K Number
K082719

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-03-30

(194 days)

Product Code
Regulation Number
876.5820
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are disposable bloodlines intended to provide extracorporeal access to a patient's blood during hemodialysis. The compatibility of available configurations is the responsibility of the physician or other licensed practitioner.

Device Description

The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS includes arterial and venous dialysis blood tubing (non-implanted blood access device) as described in 21 CFR 876.5820. The devices are packaged together for convenient use during for hemodialysis procedures. Three models of the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS (AV06A-P, AV06B-P, AV06C-P) are being manufactured for use with the Nikkiso DBB-05 Hemodialysis Delivery System (K061519). The components of the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS include pump tubing, air trap chambers, transducer filters, pressure monitoring lines, infusion lines for saline, access ports and tubing clamps which are all used to pump blood, retain and capture blood debris, infuse medications or fluids, sample blood, monitor pressure, and make connections to other devices. Major materials used for the various components are Polyvinyl chloride (PVC), polycarbonate (PC), polypropylene (PP). Materials in direct contact with blood are Polycarbonate (PC), Polyvinyl chloride (PVC), Silicone, Polypropylene (PP) and Polytetrafluoroethylene (PTFE). The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are restricted for sale by or on the orders of a physician. The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are packaged sterile and labeled for single use only and non-pyrogenic.

AI/ML Overview

The provided document describes the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS, a medical device intended for hemodialysis, and its substantial equivalence to a predicate device, the Nipro® Blood Tubing Set with Transducer Protectors and Priming Set (K010264). The acceptance criteria and the study proving conformance are detailed below.

1. Table of Acceptance Criteria and Reported Device Performance

ItemAcceptance Criteria (based on AAMI RD17:2007 and ISO10993)Reported Device Performance (NIKKLINE BLOOD TUBING LINES)
Biological SafetyMeets ANSI/AAMI RD17: 2007 and ISO10993 requirementsMet ANSI/AAMI RD17: 2007 and ISO10993 (no concerns regarding cytotoxicity, sensitization, irritation, acute systemic toxicity, hemocompatibility, genotoxicity)
Sterility (Method)N/A (Predicate method: Ethylene Oxide)Steam sterilization
Sterility (SAL)Assured SAL 10-6Assured SAL 10-6 by validation met ISO11134 and EN554
PyrogenicityNon-pyrogenicNon-pyrogenic
Single Use or ReuseSingle use onlySingle use only
Structural IntegrityMeets 4.4.1 Structural integrity AAMI RD17: 2007Met 4.4.1 Structural integrity AAMI RD17: 2007
Connection to Hemodialyzer/HemofilterMeets 4.4.2 AAMI RD17: 2007Met 4.4.2 AAMI RD17: 2007
Connection to Vascular Access DeviceMeets 4.4.3 AAMI RD17: 2007Met 4.4.3 AAMI RD17: 2007
Connection to Ancillary ComponentsMeets 4.4.4 AAMI RD17: 2007Met 4.4.4 AAMI RD17: 2007
Color CodingMeets 4.4.5 AAMI RD17: 2007Met 4.4.5 AAMI RD17: 2007 (arterial red, venous blue)
Needle Access PortsMeets 4.4.6.1 AAMI RD17: 2007Met 4.4.6.1 AAMI RD17: 2007
Transducer ProtectorMeets 4.4.9.1 Integral transducer protectors AAMI RD17: 2007Met 4.4.9.1 Integral transducer protectors AAMI RD17: 2007
Blood pathway volume(Not explicitly specified as a numerical criterion, but evaluated per AAMI RD17:2007)Performed successfully in accordance with AAMI RD17: 2007
Blood pathway flow dynamics(Not explicitly specified as a numerical criterion, but evaluated per AAMI RD17:2007)Performed successfully in accordance with AAMI RD17: 2007
Pump segment performance(Not explicitly specified as a numerical criterion, but evaluated per AAMI RD17:2007)Performed successfully in accordance with AAMI RD17: 2007
Occlusive clamping of tubing(Not explicitly specified as a numerical criterion, but evaluated per AAMI RD17:2007)Performed successfully in accordance with AAMI RD17: 2007

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a distinct "test set" in the context of a dataset for algorithmic evaluation. Instead, the testing described appears to be for verification and validation of the physical device according to established standards.

  • Sample Size: Not explicitly stated for each specific test, but the document implies that "all tests required in RD17" were performed on both the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS and the Nipro tubing set.
  • Data Provenance: The tests were performed by NIKKISO CO., LTD. for their device and were compared against the predicate device's performance which was also evaluated against the same standard by Nipro. The country of origin for NIKKISO CO., LTD. is Japan. The data is prospective in the sense that the tests were conducted specifically for the submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. The ground truth for this device's performance is established by conformance to recognized international and national standards (ANSI/AAMI RD17: 2007 and ISO10993), rather than expert consensus on a test set.

4. Adjudication Method

Not applicable. The evaluation is based on meeting pre-defined criteria in engineering and biological safety standards, not on subjective expert assessment requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This device is a physical medical accessory for hemodialysis tubing, not an AI or diagnostic imaging device that would typically undergo an MRMC study for human reader performance.

6. Standalone (Algorithm Only) Performance

Not applicable. This is a physical medical device, not an algorithm or AI.

7. Type of Ground Truth Used

The ground truth is based on established industry standards and regulatory requirements as defined in:

  • ANSI/AAMI RD17: 2007 "Cardiovascular implants and artificial organs - Extracorporeal blood circuit for hemodialyzers, hemodiafilters, and hemofitters"
  • ISO10993 (various parts for biocompatibility)
  • ISO11134 and EN554 (for steam sterilization validation)
  • ISO11135-1 (for Ethylene Oxide sterilization, referenced for the predicate)

8. Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI or machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set is relevant for this device.

{0}------------------------------------------------

K082719

NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS PREMARKET NOTIFICATION 510(k)

Section 6- 510(k) Summary

MAR 3 0 2009

  • a. Company name, address NIKKISO CO., LTD. 43-2, Ebisu 3-Chome, Shibuya-ku Tokyo, 150-8677, Japan
  • b. Contact Masashi Yoshida Manager Regulatory Affairs
  • c. Date prepared

August 18, 2008

d. Name of device

Trade Name:NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS
Common Name:Hemodialysis system and accessories
Classification Name:Hemodialysis system and accessories

e. Predicate devices

The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS is substantially equivalent to:

510(k):K010264
Trade name:Nipro® Blood Tubing Set with Transducer Protectors and
Priming Set
Product code:FIB, FJK

f. Description of the device

The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS includes arterial and venous dialysis blood tubing (non-implanted blood access device) as described in 21 CFR 876.5820.

The devices are packaged together for convenient use during for hemodialysis procedures. Three models of the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS (AV06A-P, AV06B-P, AV06C-P) are being manufactured for use with the Nikkiso DBB-05 Hemodialysis Delivery System (K061519).

The components of the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS include pump tubing, air trap chambers, transducer filters, pressure monitoring lines,

NIKKISO CO., LTD.

{1}------------------------------------------------

K082719

NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS PREMARKET NOTIFICATION 510(k)

infusion lines for saline, access ports and tubing clamps which are all used to pump blood, retain and capture blood debris, infuse medications or fluids, sample blood, monitor pressure, and make connections to other devices.

Major materials used for the various components are Polyvinyl chloride (PVC), polycarbonate (PC), polypropylene (PP). Materials in direct contact with blood are Polycarbonate (PC), Polyvinyl chloride (PVC), Silicone, Polypropylene (PP) and Polytetrafluoroethylene (PTFE).

The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are restricted for sale by or on the orders of a physician.

The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are packaged sterile and labeled for single use only and non-pyrogenic. Sterility and non-pyrogenicity were confirmed by sterilization validation and pyrogenicity testing.

Biocompatibility tests were performed in compliance with ISO10993 requirements and include acute toxicity, intracutaneous reactivity, hemolysis testing, and implantation testing. The results of biocompatibility testing did not raise any question regarding biocompatibility, and support the equivalence to the predicate device.

The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are designed for use with the Nikkiso DBB-05 Hemodialysis Delivery System (K061519).

Image /page/1/Figure/8 description: The image shows a diagram of an arterial and venous circuit. The arterial circuit includes an access port, shunt connector (red), air trap chamber, pump segment, and transducer protector line for pressure monitor. The venous circuit includes a shunt connector (blue), air trap chamber, access port, and transducer protector line for pressure monitor. The diagram also shows the connection to the hemodializer (red and blue).

Figure 1. NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS (AV06A-P)

NIKKISO CO., LTD.

{2}------------------------------------------------

K082719

NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS PREMARKET NOTIFICATION 510(k)

g. Indications for Use

Indication for Use

The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are disposable bloodlines intended to provide extracorporeal access to a patient's blood during hemodialysis. The compatibility of available configurations is the responsibility of the physician or other licensed practitioner.

h. Statement of substantial equivalence

The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS is substantially equivalent to the Nipro® Blood Tubing Set for Hemodialysis with Transducer Protectors and Priming Set (K010264) (Hereinafter "the Nipro tubing set").

Table 1 compares NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS to the Nipro tubing set. Both tubing lines have the same intended use, same configurations and same major materials. Both of the tubing lines are sterile and labeled as single use only and non-pyrogenic. Tests results from biocompatibility tests for the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS in compliance to ANSI/AAMI RD17: 2007 "Cardiovascular implants and artificial organs - Extracorporeal blood circuit for hemodialyzers, hemodiafilters, and hemofitters" (Hereinafter "RD17") and ISO10993 did not raise any concerns regarding biocompatibility. Nikkiso performed all tests required in RD17 for both NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS and the Nipro tubing set.

As shown in Table 1, both tubing lines met the requirements in RD17 including sterility, pyrogenicity and mechanical characteristic tests.

Based on the above considerations, Nikkiso concludes that the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS is substantially equivalent to the Nipro tubing set.

{3}------------------------------------------------

K082719

nikkline blood tubing lines with transducer protectors PREMARKET NOTIFICATION 510(k)

Table 1. Comparison table

.

No.ItemProposed DevicePredicate Device
1Indication for useThe NIKKLINE BLOOD TUBING LINES WITHTRANSDUCER PROTECTORS aredisposable bloodlines intended toprovide extracorporeal access to apatient's blood during hemodialysis.The compatibility of availableconfigurations is the responsibility ofthe physician or other licensedpractitioner.The Nipro Blood Tubing Set withTransducer Protector and Priming Setsare disposable blood tubing linesintended to provide extracorporealaccess to the patient's blood duringhemodialysis. The compatibility ofavailable configurations is theresponsibility of the physician incharge
2ConfigurationThe NIKKLINE BLOOD TUBING LINESWITH TRANSDUCER PROTECTORS arecomposed of various components suchas tubing, luer locks, connectors,tubing clamps and air trap chambersthat are configured for connecting apatient's vascular access system to adialyzer. The tubing lines componentsare pre-configured and packaged as anarterial line, a venous line, an infusionlines, a saline line and pressuremonitoring lines for use with theNikkiso DBB-05 HemodialysisDelivery System (K061519).The Nipro blood tubing lines arecomposed of various components suchas tubing, luer locks, connectors,clamps and drip chambers that areconfigured for connecting a patient'svascular access system to a dialyzer.The tubing lines components arepre-configured and packaged as anarterial line, a venous line, an infusionline, a saline line and pressuremonitoring lines for use with variousdialysis systems such as Fresenius,Baxter, Cobe, Althin, etc.
3Biological SafetyMeets ANSI/AAMI RD17: 2007 andISO10993.Tests results from biocompatibilitytests for The NIKKLINE BLOODTUBING LINES WITH TRANSDUCERPROTECTORS did not raise anyconcerns regarding biocompatibilityincluding:• Cytotoxicity• Sensitization• Irritation or intracutaneous reactivity• Acute Systemic toxicity• Hemocompatibility• GenotoxicityMeets ANSI/AAMI RD17 andISO10993.
4Sterility-methodSteam sterilizationEthylene Oxide sterilization
-SAL 10 $^{-6}$Assured SAL 10 $^{-6}$ by validation metISO11134 and EN554.Assured SAL 10 $^{-6}$ by validation metISO11135-1
5PyrogenicityNon- pyrogenicNon-pyrogenic
6Single use or reuseSingle use only.Single use only.
7Structural integrityMeets 4.4.1 Structural integrity AAMIRD17: 2007.Meets 4.4.1 Structural integrity AAMIRD17: 2007.
8Connection tohemodialyzer, hemodiafilteror hemofilterMeets 4.4.2 Connection tohemodialyzer, hemodiafilter orhemofilter AAMI RD17: 2007.Meets 4.4.2 Connection tohemodialyzer, hemodiafilter orhemofilter AAMI RD17: 2007.
No.ItemProposed DevicePredicate Device
9Connection to vascular access deviceNIKKLINE BLOOD TUBING LINES WITHTRANSDUCER PROTECTORSMeets 4.4.3 Connection to vascularaccess device AAMI RD17: 2007.The Nipro Tubing Set (K010264)Meets 4.4.3 Connection to vascularaccess device AAMI RD17: 2007.
10Connection to ancillary componentsMeets 4.4.4 Connection to ancillarycomponents AAMI RD17: 2007.Meets 4.4.4 Connection to ancillarycomponents AAMI RD17: 2007.
11Color codingMeets 4.4.5 Color coding AAMIRD17: 2007.The arterial patient-connection is red,the venous patient-connection is blue.Meets 4.4.5 Color coding AAMIRD17: 2007.The arterial patient-connection is red,the venous patient-connection is blue.
12Needle access portsMeets 4.4.6.1 Needle access portsAMMI RD17: 2007.Meets 4.4.6.1 Needle access portsAAMI RD17: 2007.
16Transducer protectorMeets 4.4.9.1 Integral transducerprotectors AAMI RD17: 2007.Using Borla's transducer protector(K000702).
20Raw materialsPVC, PC, PP, Silicone are used for thecomponents in the product. PBT andPTFE are used specifically for thetransducer protectors.PVC, PE, PC, PP, POM are used forthe components in the product. PBTand PTFE are used specifically forBorla's transducer protectors.

NIKKISO CO., LTD.

{4}------------------------------------------------

K082719

NIKKLINE BLOOD TUBING LINES WITH TRANSIDUCER PROTECTORS PREMARKET NOTIFICATION 510(k)

Table 1. Comparison table (continued)

PB.I. : Polybutylene terephthalate

: Polycarbonate PC

: Polyethylene PE

  • POM : Polyoxymethylene
    PP : Polypropylene

: Polytetrafluoroethylene PTFE

: Polyvinyl chloride PVC

Silicone : Silicone Rubber

Note:

Tests including blood pathway volume, blood pathway flow dynamics, pump segment performance and occlusive clamping of tubing for the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS were performed successfully in accordance with AAMI RD17: 2007. Since the Nipro Tubing Set (K010264) was cleared by FDA and subjected to the same tests by Nipro, Nikkiso concludes that the test results for the Nipro Tubing Set and the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are substantially equivalent.

{5}------------------------------------------------

K082719

NIKKIJINE BLQQID TUBING LINES WITH TRANSDUCER PROTECTORS PREMARKET NOTIFICATION 510(k)

i. Conclusion

Based on the above discussion and enclosed sections regarding substantial equivalence to predicate devices, NIKKISO CO., LTD. concludes that the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS is substantially equivalent to the Nipro® Blood Tubing Set with Transducer Protectors and Priming Set (K010264) and does not raise any new questions regarding safety or effectiveness.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines representing the wings and body. The logo is black and white.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 3 0 2009

Nikkiso Co. Ltd. c/o Dr. Fumiaki Kanai MIC International 4-1-17 Hongo Bunkyo-ku, Tokyo 113-0035 JAPAN

Re: K082719

Trade/Device Name: NIKKLINE Blood Tubing Lines with Transducer Protectors Regulation Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: KOC Dated: March 11, 2009 Received: March 17, 2009

Dear Dr. Kanai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxx(Gastroenterology/Renal/Urology)(240) 276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)(240) 276-0115
21 CFR 892.xxx(Radiology)(240) 276-0120
Other(240) 276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{8}------------------------------------------------

Indications for Use

510(k) Number (if known): K082719

Device Name: NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS

Indication for Use

The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are disposable bloodlines intended to provide extracorporcal access to a patient's blood during hemodialysis. The compatibility of available configurations is the responsibility of the physician or other licensed practitioner.

Image /page/8/Picture/5 description: The image shows the words "Prescription Use" in bold font, followed by the text "(Per 21 CFR 801 Subpart D)" in a smaller font. There is a large "X" mark above a horizontal line, which is to the right of the words "Prescription Use". The text indicates that the product is intended for prescription use and is subject to the regulations outlined in 21 CFR 801 Subpart D.

AND/OR

Over-the Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices,
510(k) NumberK082719

§ 876.5820 Hemodialysis system and accessories.

(a)
Identification. A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient.
(2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860).
(3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860).
(4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
(b)
Classification. (1) Class II (performance standards) for hemodialysis systems and all accessories directly associated with the extracorporeal blood system and the dialysate delivery system.(2) Class I for other accessories of the hemodialysis system remote from the extracorporeal blood system and the dialysate delivery system, such as the unpowered dialysis chair, hemodialysis start/stop tray, dialyzer holder set, and dialysis tie gun and ties. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.