K Number
K083086

Validate with FDA (Live)

Date Cleared
2008-12-05

(50 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The ECHELON Specialty Coils are receive-only devices that detect the MR signal used to produce transverse, coronal, sagittal, oblique, and/or curved cross-sectional images that display the internal structure of the body. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

The indications for use for the ECHELON Specialty Coils are as follows:

  • The MR-HRF-150, ECHELON High Resolution Foot Coil is a receive-only multiple ● array device used for MRI imaging of the ankle.
  • The MR-LBC-150, ECHELON Long Bone Coil is a receive-only multiple array device . used for MRI imaging of the extremities.
Device Description

The MR-HRF-150, ECHELON High Resolution Foot Coil is a receive-only multiple array device used for MRI imaging of the ankle.

The MR-LBC-150. ECHELON Long Bone Coil is a receive-only multiple array device used for MRI imaging of the extremities.

Magnetic Resonance Imaging (MRI) is based on the fact that certain atomic nuclei have electromagnetic properties that cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen. which makes it the primary nuclei currently used in magnetic resonance imaging. When placed in a static magnetic field, these nuclei assume a net orientation or alignment with the magnetic field, referred to as a net magnetization vector. The introduction of a short burst of radiofrequency (RF) excitation of a wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a re-orientation of the net magnetization vector. When the RF excitation is removed, the protons relax and return to their original vector. The rate of relaxation is exponential and varies with the character of the proton and its adjacent molecular environment. This re-orientation process is characterized by two exponential relaxation times, called T1 and T2.

A RF emission or echo that can be measured accompanies these relaxation events. The emissions are used to develop a representation of the relaxation events in a three dimensional matrix. Spatial localization is encoded into the echoes by varying the RF excitation, applying appropriate magnetic field gradients in the x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of the NMR characteristics can be reconstructed by using image processing techniques similar to those used in computed tomography.

The MR-HRF-150. ECHELON High Resolution Foot Coil is a receive-only device suitable for imaging the ankle. The coil consists of 8 elements. The signal output of each element is independently processed by the MRI system to enhance performance.

The MR-LBC-150. ECHELON Long Bone Coil is a receive-only device suitable for imaging extremities. The coil consists of 4 channels. The signal output of each element is independently processed by the MRI system to enhance performance.

AI/ML Overview

The provided text is a 510(k) summary for the ECHELON High Resolution Foot Coil and ECHELON Long Bone Coil. This document establishes substantial equivalence to predicate devices rather than undergoing a de novo approval process based on acceptance criteria and performance studies.

Therefore, the submission does not contain the information requested regarding specific acceptance criteria and a study proving the device meets those criteria. The main argument for market clearance is that "the ECHELON High Resolution Foot Coil and ECHELON Long Bone Coil are substantially equivalent to the listed predicate devices."

Here's why the requested information isn't present in this type of submission:

  • Acceptance Criteria and Reported Device Performance: These are typically established and tested for novel devices or when a device claims superior performance to existing ones. Substantial equivalence relies on demonstrating that the new device shares similar technological characteristics and intended use with already-approved predicate devices, implying similar safety and effectiveness without needing to redefine performance metrics.
  • Sample Size for Test Set and Data Provenance: Performance studies require a test set. Since a new performance study to establish new acceptance criteria was not conducted, this information is not provided.
  • Number of Experts and Qualifications: Ground truth establishment by experts is part of performance studies.
  • Adjudication Method: Related to ground truth establishment in performance studies.
  • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: These are typically conducted to quantify the impact of AI on human reader performance. This is not relevant here as the device is an MR coil, not an AI diagnostic tool.
  • Standalone Performance: This would apply to diagnostic algorithms, not an MR coil.
  • Type of Ground Truth Used: Not applicable in this context.
  • Sample Size for Training Set: Training sets are used for machine learning models and AI, which this device is not.
  • How Ground Truth for Training Set was Established: Not applicable.

In summary, the 510(k) process for substantial equivalence, as presented in this document, focuses on comparing the new device's characteristics and intended use to existing, legally marketed devices, rather than establishing and demonstrating adherence to new performance acceptance criteria through clinical studies.

{0}------------------------------------------------

510(k) Summary

Korsor6

Submitter Information

Submitter:Hitachi Medical Systems America, Inc.1959 Summit Commerce ParkTwinsburg, Ohio 44087-2371ph: (330) 425-1313fax: (330) 963-0749DEC 0 5 2008
Contact:Douglas J. Thistlethwaite
Date:September 10, 2008
Device Name
Classification Name:Coil, magnetic resonance, specialty
Classification Number:90MOS
Trade/Proprietary Name:MR-HRF-150, ECHELON High Resolution Foot Coil
Predicate Device(s):ECHELON RAPID Foot Coil
Classification Name:Coil, magnetic resonance, specialty
Classification Number:90MOS
Trade/Proprietary Name:MR-LBC-150, ECHELON Long Bone Coil
Predicate Device(s):ECHELON RAPID Foot Coil

Device Intended Use

The MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The ECHELON Specialty Coils are receive-only devices that detect the MR signal used to produce transverse, coronal, sagittal, oblique, and/or curved cross-sectional images that display the internal structure of the body. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

The indications for use for the ECHELON Specialty Coils are as follows:

  • The MR-HRF-150, ECHELON High Resolution Foot Coil is a receive-only multiple ● array device used for MRI imaging of the ankle.
  • The MR-LBC-150, ECHELON Long Bone Coil is a receive-only multiple array device . used for MRI imaging of the extremities.

{1}------------------------------------------------

Device Description

Function

The MR-HRF-150, ECHELON High Resolution Foot Coil is a receive-only multiple array device used for MRI imaging of the ankle.

The MR-LBC-150. ECHELON Long Bone Coil is a receive-only multiple array device used for MRI imaging of the extremities.

Scientific Concepts

Magnetic Resonance Imaging (MRI) is based on the fact that certain atomic nuclei have electromagnetic properties that cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen. which makes it the primary nuclei currently used in magnetic resonance imaging. When placed in a static magnetic field, these nuclei assume a net orientation or alignment with the magnetic field, referred to as a net magnetization vector. The introduction of a short burst of radiofrequency (RF) excitation of a wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a re-orientation of the net magnetization vector. When the RF excitation is removed, the protons relax and return to their original vector. The rate of relaxation is exponential and varies with the character of the proton and its adjacent molecular environment. This re-orientation process is characterized by two exponential relaxation times, called T1 and T2.

A RF emission or echo that can be measured accompanies these relaxation events. The emissions are used to develop a representation of the relaxation events in a three dimensional matrix. Spatial localization is encoded into the echoes by varying the RF excitation, applying appropriate magnetic field gradients in the x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of the NMR characteristics can be reconstructed by using image processing techniques similar to those used in computed tomography.

Physical and Performance Characteristics

The MR-HRF-150. ECHELON High Resolution Foot Coil is a receive-only device suitable for imaging the ankle. The coil consists of 8 elements. The signal output of each element is independently processed by the MRI system to enhance performance.

The MR-LBC-150. ECHELON Long Bone Coil is a receive-only device suitable for imaging extremities. The coil consists of 4 channels. The signal output of each element is independently processed by the MRI system to enhance performance.

{2}------------------------------------------------

Device Technological Characteristics

The technological characteristics of the ECHELON High Resolution Foot Coil and ECHELON Long Bone Coil are similar to the predicate devices as listed in Section 10 - Substantial Equivalence Discussion.

Conclusions

It is the opinion of Hitachi Medical Systems America, Inc. that the ECHELON High Resolution Foot Coil and ECHELON Long Bone Coil are substantially equivalent to the listed predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a stylized emblem. The emblem consists of a stylized symbol that resembles an abstract representation of a human figure or a bird in flight.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 0 5 2008

Mr. Doug Thistlethwaite Manager, Regulatory Affairs Hitachi Medical Systems America, Inc. 1959 Summit Commerce Park TWINSBURG OH 44087

Re: K083086

Trade/Device Name: Echelon High Resolution Foot Coil and Long Bone Coil Regulation Number: 21 CFR 892.1000

Regulation Name: Magnetic resonance diagnostic device

Regulatory Class: II

Product Code: MOS

Dated: October 14, 2008

Received: October 16, 2008

Dear Mr. Thistlethwaite:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device i can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)240-276-0115
21 CFR 894.xxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part. 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

hoque M. Whang

Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if KOB3086 known): ECHELON High Resolution Foot Coil and Long Bone Coil Device Name:

Indications for Use:

The MR-HRF-150, ECHELON High Resolution Foot Coil is a receive-only multiple array device used for MRI imaging of the ankle.

The MR-LBC-150, ECHELON Long Bone Coil is a receive-only multiple array device used for MRI imaging of the extremities.

Prescription Use x AND/OR

Over-the-Counter Use

(21 CFR 801 Subpart C)

(Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Halus Lewis

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.