K Number
K082140

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-09-29

(61 days)

Product Code
Regulation Number
866.1640
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cepheid Xpert™ MRSA/SA Blood Culture Assay performed on the GeneXpert® Dx System™ is a qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) DNA directly from patient positive blood cultures. The assay utilizes automated realtime polymerase chain reaction (PCR) for the amplification of MRSA/SA specific DNA targets and fluorogenic target-specific hybridization probes for the real-time detection of the amplified DNA. The assay is performed directly on positive blood culture specimens using BD BACTEC™ Plus Aerobic/F blood culture bottles that are determined as Gram Positive Cocci in Clusters (GPCC) or as Gram Positive Cocci in singles (GPC) by Gram stain. The Cepheid Xpert™ MRSA/SA Blood Culture Assay is not intended to monitor treatment for MRSA/SA infections. Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing or for epidemiological typing.

Device Description

The Cepheid Xpert™ MRSA/SA Blood Culture Assay (Xpert MRSA/SA Blood Culture Assay) is a rapid, automated DNA test for simultaneously detecting MRSA and SA Alber) I a rapids, and culture specimens. The specimens. The specimen consists of an aliquot taken from a positive blood culture bottle for testing with the Xpert MRSA/SA Blood Culture Assay. Using one of the small disposable transfer pipettes provided with the test kit, a single drop of the positive blood culture aliquot (approximately 50 µL) is transferred into the Elution Reagent. The Elution Reagent is vortexed for 10 seconds and the entire contents are transferred to "S" chamber of the disposable fluidic cartridge (the Xpert MRSA/SA Blood Culture Assay cartridge). The two single-use reagents (Reagent 1 and Reagent 2) that are provided with the assay are transferred to different, uniquelylabeled chambers of the Xpert MRSA/SA cartridge. The user initiates a test from the system user interface and places the cartridge into the GeneXpert® Dx System instrument platform, which performs hands-off real-time, multiplex polymerase chain reaction (PCR) for detection of DNA. In this platform, additional sample preparation, amplification, and real-time detection are all fully-automated and completely integrated.

The GeneXpert Dx System consists of a GeneXpert instrument, personal computer, and the multi-chambered fluidic cartridges that are designed to complete sample preparation and real-time PCR for detection of MRSA and SA in approximately 50 minutes. Each system has 2 to 16 randomly accessible modules that are each capable of performing separate sample preparation and real-time PCR tests. Each module contains a syringe drive for dispensing fluids, an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and detection.

The Xpert MRSA/SA Assay includes reagents for the simultaneous detection of the target organisms, SA and MRSA. The primers and probes in the Xpert MRSA/SA Assay detect nucleic acid sequences of the staphylococcal protein A (spa), the gene for methicillin/oxacillin resistance (mecA), and staphylococcal cassette chromosome (SCCmec) inserted in the SA chromosomal attB site.

The test includes a Sample Processing Control (SPC) to control for adequate processing of the target bacteria and to monitor the presence of inhibitor(s) in the PCR assay. The Probe Check Control (PCC) verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.

AI/ML Overview

The acceptance criteria and study proving the device meets them are detailed below for the Cepheid Xpert™ MRSA/SA Blood Culture Assay.

1. Table of Acceptance Criteria and Reported Device Performance:

The document implicitly defines acceptance criteria based on the performance observed in the predicate device and the clinical study results. The clinical study results serve as the primary demonstration of meeting these criteria.

Metric (MRSA)Acceptance Criteria (Implicit from Predicate/Standard of Care)Reported Device Performance (Xpert MRSA/SA Blood Culture Assay)
Sensitivity98.2 - 100.0% (BD GeneOhm StaphSR)100.0% (53/53) (95% CI: 93.3% - 100%)
Specificity98.2 - 100.0% (BD GeneOhm StaphSR negative % agreement)100.0% (196/196) (95% CI: 98.1% - 100%)
Metric (SA)Acceptance Criteria (Implicit from Predicate/Standard of Care)Reported Device Performance (Xpert MRSA/SA Blood Culture Assay)
Sensitivity98.8 - 100.0% (BD GeneOhm StaphSR)100.0% (77/77) (95% CI: 95.3% - 100%)
Specificity96.5 - 100.0% (BD GeneOhm StaphSR negative % agreement)99.4% (171/172) (95% CI: 96.8% - 100%)

Note: The predicate device's performance is listed as "Positive % Agreement" and "Negative % Agreement," which are equivalent to sensitivity and specificity in this context. The implicit acceptance criteria for the new device are to perform at least comparably to the predicate and the "standard of care" (reference culture results and susceptibility testing).

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Clinical Test Set: 249 specimens.
  • Data Provenance: Multi-site prospective investigation study at three US institutions. The data is prospective, collected from subjects whose routine care included blood culture testing.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

The document does not explicitly state the number of experts or their specific qualifications (e.g., years of experience as a radiologist) used to establish the ground truth for the clinical test set. It mentions "reference culture results and susceptibility testing, the current standard of care" and that susceptibility testing was performed "in accordance with the CLSI documents M2-A9 and M100-S17." This implies that qualified laboratory personnel, following established clinical microbiology guidelines, performed these reference methods.

4. Adjudication Method for the Test Set:

The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1). The ground truth was established by "reference culture results and susceptibility testing, the current standard of care." This suggests a direct comparison against the established laboratory gold standard, rather than a consensus among multiple human readers.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly done in the context of comparing human readers with and without AI assistance. The study compares the device's performance directly against "reference culture results and susceptibility testing," which is the established standard of care for identifying bacterial infections. The device is an automated nucleic acid amplification test, not an AI assistance tool for human interpretation.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

Yes, a standalone performance study was conducted. The Cepheid Xpert™ MRSA/SA Blood Culture Assay is an automated device that performs sample preparation, amplification, and real-time detection without human intervention once the sample is loaded. The "Overall Results" section (Page 12) directly reports the device's performance in identifying MRSA and SA specimens relative to culture. This refers to the algorithm's performance without a human in the loop for interpretation.

7. The Type of Ground Truth Used:

The ground truth used for the clinical study was based on:

  • Reference Culture Results: This involves standard microbiological culture techniques to isolate and identify organisms.
  • Susceptibility Testing: Performed in accordance with CLSI documents (M2-A9 and M100-S17) to determine methicillin/oxacillin resistance, using cefoxitin as a surrogate.

This represents established microbiological laboratory standards, often considered the "gold standard" for pathogen identification and antimicrobial susceptibility.

8. The Sample Size for the Training Set:

The document does not explicitly state a separate "training set" for the clinical validation studies in the way one might for a machine learning algorithm. For analytical studies, various strains were tested:

  • Analytical Inclusivity Study on CDC Staphylococcus aureus Specimens: 25 Staphylococcus aureus strains.
  • Analytical Inclusivity Study on Expanded Panel of Staphylococcus aureus Specimens: 121 additional Staphylococcus aureus strains.
  • "Empty Cassette Variants" Study: 22 Staphylococcus aureus isolates.
  • Analytical Limit of Detection: Replicates of 20 for 6 MRSA isolates and 3 MSSA isolates.
  • Cross-reactivity Study: 105 strains (98 ATCC, 7 NARSA).

9. How the Ground Truth for the Training Set Was Established:

For the analytical "training" (or rather, validation) sets mentioned above, the ground truth was established through:

  • Bacterial strain identification, PFGE type, and SCCmec type determined by the CDC (for CDC specimens).
  • Catalase, tube coagulase, and Gram stain for characterizing discordants in the expanded panel study.
  • Methicillin susceptibility assessed by disk diffusion using a cefoxitin disk.
  • Known characteristics and genetic information of the cultured strains (e.g., MRSA, SA, Cfu/test).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the word "CONFIDENTIAL" in all caps. The word is written in a simple, sans-serif font. The text is black against a white background, creating a high contrast.

Image /page/0/Picture/1 description: The image shows a handwritten string of characters. The string appears to be "K082140". The characters are written in black ink on a white background. The handwriting is somewhat messy, but the characters are still legible.

Image /page/0/Picture/11 description: The image contains the word "Cepheid." in a stylized font. To the left of the word is a graphic of three curved lines that are stacked on top of each other. The lines are thick and black, and they appear to be moving in a upward direction.

SEP 2 9 2008

5.0 510(k) Summary

As required by 21 CFR Section 807.92(c).

Submitted by:Cepheid904 Caribbean DriveSunnyvale, CA 90489Phone number: (408) 400-8230Fax number: (408) 541-6439
Contact:Russel K. Enns, Ph.D.
Date of Preparation:September 17, 2008
Device:
Trade name:Xpert™ MRSA/SA Blood Culture Assay
Common name:Methicillin-resistant Staphylococcus aureus (MRSA) andStaphylococcus aureus (SA) from positive blood culturebottles assay.
Type of Test:Nucleic Acid Amplification Test, DNA, Methicillin-resistant Staphylococcus aureus (MRSA) andStaphylococcus aureus (SA), qualitative
Classification name:Antimicrobial susceptibility test powder
Regulation number:866.1640
Procode:NQX
ClassificationAdvisory Committee:Microbiology
Panel:83
Predicate Devices:BD GeneOhm™ StaphSR Assay (510(k) #K071026),Remel Staphaurex Latex Agglutination Test (510(k)#K851949),BBL (BD) Oxacillin Screen Agar (510(k) #K863821),BD BBL CHROMagar MRSA (510(k) #K042812BD Phoenix Automated Microbiology ID/AST System(510(k) #K020322 and #K023301).

Device Description:

The Cepheid Xpert™ MRSA/SA Blood Culture Assay (Xpert MRSA/SA Blood Culture Assay) is a rapid, automated DNA test for simultaneously detecting MRSA and SA Alber) I a rapids, and culture specimens. The specimens. The specimen consists of an aliquot

{1}------------------------------------------------

taken from a positive blood culture bottle for testing with the Xpert MRSA/SA Blood Culture Assay. Using one of the small disposable transfer pipettes provided with the test kit, a single drop of the positive blood culture aliquot (approximately 50 µL) is transferred into the Elution Reagent. The Elution Reagent is vortexed for 10 seconds and the entire contents are transferred to "S" chamber of the disposable fluidic cartridge (the Xpert MRSA/SA Blood Culture Assay cartridge). The two single-use reagents (Reagent 1 and Reagent 2) that are provided with the assay are transferred to different, uniquelylabeled chambers of the Xpert MRSA/SA cartridge. The user initiates a test from the system user interface and places the cartridge into the GeneXpert® Dx System instrument platform, which performs hands-off real-time, multiplex polymerase chain reaction (PCR) for detection of DNA. In this platform, additional sample preparation, amplification, and real-time detection are all fully-automated and completely integrated.

The GeneXpert Dx System consists of a GeneXpert instrument, personal computer, and the multi-chambered fluidic cartridges that are designed to complete sample preparation and real-time PCR for detection of MRSA and SA in approximately 50 minutes. Each system has 2 to 16 randomly accessible modules that are each capable of performing separate sample preparation and real-time PCR tests. Each module contains a syringe drive for dispensing fluids, an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and detection.

The Xpert MRSA/SA Assay includes reagents for the simultaneous detection of the target organisms, SA and MRSA. The primers and probes in the Xpert MRSA/SA Assay detect nucleic acid sequences of the staphylococcal protein A (spa), the gene for methicillin/oxacillin resistance (mecA), and staphylococcal cassette chromosome (SCCmec) inserted in the SA chromosomal attB site.

The test includes a Sample Processing Control (SPC) to control for adequate processing of the target bacteria and to monitor the presence of inhibitor(s) in the PCR assay. The Probe Check Control (PCC) verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.

Device Intended Use:

The Cepheid Xpert™ MRSA/SA Blood Culture Assay performed on the GeneXpert® Dx System™ is a qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) DNA directly from patient positive blood cultures. The assay utilizes automated realtime polymerase chain reaction (PCR) for the amplification of MRSA/SA specific DNA targets and fluorogenic target-specific hybridization probes for the real-time detection of the amplified DNA. The assay is performed directly on positive blood culture specimens that are determined as Gram Positive Cocci in Clusters (GPCC) or as Gram Positive Cocci in singles (GPC) by Gram stain. The Cepheid Xpert™ MRSA/SA Blood Culture Assay is not intended to monitor treatment for MRSA/SA infections. Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing or for epidemiological typing.

{2}------------------------------------------------

In a mixed culture containing MRSA/SA and other organisms (e.g. Gram negative bacilli, yeast), results can be false negative or variable depending on the concentration of MRSA/SA present, particularly if the concentration of MRSA/SA is close to the LoD of the assay.

The performance of the Xpert MRSA/SA Blood Culture Assay was validated in Cepheid's clinical study using BD BACTEC™ Plus Aerobic/F blood culture bottles. Use of any other culture bottles and/or media would require independent validation.

Substantial Equivalence:

The Xpert MRSA/SA Blood Culture Assay is substantially equivalent to the BD GeneOhm™ StaphSR Assay (510(k) #K071026). Both assays detect SA and MRSA from positive blood cultures and determine the presence of the target organisms through real-time PCR amplification and fluorogenic target-specific hybridization detection.

Table 5.1 shows the similarities and differences between the Xpert MRSA/SA Blood Culture Assay and the BD GeneOhmTM StaphSR Assay.

The Xpert MRSA/SA Blood Culture Assay is also substantially equivalent to conventional microbiology-based predicate devices that identify (ID) and/or test for antimicrobial susceptibility (AST) of gram positive organisms, including Staphylococcus species of human origin from pure culture isolates. These conventional microbiologybased predicates accommodate multiple specimen types, including positive blood cultures. These conventional microbiology-based predicates are:

  • Remel Staphaurex Latex Agglutination Test (510(k) #K851949), .
  • BBL (BD) Oxacillin Screen Agar (510(k) #K863821), .
  • BD BBL CHROMagar MRSA (510(k) #K042812), and .
  • BD Phoenix Automated Microbiology ID/AST System (510(k) #K020322 and . #K023301).

Table 5.2 compares the new device with the conventional microbiology-based assays for SA. Table 5.3 compares the new device with the conventional microbiology-based assays for MRSA.

A multi-center study was conducted on 279 patients to determine the performance characteristics of the device relative to the reference culture results and susceptibility testing, the current standard of care. The test results showed the Xpert MRSA/SA Blood Culture Assay to be substantially equivalent to the current standard of care.

{3}------------------------------------------------

Table 5.1

Similarities and Differences Between the Xpert MRSA/SA Blood Culture Assay and the Molecular-based Predicate Device

Similarities
DevicePredicate
ItemXpert MRSA/SA Blood CultureAssayBD GeneOhm™ StaphSRAssay (510(k) #K071026)
Intended UseRapid detection of MRSA and SASame
Indication for UseIdentification of MRSA and SAcolonizationSame
Specimen TypePositive Blood CultureSame
TechnologicalPrinciplesFully-automated nucleic acidamplification (DNA); real-time PCRSame
DNA TargetSequenceSequence incorporating the insertionsite ( attBssc ) of StaphylococcalCassette Chromosome mec(SCC mec ) for detection of MRSA.Same
ClinicalComparisonResults:Xpert™ MRSA/SA Blood CultureAssay Performance vs. ReferenceCulture :MRSA:Sensitivity: 100.0 %Specificity: 100.0 %Accuracy: 100.0%PPVa: 100.0 %NPVb: 100.0 %Prevalence: 19.0 %SA:Sensitivity: 100.0%Specificity: 99.5 %Accuracy: 99.6 %PPV: 98.7 %NPV: 100.0 %Prevalence: 27.6 %aPositive predictive valuebNegative predictive valueBD GeneOhm™ StaphSRAssay Performance vs.Reference Culture methods :MRSA:Positive % Agreement: 100.0Negative % Agreement: 98.2 -100.0SA:Positive % Agreement: 98.8 -100.0Negative % Agreement: 96.5 -100.0[Data obtained from the BDGeneOhm StaphSR Assay510(k) Summary]
Differences
DevicePredicate
ItemXpert™ MRSA/SA Blood CultureAssayBD GeneOhm™ StaphSRAssay (K071026)
Test CartridgeDisposable single-use, multi-chambered fluidic cartridge.Disposable single-use PCR tube
Instrument SystemCepheid GeneXpert Dx SystemCepheid SmartCycler
SamplePreparationSelf-contained and automated after mixed specimen and two single-dose reagents are added to cartridge.Manual
ProbesTaqMan® ProbesMolecular Beacons
Internal ControlsSample processing control (SPC) and probe check control (PCC).One internal reagent control and external positive and negative controls required per run
DNA TargetSequenceSequence specific to methicillin/oxacillin resistance ( mecA gene)N/A
UsersOperators with no clinical lab experience to experienced clinical laboratory technologists.CLIA High ComplexityLaboratory Users
DNA TargetSequenceSequence specific to Staphylococcus aureus species ( spa gene)Sequence specific to Staphylococcus aureus species ( nuc gene)
Ability to identifycorrectly "EmptyCassette Variants"Yes, sequence specific to Staphylococcus aureus species ( mecA gene)No
Rapid test resultsApproximately 50 minutes to result.Approximately 60-75 minutes to result.

{4}------------------------------------------------

:

.

{5}------------------------------------------------

Table 5.2

Similarities and Differences Between the Xpert MRSA/SA Blood Culture Assay and the Conventional Microbiology-based Predicate Devices for Staphylococcus aureus (SA) only

Similarities
ItemDevicePredicates (SA only)
Xpert MRSA/SA BloodCulture AssayStaphaurex LatexAgglutinationTest for SAK851949BD PhoenixAutomatedMicrobiologySystem for SA510(k) #K020322
Intended UseDetection of SASameSame
Single useYesSameSame
Assay ControlsPositive Control: SANegative Control: S.epidermidisSameSame
Differences
ItemDevicePredicates (SA only)
Xpert MRSA/SA BloodCulture AssayStaphaurex LatexAgglutinationTest for SAK851949BD PhoenixAutomatedMicrobiologySystem for SAK020322
Mode ofDetectionSequence specific toStaphylococcus aureusspecies (spa gene)Clumping factorand protein AMicrobialutilization anddegradation ofspecificsubstrates
Specimen TypePositive Blood CulturesStaphylococcusspeciesGram Positiveorganisms
Assay formatAmplification: PCRDetection: Fluorogenic target-specific hybridizationAgglutinationwith latexparticlessensitized withfibrinogen andIgGConventional,chromogenic andfluorogenicbiochemical testsfor identification(ID) andantimicrobialresistance test(AST)

{6}------------------------------------------------

Differences
ItemDevicePredicates (SA only)
Xpert MRSA/SA BloodCulture AssayStaphaurex LatexAgglutinationTest for SAK851949BD PhoenixAutomatedMicrobiologySystem for SAK020322
Interpretation of test resultsDiagnostic software of the GeneXpert Dx SystemVisual interpretationAutomated

Table 5.3

Similarities and Differences Between the Xpert MRSA/SA Blood Culture Assay and the Conventional Microbiology-based Predicate Devices for Methicillin-Resistant Staphylococcus aureus (MRSA)

Similarities
ItemDevicePredicates (MRSA only)
XpertMRSA/SABlood CultureAssayMueller HintonAgar w/ 4% NaCland Oxacillin(Oxacillin ScreenAgar Test)K863821BBLCHROMagarMRSA(510(k)#K042812)BD PhoenixAutomatedMicrobiologySystem510(k)#K023301
IntendedUseDetection ofMRSASameSameSame
Single useYesSameSameSame
AssayControlsPositive Control:MRSANegative Control:SASameSameSame
Differences
ItemDevicePredicates (MRSA only)
XpertMRSA/SABlood CultureAssayMueller HintonAgar w/ 4% NaCland Oxacillin(Oxacillin ScreenAgar Test)K863821BBLCHROMagarMRSA(510(k)#K042812)BD PhoenixAutomatedMicrobiologySystem510(k)#K023301

{7}------------------------------------------------

Differences
ItemDevicePredicates (MRSA only)
XpertMRSA/SABlood CultureAssayMueller HintonAgar w/ 4% NaCland Oxacillin(Oxacillin ScreenAgar Test)K863821BBLCHROMagarMRSA(510(k)#K042812)BD PhoenixAutomatedMicrobiologySystem510(k)#K023301
Mode ofDetectionformethicillinresistanceSCCmec genespecific forMRSAmecA genespecific formethicillin/oxacillinresistanceGrowth onMueller HintonAgar with 4%NaCl and 6 ug/mloxacillinUse of specificChromogenicsubstrates andcefoxitin todifferentiateMRSA fromother organismsUtilizes aredox indicatorfor detection oforganismgrowth in thepresence of anantimicrobialagent
SpecimenTypePositive BloodCulturesPure cultureisolate ofStaphylococcusaureusAnterior naresPure cultureisolate ofStaphylococcusaureus
AssayformatAmplification:PCRDetection:Fluorogenictarget-specifichybridizationPhenotypicdetection basedon a 24 hourgrowth of SAinoculated onmediaPhenotypicdetection baseon a 24-48 hourgrowth of SA(mauvecolonies)inoculated onmediaAST panelscontainingMIC tests forseveralantimicrobialagents
Interpretation of testresultsDiagnosticsoftware of theGeneXpert DxSystemManual: VisualinterpretationManual: VisualinterpretationAutomated

Non-Clinical Studies:

Analytical Inclusivity Study on CDC Staphylococcus aureus Specimens

Twenty-five (25) Staphylococcus aureus strains from multiple sources provided by the CDC were tested using the Xpert™ MRSA/SA Blood Culture Assay. All strains were tested in triplicate using 100 ul of stationary phase cell suspension diluted 10 millionfold. Colony forming units per assay (CFU/test) were determined by plate counts in

{8}------------------------------------------------

triplicate. Bacterial strain identification, PFGE type and SCCmec type were determined by the CDC.

All results were reported correctly by the Xpert™ MRSA/SA Blood Culture Assay, except one specimen. Further investigation reveled that the particular specimen was actually mislabeled by the CDC.

Analytical Inclusivity Study on Expanded Panel of Staphylococcus aureus Specimens

One hundred twenty-one (121) additional Staphylococcus aureus strains were tested using the Xpert MRSA/SA Assay. Overnight cultures were grown in Brain Heart Infusion (BHI) broth and adjusted to 0.5 McFarland units. All strains were tested in triplicate using 100 uL of cultures further diluted 100 thousand to one million-fold.

MRSA (78) and SA (43) strains were selected to broadly represent the range of genetic diversity found in the species Staphylococcus aureus based on phylogenetic structure. Selections represent primary lineages with emphasis on specific clonal complexes within which MRSA is predominantly observed. Lineages that contain MRSA and SA, as well as those that contain SA exclusively were included.

The Xpert MRSA/SA Assay correctly identified 116 of 121 strains. The 5 discordants were characterized by catalase, tube coagulase, and Gram stain. Methicillin susceptibility was assessed by disk diffusion using a 30 µg cefoxitin disk and a diameter cut-off of 21/22 mm.

Three of 78 MRSA strains were reported MRSA negative; SA positive using the Xpert assay. Further characterization indicates these strains are not resistant and were correctly reported MRSA negative; SA positive.

Two of 43 SA strains were reported MRSA positive using the Xpert assay. Further characterization indicates these strains are resistant and were correctly reported MRSA positive; SA positive.

Each of the 12 known USA300 isolates were correctly reported MRSA positive and SA positive as expected.

In another study, 22 Staphylococcus aureus isolates identified as "empty cassette variants" were tested using the Xpert MRSA/SA Assay. Overnight cultures were adjusted to 0.5 McFarland units. All strains were tested from cultures further diluted 100-fold (high) and 100 thousand-fold (low), providing a range of 300 - 3000 CFU/mL. The Xpert MRSA/SA Assay correctly identified all 22 isolates as MRSA negative and SA positive. At both cell concentrations tested, only Cts for the spa and SCCmec targets were reported. No mecA Cts were reported.

Analytical Limit of Detection

{9}------------------------------------------------

Studies were performed to determine the 95% confidence intervals for the analytical limit of detection (LoD) of Staphylococcus aureus (SA) cells and methicillin-resistant Staphylococcus aureus (MRSA) cells diluted into a blood and blood culture medium that can be detected using the Xpert MRSA/SA Blood Culture Assay. The limit of detection is defined as the lowest number of colony forming units (CFU) per test that can be reproducibly distinguished from negative samples with 95% confidence or the lowest concentration at which 19 of 20 replicates are positive.

For MRSA, replicates of 20 were evaluated at each MRSA concentration tested (CFU/test) for 6 individual isolates representing SCCmec types I, II, III, IVa, V, and VI. When characterized by pulsed-field gel electrophoresis (PFGE), USA100, the most common healthcare-acquired strain and USA400, one of the most common communityacquired strains were represented.

For SA, replicates of 20 were evaluated at each MSSA concentration (CFU/test) for 3 individual MSSA isolates. USA types USA900 and USA1200 were represented.

The estimate and confidence intervals were determined using logistic regression with data (number of positive results per number of replicates at each level) over the range of CFU/test loadings. The confidence intervals were determined using maximum likelihood estimates on the logistic model parameters using the large sample variance-covariance matrix.

The results of this study indicate that the Xpert MRSA/SA Blood Culture Assay will produce a positive SA result 95% of the time with 95% confidence for a positive blood culture aliquot (1 drop or 50uL) containing 100 CFU and a positive MRSA result 95% of the time with 95% confidence for a blood culture aliquot (1 drop or 50μL) containing 250 CFU.

An auxiliary study evaluated MRSA LOD in the presence of up to 1 x 106 MSSA cells. Under the conditions of this study, no significant competitive inhibitory effects were observed on the analytical LoD of MRSA SCCmec types I, II, III, IVa, V, or VI in the presence of competing MSSA cells as high as 1:1x103.

Linearity

A study was conducted to define the reportable range of the Xpert MRSA/SA Assay and demonstrate a linear relationship between SA and MRSA input and assay output (Ct). Linearity was evaluated using ten-fold serial dilutions (1e8 CFU/sample - 10 CFU/sample) of SA and MRSA isolates.

The Xpert MRSA/SA Assay responds linearly (r2 = 0.998) with respect to spa detection as a function of SA cell input over 6 logs. PCR efficiency for the spa reaction is 100%.

The Xpert MRSA/SA Assay responds linearly (r2 = 0.999) with respect to spa detection as a function of MRSA cell input over 6 logs. PCR efficiency for the spa reaction is 95.4%.

{10}------------------------------------------------

The Xpert MRSA/SA Assay responds linearly (r2 = 0.999) with respect to mecA detection as a function of MRSA cell input over 6 logs. PCR efficiency for the mecA reaction is 93.3%.

The Xpert MRSA/SA Assay responds linearly (r2 = 0.999) with respect to SCCmec detection as a function of MRSA cell input over 5 logs. PCR efficiency for the mec reaction is 94.6%

Analytical Specificity

Cross-reactivity Study

One hundred five 105 strains were collected, quantitated and tested using the Xpert MRSA/SA Assay. The 98 cultures from the American Type Culture Collection (ATCC) and 7 strains from the Network on Antimicrobial Resistance in Staphylococcus aureus (NARSA) represent species phylogenetically related to Staphylococcus aureus or those potentially encountered in a hospital environment.

Of these, methicillin-sensitive coagulase negative staphylococci (29) and methicillinresistant coagulase negative staphylococci (9) were included. The organisms tested were identified as either Gram positive (74), Gram negative (28), or yeast (3). The organisms were further classified as either aerobic (95) or anaerobic (10).

Two or more replicates of each isolate were tested at 1.7 - 3.2 McFarland units. Under the conditions of the study, all isolates were reported MRSA negative and SA negative; none of the isolates were detected by the Xpert MRSA/SA Assay. Positive and Negative controls were included in the study. The analytical specificity was 100%.

Interfering Substances Study

Substances that may be present in blood cultures with potential to interfere with the Xpert MRSA/SA Blood Culture Assay were evaluated directly relative to the performance of the Xpert MRSA/SA Assay. Potentially interfering substances (IS) tested are shown in Table 5.6. They included anticoagulated whole blood and blood culture media components containing the anticoagulant sodium polyanetholesulfonate (SPS) or ion exchange and nonionic adsorbent resins to remove antimicrobials as listed in Table 5.4 with the active ingredients and concentrations shown. Negative samples were tested in each substance to determine the effect on the performance of the sample processing control (SPC). Positive samples were tested with MRSA cells spiked near the analytical LoD. All results were compared to positive and negative buffer controls.

Statistical analysis was conducted on the data generated.

Table 5.4: Potentially Interfering Blood Culture Substances Tested
------------------------------------------------------------------------------------
SubstanceActive IngredientConc.Level*
TET Buffer (control)Control100% (v/v)1
Whole Blood (ACD)N/A50% (v/v)2

{11}------------------------------------------------

Whole Blood (EDTA)N/A50% (v/v)3
Whole Blood (Heparin)N/A50% (v/v)4
Whole Blood (SodiumCitrate)N/A50% (v/v)5
Human PlasmaN/A100% (v/v)6
BacT/Alert SA AerobicSupplementedTryptic Soy,0.035% SPS100% (v/v)7
BacT/Alert SN AnaerobicSupplementedTryptic Soy,0.035% SPS100% (v/v)8
BACTEC STDAnaerobic/F3% Soybean-Casein DigestBroth, 0.025% SPS100% (v/v)9
BACTEC STD Aerobic/F3% Soybean-Casein DigestBroth, 0.025% SPS100% (v/v)10
BACTEC PlusAnaerobic/F2.75% Soybean-Casein DigestBroth, 0.05% SPS,17% resins100% (v/v)11
BACTEC Plus Aerobic/F2.75% Soybean-Casein DigestBroth, 0.05% SPS,17% resins100% (v/v)12

(*) Level identifies each substance in the statistical analysis (ANOVA and Dunnett's multiple comparison method).

Statistical analysis of the data generated with and without these potentially interfering substances demonstrates that under the conditions of this study their presence did not affect assay performance. The inability to induce either MRSA or SPC failures in the presence of these substances is attributed to adequate removal of potential inhibitors during sample processing.

Clinical Studies

Clinical Comparison Study

Performance characteristics of the Xpert MRSA/SA Blood Culture Assay were determined in a multi-site prospective investigation study at three US institutions by comparing the Xpert MRSA/SA Blood Culture Assay with culture.

Subjects included individuals whose routine care called for blood culture testing. If the blood culture sample was positive for microbial growth and the Gram stain showed Gram positive cocci (singles or in clusters), the sample was eligible for inclusion in the clinical study, and aliquots of leftover culture material were obtained for testing by the Xpert MRSA/SA Blood Culture Assay. Culture and Gram stain procedures, and patient management continued at the site per the standard practice. Susceptibility testing was

{12}------------------------------------------------

performed in accordance with the CLSI documents M2-A9 and M100-S17.89 Cefoxitin was used as a surrogate for detecting methicillin/oxacillin resistance.

Performance of the Xpert MRSA/SA Blood Culture Assay was calculated relative to the reference culture results.

Overall Results

A total of 249 specimens were tested for MRSA and SA by Xpert MRSA/SA Blood Culture Assay and culture.

The Xpert MRSA/SA Blood Culture Assay identified 100 % of the specimens positive for MRSA and 100 % of the specimens negative for MRSA relative to culture.

The Xpert MRSA/SA Blood Culture Assay identified 100 % of the specimens positive for SA and 99.4 % of the specimens negative for SA relative to the reference culture method.

The performance of the Xpert MRSA/SA Blood Culture Assay is summarized in Table 2.

Culture
MRSA+SA+/MRSA-Neg/NoGrowthTotal
XpertMRSA+530053
SA+/MRSA-024125
SA-00171171
Total5324172249
Xpert PerformanceMRSA:
Positive Percent Agreement: 53/53100%95% CI: 93.3% - 100%
Negative Percent Agreement: 196/196100%95% CI: 98.1% - 100%
SA:
Positive Percent Agreement: 77/77100%95% CI: 95.3% - 100%
Negative Percent Agreement: 171/17299.4%95% CI: 96.8% - 100%

Table 2: MRSA/SA Performance vs. Reference Culture

Of the Xpert MRSA/SA Blood Culture Assays run on eligible specimens, 92.8% (233/251) of these specimens were successful on the first attempt. The remaining 18 gave indeterminate results on the first attempt (10 "INVALID", 7 "ERROR" and 1 "NO RESULT"). One of the indeterminate specimens could not be retested due to insufficient reagents available at the site to perform the retest. Of the 17 indeterminate on the first attempt with sufficient sample for retest, 94.1% (16/17) gave a result on the second attempt; one was indeterminate on the second attempt.

{13}------------------------------------------------

Antibiotic Usage

Among the 249 cases in the eligible dataset, antibiotic use within the 3 weeks prior to sample collection was reported for 65 and no antibiotic use was confirmed for 159; for 25 cases, antibiotic status was unknown. Antibiotic use did not cause a statistically significant difference in assay positive percent agreement or negative percent agreement.

Reproducibility Study

A panel of 10 specimens with varying concentrations of SA, MRSA and Staphylococcus epidermidis (negative) were tested in duplicate on 10 different days at each of the three sites (10 specimens x 2 times/ day x 10 days x 3 sites). One lot of Xpert MRSA/SA Blood Culture was used at each of the 3 testing sites. Xpert MRSA/SA Blood Culture Assays were performed according to the Xpert MRSA/SA procedure. Results are summarized in Table 5.6. Table 6 provides mean cycle threshold (Ct) values with variance components (SD and %CV) for each concentration level tested.

Specimen IDSite 1Site 2Site 3% Total Agreement
Neg (MSSE)100%(20/20)100%(20/20)100%(20/20)100% (60/60)
SA High Neg100%(20/20)100%(20/20)100%(20/20)100% (60/60)
SA Low Pos100%(20/20)100%(20/20)100%(20/20)100% (60/60)
SA Moderate Pos100%(20/20)100%(20/20)100%(20/20)100% (60/60)
MRSA1 High Neg100%(20/20)100%(20/20)100%(20/20)100% (60/60)
MRSA1 Low Pos100%(20/20)100%(20/20)100%(20/20)100% (60/60)
MRSA1 Moderate Pos100%(20/20)100%(20/20)100%(20/20)100% (60/60)
MRSA2 High Neg100%(20/20)100%(20/20)100%(20/20)100% (60/60)
MRSA2 Low Pos100%(20/20)100%(20/20)100%(20/20)100% (60/60)
MRSA2 Moderate Pos100%(20/20)100%(20/20)100%(20/20)100% (60/60)
% Total Agreementby Site100%(200/200)100%(200/200)100%(200/200)100% (600/600)

Table 5.6 - Summary of Reproducibility Results

{14}------------------------------------------------

BG
LevelSiteMeanStdDevCV
MRSA1 neg134.581.654.8%
MRSA1 neg234.270.872.5%
MRSA1 neg334.571.143.3%
MRSA1 negAll34.471.253.6%
MRSA2 neg134.440.782.3%
MRSA2 neg234.380.671.9%
MRSA2 neg334.601.022.9%
MRSA2 negAll34.470.822.4%
SA neg134.090.802.3%
SA neg234.390.521.5%
SA neg334.230.882.6%
SA negAll34.230.752.2%
NEG134.371.043.0%
NEG234.350.561.6%
NEG334.270.832.4%
NEGAll34.330.822.4%

Table 6 – Mean Cycle Threshold (Ct) Value With Variance Components (SD And %CV) For Each Concentration Level Tested In The Reproducibility Study

spa
LevelSiteMeanStdDevCV
MRSAI low131.140.581.9%
230.620.381.2%
330.530.311.0%
All30.760.511.7%
MRSA1 mod129.230.341.2%
229.030.531.8%
328.880.331.2%
All29.050.431.5%
MRSA2 low130.940.471.5%
230.540.451.5%
330.770.391.3%
All30.750.461.5%
MRSA2 mod129.560.521.8%
229.280.411.4%
329.490.461.5%
All29.440.471.6%
SA low132.930.772.3%

.

{15}------------------------------------------------

232.850.712.2%
332.570.401.2%
All32.780.652.0%
SA mod132.250.722.2%
231.950.571.8%
331.810.441.4%
All32.000.611.9%
mecA
LevelSiteMeanStdDevCV
MRSA1 low131.370.501.6%
MRSA1 low230.920.461.5%
MRSA1 low330.780.270.9%
MRSA1 lowAll31.020.481.6%
MRSA1 mod129.480.341.2%
MRSA1 mod229.270.411.4%
MRSA1 mod329.230.351.2%
MRSA1 modAll29.330.381.3%
MRSA2 low131.220.501.6%
MRSA2 low231.020.652.1%
MRSA2 low331.260.601.9%
MRSA2 lowAll31.170.591.9%
MRSA2 mod130.010.571.9%
MRSA2 mod229.620.361.2%
MRSA2 mod329.790.441.5%
MRSA2 modAll29.800.481.6%
SCCmec
LevelSiteMeanStdDevCV
MRSA1 low133.130.672.0%
232.650.571.7%
332.470.300.9%
All32.750.601.8%
MRSA1 mod131.150.421.3%
230.990.431.4%
331.040.521.7%
All31.060.461.5%
MRSA2 low133.010.451.4%
232.650.712.2%

·

{16}------------------------------------------------

333.030.571.7%
All32.900.601.8%
MRSA2 mod131.740.551.7%
231.370.431.4%
331.620.421.3%
All31.570.491.5%

Conclusions

The results of the nonclinical analytical and clinical performance studies summarized above demonstrate that the MRSA/SA Blood Culture Assay is as safe, as effective, and performs as well as or better than the predicate device.

:

:

{17}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/17/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The logo is black and white.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

SEP 2 9 2008

Russel K. Enns, Ph.D. Senior Vice President Regulatory & Clinical Affairs, Quality System and Reimbursement Cepheid, Inc. 904 Caribbcan Drive Sunnyvale, CA 94089

K082140 Trade/Device Name: Cepheid Xpert™ MRSA/SA Blood Culture Assay Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: Class II Product Code: NQX Dated: September 17, 2008 Received: September 19, 2008

Dear Dr. Enns:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act (Act do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labelies roquilibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS), and good manufacturing practi

{18}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Sally attaym

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{19}------------------------------------------------

4.0 Indications for Use Statement

510(k) Number (if known): K082140

Device Name: Xpert™ MRSA/SA Blood Culture Assay

Indications for Use:

The Cepheid Xpert™ MRSA/SA Blood Culture Assay performed on the GeneXpert® Dx System™ is a qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) DNA directly from patient positive blood cultures. The assay utilizes automated realtime polymerase chain reaction (PCR) for the amplification of MRSA/SA specific DNA targets and fluorogenic target-specific hybridization probes for the real-time detection of the amplified DNA. The assay is performed directly on positive blood culture specimens using BD BACTEC™ Plus Aerobic/F blood culture bottles that are determined as Gram Positive Cocci in Clusters (GPCC) or as Gram Positive Cocci in singles (GPC) by Gram stain. The Cepheid Xpert™ MRSA/SA Blood Culture Assay is not intended to monitor treatment for MRSA/SA infections. Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing or for epidemiological typing.

Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Freddie Lee Cooke
Division Sign Off

sion Sign-C

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K082140

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).