K Number
K082756

Validate with FDA (Live)

Date Cleared
2009-10-21

(397 days)

Product Code
Regulation Number
880.2800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Raven ProTest - Steam is a self-contained Biological Indicator inoculated with Geobacillus stearothermophilus spores and is intended for monitoring the efficacy of steam sterilization cycles (10 minute 132°C gravity displacement and 3 minute 132°C flash gravity displacement cycles). Raven ProTest has a validated reduced incubation time of 24 hours.

Device Description

The biological indicator consists of a self-contained unit that includes bacterial spores of Geobacillus stearothermophilus ATCC #7953 inoculated onto a paper filter carrier and a small glass ampoule containing modified Tryptic Soy Broth with Bromocresol Purple acting as a pH indicator encased in a plastic vial that serves as the culture tube. ProTest - Steam is intended for use in 10 minute 132°C gravity displacement and 3 minute 132°C flash gravity displacement cycles.

AI/ML Overview

This document describes a 510(k) submission (K082756) for the Raven ProTest - Steam Biological Indicator, which is intended for monitoring steam sterilization cycles. The submission seeks to expand the label claims to include 132°C gravity/flash gravity cycles. The key information is structured as follows:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Performance Characteristic)Reported Device Performance
Resistance to sterilization processes (efficacy in high-temperature gravity cycles)Demonstrated per AAMI/ISO 11138-1:2006
Spore populationDemonstrated per AAMI/ISO 11138-1:2006
Recovery of low numbers of injured sporesDemonstrated per AAMI/ISO 11138-1:2006
Reduced incubation period (24 hours)Validated 24-hour reduced incubation time
Overall effectiveness in monitoring routine steam sterilization cyclesDemonstrated for all lots tested

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Testing was performed on "three separate lots of product manufactured from three different primary spore crops." The exact number of individual biological indicators tested within these lots is not specified.
  • Data Provenance: Not explicitly stated, but assumed to be from laboratory testing conducted by Raven Biological Laboratories. The data is retrospective in relation to the submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This device is a biological indicator, and its function is to change color in the presence or absence of viable spores after a sterilization cycle. The "ground truth" is determined by the biological activity of the Geobacillus stearothermophilus spores and the effectiveness of the sterilization process itself, not by human expert opinion interpreting an output.

4. Adjudication Method for the Test Set

Not applicable. The outcome of the biological indicator (color change, turbidity) is a direct biological and chemical reaction, not subject to human adjudication in the context of establishing ground truth for the device's performance. The "ground truth" for each test condition (e.g., sterilization or no sterilization) would be established by physical/chemical parameters of the sterilizer or deliberate under-processing. A "control unit" is used to confirm spore viability.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. This is a biological indicator, not an imaging or diagnostic device requiring human interpretation of complex data. Therefore, an MRMC study is not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

Yes. The described testing focuses solely on the performance of the biological indicator itself, independent of a human interpreter's direct involvement in its primary function. The indicator's performance (color change, spore recovery, resistance) is intrinsic to the device.

7. The Type of Ground Truth Used

The ground truth is based on:

  • Biological Activity: The viability and growth of Geobacillus stearothermophilus spores, indicated by metabolic activity leading to a pH change.
  • Chemical Change: The color change of the bromocresol purple indicator from purple to yellow, signifying acid production from spore metabolism.
  • Physical Parameters of Sterilization: The conditions of the 132°C gravity displacement and 132°C flash gravity displacement steam sterilization cycles, which are designed to kill microorganisms. A successful cycle should result in no growth, while an unsuccessful one should (ideally) show growth.

8. The Sample Size for the Training Set

Not applicable in the typical sense of machine learning training sets. This device is a biological indicator, not an AI/ML algorithm that requires a training set. Its "training" is in its design and manufacturing process to consistently meet biological and chemical specifications.

9. How the Ground Truth for the Training Set Was Established

Not applicable. (See point 8). The "ground truth" for manufacturing control and device specification adherence would be established through a robust Quality Management System and adherence to relevant standards (e.g., AAMI/ISO 11138-1:2006). For example, spore counts and D-values (decimal reduction time) are verified during the manufacturing of biological indicators.

{0}------------------------------------------------

K082756

OCT 21 2009

510(k) Summary

510(k) Notification ProTest - Steam Biological Indicator

Submitted by:

Contact:

Raven Biological Laboratories 8607 Park Drive Omaha, NE 68127

Wendy Royalty-Hann Quality Assurance/Regulatory Affairs Manager

Or -

Robert V. Dwyer, Jr. President

Biological Indicator

Indicator for Steam

Phone: (402) 593-0781 Fax: (402) 593-0921

Prepared on:

Device:

Classification:

Predicate Device:

Class II

ProTest Steam (K041386)

Trade name: ProTest - Steam

Common name: Self-contained Biological

DESCRIPTION

The biological indicator consists of a self-contained unit that includes bacterial spores of Geobacillus stearothermophilus ATCC #7953 inoculated onto a paper filter carrier and a small glass ampoule containing modified Tryptic Soy Broth with Bromocresol Purple acting as a pH indicator encased in a plastic vial that serves as the culture tube. ProTest - Steam is intended for use in 10 minute 132°C gravity displacement and 3 minute 132°C flash gravity displacement cycles.

OPERATIONAL PRINCIPALS

The ProTest - Steam Biological Indicator is placed with a load in the sterilization chamber and subjected to a normal steam sterilization cycle. The unit is then removed

{1}------------------------------------------------

and activated by crushing the glass media ampoule inside. The processed unit and an unprocessed (control) unit are incubated at 55-60°C for 24 hours.

During incubation, the available food supply (Tryptic Soy Broth) and temperature promote growth of any viable spores. As viable spores germinate and consume the provided nutrients waste products are released, increasing the acidity of the media which lowers the pH and causes a color change from purple to yellow.

Evidence of growth by color change and/or turbidity within 24 hours should be interpreted as a failure to meet the conditions necessary for sterilizations, provided signs of growth are present in the control unit.

STATEMENT OF SIMILARITY TO LEGALLY MARKETED PREDICATE DEVICE

The subject device ProTest - Steam is identical in composition and function to the legally marketed predicate device ProTest - Steam. This submission is to expand the label claims for the device to include 132ºC gravity/flash gravity cycles.

  • Both are intended for use in moniforing steam sterilization cycles O
  • ם Utilize the same strain of bacterial spores.
  • Utilize the same carrier material. ם
  • The same size and shape.
  • Activated in the same manner.

DESCRIPTION OF TESTING

Testing was performed in accordance with AAMVISO 11138-1:2006 to validate the labeled claims and performance characteristics of ProTest - Steam.

Three separate lots of product manufactured from three different primary spore crops were tested for resistance, spore population, effectiveness in high temperature gravity cycles, recovery of low numbers of injured spores, and a reduced incubation period of 24 hours. For all lots tested, the above parameters and overall effectiveness in monitoring routine steam sterilization cycles has been demonstrated.

STATEMENT OF SAFETY AND EFFECTIVENESS

Based on the similar claims, design and results from the above mentioned testing, the ProTest - Steam Biological Indicator has been demonstrated to be substantially equivalent to and therefore, as safe and effect as, the legally marketed device ProTest -Steam.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, along with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the eagle. The text is in all caps and is evenly spaced around the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

Ms. Wendy Royalty-Hann Director of Quality and Regulatory Affairs Raven Biological Laboratories, Incorporated 8607 Park Drive Omaha, Nebraska 68127

OCT 21 2009

Re: K082756

Trade/Device Name: Raven ProTest - Steam Regulation Number: 21CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: FRC Dated: June 25, 2009 Received: October 2, 2009

Dear Ms. Royalty-Hann:

. 1998

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Ms. Royalty-Hann

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Arthen D. Anster fer

Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Indications for Use Statement

510(k) Number

Device Name

Raven ProTest - Steam

Indications for Use

Raven ProTest - Steam is a self-contained Biological Indicator inoculated with Geobacillus stearothermophilus spores and is intended for monitoring the efficacy of steam sterilization cycles (10 minute 132°C gravity displacement and 3 minute 132°C flash gravity displacement cycles). Raven ProTest has a validated reduced incubation time of 24 hours.

AND/OR Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Myleskey. 10

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K082256

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).