K Number
K083332

Validate with FDA (Live)

Date Cleared
2009-01-09

(58 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The L218TK Color display is intended to be used for displaying and viewing of digital image diagnosis by trained physicians.

To guarantee the display performance as specified, it must only be used for in Conjunction with NEC approved display cards.

TheL218TK cannot be used for a life-support system.

This unit is designed for exclusive interconnection with IEC60601 -1 certified equipment.

Device Description

Medical Display, L218TK is a 21.3" Color LCD monitor that displays image for medical use. It provides 2 mega pixel (1600*1200) resolution with adjustable gamma gray scale for more precise diagnose use in CT, MRI, HIS, and PACS.

AI/ML Overview

This document is a 510(k) Summary of Safety and Effectiveness for a medical display, the NEC MD21M 21.3" Diagnostic Imaging LCD monitor (Model L218TK). It's a submission to the FDA to demonstrate substantial equivalence to a legally marketed predicate device. As such, it does not contain a detailed study proving the device meets specific acceptance criteria in the way a clinical trial or performance study for an AI algorithm would.

Instead, the "acceptance criteria" here are implicitly related to demonstrating substantial equivalence to the predicate device and meeting general performance expectations for a diagnostic display. The "study" proving it meets these is the comparison to the predicate device and the inherent design specifications of the display unit.

Here's the breakdown based on the provided text, recognizing the limitations of a 510(k) for a display device vs. an AI algorithm:

1. Table of Acceptance Criteria and Reported Device Performance

Given that this is a 510(k) for a medical display and not an AI algorithm, the "acceptance criteria" are not explicitly listed in a quantitative fashion that would typically be found in an AI performance study. Instead, the general acceptance criteria are implicit:

  • Intended Use Equivalence: The device's intended use must be substantially equivalent to the predicate device.
  • Technological Characteristics Equivalence: The device must share similar technological characteristics, or any differences must not raise new questions of safety or effectiveness.
  • Display Performance: The display must be capable of showing digital images for diagnosis by trained physicians, including specific resolution and adjustable gamma grayscale, as stated in the description.
  • Safety Standards: Compliance with relevant safety standards (e.g., IEC60601-1 for interconnection).
Acceptance Criteria (Implicit from 510(k) Purpose)Reported Device Performance
Intended UseSame as predicate device (displaying digital images for diagnosis by trained physicians)
Resolution2 mega pixel (1600*1200)
Grayscale AdjustmentAdjustable gamma grayscale
Image Display CapabilityDisplays images for medical use (CT, MRI, HIS, PACS)
Safety & EffectivenessConcluded to be substantially equivalent to predicate device; minor differences do not raise new questions of safety and effectiveness. Does not contact patient, does not control life-sustaining device.
Interconnection SafetyDesigned for exclusive interconnection with IEC60601-1 certified equipment

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size for Test Set: Not applicable in the context of this 510(k) for a display device. There isn't a "test set" of medical images or patient data in the way there would be for an AI algorithm. The evaluation is on the hardware's specifications and its equivalence to a predicate device.
  • Data Provenance: Not applicable.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • Not applicable. Ground truth for medical image interpretation by experts is not established for a display monitor's 510(k). The regulatory body (FDA) reviews technical specifications and equivalence claims.

4. Adjudication Method for the Test Set

  • Not applicable. There is no "test set" requiring adjudication in this context.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, an MRMC comparative effectiveness study was not done. This type of study is for evaluating the impact of a new diagnostic method or AI on human reader performance, which is not relevant for a display screen's 510(k) submission.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not applicable. This device is a display monitor, not an algorithm.

7. The Type of Ground Truth Used

  • Not applicable. For a display device, the "ground truth" would relate to its physical and performance specifications (e.g., luminance, contrast, resolution, color accuracy), which are verified through engineering testing and comparison to industry standards and the predicate device, not necessarily against clinical ground truth like pathology for an AI.

8. The Sample Size for the Training Set

  • Not applicable. This device does not involve a "training set" as it is a hardware display, not an AI algorithm.

9. How the Ground Truth for the Training Set was Established

  • Not applicable.

{0}------------------------------------------------

1083332

JAN - 9 2009

510(K) Summary of Safety and Effectiveness

As required by 807.92

DEVICE ESTABLISHMENT AND CONTACT PERSON 1.

Mr. Shuichi Kino

Manager

NEC Display Solutions Ltd.

4-13-23 Shibaura, Minato-ku, Tokyo, 108-0023 Japan

Ph: +81-465-85-2376

2. COMPANY REISTRATION NUMBER

3003623028

DATE SUMMARY PREPARED 3.

24 October 2008

DEVICE NAME 4.

Trade Name:MD21M 21.3" Diagnostic Imaging LCD monitor
Model Number:L218TK
Common Name:Color LCD Monitor, Color Diagnostic Display, etc.
Classification Name:System, Image Processing, Radiological (CLASS II CFR 892-2050)

4. PREDICATE DEVICE

MDC2130-2HC 21.3" 2MP Color LCD Monitor by CHILIN Technology Co., Ltd. (K063579).

DEVICE DESCRIPTION 5.

Medical Display, L218TK is a 21.3" Color LCD monitor that displays image for medical use. It provides 2 mega pixel (1600*1200) resolution with adjustable gamma gray scale for more precise diagnose use in CT, MRI, HIS, and PACS.

{1}------------------------------------------------

DEVICE OF INTENDRD USE 6.

The L218TK color display is intended to be used for displaying of digital image diagnosis by trained physicians.

To guarantee the display performance as specified, it must only be used for in conjunction with NEC approved display cards.

The L218TK cannot be used for a life-support system.

This unit is designed for exclusive interconnection with IEC60601-1 certified equipment.

7. CONCLUSION

vi

2MP Medical Color Display, L218TK has the same intended use as the predicate device MDC2130-2HC, and they both share the similar characteristics except some minor differences which do not raise new questions of safety and effectiveness. The device does not contact with the patient nor does it control any life sustaining device. Therefore we concluded that it is substantially equivalent to MDC2130-2HC by Chi Lin Technology Co., Ltd. (K063579).

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services (USA). The seal features an abstract eagle design with three stylized wing segments. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" are arranged in a circular pattern around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

JAN - 9 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NEC Display Solution Ltd. % Mr. Angus Hsu Prodigy Technology Consultant Co., Ltd. Safety Certification Service 1F, No. 181, Sec.2, Wunhua 1st Rd., Linkou Township Taipei 244 TAIWAN CHINA

Re: K083332

Trade/Device Name: Medical Display, L218TK Regulation Number: 21 CFR 892,2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 12, 2008 Received: November 12, 2008

Dear Mr. Hsu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxx(Gastroenterology/Renal/Urology)(240) 276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)(240) 276-0115
21 CFR 892.xxx(Radiology)(240) 276-0120
Other(240) 276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.

Sincerely yours,

Laura B. Morris

anine M. Morris Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Niame: Medical Display, L218TK

Indications For USe: The L218TK Color display is intended to be used for displaying and viewing of digital image diagnosis by trained physicians.

To guarantee the display performance as specified, it must only be used for in Conjunction with NEC approved display cards.

TheL218TK cannot be used for a life-support system.

This unit is designed for exclusive interconnection with IEC60601 -1 certified equipment.

Prescription Use V (Part 21 CFR 801 Subpart D)

vii

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Stephan Remmler

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).