927-0446 William F. Tapia December 5, 2008 Tactile Guidance System - Hip (TGS - Hip) 21 CFR 882.4560 Stereotaxic Instrument Class II HAW

Substantial Equivalence Claimed To: The TGS – Hip is substantially equivalent to MAKO Surgical's Tactile Guidance System (K072806), Brainlab's Vectorvision Hip (K010602, K040368, K052213, K072716) and Orthosoft's Navitrack System - Total Hip Replacement (K022364).

Description: The TGS - Hip is a stereotaxic instrument that includes an optical detector, computer, dedicated instrumentation, operating software, tools and accessories, and a robotic arm. The TGS – Hip is enabled for use in conjunction with a 30 party drill system (e.g., Stryker drill system) in order to support the surgeon's preparation of the acetabulum during total hip arthroplasty. The TGS – Hip uses patient CT data to assist the physician with presurgical planning and interpretive navigation. The TGS - Hip robotic arm serves as an "intelligent" tool holder or tool guide used by a surgeon for stereotactic guidance during minimally invasive orthopedic surgical procedures. The TGS - Hip robotic arm, an electromechanical arm, is passively constrained by software-defined spatial boundaries implemented through the use of the robotic arm and is designed to support a surgeon's preparation of an anatomical site for an orthopedic implant with standard surgical tools such as burrs and reamers.

Summary of Technological Characteristics Compared to Predicate Devices:

The technological characteristics of the TGS-Hip compared to the predicate devices is listed below.

Technological CharacteristicsTGS-HipBrainlab Vectorvision (VV)HipOrthosoft NavitrackSystem - Total HipReplacement
Major ComponentsGuidance Module, roboticarm, camera standAvailable in several differentconfigurations (VV-Compact,VV-Sky, VV-2)Computer cart, camerastand
Tools/accessoriesVarious probes, arrays trackedby optical cameraVarious probes, arraystracked by optical cameraVarious probes, arraystracked by optical camera
Images UseCTCT, CT-freeCT

Performance Data:

System level verification testing was performed in the laboratory with TGS-Hip using sawbone models to evaluate setup, registration, and overall accuracy and functionality of the system in supporting acetabular reaming during THA. Further testing was performed with TGS-Hip using cadaveric material where postoperative x-rays/CT scans were obtained and evaluated in order to validate the system's intended use. The results of these tests satisfied all required acceptance criteria and were found to support substantial equivalence of the TGS-Hip to the predicate devices.

Intended Use/Indications for Use: The Tactile Guidance System - Hip is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical

{1}------------------------------------------------

K0Y3644

Image /page/1/Picture/1 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold, sans-serif font, with the words "SURGICAL CORP." in a smaller font size underneath. Above the word "MAKO" is a stylized graphic element that resembles a wave or a fin, adding a visual element to the logo.

255 davie road

2555 davie road

fax 954 927 4

structures during orthopedic procedures. The Tactile Guidance System - Hip is indicated for use in Stravaros dailing of thopours procedures. In the reserv may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include:

Acetabular reaming during total hip arthoplasty (THA) ்

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with outstretched wings.

Public Health Service

JUN 19 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mako Surgical Corporation % Mr. William F. Tapia Director, Regulatory Affairs 2555 Davie Road Davie, Florida 33317

Re: K083644

Trade/Device Name: Tactile Guidance System - Hip Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: OLO Dated: May 28, 2009 Received: June 2, 2009

Dear Mr. Tapia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. William F. Tapia

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address 1 http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Per

Per

Per D.N

Per

Mark N. Melkersony Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

683644

Image /page/4/Picture/1 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold letters, with the words "SURGICAL CORP." in smaller letters underneath. Above the word "MAKO" is a stylized graphic that resembles a shark fin or a mountain peak. The logo is simple and modern, and it conveys a sense of strength and precision.

fort lauderdale, florida 33317 tel 954.927.2044 fax 954.927.0446 2555 davie road

INDICATIONS FOR USE

510(k) Number (if known): K083644

Device Name: Tactile Guidance System - Hip

Indications for Use:

The Tactile Guidance System - Hip is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

The Tactile Guidance System - Hip is indicated for use in surgical hip procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include: ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Prescription Use X

OR

Over-the-Counter Use

(Per 21 CFR 801.109)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nil keogh for mxm
(Division Sign Off)

Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K083644

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).