K Number
K082263

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-10-24

(77 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ceralas E15/980, E30/980 is intended for delivery of laser light to soft tissue in the contact or non-contact mode during surgical procedures, including via endoscopes, introducers, or catheters. The Ceralas E15/980, E30/980 is generally indicated for incision, excision, vaporization, ablation, hemostasis, or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, cardiothoracic surgery, and ophthalmology. This Ceralas E15/980, E30/980 is specifically indicated for hemostasis and coagulation of soft tissue (including cardiac tissue) and laser assisted lipolysis.

Device Description

The Ceralas E15/980, E30/980 has the same technological characteristics as the cleared Ceralas D. The laser has an additionally incorporated RFID sensor/antenna in the fiber port so as to recognize fibers with an RFID chip in their connectors. The fibers used with the Ceralas E have been previously cleared by FDA (K951775, K964497, K060050) for use with its other laser products and have an RFID chip incorporated into the proximal connector end.

AI/ML Overview

The provided document is a 510(k) summary for a medical device (Ceralas E 980nm Diode Laser). It does not contain information about acceptance criteria or a study proving device performance against such criteria.

The document primarily focuses on:

  • Device Identification and Regulatory Information: Name, manufacturer, classification, predicate device.
  • Purpose of the Submission: Explaining that the Ceralas E is a modification of a previously cleared device (Ceralas D).
  • Intended Use and Specific Indications for Use: A comprehensive list of surgical procedures and tissues for which the device is intended (e.g., incision, excision, vaporization, hemostasis in various medical specialties).
  • Technological Characteristics: Stating that it has the same characteristics as the cleared Ceralas D, with an added RFID sensor to recognize fibers.
  • Substantial Equivalence Claim: Asserting that the minor differences do not raise new safety or effectiveness questions, making it substantially equivalent to its predicate.
  • FDA Clearance Letter: Confirming the FDA's review and their determination of substantial equivalence.

Therefore, I cannot fulfill your request to describe acceptance criteria and a study proving the device meets them, as this information is not present in the provided text.

In absence of the requested information, the following response will indicate that the information is not available in the document.


1. A table of acceptance criteria and the reported device performance
Not provided in the document. The document lists the intended uses and indications for the device, but does not specify quantifiable acceptance criteria or performance metrics.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
Not provided in the document. The submission is for substantial equivalence to a predicate device, and it does not detail new clinical or performance studies with specific test sets.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable/Not provided in the document. Ground truth establishment is typically relevant for studies validating diagnostic or AI-driven devices, which is not the case for this laser device submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable/Not provided in the document. Adjudication methods are typically relevant for studies validating diagnostic or AI-driven devices, which is not the case for this laser device submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable/Not provided in the document. This is not an AI-assisted diagnostic device, so an MRMC study is not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable/Not provided in the document. This is not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable/Not provided in the document. Ground truth is typically relevant for studies validating diagnostic or AI-driven devices.

8. The sample size for the training set
Not applicable/Not provided in the document. This is not an AI algorithm, and no specific training set for performance evaluation is mentioned.

9. How the ground truth for the training set was established
Not applicable/Not provided in the document. Ground truth establishment for a training set is typically relevant for AI/ML device development.

{0}------------------------------------------------

510(k) SUMMARY

Biolitec, Inc.'s Ceralas E 980nm Diode Laser (Models E15/980,30)

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

OCT 2 4 2008

Co

Hogan & Hartson 555 13th Street NW Washington DC 20004

202 637-5794 Phone: Facsimile: (202) 637-5910

Contact Person: Jonathan S. Kahan

Date Prepared:July 24, 2008

Name of Device and Name/Address of Sponsor

Ceralas E 980nm Diode Laser (Models E15/980,E30/980)

Biolitec, Inc. 515 Shaker Road East Longmeadow, MA 01028

Common or Usual Name

Diode Laser

Classification Name

Laser, Surgical Diode Laser System

CFR Provision and Product Code

21 C.F.R. 878.4810, GEX

Predicate Devices

Biolitec Ceralas D 980 Diode Laser (Models D15, D25) (K032863, K072779, K081015)

Purpose of the Special 510(k) notice.

The Ceralas E15/980, E30/980 is a modification to Ceralas D 980.

{1}------------------------------------------------

Intended Use

The Ceralas E15/980, E30/980 is intended for delivery of laser light to soft tissue in the contact or non-contact mode during surgical procedures, including via endoscopes, introducers, or catheters. The Ceralas E15/980, E30/980 is generally indicated for incision, excision, vaporization, ablation, hemostasis, or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, cardiothoracic surgery, and ophthalmology. This Ceralas E15/980, E30/980 is specifically indicated for hemostasis and coagulation of soft tissue (including cardiac tissue) and laser assisted lipolysis.

6

Specific Indications for Use

Ear, Nose and Throat and Oral Surgery:

Hemostasis, incision, excision, ablation, coagulation, and vaporization of tissue from the ear, nose, throat and adjacent areas including soft tissue in the oral cavity. Examples include:

  • Removal of benign lesions from the ear, nose and throat .
  • . Excision and vaporization of vocal cord nodules and polyps
  • Incision and excision of carcinoma in situ .
  • Ablation and vaporization of hyperkeratosis ●
  • Excision of carcinoma of the larynx .
  • Laryngeal papillomectomy ●
  • Excision and vaporization of herpes simplex I and II .
  • Neck dissection .
  • Tonsillectomy .
  • Thyroidectomy ●
  • Vocal cord polypectomy ●
  • Hemiglossectomy .
  • Tracheal stenosis ●
  • Oral cavity lesions ●

Dental Applications:

Intraoral and extraoral soft tissue (including marginal and interdental gingiva and epithelial lining of free gingiva). Examples include:

  • Frenectomy �
  • Frenotomy ●
  • . Biopsy
  • Operculectomy .
  • Implant recovery ●
  • . gingivectomy
  • gingivoplasty ●
  • gingival troughing .
  • crown lengthening .

{2}------------------------------------------------

  • hemostasis of donor site
  • Removal of granulation tissue .
  • laser assisted flap surgery .
  • debridement of diseased epithelial lining
  • . incisions and draining of abscesses
  • ◆ tissue retraction for impressions
  • . papillectomy
  • vestibuloplasty
  • . excision of lesions
  • . exposure of unerupted/partially erupted teeth
  • . leukoplakia
  • . removal of hyperplastic tissues
  • . treatment of aphthous ulcers
  • . Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket)
  • . pulpotomy
  • . pulpotomy as an adjunct to root canal therapy
  • light activation of bleaching materials for teeth whitening .

Arthroscopy:

Hemostasis, incision, excision, coagulation, vaporization and ablation of joint tissues during arthroscopic surgery. Examples include:

  • Menisectomy ●
  • . Synovectomy
  • Chondromalacia .

Gastroenterology:

Hemostasis, incision, excision, ablation, coagulation and vaporization of tissue in the upper and lower gastrointestinal tracts and also with endoscopic procedures. Examples include:

  • Hemostasis of upper and lower GI bleeding
  • Excision and vaporization of colorectal carcinoma
  • . Excision of polyps

General Surgery, Dermatology, Plastic Surgery and Podiatry:

Excision, ablation, vaporization and photocoagulation of skin lesions, hemostasis, incision, excision, vaporization, ablation and debulking of soft tissue, abdominal, rectal, skin, fat or muscle tissue and dermabrasion. Examples include:

  • Matrixectomy ●
  • . Excision of neuromas
  • . Excision of periungual and subungual warts
  • Excision of plantar warts .

{3}------------------------------------------------

  • Excision of keloids
  • · Liver resection
  • · Excision of cutaneous lesions
  • · Hemorrhoidectomy
  • · Appendectomy
  • Debridement of decubitus ulcers
  • · Hepatobiliary tumors
  • · Mastectomy
  • · Dermabrasion
  • · Vaporization and hemostasis of capillary hemangioma
  • · Excision, vaporization and hemostasis of abdominal tumors
  • · Excision, vaporization and hemostasis of rectal pathology
  • · Pilonidal cystectomy
  • · Herniorapphy
  • · Adhesiolysis
  • · Parathyroidectomy
  • · Laparoscopic cholecystectomy
  • · Thyroidectomy
  • · Resection of organs
  • · Debridement of wounds
  • · Phototcoagulation of teleangectasia of the legs and face
  • · Photocoagulation of vascular lesions of the face and extremities
  • · Endovascular coagulation of the greater saphenous vein of the thigh in patients with superficial vein reflux

Urology:

Excision, vaporization, incision, coagulation, ablation and hemostasis of urological tissues. Examples include:

  • Vaporization of urethral tumors .
  • Release of urethral stricture �
  • Removal of bladder neck obstruction ♥
  • Excision and vaporization of condyloma .
  • Lesions of external genitalia .

Gynecology:

Ablation, excision, incision, coagulation, hemostasis and vaporization of gynecological tissue. Examples include:

  • Endometrial ablation �
  • t Excision or vaporization of condylomata acuminata
  • Vaporization of cervical intraepithelial neoplasia �
  • . Cervical conization
  • . Menorrhagia

{4}------------------------------------------------

Neurosurgery:

Vaporization, coagulation, excision, incision, ablation and hemostasis of soft tissue. Example includes:

Hemostasis in conjunction with menigiomas

Pulmonary Surgery:

Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system. Examples include:

  • Tracheobronchial malignancy or stricture .
  • . Benign and malignant pulmonary obstruction
  • Endoscopic pulmonary applications .

Cardiothoracic Surgery:

Incision, excision, vaporization, ablation, hemostasis, or coagulation of soft tissue, including cardiac tissues.

Laser Assisted Lipolysis

Ophthalmology:

The ablation, vaporization, excision, incision, and coagulation of soft tissue in ophthalmology. Examples include:

  • Open Dacryocystorhinostomy
  • . Endonasal Dacryocystorhinostomy
  • Blepharoplasty .
  • Oculoplastics .
  • Tumor excision and biopsy .
  • Eyelid reconstruction .

Technological Characteristics

The Ceralas E15/980, E30/980 has the same technological characteristics as the cleared Ceralas D. The laser has an additionally incorporated RFID sensor/antenna in the fiber port so as to recognize fibers with an RFID chip in their connectors. The fibers used with the Ceralas E have been previously cleared by FDA (K951775, K964497, K060050) for use with its other laser products and have an RFID chip incorporated into the proximal connector end.

Substantial Equivalence

The Ceralas E15/980, E30/980 has the same intended use and similar indications, principles of operation, and technological characteristics as the cleared Ceralas D 980 (Models D15, D25), The minor differences in the Ceralas E15/980, E30/980 technological characteristics do not raise

{5}------------------------------------------------

Koy 22 2i3 P3 6 of 6

any new questions of safety or effectiveness. Thus, the Ceralas E15/980, E30/980 is substantially equivalent to its predicate devices.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle faces to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Biolitec, Inc. % Hogan & Hartson, LLP Mr. Jonathan S. Kahan Columbia Square 555 Thirteenth Street, Northwest Washington, District of Columbia 20004

OCT 2 4 2008

Re: K082263 Trade/Device Name: Ceralas E 980nm Diode Laser (Models E15/980,E30/980) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: September 30, 2008 Received: September 30, 2008

Dear Mr. Kahan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{7}------------------------------------------------

Page 2 - Mr. Jonathan S. Kahan

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events. (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Mark M. Mulhausen

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Ko 82263

Indications for Use Statement

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Ceralas E 980nm Diode Laser (Models E15/980,E30/980)

Indications for Use:

The Ceralas E15/980, E30/980 is intended for delivery of laser light to soft tissue in the contact or non-contact mode during surgical procedures, including via endoscopes, introducers, or catheters. The Ceralas E15/980, E30/980 is generally indicated for incision, excision, vaporization, ablation, hemostasis, or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), dental procedures, arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, cardiothoracic surgery, and ophthalmology. This Ceralas E15/980, E30/980 is specifically indicated for hemostasis and coagulation of soft tissue (including cardiac tissue) and laser assisted lipolysis.

Specific Indications for Use

Ear, Nose and Throat and Oral Surgery:

Hemostasis, incision, excision, ablation, coagulation, and vaporization of tissue from the ear, nose, throat and adjacent areas including soft tissue in the oral cavity. Examples include:

  • Removal of benign lesions from the ear, nose and throat ↓
  • Excision and vaporization of vocal cord nodules and polyps .
  • Incision and excision of carcinoma in situ ◆
  • Ablation and vaporization of hyperkeratosis t
  • Excision of carcinoma of the larynx �
  • ◆ Laryngeal papillomectomy
  • Excision and vaporization of herpes simplex I and II ♥
  • . Neck dissection
  • Tonsillectomy .
  • Thyroidectomy .
  • Vocal cord polypectomy .
  • Hemiglossectomy ●
  • Tracheal stenosis .
  • Oral cavity lesions .

Dental Applications:

Intraoral and extraoral soft tissue (including marginal and interdental gingiva and epithelial lining of free gingiva). Examples include:

  • Frenectomy ◆
  • Frenotomy
  • Biopsy ●

{9}------------------------------------------------

082263

pg 2 of 5

  • Operculectomy
  • Implant recovery
  • gingivectom y �
  • Gingivoplasty ◆
  • . Gingival troughing
  • . Crown lengthening
  • . Hemostasis of donor site
  • . Removal of granulation tissue
  • . Laser assisted flap surgery
  • . Debridement of diseased epithelial lining
  • Incisions and draining of abscesses
  • . Tissue retraction for impressions
  • Papillectomy .
  • . Vestibuloplasty
  • . Excision of lesions
  • Exposure of unerupted/partially erupted teeth .
  • . Leukoplakia
  • Removal of hyperplastic tissues .
  • Treatment of aphthous ulcers .
  • Sulcular debridement (removal of diseased or inflamed soft tissue in the ◆ periodontal pocket)
  • Pulpotomy .
  • Pulpotomy as an adjunct to root canal therapy �
  • Light activation of bleaching materials for teeth whitening

Arthroscopy:

Hemostasis, incision, excision, coagulation, vaporization and ablation of joint tissues during arthroscopic surgery. Examples include:

  • . Menisectomy
  • Synovectomy
  • Chondromalacia

Gastroenterology:

Hemostasis, incision, excision, ablation, coagulation and vaporization of tissue in the upper and lower gastrointestinal tracts and also with endoscopic procedures. Examples include:

  • Hemostasis of upper and lower GI bleeding
  • Excision and vaporization of colorectal carcinoma
  • Excision of polyps

{10}------------------------------------------------

General Surgery, Dermatology, Plastic Surgery and Podiatry:

Excision, ablation, vaporization and photocoagulation of skin lesions, hemostasis, incision, excision, vaporization, ablation and debulking of soft tissue, abdominal, rectal, skin, fat or muscle tissue and dermabrasion. Examples include:

  • Matrixectomy .
  • Excision of neuromas .
  • Excision of periungual and subungual warts .
  • Excision of plantar warts .
  • Excision of keloids .
  • Liver resection .
  • Excision of cutaneous lesions .
  • Hemorrhoidectomy .
  • Appendectomy .
  • Debridement of decubitus ulcers ●
  • t Hepatobiliary tumors
  • Mastectomy .
  • Dermabrasion .
  • Vaporization and hemostasis of capillary hemangioma .
  • Excision, vaporization and hemostasis of abdominal tumors .
  • Excision, vaporization and hemostasis of rectal pathology ●
  • Pilonidal cystectomy .
  • Herniorapphy .
  • Adhesiolysis .
  • . Parathyroidectomy
  • Laparoscopic cholecystectomy �
  • Thyroidectomy .
  • Resection of organs .
  • Debridement of wounds .
  • Phototcoagulation of teleangectasia of the legs and face .
  • Photocoagulation of vascular lesions of the face and extremities �
  • Endovascular coagulation of the greater saphenous vein of the thigh in patients with superficial vein reflux

Urology:

Excision, vaporization, incision, coagulation, ablation and hemostasis of urological tissues. Examples include:

  • Vaporization of urethral tumors .
  • Release of urethral stricture t
  • Removal of bladder neck obstruction .
  • Excision and vaporization of condyloma .
  • Lesions of external genitalia

{11}------------------------------------------------

Gynecology:

Ablation, excision, incision, coagulation, hemostasis and vaporization of gynecological tissue. Examples include:

  • Endometrial ablation .
  • Excision or vaporization of condylomata acuminata .
  • Vaporization of cervical intraepithelial neoplasia �
  • Cervical conization ●
  • Menorrhagia .

Neurosurgery:

Vaporization, coagulation, excision, incision, ablation and hemostasis of soft tissue. Example includes:

Hemostasis in conjunction with menigiomas

Pulmonary Surgery:

Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system. Examples include:

  • Tracheobronchial malignancy or stricture ●
  • Benign and malignant pulmonary obstruction .
  • Endoscopic pulmonary applications ●

Cardiothoracic Surgery:

Incision, excision, vaporization, ablation, hemostasis, or coagulation of soft tissue, including cardiac tissues.

Laser Assisted Lipolysis

Ophthalmology:

The ablation, vaporization, excision, incision, and coagulation of soft tissue in ophthalmology. Examples include:

  • Open Dacryocystorhinostomy .
  • Endonasal Dacryocystorhinostomy .
  • Blepharoplasty t
  • Oculoplastics �
  • Tumor excision and biopsy .
  • Eyelid reconstruction .

{12}------------------------------------------------

K22(3

Prescription Use X (Per 21 C.F.R. 801.109) AND/OR

Over-The-Counter Use (Per 21 C.F.R. 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nil RPOgl for mxm

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

K082263 510(k) Number

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.