K Number
K083169

Validate with FDA (Live)

Date Cleared
2009-03-30

(154 days)

Product Code
Regulation Number
870.3450
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ePTFE SLF™ Spiral Vascular Grafts are indicated for use as vascular prostheses.

The device is intended for use in the bypass or reconstruction of occluded or diseased peripheral arterial blood vessels above or below the knee. Graft configurations are intended for use as arterial conduits for bypass, or reconstruction of peripheral arterial blood vessels.

Device Description

The device is a vascular graft which has a specially designed section intended to induce spiral laminar flow. The grafts are manufactured from a straight tubular expanded polytetrafluoroethylene (ePTFE) vascular graft, supplied by Vascutek Ltd.

The graft includes a helical overlay of polytetrafluoroethylene (PTFE) beading, injection moulded onto the external surface of the ePTFE tube. The function of the beading is to provide reinforcement for the tube.

AI/ML Overview

This device, the ePTFE SLF™ Spiral Vascular Graft, is a vascular prosthesis. The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a clinical study with a detailed performance table.

Therefore, many of the requested elements regarding acceptance criteria, sample sizes, expert ground truth, adjudication, and comparative effectiveness studies are not applicable in this context.

Here's an analysis based on the provided text:

1. A table of acceptance criteria and the reported device performance

Not Applicable. The 510(k) summary for the ePTFE SLF™ Spiral Vascular Graft does not present specific quantitative acceptance criteria or a direct performance table in the way a clinical trial report would. Instead, it asserts "substantial equivalence" to predicate devices based on technological characteristics and preclinical data.

The performance claim is qualitative:
"Bench testing and animal data demonstrated that the safety and effectiveness of the ePTFE SLF™ Spiral Vascular Grafts is equivalent to the predicate devices."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not specified. The document mentions "bench testing and animal data" but does not provide details on the sample sizes, study design (retrospective/prospective), or data provenance for these tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not Applicable. Ground truth, in the context of clinical studies and expert review for diagnostic devices, is not relevant here as this is a medical device for implantation, and its safety and effectiveness were established through non-clinical testing.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not Applicable. See point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not Applicable. This is an implanted medical device, not a diagnostic or AI-assisted interpretation tool. Therefore, MRMC studies and the concept of human readers improving with AI assistance are irrelevant in this context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not Applicable. See point 5.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the safety and effectiveness determination, the "ground truth" was established through bench testing and animal data. This would typically involve objective measurements of physical properties (e.g., burst strength, flexibility, thrombogenicity) and biological responses in animal models (e.g., patency, tissue integration, absence of adverse reactions), compared against the predicate devices. The document does not detail specific metrics or outcomes data.

8. The sample size for the training set

Not Applicable. There is no "training set" in the context of this 510(k) submission. This is not an AI/ML device that requires a training phase.

9. How the ground truth for the training set was established

Not Applicable. See point 8.

{0}------------------------------------------------

K083169

MAR 3 0 2009

Tayside Flow Technologies Ltd.

For the ePTFE SLF

Spiral Vascular Graft

510(k) Summary

This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92.

Submitter's Name:

Tayside Flow Technologies Limited

Submitter's Address:

Tayside Flow Technologies Ltd: Unit 22, Prospect Business Centre Technology Park, Dundee Scotland, DD2 1TY

+44 (0) 1382 598 532 Telephone +44 (0) 1382 598 528 Fax

Establishment Registration Number:

Still to be obtained

Contact Person:

Edwin Lindsay

+44 (0) 7917134922 Telephone

Date Prepared:

20th October 2008

{1}------------------------------------------------

Tayside Flow Technologies Ltd. Traditional 510(k)

For the ePTFE SLF™ Spiral Vascular Graft

510(k) Summary

Device Classification Information:

RegulationNumberDevice NameDeviceClassProductCodeClassificationPanel
870.3450Prosthesis, Vascular Graft,Of 6mm And GreaterDiameterClass 2DSYCardiovascular
870.3450Prosthesis, Vascular Graft,Of Less Then 6mmDiameterClass 2DYFCardiovascular

Device Trade Name:

ePTFE SLF™ Spiral Vascular Graft

Device Common Name:

Vascular Graft with SLF™

Intended Use:

The ePTFE SLF™ Spiral Vascular Grafts are indicated for use as vascular prostheses.

The device is intended for use in the bypass or reconstruction of occluded or diseased peripheral arterial blood vessels above or below the knee. Graft configurations are intended for use as arterial conduits for bypass, or reconstruction of peripheral arterial blood vessels.

ONLY trained and qualified physicians and/or surgeons, under the controlled conditions of a hospital operating theatre environment are indicated for use of this device for implantation.

Summary of Substantial Equivalence:

The ePTFE SLF™ Spiral Vascular Grafts are substantially equivalent to Veryan Medical Limited SwirlGraft Graft (K051312), Bard IMPRA DistaFlo® Bypass Graft (K983861), DynaFlo™ Bypass Graft (K050049) and Vascutek PTFE Supported ePTFE Vascular Prostheses (K043552)

Device Description:

The device is a vascular graft which has a specially designed section intended to induce spiral laminar flow. The grafts are manufactured from a straight tubular expanded polytetrafluoroethylene (ePTFE) vascular graft, supplied by Vascutek Ltd.

The graft includes a helical overlay of polytetrafluoroethylene (PTFE) beading, injection moulded onto the external surface of the ePTFE tube. The function of the beading is to provide reinforcement for the tube.

{2}------------------------------------------------

Tayside Flow Technologies Ltd. Traditional 510(k)

For the ePTFE SLF™ Spiral Vascular Graft

510(k) Summary

Technological Characteristics:

A comparative review of the ePTFE SLF™ Spiral Vascular Grafts with the predicate device found that the technological characteristics, performance and principle of operation were substantially equivalent.

Performance Data:

Bench testing and animal data demonstrated that the safety and effectiveness of the
ePTFE SLF™ Spiral Vascular Grafts is equivalent to the predicate devices.

Safety and Effectiveness:

The ePTFE SLF™ Spiral Vascular Grafts utilises similar technology currently found in legally marketed predicate devices. Based on testing and comparison with the predicate device, the ePTFE SLF™ Spiral Vascular Grafts indicated no adverse indications or results. It is our determination that the ePTFE SLF™ Spiral Vascular Grafts is safe, effective and performs within its design specifications and is substantially equivalent to the predicate device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features an eagle with its wings spread, symbolizing protection and strength. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

MAR 3 0 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Tayside Flow Technologies LTD c/o Tayside Flow Technologies LTD 1141 E. Hawken Way Chandler, Arizona 85286 Attention: Mr. Edwin Lindsay

Re: K083169

Trade Name: ePTFE SLF™ Spiral Vascular Graft Regulation Number: 21 CFR 870.3450 Regulation Name: Vascular graft prosthesis Regulatory Class: II (two) Product Code: DSY Dated: February 6, 2009 Received: February 11, 2009

Dear Mr. Lindsay:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set.

{4}------------------------------------------------

Page 2 - Mr. Edwin Lindsay

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Duna R. Vachner

Bram D. Zuckerman, M.D. - Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):_K083169

Device Name: ePTFE SLFTM Spiral Vascular Graft

Indications For Use:

The ePTFE SLF™ Spiral Vascular Graft is indicated for ruse as a vascular prosthesis. It is intended for use in the bypass of reconstruction of occluded or diseased peripheral arterial blood vessels above or below the knee. Graft configurations are intended for use as arterial conduits for bypass, or reconstruction of peripheral arterial blood vessels.

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Duver & Vochmer
(Division Sign-Off)

Page 1 of 1

Division of Cardiovascular Devices

510(K) Nell, ber_

§ 870.3450 Vascular graft prosthesis.

(a)
Identification. A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”