K Number
K083255

Validate with FDA (Live)

Date Cleared
2008-11-21

(17 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 16 Ch AI Breast Coil 1.5 T and the 16 Ch AI Breast Coil 3 T are indicated for use as a diagnostic imaging device accessory to produce transverse, sagittal, coronal and oblique images, spectroscopic images and/or spectra, and that displays the internal structure of the breast.

Device Description

The 16 Ch AI Breast Coil 1.5 T and the 16 Ch AI Breast Coil 3 T are receive only MR coils for imaging the human breast. Both coils consist of 16 coil elements. Two housing halves each contain eight independent coil elements. Seven of each are arranged in a cylindrical geometry while one element is positioned in a horizontal plane at top of the cylinder faced outwards.

AI/ML Overview

The provided document describes the K083253 510(k) submission for the "16 Ch AI Breast Coil 1.5 T" and "16 Ch AI Breast Coil 3 T." This submission is for an accessory to an existing Magnetic Resonance Diagnostic Device, specifically a specialized coil for breast imaging. It claims substantial equivalence to a previously cleared predicate device, the "CP Breast Array Coil."

The document details safety and performance characteristics that were considered during the evaluation for substantial equivalence but does not describe a clinical study or specify acceptance criteria in the context of device performance metrics with numerical targets and achieved results. Instead, it focuses on laboratory testing to demonstrate equivalence based on certain technical parameters.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not provide a table with explicit acceptance criteria (e.g., "SNR must be > X dB") and corresponding reported device performance values. It states that "Signal to Noise Ratio (SNR) and image uniformity tests according to IEC 62464-1 were performed... and the results presented in this submission show that they are equivalent with the predicate devices." This implies that the acceptance criterion was likely equivalence to the predicate device's performance for these metrics, rather than a specific numeric threshold.

Performance ParameterAcceptance Criterion (Implied)Reported Device Performance (Implied)
Signal to Noise Ratio (SNR)Equivalent to predicate device (CP Breast Array Coil)Equivalent to predicate device
Image UniformityEquivalent to predicate device (CP Breast Array Coil)Equivalent to predicate device
Spectral ResolutionEquivalent to predicate device (CP Breast Array Coil)Equivalent to predicate device

2. Sample Size Used for the Test Set and Data Provenance

The document only mentions "Laboratory testing" and does not specify a sample size for a test set in the context of human subjects or clinical data. The tests performed ("Signal to Noise Ratio (SNR) and image uniformity tests according to IEC 62464-1" and "spectroscopic tests on SNR and spectral resolution") are typically conducted on phantoms or test objects rather than human subjects for pre-market clearance of an MR coil accessory. Therefore, "data provenance" (country of origin, retrospective/prospective) is not applicable in the usual sense of clinical trial data.

3. Number of Experts Used to Establish Ground Truth and Qualifications

Not applicable. The evaluation was based on laboratory engineering tests (e.g., SNR, image uniformity, spectral resolution) rather than clinical interpretation requiring expert ground truth.

4. Adjudication Method for the Test Set

Not applicable. There was no clinical test set requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The document describes laboratory performance testing of an MR coil accessory, not a comparative effectiveness study involving human readers and AI assistance for diagnostic interpretation.

6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

Yes, in a sense. The testing described focuses on the intrinsic performance characteristics of the coil itself (e.g., its SNR, uniformity). This can be considered "standalone" as it evaluates the device's physical capabilities independently of human interaction for interpretation. However, it's not "algorithm only" as there's no diagnostic algorithm being evaluated.

7. Type of Ground Truth Used

The "ground truth" for the laboratory tests mentioned would be the physical properties and measurements obtained from test phantoms or standard measurement protocols (e.g., IEC 62464-1). It's not expert consensus, pathology, or outcomes data.

8. Sample Size for the Training Set

Not applicable. This device is an MR coil, not an AI or machine learning algorithm that requires a training set of data.

9. How Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

K083253

NOV 21 2008

Section 8 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.

I. General Information

Date of summary preparation: October 08, 2008

Manufacturer Rapid Biomedical GmbH Technologiepark Wuerzburg-Rimpar Kettelerstrasse 3-11 D-97222 Rimpar, Germany Germany

Registration number: 3005049692

Importer/Distributor

Siemens AG Healthcare Sector Henkestrasse 127 D-91052 Erlangen, Germany Germany

Registration number: 8010024

Contact Person

Mr. Armin Purea Rapid Biomedical GmbH Kettelerstrasse 3-11 D-97222 Rimpar, Germany

Phone: +49 (9365) 8826-48 Fax: +49 (9365) 8826-99 e-mail: armin.purea@rapidbiomed.de

II. Classification and Device Name

Classification Panel:Radiology
Classification Name:Magnetic Resonance Diagnostic Device Accessory
Device Class:Class II [21 CFR § 892.1000]
Product Code:MOS
Product Nomenclature:Coil, Magnetic Resonance, Specialty
Common Name:Special Purpose Coil
Trade Name:16 Ch AI Breast Coil 1.5 T16 Ch Al Breast Coil 3 T

{1}------------------------------------------------

III. Safety and Effectiveness Information Supporting Substantial Equivalence

Intended Use

The 16 Ch AI Breast Coil 1.5 T and the 16 Ch AI Breast Coil 3 T are indicated for use as a diagnostic imaging device accessory to produce transverse, sagittal, coronal and oblique images, spectroscopic images and/or spectra, and that displays the internal structure of the breast.

Device Description

The 16 Ch AI Breast Coil 1.5 T and the 16 Ch AI Breast Coil 3 T are receive only MR coils for imaging the human breast. Both coils consist of 16 coil elements. Two housing halves each contain eight independent coil elements. Seven of each are arranged in a cylindrical geometry while one element is positioned in a horizontal plane at top of the cylinder faced outwards.

Equivalency Information

Rapid Biomedical believes that the 16 Ch AJ Breast Coil 1.5 T and 3 T is substantially equivalent to the cleared CP Breast Array Coil for MAGNETOM Harmony and Symphony systems and the described in the following submission:

Predicate Device NameFDA Clearance NumberFDA Clearance Date
CP BREAST ARRAY COILK97363012/05/1997

Summary of Technological Characteristics of the Principal Device as compared with the Predicate Device

In contrast to the predicate device, the 16 Ch AI Breast Coils 1.5 T and 3 T consist of 16 coil elements as opposed to 4. Furthermore, the 16 Ch AI Breast Coil 3 T is designed for a field strength of 3 T. We believe, however, that both coils are substantially equivalent to the predicate device.

General Safety and Effectiveness Concerns

The following safety and performance parameters:

[Safety]

  • Maximum Static Field
  • Rate of Change of Magnetic Field
  • RF Power Deposition
  • Acoustic Noise Level

[Performance-Imaging]

  • Geometric Distortion
  • Slice Profile, Thickness and Gap
  • High Contrast Spatial Resolution

{2}------------------------------------------------

[Performance-Spectroscopy]

  • Spatial Localization Accuracy
  • Peak Assignment Accuracy
  • Solvent Suppression
  • Decoupling

specified by the FDA Guidance document for MR Diagnostic Devices are unaffected by the modifications described within this notification.

The following parameters were considered for the new 16 Ch AI Breast Coil Kit 1.5 T and 3 T:

[Safety] - Biocompatibility

[Performance-Imaging]

  • Signal to Noise Ratio
  • Image Uniformity

[Performance-Spectroscopy]

  • Spectral Resolution
  • Signal to Noise Ratio

No new materials were used for the new 16 Ch AI Breast Coil Kit 1.5 T and 3 T compared to the predicate device. Therefore no biocompatibility tests were performed. Signal to Noise Ratio (SNR) and image uniformity tests according to IEC 62464-1 were performed for the new 16 Ch AI Breast Coil Kit 1.5 T and 3 T and the results presented in this submission show that they are equivalent with the predicate devices. Furthermore, spectroscopic tests on SNR and spectral resolution were carried out.

Conclusion as to Substantial Equivalence

Laboratory testing was performed to support this claim of substantial equivalence and to show that the technological differences do not raise any new questions pertaining to safety and effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 21 2008

RAPID Biomedical GmbH % Mr. Tamas Borsai Division Manager, Medical Division TÜV Rheinland of North America 12 Commerce Road NEWTOWN CT 06470

Re: K083255

Trade/Device Name: 16 Ch AI Breast Coil 1.5 T and 16 Ch AI Breast Coil 3 T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: October 31, 2008 Received: November 4, 2008

Dear Mr. Borsai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the iddictions for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Found, or ug and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radion control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to 1egally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation mumber at the top of this letter:

21 CFR 876.xxx(Gastroenterology/Renal/Urology240-276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)240-276-0115
21 CFR 894.xxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Development (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 633-241 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html .

Sincerely vours.

hoque Mr. Whang

Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Section 2 Indications for Use Statement

Indications for Use

510(k) Number (if known)

Device Name: 16 Ch AI Breast Coil 1.5 T and 16 Ch AI Breast Coil 3 T

Indications for Use:

The 16 Ch AI Breast Coil 1.5 T and the 16 Ch AI Breast Coil 3 T are indicated for use as a diagnostic imaging device accessory to produce transverse, sagittal, coronal and oblique images, spectroscopic images and/or spectra, and that displays the internal structure of the breast.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(please do not write below this line- continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

N/A