VORICONAZOLE 1UG BBL SENSI-DISC

K082272 · Becton, Dickinson & CO · JTN · Oct 2, 2008 · Microbiology

Device Facts

Record IDK082272
Device NameVORICONAZOLE 1UG BBL SENSI-DISC
ApplicantBecton, Dickinson & CO
Product CodeJTN · Microbiology
Decision DateOct 2, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1620
Device ClassClass 2

Intended Use

Antimicrobial Susceptibility Test Disks are used for semi-quantitative in vitro susceptibility testing by standardized agar diffusion test procedures. Voriconazole 1ug BBL" Sensi-Disc " is intended for use in determining the susceptibility to Voriconazole of a wide range of pathogens, as described in the "Indications for Use" section. Zone sizes used for interpretation of tests, including control organism limits, were determined by the antimicrobic manufacturer and received FDA approval under NDA Number 21-266.

Device Story

Device consists of high-quality paper disks impregnated with 1µg Voriconazole. Used in clinical microbiology laboratories for semi-quantitative in vitro susceptibility testing via agar diffusion (Bauer-Kirby method). Disks are placed on Mueller Hinton Agar plates inoculated with pure cultures of clinical isolates. Following incubation, healthcare professionals measure zones of inhibition surrounding disks. Measured diameters are compared against established zone size ranges from the pharmaceutical package insert to categorize organisms as susceptible, susceptible-dose dependent, or resistant. Results assist clinicians in selecting appropriate antifungal therapy for patients with Candida infections.

Clinical Evidence

No clinical data. Substantial equivalence is based on the established agar diffusion methodology and the performance of the Voriconazole agent as documented in the FDA-approved drug insert (NDA 21-266).

Technological Characteristics

High-quality paper disks impregnated with 1µg Voriconazole. Standardized agar diffusion (Bauer-Kirby) principle. Utilizes Mueller Hinton Agar as the test medium per CLSI standards (M2-A9, M100-S18). Disks are packaged in cartridges of 50. No electronic or software components.

Indications for Use

Indicated for in vitro agar diffusion susceptibility testing to determine pathogen susceptibility to Voriconazole. Active against Candida albicans, C. glabrata, C. krusei, C. parapsilosis, C. tropicalis (in vitro and clinical infections), and C. lusitaniae, C. guilliermondii (in vitro only).

Regulatory Classification

Identification

An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Submission ## 510(k) SUMMARY KO 32272 | SUBMITTED BY: | Becton, Dickinson and Company<br>7 Loveton Circle<br>Sparks, MD 21152<br>Phone 410-316-4938<br>Fax: 410-316-4499 | OCT 02 2008 | |------------------------|------------------------------------------------------------------------------------------------------------------|-------------| | CONTACT NAME: | Janine Matlak, Regulatory Affairs Specialist | | | DATE PREPARED: | August 8, 2008 | | | DEVICE TRADE NAME: | Voriconazole 1µg, BBL™ Sensi-Disc™ Antimicrobial<br>Susceptibility Test Disks | | | DEVICE COMMON NAME: | Antimicrobial Susceptibility Test Disks | | | DEVICE CLASSIFICATION: | 21 CFR§866.1620, Class II (Product Code JTN), Susceptibility<br>Test Disks, Antimicrobial | | | PREDICATE DEVICE: | Other BBL™ Sensi-Disc™<br>(eg, Ciprofloxacin 5 µg, BBL™ Sensi-Disc™) | | ## INTENDED USE: Antimicrobial Susceptibility Test Disks are used for semi-quantitative in vitro susceptibility testing by standardized agar diffusion test procedures. Voriconazole 1ug BBL" Sensi-Disc " is intended for use in determining the susceptibility to Voriconazole of a wide range of pathogens, as described in the "Indications for Use" section. Zone sizes used for interpretation of tests, including control organism limits, were determined by the antimicrobic manufacturer and received FDA approval under NDA Number 21-266. {1}------------------------------------------------ ## 510(k) SUMMARY ## Indications for Use: Use of Voriconazole 1μg, BBL™ Sensi-Disc™ for in vitro agar diffusion susceptibility testing is indicated when there is a need to determine the susceptibility of pathogens to Voriconazole. The concentration of 1 ug has been shown to be active in vitro against most strains of Candida species listed below, as described in the FDA approved drug insert for this agent. ## Active In Vitro and in Clinical Infections Against: Candida albicans Candida glabrata Candida krusei Candida parapsilosis Candida tropicalis ## Active In Vitro Against: Candida lusitaniae Candida guilliermondii {2}------------------------------------------------ ## DEVICE DESCRIPTION: Voriconazole 1ug BBL™ Sensi-Disc™ is prepared by impregnating high quality paper with accurately determined amounts of Voriconazole supplied by the drug manufacturer. Each Voriconazole disk is clearly marked on both sides with the agent and drug content. Voriconazole cartridges each contain 50 impregnated disks that are packed as either a single cartridge in a single box, or in a package containing ten cartridges. Voriconazole disks are used for semi-quantitative in vitro susceptibility evaluations by the agar diffusion test method. Agar diffusion susceptibility methods employing dried filter paper disks impregnated with specific concentrations of antimicrobial agents were developed in the 1940s. In order to eliminate or minimize variability in the testing, Bauer et al. developed a standardized procedure in which Mueller Hinton Agar was selected as the test medium. Various regulatory agencies and standards-writing organizations subsequently published standardized reference procedures based on the Bauer-Kirby method. Among the earliest and most widely accepted of these standardized procedures were those published by the U.S. Food and Drug Administration (FDA) and the World Health Organization (WHO). The procedure was adopted as a consensus standard by the Clinical and Laboratory Standards Institute (CLSI) [Formerly National Committee for Clinical Laboratory Standards (NCCLS)] and is periodically updated. ## DEVICE PRINCIPLE: Disks containing a wide variety of selected agents are applied to the surface of Mueller Hinton Agar plates, supplemented as needed and inoculated with pure cultures of clinical isolates. Following incubation, the plates are examined and the zones of inhibition surrounding the disks are measured and compared with established zone size ranges for individual agents in order to determine the agent(s) most suitable for use in therapy. The categorical interpretation [susceptible (S), susceptible-dose dependent (S-DD), or resistant (R)] for the organism being tested is made by comparing zone diameters to those found in the approved pharmaceutical package insert. {3}------------------------------------------------ ## DEVICE COMPARISON: The BBL™ Sensi-Disc" Antimicrobial Susceptibility Test Disks - Voriconazole 1 ug is similar to the BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks - Ciprofloxacin 5 µg in that: - Both methods are for susceptibility testing using paper disks impregnated with an agent. . - Both methods are intended to test susceptibility to pathogenic isolates. . - Both methods provide the user with minimum inhibitory concentration (MIC) results based on . measurements of zone diameters. - Both methods require the user to determine categorical interpretations using the measured zone . diameters against CLSI Approved Standards. - Both methods use pure cultures of isolates. t The BBL" Sensi-Disc" Antimicrobial Susceptibility Test Disks - Voriconazole 1 µg differs from the BBL " Sensi-Disc " Antimicrobial Susceptibility Test Disks - Ciprofloxacin 5 µg in that: - BBL" Sensi-Disc" Antimicrobial Susceptibility Test Disks Voriconazole 1μg is a susceptibility test . that uses disks impregnated with Voriconazole at a concentration of 1 µg while the BBL™ Sensi-Disc Antimicrobial Susceptibility Test Disks - Ciprofloxacin 5 µg is a susceptibility test that uses disks impregnated with Ciprofloxacin at a concentration of 5 µg. - BBL " Sensi-Disc" Antimicrobial Susceptibility Test Disk Voriconazole 1 ug is a susceptibility test . used to test a different battery of isolates than the BBL™ Antimicrobial Susceptibility Test Disk - Ciprofloxacin 5 ug. ## SUBSTANTIAL EQUIVALENCE TESTING DATA: See the Voriconazole drug package insert, "Susceptibility Testing Methods: Diffusion Techniques" (Appendix 1). {4}------------------------------------------------ ### 1.0 INTRODUCTION Becton, Dickinson and Company, BD Diagnostic Systems, is submitting this premarket notification in accordance with 21CFR5807.81 for a device being introduced into commercial distribution for the first time. BD Diagnostic Systems is adding Voriconazole 1 µg to the currently available BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks product. Ciprofloxacin (5 µg), was added to the BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks product line when FDA clearance was granted per premarket notification #K874425. Disks containing Ciprofloxacin are being used as one example of this product line and are the predicate device for this submission. ### 1.1 Intended Use Antimicrobial Susceptibility Test Disks are used for semi-quantitative in vitro susceptibility testing by standardized agar diffusion test procedures. Voriconazole lug BBL™ Sensi-Disc™ is intended for use in determining the susceptibility to Voriconazole of a wide range of pathogens, as described under Indications for Use (Section 1.2). Zone diameters used for interpretation of tests, including control organism limits, were determined by the antimicrobic manufacturer, and received FDA approval under NDA 21-266. {5}------------------------------------------------ #### 1.2 Indications for Use Use of Voriconazole 1µg, BBL™ Sensi-Disc™ for in vitro agar diffusion susceptibility testing is indicated when there is a need to determine the susceptibility of pathogens to Voriconazole. The concentration of 1 ug has been shown to be active in vitro against most strains of Candida species listed below, as described in the FDA approved drug insert for this agent. ## Active In Vitro and in Clinical Infections Against: Candida albicans Candida glabrata Candida krusei Candida parapsilosis Candida tropicalis ## Active In Vitro Against: Candida lusitaniae Candida guilliermondii {6}------------------------------------------------ ### 2.0 DEVICE INFORMATION #### 2.1 Device Description Voriconazole I ug BBL™ Sensi-Disc" is prepared by impregnating high quality paper with accurately determined amounts of Voriconazole supplied by the drug manufacturer. Each Voriconazole disk is clearly marked on both sides with the agent and drug content. Voriconazole cartridges each contain 50 impregnated disks that are packed as either a single cartridge in a single box, or in a package containing ten cartridges. Voriconazole disks are used for semi-quantitative in vitro susceptibility evaluations by the agar diffusion test method. Agar diffusion susceptibility methods employing dried filter paper disks impregnated with specific concentrations of antimicrobial agents were developed in the 1940s. In order to eliminate or minimize variability in the testing, Bauer et al. developed a standardized procedure in which Mueller Hinton Agar was selected as the test medium. 12 Various regulatory agencies and standards-writing organizations subsequently published standardized reference procedures based on the Bauer-Kirby method. Among the earliest and most widely accepted of these standardized procedures were those published by the U.S. Food and Drug Administration (FDA) and the World Health Organization (WHO).45 The procedure was adopted as a consensus standard by the Clinical Laboratory Standard Institute (CLSI) [Formerly National Committee for Clinical Laboratory Standards (NCCLS)] and is periodically updated. 6.7,8 Becton, Dickinson and Company <sup>1</sup> Bauer, A.W., W.M.M. Kirby, J.C. Sherris, and M. Turck. 1966. Antibiotic susceptbillity testing by a standardized single disk method. Am. J Clin. Pathol. 45:493-496. Allin Padler 18. 100 100. <sup>3</sup> Federal Register. 1972. Rules and regulations. Antibiotic susceptibility discs. Fed. Regist. 37:20526-20529. Erratum, 38:2756, 1973. * Encsson, H.M., and J.C. Sherris. 1971. Antibitic sensitivity testing. Report of an international collaborative study. Acta Pathol. Microbiol. Scand Sec. B. Sunnt 217:1-90. Stand. Del E. Soppit Litter on Biological Standardization. 1977. Technical report series 610. W.H.O., Geneva. <sup>8</sup> Clinical and Laboratory Standards Institute (Fornerly NCCLS). 2006. Approved standard M2-A9. Performance standards for antimicrobial disk susceptibility tests, 90 ed. CLSI, Wayne, Pa. Clinical and Laboratory Standards Institute (Formerly NCCLS). 2008. M100-S18 (M2). Performance Standards for Artimicrobial Susceptibility Testing; Eighteenth Informational Supplement. CLSI, Wayne, Pa. Susceptiblity Testing, Lightenin mornations Ouplement. OLST, Varifungal Disk Diffusion Susceptibility Testing of Yeasts; Approved Guideline. CLSI, Wayne, Pa. {7}------------------------------------------------ ### 2.2 Device Principle Disks containing a wide variety of selected agents are applied to the surface of Mueller Hinton Agar plates, supplemented as needed and inoculated with pure cultures of clinical isolates. Following incubation, the plates are examined and the zones of inhibition surrounding the disks are measured and compared with established zone size ranges for individual agents in order to determine the agent(s) most suitable for use in therapy. The categorical interpretation [susceptible (S), susceptible-dose dependent (S-DD), or resistant (R)] for the organism being tested is made by comparing zone diameters to those found in the approved pharmaceutical package insert. {8}------------------------------------------------ ### 3.0 DEVICE COMPARISON The BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks, Voriconazole 1 ug as described in this submission, is substantially equivalent to a legally marketed device for the susceptibility testing by agar disk diffusion method. The BBLTM Sensi-Disc™ Antimicrobial Susceptibility Test Disks, Voriconazole 1 ug is substantially equivalent to all of the BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks manufactured and marketed by Becton, Dickinson and Company for over 30 years. Specifically, the BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks, Voriconazole 1 ug is substantially equivalent to BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks, Ciprofloxacin 5 µg (K874425). The Voriconazole 1 ug BBL™ Sensi-Disc™ has the same technological characteristics as its predicate, Ciprofloxacin 5 ug, BBLTM Sensi-DiscTM. The BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disk - Voriconazole 1 ug is similar to the BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disk -- Ciprofloxacin 5 ug in that: - Both methods are for susceptibility testing using paper disks impregnated with an agent. . - Both methods are intended to test susceptibility to pathogenic isolates. . - Both methods provide the user with minimum inhibitory concentration (MIC) results based on . measurements of zone diameters. - Both methods require the user to determine categorical interpretations using the measured zone . diameters against CLSI Approved Standards. - . Both methods use pure cultures of isolates. The BBL " Sensi-Disc" Antimicrobial Susceptibility Test Disks - Voriconazole 1 ug differs from the BBL " Sensi-Disc" Antimicrobial Susceptibility Test Disks - Ciprofloxacin 5 ug in that: - BBL " Sensi-Disc" Antimicrobial Susceptibility Test Disks Voriconazole lug is a susceptibility test . that uses disks impregnated with Voriconazole at a concentration of I ug while the BBL " Sensi-Disc Antimicrobial Susceptibility Test Disks - Ciprofloxacin 5 µg is a susceptibility test that uses disks impregnated with Ciprofloxacin at a concentration of 5 ug. - BBL" Sensi-Disc" Antimicrobial Susceptibility Test Disk -- Voriconazole 1 ug is a susceptibility test . used to test a different battery of isolates than the BBL™ Sensi-Disc" Antimicrobial Susceptibility Test Disk - Ciprofloxacin 5 ug. <sup>1</sup> The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence as found in the Federal Food, Drug and Cosmetic Ad, as applied under 21 CFR 807, Subpat E under which a device can be marketed r our re-out read approval or reclassification of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the U.S. Patent Laws or the courts. {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular emblem with an eagle in the center. The eagle is facing to the right and has its wings spread. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are written around the edge of the circle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 # OCT 02 2008 Becton, Dickinson and Company c/o Ms. Janine Matlak Regulatory Affairs Specialist 7 Loveton Circle Sparks, MD 21152 k082272 Re: > Trade/Device Name: BBLTM Sensi-Disc™ Antimicrobial Susceptibility Test Disks, Voriconazole 1ug Regulation Number: 21 CFR§ 866.1620 Regulation Name: Antimicrobial Susceptibility Test Disc. Regulatory Class: Class II Product Code: JTN Dated: August 8, 2008 Received: August 11, 2008 Dear Ms. Matlak: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {10}------------------------------------------------ ## Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Sale, a For Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {11}------------------------------------------------ ## INDICATIONS FOR USE 14082272 510(k) Number (if known): Device Name: BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks, Voriconazolel Lg ### Indications for Use: Use of Voriconazole 1μg, BBL™ Sensi-Disc™ for in vitro agar diffusion susceptibility testing is indicated when there is a need to determine the susceptibility of pathogens to Voriconazole. The concentration of 1 ug has been shown to be active in vitro against most strains of Candida species listed below, as described in the FDA approved drug insert for this agent. ## Active In Vitro and in Clinical Infections Against: Candida albicans Candida glabrata Candida krusei Candida parupsilosis Candida tropicalis Active In Vitro Against: Cundida lusitaniae Candida guilliermondii Prescription Use V (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (Per 21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Freddie K. Poole ion Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 31002) K08 2272 Becton, Dickinson and Company
Innolitics
510(k) Summary
Decision Summary
Classification Order