K Number
K082571

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-09-30

(25 days)

Product Code
Regulation Number
884.5980
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Avaulta™ Support System is indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended either as mechanical support or bridging material for the fascial defect.

Device Description

The Avaulta™ Support System includes a sterile, single use, permanent implant that provides long term reinforcement to support structures in the correction of anterior or posterior vaginal wall prolapse. The central soft knit section provides compliant organ support while the strong knit arms provide improved strength for tension free fixation of the implant.

The Avaulta™ Plus Biosynthetic Support System and Avaulta™ Solo support system both utilize a nonabsorbable monofilament, polypropylene mesh to provide long-term reinforcement for support structures. The Avaulta™ Plus Biosynthetic Support System adds a porous, acellular, ultrathin sheet of crosslinked collagen attached to the polypropylene mesh which serves to establish a protective barrier between mucosal tissue and the polypropylene mesh and contains apertures uniformly sized to allow for ingrowth of host tissue and capillary vessels.

AI/ML Overview

The provided text is a 510(k) summary for the Avaulta™ Solo Support System and Avaulta™ Plus Biosynthetic Support System (K082571). It describes a surgical mesh device for vaginal wall prolapse.

However, the document does not contain any information regarding acceptance criteria or the study that proves the device meets specific performance criteria through an AI/algorithm-based evaluation. The entire submission is focused on establishing substantial equivalence to a predicate device based on similar intended use, general design, fundamental scientific technology, and performance data from physical testing in accordance with guidance for surgical mesh.

Therefore, I cannot provide the requested information. The document details the device's description, intended use, and substantial equivalence to a predicate device, but it lacks any mention of an AI or algorithm-based system, its performance metrics, or related studies.

To illustrate what a comprehensive response would look like if the information were present, I'll use placeholders for each section, emphasizing that this information is missing from the provided text.


1. Table of Acceptance Criteria and Reported Device Performance

Performance Metric (Hypothetical)Acceptance Criteria (Hypothetical)Reported Device Performance (Hypothetical)
[Specific AI Performance Metric 1][Target Value or Range][Achieved Value]
[Specific AI Performance Metric 2][Target Value or Range][Achieved Value]
.........

2. Sample Size for Test Set and Data Provenance

  • Test Set Sample Size: Not provided (no AI/algorithm study described).
  • Data Provenance: Not provided (no AI/algorithm study described).

3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)

  • Number of Experts: Not provided (no AI/algorithm study described).
  • Qualifications of Experts: Not provided (no AI/algorithm study described).

4. Adjudication Method for Test Set

  • Adjudication Method: Not provided (no AI/algorithm study described).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? No, an MRMC study is not mentioned as part of this 510(k) submission.
  • Effect Size of Human Readers with vs. without AI assistance: Not applicable, as no AI-assisted study is described.

6. Standalone (Algorithm Only) Performance Study

  • Was a standalone study done? No, a standalone algorithm performance study is not mentioned. The submission focuses on the performance of the physical surgical mesh itself.

7. Type of Ground Truth Used

  • Type of Ground Truth: Not provided (no AI/algorithm study described). For the physical device, "substantial equivalence" to a predicate device is the primary "ground truth" for regulatory approval, supported by physical testing data, not clinical ground truth in the AI sense.

8. Sample Size for Training Set

  • Training Set Sample Size: Not applicable, as no AI/algorithm is described.

9. How Ground Truth for Training Set Was Established

  • Ground Truth Establishment (Training Set): Not applicable, as no AI/algorithm is described.

Conclusion based on provided text:

The provided 510(k) summary for the Avaulta™ Solo Support System and Avaulta™ Plus Biosynthetic Support System does not describe an AI/algorithm-based device nor any studies related to its performance against acceptance criteria typically associated with AI systems. The document focuses on the mechanical and biological properties of a surgical mesh and its substantial equivalence to a predicate device, as per the requirements for surgical mesh products. Therefore, all the requested information pertaining to AI/algorithm performance studies is absent.

{0}------------------------------------------------

Bard Urological Division C. R. Bard, Inc. 13183 Harland Dr. Covington, GA 30014

SEP 3 0 2008

Image /page/0/Picture/2 description: The image shows the text "K082571 1/2 BARD". The text is handwritten and appears to be a label or identifier. The word "BARD" is written in block letters.

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATIC

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

  • Submitter Information: A.
    Submitter's Name:

Address:

C. R. Bard, Inc. Bard Urological Division 13183 Harland Drive Covington, GA 30014

Contact Person:John C. Knorpp
Contact Person's Telephone Number:678-342-4920
Contact Person's Fax:770-788-5513
  • в. DEVICE NAME:
Trade Name(s):Avaulta™ Solo Support SystemAvaulta™ Plus Biosynthetic Support System
Common/Usual Name:Surgical Mesh
Classification Names:79 - Mesh, Surgical, Polymeric (OTP, PAI)
CFR Reference:21 CFR 878.3300
Classification Panel:General and Plastic Surgery
  • C. PREDICATE DEVICE NAME:
    Trade Names: Avaulta™ Solo Support System Avaulta™ Plus Biosynthetic Support System K063712

  • D. Device Description:
    The Avaulta™ Support System includes a sterile, single use, permanent implant that provides long term reinforcement to support structures in the correction of anterior or posterior vaginal wall prolapse. The central soft knit section provides compliant organ support while the strong knit arms provide improved strength for tension free fixation of the implant.

The Avaulta™ Plus Biosynthetic Support System and Avaulta™ Solo support system both utilize a nonabsorbable monofilament, polypropylene mesh to provide long-term reinforcement for support structures. The Avaulta™ Plus Biosynthetic Support System adds a porous, acellular, ultrathin sheet of crosslinked collagen attached to the polypropylene mesh which serves to establish a protective barrier between mucosal tissue and

{1}------------------------------------------------

K08JΓ71
2/2

the polypropylene mesh and contains apertures uniformly sized to allow for ingrowth of host tissue and capillary vessels.

  • E. INTENDED USE:
    The Avaulta™ Support System is indicated for tissue reinforcement and longlasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended either as mechanical support or bridging material for the fascial defect.

  • TECHNOLOGICAL CHARACTERISTICS SUMMARY: F.
    The subject Avaulta™ Support System has the same intended use, general design and fundamental scientific technology as the predicate device.

  • PERFORMANCE DATA SUMMARY: ુ.
    The appropriate testing to determine substantial equivalence was completed. This includes testing in accordance with Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh (March 22, 1999).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its wings and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SEP 2 8 2012

C.R. Bard, Incorporated % John C. Knorpp, RAC Director, Regulatory Affairs Bard Urological Division 13183 Harland Drive COVINGTON GA 30014

Re: K082571 Trade/Device Name: Avaulta Solo™ Support System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OTP, PAI Dated: September 3, 2008 Received: September 5, 2008

Dear Mr. Knorpp:

This letter corrects our substantially equivalent letter of September 30, 2008.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Benjamin R. Weeks

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

R. Bard, Inc., Bard Urological Division Avaulta™ Support System Premarket Notification [510(k)]

1.4 Indications for Use Statement

510(k) Number (if known):K082571
Device Name:Avaulta™ Support System

Indications for Use:

The Avaulta™ Support System is indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended either as mechanical support or bridging material for the fascial defect.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

Milfre dyl. Forman

rivision of Anesthesiology, General Hospital ifection Control, Dental Devices

510(k) Number

(Recommended Format 11/13/2003)

§ 884.5980 Surgical mesh for transvaginal pelvic organ prolapse repair.

(a)
Identification. Surgical mesh for transvaginal pelvic organ prolapse repair is a prescription device intended to reinforce soft tissue in the pelvic floor. This device is a porous implant that is made of synthetic material, non-synthetic material, or a combination of synthetic and non-synthetic materials. This device does not include surgical mesh for other intended uses (§ 878.3300 of this chapter).(b)
Classification. Class III (premarket approval).(c)
Date premarket application approval or notice of completion of a product development protocol is required. A premarket application approval or notice of completion of a product development protocol for a device is required to be filed with the Food and Drug Administration on or before July 5, 2018, for any surgical mesh described in paragraph (a) of this section that was in commercial distribution before May 28, 1976, or that has, on or before July 5, 2018, been found substantially equivalent to a surgical mesh described in paragraph (a) of this section that was in commercial distribution before May 28, 1976. Any other surgical mesh for transvaginal pelvic organ prolapse repair shall have an approved premarket application or declared completed product development protocol in effect before being placed in commercial distribution.