K Number
K082540

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-11-28

(87 days)

Product Code
Regulation Number
882.1320
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This GL Entropy Sensor is intended to be used for adults and pediatric patients older than 2 years with GE Entropy measurement devices to enable recording of physiological signals (such as EEG). To connect this sensor to the measurement device, use the GE Entropy Cable.

Device Description

The GF. Entropy Sensor (M1038681) is a sensor ussembly with three (3) pre-gelled EEG electrodes. The GE Entropy Sensor is applied to the skin of the patient to record electrophysiological (such as EEG and FEMO) signals. It is a low impedance, single patient use, non-sterile disposable clectrode sensor that is designed for application to the frontal/temporal area. The GE Entropy Sensor is designed to provide ease of use and electrode placement accuracy. It is used in conjunction with the M-ENTROPY and E-ENTROPY module. The Entropy sensor collects EEG and facial EMG signals from these arcas, and the differential signal from the temple to the center of the forehead is used to calculate the Entropy variables. To obtain the Jowest possible skin impedance, a preparation pad is used to lightly abrade the skin to remove the insulating outer layer. One preparation pad is included with each sensor. The GE Entropy Sensor is individually packaged inside a moisture tight foil pouch. One preparation pad is included inside the same foil pouch. A selling box for the GE Entropy Sensor contains 25 sensor pouches and an instruction insert. The GE Entropy Sensor is connected directly to the sensor-end of the GL Entropy Cable and the device-end of the cable is connected directly to the M-L'NTROPY or E-ENTROPY module.

AI/ML Overview

The provided document is a 510(k) Premarket Notification summary for the GE Entropy Sensor (REF M1038681). This document does not contain a detailed study proving the device meets specific acceptance criteria in the way a clinical trial or performance study report typically would.

Instead, this 510(k) summary focuses on demonstrating substantial equivalence to a predicate device (GE Entropy Sensor REF M1123614, cleared Dec. 21, 2007 (K062580)). The primary changes described for the new device are:

  1. Labeling modification to incorporate pediatric use in patients older than 2 years. This claim is based on the predicate device's associated GE Entropy measurement device (K061907) being cleared for pediatric use.
  2. Internal wire rerouting and removal of a shrink wrap/flag sticker to change electrode order from 2-1-3 to 1-2-3, solely for improved usability, with no impact on form, fit, function, safety, or effectiveness.

Therefore, the document does not present a study with specific acceptance criteria, sample sizes, expert ground truth, or MRMC studies that would typically be associated with proving a device's performance against new clinical metrics. The "acceptance criteria" are instead compliance with recognized standards and a demonstration that the minor changes made to the sensor do not introduce new questions of safety or effectiveness compared to the predicate.

Here's the information based on the provided text, categorized to match your request, while highlighting the absence of typical performance study details:


Acceptance Criteria and Device Performance (as inferred from the 510(k) document)

Acceptance Criteria (Inferred)Reported Device Performance
Safety and Effectiveness: No new questions of safety and effectiveness compared to the predicate device.The GE Entropy Sensor (REF M1038681) is "substantially equivalent in safety and effectiveness" to the predicate GE Entropy Sensor (REF M1123614). The changes (labeling for pediatric use, internal wire rerouting for usability) do not affect form, fit, function, safety, or effectiveness. The basis for pediatric use is the FDA clearance of the associated measurement device (K061907) for pediatric patients older than 2 years.
Compliance with Recognized Standards: Adherence to relevant national and international medical device standards.The device has been assessed against and found to comply with the following standards (no differences in standards compared to the predicate): - FDA 21 CFR Part 898, § 898.12 (Performance standard for electrode lead wires and cables) - EN 60601-1:2005 (Part 1: General requirements for safety) - IEC 601-2-26:1994 (Particular requirements for electroencephalographs) - ANSI/AAMI EC12-2000 - ANSI/AAMI ES1:1993 - UL 2601-1 - CAN/CSA C22.2NO601.1 - IEC 664-1 (Insulation coordination for equipment within low-voltage systems) - ANSI/AAMI EC53-1995 (ECG cables and leadwires) - FDA /ODE Guidance for Content of Premarket Submission for Software Contained in Medical Devices, May 11, 2005 - ISO 15223:2000 (Medical Devices - Symbols to be used with medical device labeling and information to be supplied) - EN 980+AI+A2 (Graphical symbols for use in the labeling of medical devices) - ISO 10993-1, -5, -10 (Biological evaluation of medical devices) - ISO 14971:2000 (Medical devices -- Application of risk management to medical devices)
Functionality: Ability to record physiological signals (EEG, FEMG) via 3 pre-gelled electrodes.The device is a "sensor assembly with three (3) pre-gelled EEG electrodes" designed to "record electrophysiological (such as EEG and FEMG) signals." The changes made (electrode order) do not affect this core function and "improves usability."
Biocompatibility: Safe for skin contact.Assessed against ISO 10993-1, -5, -10 "Biological evaluation of medical devices."

Since this is a 510(k) for substantial equivalence based on minor non-clinical changes and leveraging prior clearances of associated devices, the following requested information is not applicable or not provided in this type of submission:

  1. Sample size used for the test set and the data provenance: Not applicable; no specific clinical test set for performance evaluation is described. The assessment is based on design changes and standard compliance.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable; no ground truth establishment for a test set is described.
  3. Adjudication method for the test set: Not applicable.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable; this device is an EEG sensor, not an AI diagnostic algorithm, and no MRMC study is mentioned.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable; this is a sensor, not an algorithm.
  6. The type of ground truth used: Not applicable; no new ground truth was established for performance evaluation of the sensor itself, beyond compliance with standards and the predicate device's existing performance.
  7. The sample size for the training set: Not applicable; this is a physical sensor, not a machine learning model requiring a training set.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Page I of 3

K082540

Premarket Notification 510(k) Summary As required by section 807.92 GE Entropy Sensor (REF M1038681)

NOV 2 8 2008

GENERAL COMPANY INFORMATION as required by 807.92(a)(1)

COMPANY NAME/ADDRESS/PHONF/FAX:

GF. Healthcarc 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344

NAME OF CONTACT:

Mr. Joel Kent

DATE:

July 30, 2008

DEVICE NAME as required by 807.92(a)(2)

TRADE NAME:

GE Entropy Sensor (REF M1038681)

COMMON NAME:

--

Entropy sensor

CLASSIFICATION NAME:

The following Class II classifications appear applicable:

FDA Product Code ___ Classification Name 21 CFR Section

GXY Liectrode, cutaneous 882.1320

NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)

The disposable GE Entropy Sensor (REF M1038681) is substantially equivalent in safety and effectiveness to the predicate GE Entropy Sensor (RFF MI 123614) cleared Dec. 21, 2007 (K062580).

Approved 2008-8-6 15:58 EET DST M1149634 002

Antti Ylöstalo PROJ-REPORT, GE Entropy Sensor Premarket Notification Summary (Pediatric Claim)

Reproduced from the clactronic master in MATRIX

{1}------------------------------------------------

DEVICE DESCRIPTION as required by 807.92(a)(4)

The GF. Entropy Sensor (M1038681) is a sensor ussembly with three (3) pre-gelled EEG electrodes. The GE Entropy Sensor is applied to the skin of the patient to record electrophysiological (such as EEG and FEMO) signals. It is a low impedance, single patient use, non-sterile disposable clectrode sensor that is designed for application to the frontal/temporal area. The GE Entropy Sensor is designed to provide ease of use and electrode placement accuracy. It is used in conjunction with the M-ENTROPY and E-ENTROPY module. The Entropy sensor collects EEG and facial EMG signals from these arcas, and the differential signal from the temple to the center of the forehead is used to calculate the Entropy variables. To obtain the Jowest possible skin impedance, a preparation pad is used to lightly abrade the skin to remove the insulating outer layer. One preparation pad is included with each sensor. The GE Entropy Sensor is individually packaged inside a moisture tight foil pouch. One preparation pad is included inside the same foil pouch. A selling box for the GE Entropy Sensor contains 25 sensor pouches and an instruction insert. The GE Entropy Sensor is connected directly to the sensor-end of the GL Entropy Cable and the device-end of the cable is connected directly to the M-L'NTROPY or E-ENTROPY module.

INTENDED USF. as required by 807.92(a)(5)

Indication for use for GL Entropy Sensor (REF M1038681)

This GE Entropy Sensor is intended to be used for adults and pcdiatric patients older than 2 years with GE Entropy measurement devices to enable recording of physiological signals (such as EEG). To connect this sensor to the measurement device, use the GE Entropy Cable.

SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)

The disposable GE Entropy Sensor (REF M1038681) is substantially equivalent in safety and effectiveness to the predicate GE Entropy Sensor (REF Ml 123614) cleared Dec. 21, 2007 (K062580). The GE Entropy Sensor (REF M1038681) is physically cquivalent to the predicate GE Entropy Sensor (REF M1123614).

There are only two changes between the proposed GE Entropy Sensor (M1038681) and the predicate GE Entropy Sensor (REF MI 123614).

The proposed disposable GE Entropy Sensor (REF M1038681) has changed laheling to incorporate pediatric use in patients older than 2 years. Only the labeling, artwork and different wording of the instruction for use insert have changed.

The proposed disposable GE Entropy Sensor (RBF M1038681) wires has been rerouted inside the connector and shrink wrap/flag sticker has been removed to enable a change in the electrode order from 2-1-3 to 1-2-3. This improves the usability of the product. This change does not affect to form, fit or function nor safety and effectiveness of the product.

The basis for the new claim is the fact that the GL Entropy measurement device with which this sensor can be used has now been cleared by FDA for use in pediatric patients older than 2 years · (K061907) cleared March 27, 2008.

SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCI.USIONS as required by 807.92(b)(1)(3)

M1149634

Reproduced from the electronic master in MATRIX

{2}------------------------------------------------

The GE Entropy Sensor (KEF M1038681) have been assessed against the standards below. The devices have been thoroughly tested through validation and verification of specifications. There are no differences in standards compared to the predicate device (REF MI123614).

  • . FDA 21 CFR Part 898, § 898.12 (Performance standard for electrode lead wires and cables)
  • . EN 60601-1:2005 (Part 1: General requirements for safety)
  • . IEC 601-2-26:1994 Particular requirements for electroencephalographs
  • ANSI/AAMI EC12-2000
  • . ANSI/AAMI ES1:1993
  • . UL 2601-1
  • . CAN/CSA C22.2NO601.1
  • IEC 664-1 Insulation coordination for equipment within low-voltage systems
  • . ANSI/AAMI EC53-1995 ECG cables and leadwires
  • FDA /ODE Guidance for Content of Premarket Submission for Software Contained in Medical . Devices, May 11, 2005
  • ISO 15223:2000 Medical Devices - Symbols to be used with medical device labeling and information to be supplied
  • EN 980+AI+A2 Graphical symbols for use in the labeling of medical devices ●
  • . ISO 10993-1, -5, -10Biological evaluation of medical devices
  • ISO 14971:2000 Medical devices -- Application of risk management to medical devices .

CONCLUSION:

The summary above shows that there are no new questions of safety and effectiveness for the disposable GE Entropy Sensor (REF M1038681) and it is substantially equivalent in safety and effectiveness to the predicate GE Entropy Sensor (KEF MI 123614) cleared Dec. 21, 2007 (K062580).

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure in profile. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

GE Healthcare c/o Mr. Joel C. Kent Manager, Quality and Regulatory Affairs 86 Pilgrim Road Needham, MA 02492

Re: K082540

Trade/Device Name: Entropy Sensor (REF M1038681) Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: II Product Code: GXY Dated (Date on orig SE ltr): August 12, 2008 Received (Date on orig SE ltr): September 2, 2008

Dear Mr. Kent:

This letter corrects our substantially equivalent letter of November 28, 2008.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, oood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

APR - 9 2012

{4}------------------------------------------------

Page 2 - Mr. Joel C. Kent

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Kesia Alexander

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): _

Device Name: GE Entropy Sensor (REF M1038681)

Indications for Use:

Indication for use for GE Entropy Sensor (REF M1038681)

This GL Entropy Sensor is intended to be used for adults and pediatric patients older than 2 years with GE Entropy measurement devices to enable recording of physiological signals (such as EEG). To connect this sensor to the measurement device, use the GE Entropy Cable.

Prescription Use _ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page

(Division Sign-Of Division of General, Restorative, and Neurological Devices

Kort

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).