K Number
K082497

Validate with FDA (Live)

Date Cleared
2009-05-07

(251 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Intuitive Surgical EndoWrist One Hot Shears Instrument is intended to be used with the da Vinci and da Vinci S Surgical Systems for endoscopic cutting and coagulation of tissue during surgery,

Device Description

This 510(k) is being submitted for an endoscopic instrument, the EndoWrist One Hot Shears Instrument, to be used with the Intuitive Surgical da Vinci and da Vinci S Surgical Systems. The Intuitive Surgical EndoWrist® instrument family currently consists of resposable endoscopic tools including the predicate EndoWrist® Monopolar Curved Scissors Instrument. The subject Hot Shears instrument has the same intended use as the Monopolar Curved Scissors but is a single-use disposable endoscopic tool.

This Intuitive Surgical instrument will be compatible with the previously cleared da Vinci and da Vinci S Endoscopic Instrument Control Systems Models IS1200 and IS2000.

AI/ML Overview

The provided text describes a 510(k) summary for the Intuitive Surgical EndoWrist One Hot Shears Instrument, which is a single-use endoscopic tool intended for cutting and coagulation of tissue during surgery with the da Vinci and da Vinci S Surgical Systems. The submission aims to demonstrate substantial equivalence to a predicate device, the EndoWrist Monopolar Curved Scissors Instrument.

Here's an analysis based on your requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state specific acceptance criteria (e.g., numerical thresholds for performance). Instead, it relies on the concept of "substantial equivalence" to a predicate device. The performance data section describes the types of studies conducted rather than specific criteria or results.

Acceptance Criteria CategoryReported Device Performance
Functional Equivalence"Design analysis and comparison, as well as bench testing, have been conducted to confirm that basic functional characteristics of the subject device are substantially equivalent to the predicate device cited, and that design output meets the design input requirements."
Safety & Effectiveness"Based upon the technical information, Indications for Use and performance information provided in this pre-market notification, the EndoWrist One Hot Shears Instrument described herein has been shown to be substantially equivalent to the current legally marketed predicate device."
Single-Use IndicationThe device is assigned only one life by the da Vinci systems, ensuring it cannot be used again after expiration, which is "identical to the manner in which a reusable instrument expires" and ensures its single-use characteristic.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a sample size for any test set or the data provenance (e.g., country of origin, retrospective or prospective). It only mentions "bench testing" was conducted.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not mention the use of experts to establish a ground truth for a test set. The evaluation is based on comparison to a predicate device and bench testing.

4. Adjudication Method for the Test Set

No adjudication method is mentioned as there's no described human-based assessment of a test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size

No MRMC study is mentioned. This device is a surgical instrument, not an AI diagnostic tool, so such a study would not be applicable in this context.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This question is not applicable. The device is a surgical instrument and not an algorithm. Performance assessment is focused on its mechanical and electrical function in conjunction with the surgical system.

7. The Type of Ground Truth Used

The "ground truth" used for proving substantial equivalence is the established performance and safety of the predicate device (EndoWrist Monopolar Curved Scissors Instrument). The new device's performance is compared against the known characteristics and functionality of this already cleared device through "design analysis and comparison, as well as bench testing."

8. The Sample Size for the Training Set

This question is not applicable. The device is a physical surgical instrument and does not involve AI algorithms with training sets.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable as there is no training set mentioned for this device.

{0}------------------------------------------------

082497

Section IV

MAY - 7 2009

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

510(k) Number:

DateAugust 28, 2008
SubmitterIntuitive Surgical, Inc.
950 Kifer Road
Sunnyvale, CA 94086
ER Number2955842
ContactJames Farnworth
Director, Regulatory Affairs
Telephone: (408) 523 - 8687
Fax: (408) 523 - 1390
E-mail: james.farnworth@intusurg.com
Subject DeviceTrade Name(s):
Intuitive Surgical® EndoWrist® One Hot Shears Instrument
Classification Name:
System, Surgical, Computer Controlled Instrument (21 CFR 876.1500)
Common Name:
Endoscopic Instrument Control System,
Endoscopic Instruments and Accessories
Device Class:
Intuitive Surgical da Vinci Surgical System and Endoscopic Instruments: Class II, NAY, System, Surgical, Computer Controlled Instrument
Predicate DevicesIntuitive Surgical EndoWrist Monopolar Curved Scissors Instrument, K050005
DeviceThis 510(k) is being submitted for an endoscopic instrument, the

P. 1/3

{1}------------------------------------------------

Description

EndoWrist One Hot Shears Instrument, to be used with the Intuitive Surgical da Vinci and da Vinci S Surgical Systems. The Intuitive Surgical EndoWrist® instrument family currently consists of resposable endoscopic tools including the predicate EndoWrist® Monopolar Curved Scissors Instrument. The subject Hot Shears instrument has the same intended use as the Monopolar Curved Scissors but is a single-use disposable endoscopic tool.

This Intuitive Surgical instrument will be compatible with the previously cleared da Vinci and da Vinci S Endoscopic Instrument Control Systems Models IS1200 and IS2000.

Indications for The Intuitive Surgical EndoWrist One Hot Shears Instrument is Use intended to be used with the da Vinci and da Vinci S Surgical Systems for endoscopic cutting and coagulation of tissue during surgery.

Comparison to Predicate Device

The EndoWrist One Hot Shears Instrument is identical in shape, function, activation, and use as the predicate reusable Monopolar Curved Scissors instrument. Specific design and material changes were made in order to create a functionally equivalent instrument intended only for a single surgical procedure. Additionally, the trade name has been updated from EndoWrist to EndoWrist One to hight the single-use characteristic of the new instrument.

The overall scissor shape and size of the Hot Shears are essentially identical to that of the Monopolar Curved Scissors. The proximal housing and chassis have the same physical dimensions as the predicate instrument, but are provided with a different color scheme and labeling to aid in identification.

The predicate Monopolar Curved Scissors instrument uses a sterile removable and disposable Tip Cover accessory which insulates the clevis portion of the distal instrument tip. This Tip Cover is provided as an accessory sciparate from the instrument. The subject Hot Shears incorporates a Tip Cover with the same function and insulating material directly into its assembly and the user receives the sterile instrument with the tip cover already in place on the distal tip.

The single-use Hot Shears Instrument is compatible for use with all electrosurgical generators already cleared for use with the reusable

{2}------------------------------------------------

Monopolar Curved Scissors instrument.

The Hot Shears instrument interfaces with the da Vinci and da Vinci S surgical systems in an identical manner to the reusable instruments. Both reusable EndoWrist instruments and the single use Hot Shears are assigned a specific number of working lives, and the lives are decremented in exactly the same manner by the da Vinci Systems. The Hot Shears are assigned only one life, and the systems will not recognize the instrument after that one life is expired, ensuring that it cannot be used again with the da Vinci This is identical to the manner in which a reusable systems. instrument expires.

The EndoWrist One Hot Shears is provided to the customer sterile and packaged in a sealed tray and outer carton. The predicate Monopolar Curved Scissors is provided non-sterile, and the user is responsible for cleaning and sterilizing the instrument prior to each use.

Labeling updates for the Hot Shears Instrument consist of new instrument housing markings, new tray and box labeling and a new Instructions For Use.

TechnologicalCharacteristicsThe technological characteristics of the subject devices are identicalto the predicate devices.
PerformanceDataDesign analysis and comparison, as well as bench testing, havebeen conducted to confirm that basic functional characteristics ofthe subject device are substantially equivalent to the predicatedevice cited, and that design output meets the design inputrequirements.
ConclusionBased upon the technical information, Indications for Use andperformance information provided in this pre-market notification,the EndoWrist One Hot Shears Instrument described herein hasbeen shown to be substantially equivalent to the current legallymarketed predicate device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle faces to the right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular arrangement.

Public Health Service

MAY - 7 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Intuitive Surgical, Inc. % Ms. James Farnworth Director, Regulatory Affairs 950 Kifer Road Sunnyvale, California 94086

Re: K082497

Trade/Device Name: Intuitive Surgical® EndoWrist® One Hot Shears Instrument Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: NAY Dated: March 26, 2009 Received: March 27, 2009

Dear Ms. Farnworth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at

{4}------------------------------------------------

Page 2 - Ms. James Farnworth

(240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at (240) 276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Daniel Keane

Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K082497

Intuitive Surgical® EndoWrist® One Hot Shears Instrument Device Name:

Indications For Use:

The Intuitive Surgical EndoWrist One Hot Shears Instrument is intended to be used with the da Vinci and da Vinci S Surgical Systems for endoscopic cutting and coagulation of tissue during surgery,

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neike Oden for mkm

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K082497

EndoWrist® One Hot Shears Instrument

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.