MSC LIPID CONTROL

K082067 · Maine Standards Co. · JJY · Aug 22, 2008 · Clinical Chemistry

Device Facts

Record IDK082067
Device NameMSC LIPID CONTROL
ApplicantMaine Standards Co.
Product CodeJJY · Clinical Chemistry
Decision DateAug 22, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

The MSC Lipid Control is intended for use as an assayed control material to monitor the ongoing precision of clinical laboratory analysis for Cholesterol, Triglycerides, HDL Cholesterol, and LDL Cholesterol. This material is intended for use on automated, semi-automated, and manual clinical chemistry analyzer systems.

Device Story

MSC Lipid Control is a human serum-based liquid control; contains stabilized Cholesterol, Triglycerides, HDL Cholesterol, and LDL Cholesterol. Used in clinical laboratories to monitor precision of chemistry analyzer systems. Laboratory personnel use the control to verify analytical performance; results help ensure accuracy of patient lipid testing. No patient-facing interface; purely a diagnostic quality control reagent.

Clinical Evidence

No clinical data. Bench testing only.

Technological Characteristics

Human serum-based liquid control; contains stabilized Cholesterol, Triglycerides, HDL Cholesterol, and LDL Cholesterol. Compatible with automated, semi-automated, and manual clinical chemistry analyzers.

Indications for Use

Indicated for use as an assayed quality control material in clinical laboratory settings to monitor the precision of Cholesterol, Triglycerides, HDL Cholesterol, and LDL Cholesterol assays on automated, semi-automated, or manual chemistry analyzers.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ AUG 2 2 2008 Image /page/0/Picture/1 description: The image shows the logo for Maine Standards Company. The logo is black and white and features the word "maine" on top of the word "standards". A diagonal line separates the two words. The words "COMPANY INC" are written in small letters below the word "standards". ## 510(k) SUMMARY "This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807 92." "The assigned 510(k) number is: __ Ko82067_" Submitter: Maine Standards Company Address: 765 Roosevelt Trail Windham, ME 04062 Telephone: 207-892-1300 Fax: 207-892-2266 Holly A. Cressman, Mgr. QA/RA Contact: Summary prepared on: July 10, 2008 Multi-Analyte Controls, All Kinds (Assayed and Unassayed) Device classification name: Quality control material (assayed and unassayed) Device description: Proprietary Name: MSC Lipid Control Regulation Number: 21 CFR 862.1660 Product Code: JJY Regulatory Class: Class I #### Predicate Device: Bio-Rad Liquichek™ Lipids Control Level 1 (K012513). Device description: The MSC Lipid Control is a human serum based liquid control containing stabilized Cholesterol, Triglycerides, HDL Cholesterol and LDL Cholesterol of human origin. Intended use: The MSC Lipid Control is intended for use as an assayed control material to monitor the ongoing precision of clinical laboratory analysis for Cholesterol, Triglycerides, HDL Cholesterol, and LDL Cholesterol. This material is intended for use on automated, semi-automated, and manual clinical chemistry analyzer systems. ### Summary: The information provided in this pre-market notification demonstrates that the performance of MSC Lipid Control is substantially equivalent in form and function to the predicate devices for its stated intended use. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows a logo with a circular border containing text, and a central graphic. The text around the border is in a circular arrangement. The central graphic features a stylized depiction of three curved shapes, resembling birds in flight or abstract waves. The logo appears to be a symbol for an organization or agency, possibly related to health or public service. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Maine Standards Co. c/o Ms. Holly Cressman 765 Roosevelt Trail Windham, ME 04062-5365 AUG 2 2 2008 Re: k082067 Trade Name: MSC Lipid Control Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I (reserved) Product Codes: JJY Dated: July 17, 2008 Received: July 22, 2008 Dear Ms. Cressman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {2}------------------------------------------------ #### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Jean M. Cooper, M.S., D.V.M. Jéan M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use K082067 510(k) Number: Device Name: MSC Lipid Control Indications For Use: The MSC Lipid Control is intended for use as an assayed control material to monitor the ongoing precision of clinical laboratory analysis for Cholesterol, Triglycerides, HDL Cholesterol, and LDL Cholesterol. This material is intended for use on automated, semi-automated, and manual clinical chemistry analyzer systems. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) signature Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K082067 Page 1 of /
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%