Intended Use

The Synthes USS (including USS Side-Opening, USS Dual-Opening, USS Small Stature (which includes small stature and pediatric patients), USS VAS variable axis components, USS Fracture, Click' X Monoaxial, Pangea, Pangea Monoaxial, and ClampFix) are non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation systems (T1-S2/ilium), a posterior hook fixation system (T1-L5), or as an anterolateral fixation system (18-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheucrmann's Disease), tumor, stenosis, and failed previous fusion (pseudoarthrosis).

When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.

When used with the 3.5/6.0-mm parallel connectors, the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, and ClampFix) can be linked to the CerviFix System. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix System. When used with the 5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, and ClampFix),

In addition, Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea Monoaxial, and ClampFix) can be interchanged with all USS 6.0 mm rods and transconnectors.

Device Description

The Synthes ClampFix System is a result of design modifications to the predicate devices. It is substantially equivalent to the predicates in design, function, material, and intended use.

AI/ML Overview

This document describes the Synthes ClampFix System, a medical device. Based on the provided text, here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Substantial Equivalence to Predicate DevicesBench testing results demonstrate that the Synthes ClampFix System is substantially equivalent to the predicate devices in design, function, material, and intended use.

2. Sample Size Used for the Test Set and Data Provenance

The document explicitly states: "Clinical data and conclusions were not needed for this device." This indicates that no clinical test set was used, and therefore, no specific sample size or data provenance details are available. The evaluation relied on non-clinical bench testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Since no clinical test set was used, no experts were involved in establishing ground truth for a test set related to clinical performance. The "ground truth" for substantial equivalence was established through non-clinical bench testing, likely by engineers and product specialists adhering to relevant standards.

4. Adjudication Method for the Test Set

As no clinical test set was conducted, no adjudication method was applied.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was conducted. The document states that clinical data and conclusions were not needed.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This device is a physical spinal fixation system, not an algorithm or AI. Therefore, the concept of a "standalone (algorithm only without human-in-the-loop performance)" study does not apply.

7. Type of Ground Truth Used

The ground truth used was based on bench testing results and comparison of the device's design, function, and materials to legally marketed predicate devices, demonstrating substantial equivalence.

8. Sample Size for the Training Set

Since this is a physical medical device and not an AI/algorithm, the concept of a "training set" in the context of machine learning does not apply. The development of the device would have involved engineering design, prototyping, and testing, but not a "training set" in the computational sense.

9. How the Ground Truth for the Training Set Was Established

As the concept of a "training set" is not applicable, there is no information on how its ground truth was established. The device's design and performance are validated through engineering principles and compliance with medical device regulations.

{0}------------------------------------------------

Spine

510(k) Summary 6

Traditional 510(k) - Synthes ClampFix System

Page 20 of 51

Page 1 of 2

{1}------------------------------------------------

@ SYNTHES " Spine

510(k) Summary
USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, and ClampFix) can be linked to the CerviFix System. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix System. When used with the 5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, and ClampFix).In addition, Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, and ClampFix) can be interchanged with all USS 6.0 mm rods and transconnectors.
Comparison of the device to predicate device(s):The Synthes ClampFix System is a result of design modifications to the predicate devices. It is substantially equivalent to the predicates in design, function, material, and intended use.
Performance Date (Non-Clinical and/or Clinical):Non-Clinical Performance and Conclusions:Bench testing results demonstrate that the Synthes ClampFix System is substantially equivalent to the predicate devices.Clinical Performance and Conclusions:Clinical data and conclusions were not needed for this device.

Traditional 510(k) - Synthcs ClampFix System ·

1

and the comments of the comments of

.

.

page 2 of 2

:

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is positioned to the right of a circular text element. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" and is arranged around the circumference of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Synthes Spine % Mr. Jason Lipman 1302 Wrights Lane East West Chester, Pennsylvania 19380

DEC 1 1 2008

Re: K082914

Trade/Device Name: Synthes ClampFix System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: III Product Code: NKB, MNH, MNI, KWP, KWQ Dated: September 29, 2008 Received: September 30, 2008

Dear Mr. Lipman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 -- Mr. Jason Lipman

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N Mikkelson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

SOIne

5 Indications for Use Statement

510(k) Number: K082914 Device Name: Synthes ClampFix System

Indications for Use:

The Synthes USS (including USS Side-Opening, USS Dual-Opening, USS Small Stature (which includes small stature and pediatric patients), USS VAS variable axis components, USS Fracture, Click' X Monoaxial, Pangea, Pangea Monoaxial, and ClampFix) are non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation systems (T1-S2/ilium), a posterior hook fixation system (T1-L5), or as an anterolateral fixation system (18-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheucrmann's Disease), tumor, stenosis, and failed previous fusion (pseudoarthrosis).

When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.

When used with the 3.5/6.0-mm parallel connectors, the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, and ClampFix) can be linked to the CerviFix System. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix System. When used with the 5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, and ClampFix),

In addition, Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea Monoaxial, and ClampFix) can be interchanged with all USS 6.0 mm rods and transconnectors.

Prescription Use X (21 CFR 801 Subpart D) AND / OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
A N I .

(
Division Sig

Division of General, Restorative,
and Neurological Devices

510(k) Number K022914
Traditional 510(k) - Synthes ClampFix System

Page 19 of 51

N/A