K Number
K082879

Validate with FDA (Live)

Date Cleared
2009-06-16

(260 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes who may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity..

Eye care practitioners may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement as prescribed by the eyecare professional. The contact lens may be disinfected using chemical (not heat) disinfection system.

Device Description

EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear are available as spherical lenses manufactured by cast molded method. The lens is made from a lens material that is approximately 55% water. The hydrogel lens' material is a copolymer composed of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid (MAA), which was cross-linked with Ethylene Glycol Dimethacrylate (EGDMA) via UV photopolymerization. The EZvue (Ocufilcon D) lens with visible tint is tinted blue using Reactive Blue Dye #19 to make the lens more visible for handling. Lenses are supplied sterile in sealed blister packs containing sterile isotonic phosphate buffered saline.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria/TestDevice Performance / Result
Clinical PerformanceVision CorrectionNearly a hundred percent of participants' vision was corrected.
Lens SatisfactionNearly all participants were satisfied with lens wearing and care.
Overall Safety & EfficacyProducts are "good and safe for customers."
Technological Characteristics Studies (Comparison to Predicate)
Diameter range13.80~14.40mm (Implicitly meets predicate, as "equivalent and comparable")
Power range+8.00D~-12.00D (Implicitly meets predicate, as "equivalent and comparable")
Center thicknessVaries with power (0.080~0.150mm) (Implicitly meets predicate, as "equivalent and comparable")
Refractive index1.410 (hydrated) (Implicitly meets predicate, as "equivalent and comparable")
Light transmittance>93% (Implicitly meets predicate, as "equivalent and comparable")
Water content54 to 56 % (Implicitly meets predicate, as "equivalent and comparable")
Oxygen permeability16×10^-15 (cm²/sec)(mlO₂/ml-mmHg)]@ 35℃ (Implicitly meets predicate, as "equivalent and comparable")
BiocompatibilityIrritation Test (Rabbit Eye)Extracts would be considered non-irritated.
Systemic ToxicityExtracts would be considered non-toxic.
CytotoxicityLens is not cytotoxic under the conditions of the study.
MicrobiologySterilization (SAL)Steam sterilization process validated to deliver a minimum SAL of 10⁻⁶, complying with FDA Group IV.
Shelf-life StabilityData supports lens remains sterile through the expiration date.
LeachabilityLeachable MaterialsNo leachable monomers and addictive residues detected at reported levels.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 64 human eyes.
  • Data Provenance: The document does not explicitly state the country of origin for the clinical study participants. It mentions the company is located in Taiwan, but this doesn't confirm where the study was conducted. The study was prospective as it involved testing the device on human eyes for a duration of 6 months.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts or their qualifications for establishing ground truth in the clinical study. It only states that "Nearly a hundred percent of the participants' vision was corrected and nearly all were satisfied with the lens wearing and care of lenses," implying subjective and objective assessments were made, likely by participating clinicians/optometrists.

4. Adjudication Method (e.g. 2+1, 3+1, none) for the Test Set

The document does not describe any specific adjudication method for the test set.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No such study was conducted or mentioned in the provided text. This device is a contact lens, not an AI-assisted diagnostic tool.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

This question is not applicable. The device is a medical device (contact lens) for vision correction, not an algorithm. Therefore, "standalone algorithm performance" is not relevant.

7. The Type of Ground Truth Used

For the clinical study, the ground truth was based on:

  • Clinical Outcomes/Performance: Measured vision correction.
  • Patient Reported Outcomes: Satisfaction with lens wearing and care.
  • Expert Clinical Assessment: Implied assessment by the clinical practitioners involved in monitoring the 64 eyes.

For the non-clinical studies (technological characteristics, biocompatibility, microbiology, leachability), the ground truth was established through standardized laboratory testing and comparison to the predicate device's established characteristics.

8. The Sample Size for the Training Set

This question is not applicable. There is no concept of a "training set" for this type of medical device. The "training" for manufacturing would be quality control and process validation, not data-driven machine learning.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable for the reasons stated above.

{0}------------------------------------------------

510 (K) Summary

JUN 1 6 2009

SUMMARY OF SAFETY AND EFFECTIVENESS FOR EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear

Submitter Information:
Company:I-SEE VISION TECHNOLOGY INC.
1F, No. 45, Lane 2, Section 2, Kuang-Fu Rd., Hisnchu, Taiwan,
30071
Contact Person:Mr. Tom Lin,
Special Assistant of President Office
Phone:886-3-5752822
FAX:886-3-5752823
E-Mail:tomlin@iseevision.com

Identification of Device:

Classification Name:Soft hydrophilic contact lens, per 21 CFR. 886.5925
Trade Name:EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens forDaily Wear
Common or usual Name:Soft (hydrophilic) Contact lens (daily wear)
FDA Classification:Class II

Predicate Device:

Hydron Biomedics 55 (Ocufilcon D) Visibility Tint Molded daily wear soft contact lens cleared via K972303 Ocular Sciences / American Hydron Inc.

Indications for Use

EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes who may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity..

Eye care practitioners may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement as prescribed by the eyecare professional. The contact lens may be disinfected using chemical (not heat) disinfection system.

Description of Device

EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear are available as spherical lenses manufactured by cast molded method. The lens is made from a lens material that is approximately 55% water. The hydrogel lens' material is a

Page 1/3 【B】

{1}------------------------------------------------

copolymer composed of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid (MAA), which was

cross-linked with Ethylene Glycol Dimethacrylate (EGDMA) via UV photopolymerization. The EZvue (Ocufilcon D) lens with visible tint is tinted blue using Reactive Blue Dye #19 to make the lens more visible for handling. Lenses are supplied sterile in sealed blister packs containing sterile isotonic phosphate buffered saline.

Summary of Clinical Study:

The EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear were tested in 64 human eyes separately within 6 months. Nearly a hundred percent of the participants' vision was corrected and nearly all were satisfied with the lens wearing and care of lenses. In general, these products are good and safe for customers.

Nonclinical Studies:

A series of nonclinical performance tests were performed to demonstrate the safety and effectiveness of the EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear, and establish substantial equivalence to predicate lenses - Hydron Biomedics 55 (Ocufilcon D) Visibility Tint Molded daily wear soft contact lens (K972303). All testing was conducted in accordance with the May 1994 FDA guideline titled Premarket Notification 510(K) Guidance Document for Class IV Contact lenses, and in conformance to applicable device regulations. The evidence of substantial equivalent to the predicate lens described as follow:

  • a) Technological characteristics studies
    EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear designs in the following parameter ranges:

Diameter range; 13.80~14.40mm

Power range: +8.00D~-12.00D

Center thickness: varies with power (0.080~0.150mm)

Lenses have the following properties:

Refractive index: 1.410 (hydrated)

Light transmittance: >93%

Water content: 54 to 56 %

Oxygen permeability (edged corrected): 16×10~15(cm²/sec)(mlQz/ml-mmHg)]@ 35℃

There characterizations of EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear s are equivalent and comparable to those of predicate lenses.

  • b) Biocompatibility
    In accordance with the May 1994 Guidance Document for daily wear contact lenses, toxicity studies have been conducted on the model: EZvue (Ocufilcon D) Soft

{2}------------------------------------------------

(hydrophilic) Contact Lens for Daily Wear. The Irritation test in the rabbit eye and Systemic toxicity studies indicate the extracts would be considered as non-toxic and nor irritated. The Cytotoxicity testing demonstrates the lens is not cytotoxic under the conditions of the study.

c) Microbiology

Steam sterilization process has been validated to deliver a minimum SAL of 10°, thereby complying with the requirement of FDA Group IV. There is shelf-life stability data supporting that the lens remains sterile through the expiration date claimed for the product.

d) Leachability

Studies were conducted to determine the leachable materials from the finished lens. The results show that, at the levels of the detection reported, there are no leachable monomers and addictive residues.

Substantial equivalence Statement:

Testing performed on the EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear indicated that it can support the efficiency and security use as well as the predicate devices- Hydron Biomedics 55 (Ocufilcon D) Visibility Tint Molded daily wear soft contact lens (K972303), when used in accordance with the instructions for use. It is due to the facts that The risks and benefits of the subject device are the same as soft contact lenses for to the daily wear.

In conclusion, it is I-SEE's conviction that data submitted in this 510(k) to validate the claim of substantial equivalency, substantiates our ability to manufacture a soft contact lens, the EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear, with the same established safety profile and effectiveness as the predicate device -- Hydron Biomedics 55 (Ocufilcon D) Visibility Tint Molded daily wear soft contact lens (K972303).

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three stripes across its wings.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

I-See Vision Technology Inc. c/o Ms. Jennifer Reich Senior Consultant Harvest Consulting Corp. 2904 N. Boldt Drive Flagstaff, AZ 86001

JUN 1 6 2009

Re: K082879

Trade/Device Name: EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lenses Regulatory Class: Class II Product Code: LPL Dated: June 9, 2009 Received: June 11, 2009

Dear Ms. Reich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Ms. Jennifer Reich

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Kesia Alexander for

Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K082874

Device Name: EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear I-SEE Vision Technnology Inc.

Indications For Use:

EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes who may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity.

Eye care practitioners may prescribe the contact lens for frequent replacement wear, with cleaning, rinsing and disinfection and scheduled replacement as prescribed by the eyecare professional. The contact lens may be disinfected using chemical (not heat) disinfection system.

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Netref

Division of Ophthalmic and Ear. Nose and Throat Devices

Page 1 of 1

510(k) Number K092879

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.