K Number
K082049

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-12-05

(140 days)

Product Code
Regulation Number
866.3310
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LIAISON® Anti-HAV assay is an in vitro chemiluminescent immunoassay intended for the qualitative detection of total antibodies to hepatitis A (anti-HAV) in human serum and sodium heparinized plasma samples using the automated LIAISON® Analyzer. The assay is indicated as an aid in the laboratory diagnosis of current or previous HAV Infections in conjunction with other serological and clinical information and to determine the presence of an antibody response to HAV in vaccine recipients.

This assay is not intended for screening blood or solid or soft tissue donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients. The user is responsible for establishing their own assay performance characteristics in these populations.

The LIAISON® Control Anti-HAV (negative and positive) is intended for use as assayed quality control samples to monitor the performance of the LIAISON® Anti-HAV assay.

Device Description

The method for qualitative determination of anti-HAV is a competitive sandwich chemiluminescence immunoassay (CLIA) based on neutralization. The assay uses magnetic particles (solid phase) coated with IgG antibodies to HAV (mouse monoclonal), and a mouse monoclonal anti- HAV antibody conjugate linked to an isoluminol derivative (isoluminol-antibody conjugate).

The first incubation step consists of adding the HAV antigen to calibrators, samples or controls, during which anti-HAV present in calibrators, samples or controls binds to a fixed and limited amount of HAV, thus forming an HAV-anti-HAV immune complex.

After this step the second incubation follows and it involves addition of magnetic microparticles and conjugate into the reaction module, during which the antibody conjugate and the solid-phase antibody compete with anti-HAV present in the specimen for HAV. This allows the conjugate to bind to the solid phase and to form a sandwich. If all HAV added is sequestered in an HAV-anti-HAV immune complex during the first incubation, no sandwich is formed during the second incubation. After the second incubation, the unbound material is removed with a wash cycle.

Subsequently, the starter reagents are added and a flash chemiluminescence reaction is thus induced. The light signal, and hence the amount of isoluminolantibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is inversely indicative of anti-HAV present in calibrators, samples or controls.

AI/ML Overview

The provided text describes the performance data for the DiaSorin LIAISON® Anti-HAV assay. However, it does not explicitly state specific acceptance criteria (e.g., "Positive agreement must be >= 95%"). Instead, it reports the observed performance (percent agreement and confidence intervals) which implies these results were deemed acceptable by the regulatory body for 510(k) clearance based on substantial equivalence to a predicate device.

Therefore, the table below will present the reported device performance, and the description will elaborate on the various studies conducted to demonstrate this performance.


Acceptance Criteria and Study Details for DiaSorin LIAISON® Anti-HAV Assay

1. Table of Acceptance Criteria and Reported Device Performance

As explicit quantitative acceptance criteria (e.g., "The positive agreement must be at least X%") are not stated in the provided document, the table will reflect the reported performance that was found to be sufficient for substantial equivalence. The document presents "Percent Agreement" values along with their 95% Confidence Intervals as the primary performance metrics. The underlying acceptance for these values is implicit through the 510(k) clearance process, which confirms the device is safe and effective as a predicate device.

Performance Metric CategoryReported Device Performance (Percent Agreement)95% Confidence Interval
Prospective Population
At Risk / HAV Testing (Positive)96.6%93.5% - 98.0%
At Risk / HAV Testing (Negative)99.0%97.9% - 99.6%
Pediatric Population (Positive)78.6%49.2% - 95.3%
Pediatric Population (Negative)96.8%92.0% - 99.1%
Retrospective Population
Current/Previous HAV Infection100.0%98.0% - 100%
Pediatric Current/Previous HAV Inf.100.0%98.0% - 100%

Note: The low positive agreement for the Pediatric Population (78.6%) with a wide confidence interval (49.2 – 95.3%) indicates a smaller number of positive samples in that specific cohort, but was still accepted for clearance.

2. Sample Sizes and Data Provenance

  • Prospective Studies:
    • HAV Testing and At Risk Populations: 739 samples.
      • 500 excess serum samples from individuals in the Northeastern U.S. sent for HAV testing.
      • 239 samples from individuals at risk for viral hepatitis (homosexual males, healthcare workers, commercial sex workers, drug users, prison inmates, dialysis patients, hemophiliacs).
    • Pediatric Population: 108 samples from children in the United States.
    • Vaccine Study: 73 individuals (9 TWINRIX sets, 32 HAVRIX sets, 32 VAQTA sets of pre- and post-vaccine samples).
  • Retrospective Studies:
    • Current/Previous HAV Infection (Adults): 109 samples from adults in Eastern U.S. and Egypt.
    • Current HAV Infection (Pediatric): 42 samples from pediatric patients in Egypt.
  • Expected Values (Prevalence Study): 802 apparently healthy adults (301 from Western U.S., 501 from Eastern U.S.).

Data Provenance: The data is a mix of prospective and retrospective collections. Geographically, samples were from the Northeastern U.S., other unspecified regions of the U.S. (for at-risk groups and pediatric prospective), Eastern U.S., Western U.S., and Egypt.

3. Number of Experts and Qualifications for Ground Truth

The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the test sets. Instead, the "Comparator ELISA" assay is consistently referred to as the reference method for determining the true positive, negative, or equivocal status of samples. This implies that the accepted clinical standard for Hepatitis A testing at the time (the predicate ELISA assay) served as the ground truth.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method involving multiple experts. The comparator ELISA assay results were the reference. For samples that yielded "Equivocal" or "Borderline" results by the comparator ELISA, these were sometimes retested per the Instructions for Use of the comparator method. Samples that remained equivocal or borderline after retesting were handled as distinct categories in the comparison tables.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not done. This device is an in vitro diagnostic (IVD) assay designed to be read by automated equipment (LIAISON® Analyzer) and interpreted by laboratory professionals, not primarily by human "readers" interpreting images or complex data in the same way an AI for medical imaging would. The performance is compared to a predicate assay, not human interpretation.

6. Standalone (Algorithm Only) Performance

Yes, a standalone performance study was done. The entire performance evaluation section ("Performance Data") describes the performance of the LIAISON® Anti-HAV assay as a standalone algorithm/device, comparing its results directly against those of a predicate ELISA assay on various sample populations. There is no human-in-the-loop component described for these performance evaluations.

7. Type of Ground Truth Used

The ground truth was established by a comparator ELISA assay, which is referred to throughout the document as the reference method. Specifically, the predicate device, DiaSorin Inc. ETI-AB-HAVK Plus assay (PMA #P890019/S05), served as the standard for comparison.

8. Sample Size for the Training Set

The document does not specify a separate "training set" size. For in vitro diagnostic (IVD) assays, the development process typically involves internal method validation and optimization by the manufacturer, rather than a distinct "training set" in the machine learning sense. The samples described in the "Performance Data" section are effectively the "test set" or clinical validation set used to demonstrate performance for regulatory submission.

9. How Ground Truth for the Training Set Was Established

As no explicit training set is mentioned in the context of machine learning, this question is not directly applicable. For the overall assay development and validation, the ground truth was established by comparison to existing, clinically accepted methods, presumably the predicate ELISA assay during earlier stages of development and optimization.

{0}------------------------------------------------

KOrzoy9

DiaSorin LIAISON® anti-HAV Premarket Notification

5.0 510(k) SUMMARY

DEC 0 5 2008

Carol A. DePouw Regulatory Affairs Specialist DiaSorin Inc. 1951 Northwestern Avenue P.O. Box 285 Stillwater, MN 55082-0285 Phone (651) 351-5850 Fax (651) 351-5669 Email: carol.depouw@diasorin.com

NAME OF DEVICE:

SUBMITTED BY:

Trade Name:

Common Names/Descriptions:

Classification Names:

Product Code:

Hepatitis A Test (Antibody and IgM Antibody)

Hepatitis A Virus (HAV Serological Reagents)

LOL. JJX

LIAISON® Anti-HAV

LIAISON® Control Anti-HAV

PREDICATE DEVICES

DiaSorin Inc. ETI-AB-HAVK Plus assay (PMA #P890019/S05)

DEVICE DESCRIPTION:

INTENDED USE: The LIAISON® Anti-HAV assay is an in vitro chemiluminescent immunoassay intended for the qualitative detection of total antibodies to hepatitis A (anti-HAV) in human serum and sodium heparinized plasma samples using the automated LIAISON® Analyzer. The assay is indicated as an aid in the laboratory diagnosis of current or previous HAV Infections in conjunction with other serological and clinical information and to determine the presence of an antibody response to HAV in vaccine recipients.

This assay is not intended for screening blood or solid or soft tissue donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients. The user is responsible for establishing their own assay performance characteristics in these populations.

The LIAISON® Control Anti-HAV (negative and positive) is intended for use as assayed quality control samples to monitor the performance of the LIAISON® Anti-HAV assay.

KIT DESCRIPTION: The method for qualitative determination of anti-HAV is a competitive sandwich chemiluminescence immunoassay (CLIA) based on neutralization. The assay uses magnetic particles (solid phase) coated with IgG antibodies to HAV (mouse monoclonal), and a mouse monoclonal anti- HAV

{1}------------------------------------------------

antibody conjugate linked to an isoluminol derivative (isoluminol-antibody conjugate).

The first incubation step consists of adding the HAV antigen to calibrators, samples or controls, during which anti-HAV present in calibrators, samples or controls binds to a fixed and limited amount of HAV, thus forming an HAV-anti-HAV immune complex.

After this step the second incubation follows and it involves addition of magnetic microparticles and conjugate into the reaction module, during which the antibody conjugate and the solid-phase antibody compete with anti-HAV present in the specimen for HAV. This allows the conjugate to bind to the solid phase and to form a sandwich. If all HAV added is sequestered in an HAV-anti-HAV immune complex during the first incubation, no sandwich is formed during the second incubation. After the second incubation, the unbound material is removed with a wash cycle.

Subsequently, the starter reagents are added and a flash chemiluminescence reaction is thus induced. The light signal, and hence the amount of isoluminolantibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is inversely indicative of anti-HAV present in calibrators, samples or controls.

PERFORMANCE DATA:

COMPARATIVE TESTING: Prospective and Retrospective studies were performed to evaluate the performance of the LIAISON® Anti-HAV assay among individuals who were sent to the lab for Hepatitis A testing and those at high risk for viral hepatitis.

The prospective study consisted of 739 samples from individuals who were sent to the lab for HAV testing or at risk for viral hepatitis, 108 samples from a pediatric population and 73 individuals who participated in a vaccine study. The retrospective study consisted of 109 samples from adults with a current or previous hepatitis A infection and 42 pediatric patients with current hepatitis A infection.

Prospective

HAV Testing and At Risk Populations

Of the 739 samples in this study 500 were excess serum samples from individuals in the Northeastern U.S. sent to the laboratory for HAV testing. In this group 59.8% were female (n=299) ranging in age from 20 - 101 yrs. and 40.2% were male (n=201) ranging in age from 17 to 89. The remaining 239 samples were from individuals at risk for viral hepatitis due to lifestyle, behaviour or occupation. This group consisted of the following: homosexual males (n=38), healthcare workers (n=10), commercial sex workers (n=34), druq users (n=77), prison inmates (n=49), dialysis patients (n=25) and hemophiliacs (n=6). Of these subjects 29.7% were females (n=71) ranging in age from 17 to 79, and 43.1% were males (n=103) ranging in age from 16 to 79. The age and gender were unknown for the remaining 27.2% (n=65).

Two samples were Equivocal by the Comparator ELISA assay after repeat testing per the Instructions for use.

{2}------------------------------------------------

The data for the combined populations are shown in Table 1.

Table 1: HAV testing population and At risk population comparison of LIAISON® Anti-
HAV and the Comparator ELISA
LIAISON®Anti-HAVComparator ELISATotal
PositiveBorderlineNegative
Positive23003233
Equivocal1**124*
Negative71494502
Total2382*499739
  • Repeat equivocal or borderline results

**sample not retested

Percent AgreementExact 95% Confidence Interval
Positive230/23996.6%93.5 - 98.0%
Negative494/49999.0%97.9 - 99.6%

Specimens that were Borderline with the comparison method and Negative with 피 the LIAISON® Anti-HAV assay were considered to be False Negative on the LIAISON® Anti-HAV assay.

Pediatric Population

One hundred eight (108) pediatric samples were prospectively collected tested from children in the United States. Of the 108 pediatric samples 57.4% were female (n=62) and 42.6% were ale (n=46), ranging in age from 2 to 17. The results are presented in the Table 2.

Table 2: Pediatric Population Comparison of LIAISON® Anti-HAV and Comparator ELISA

LIAISON®Anti-HAVComparator ELISATotal
PositiveEquivocalNegative
Positive110213
Equivocal1012*
Negative119193
Total131*94108
  • Repeat equivocal or borderline results
Percent AgreementExact 95% Confidence Interval
Positive11/1478.6%49.2 – 95.3%
Negative91/9496.8%92.0 – 99.1%

Section 5

{3}------------------------------------------------

  • Specimens that were Borderline with the companson method and Negative with I the LIAISON® Anti-HAV assay were considered to be False Negative on the LIAISON® Anti-HAV assay.

Retrospective

Current/Previous HAV Infection;

Of the 151 retrospective samples 109 were from adults with a current or previous hepatitis A infection collected in Eastern U.S. and Egypt. They consisted of 31.2% females (n=34) 35.8% males (n=39) ranging in age from 18 to 51. For 32.1% (n=35) of the samples gender and age were unknown. One sample (0.9%) was from an individual18 years of age and gender unknown. The other 42 samples were from pediatric patients with current hepatitis A infection collected in Eqypt. They consisted of 31% female (n=13) and 69% male (n=29), ranging in age from 4-17 years.

Table 3: "Current/Previous" HAV Infection Comparison of LIAISON® Anti-HAV and the Comparator ELISA

LIAISON®Anti-HAVComparator ELISATotal
PositiveEquivocalNegative
Positive10900109
Equivocal0000
Negative0000
Total10900109
Percent AgreementExact 95% Confidence Interval
Positive109/109100.0%98.0 – 100%

Table 4: "Current/Previous" HAV Infection Pediatric Population Comparison with the LIAISON® Anti-HAV and the Comparator ELISA

LIAISON®Anti-HAVComparator ELISATotal
PositiveEquivocalNegative
Positive420042
Equivocal0000
Negative0000
Total420042
Percent AgreementExact 95% Confidence Interval
Positive42/42100.0%98.0 – 100%

{4}------------------------------------------------

Vaccine study

The HAV antibody response to vaccination was evaluated with the three different vaccines currently licensed in the United States: TWINRIX® (Hepatitis A, Inactivated and Hepatitis B (recombinant)), HAVRIX® (Hepatitis A, Inactivated) both manufactured by GlaxoSmithKline Biologicals and VAQTA® (Hepatitis A, Inactivated) manufactured by MERCK & CO., INC.

For TWINRIX vaccine. 9 matched sets of pre- and post- vaccine samples were available.

For HAVRIX® vaccine, 32 matched sets of pre- and post- vaccine samples were available.

For VAQTA® vaccine, 32 matched sets of pre- and post- vaccine samples were available.

The data are shown in the table below.

Table 5: Comparison of Vaccine samples on the LIAISON® Anti-HAV and Comparator assay

ComparatorLIAISON® Anti-HAV
PosEqvNegPosEqvNeg
Prevaccine009009
TWINRIXPost 2nd dose900900
Post 3rd dose900900
HAVRIXPre-vaccine00320032
4wk Post vaccine31012912
VAQTAPre-vaccine00320032
4 wk Post vaccine31013101

The LIAISON® Anti-HAV demonstrated agreement with the Comparator ELISA as follows:

Prospective Population

"At Risk" and "HAV Testing" Positive agreement - 96.6% (95% Cl = 93.5-98.0%) Negative agreement - 99.0% (95% CI = 97.9 - 99.6%)

"Pediatric Population"

Positive agreement - 78.6% (95% Cl = 49.2-95.3%) Negative agreement - 96.8% (95% Cl = 92.0 - 99.1%)

Retrospective Population

"Adult and Pediatric Current/Previous HAV Infection" Positive agreement - 100% (95% Cl = 98.0 - 100%)

{5}------------------------------------------------

The results demonstrate that the LIAISON® Anti-HAV assay can be used with the LIAISON® Analyzer for the qualitative detection of total antibodies to hepatitis A.

EXPECTED VALUES:

Prevalence

The expected prevalence results of the LIAISON® Anti-HAV assay were determined in 802 apparently healthy adults from the Western and the Eastern regions of the U.S. Three hundred one (301) samples were from the Western U.S. and 501 were samples from the Eastern U.S.

Of the Western U.S. individuals 53.8% were Females (n=162) ranging in age from 9 to 87 and 46.2% were Males (n=139) ranging in age from 16 to 76. The majority of the individuals were Caucasian (60.8%), with other ethnic groups represented as follows: Hispanic (17.6%), African Americans (15.3%), Asian (6.0%) and Middle Eastern (0.3%). In the study group from the Western region, 26.3% of the individuals were found to be positive for anti-HAV antibodies.

Of the Eastern U.S. individuals 46.5% were Females (n=233) ranging in age from 17 to 83, and 53.5% were Males (n=268) ranging in age from 17 to 82. The majority of the individuals were Caucasian (69.9%), with other ethnic groups represented as follows: Hispanic (14.0%), African American (12.1%) and Asian (4.0%). In the study group from the Eastern region 20% of the individuals were found to be positive for Anti-HAV antibodies.

The Expected results for the Western and Eastern regions of the U.S. are presented in the tables below.

One sample from each of the regions gave an Equivocal result after repeat testing per Instructions for use.

NNegativeEquivocalPositivePositivePrevalence
Total30122117926.3%
Gender
Female16211504729.0%
Male13910613223.0%
Age range (years)N(-)(Eqv)(+)
≤181290325.0%
<101001
10 - 1915100533.3%
20 - 29815602530.9%
30 - 39684702130.9%
40 - 4952421917.3%
50 - 5948410714.6%
60 - 6931220929.0%
≥ 70530240.0%

Expected results for the LIAISON® Anti-HAV assay from the Western U.S. (High Prevalence - n=301)

Section 5

{6}------------------------------------------------

NNegativeEquivocalPositivePositivePrevalence
Total501400110020.0%
Gender
Female23318814418.9%
Male26821205620.9%
Age range (years)N(-)(Eqv)(+)
≤18460000.0%
<100NANANANA
10 - 194949000.0%
20 - 2939330615.4%
30 - 39786311418.0%
40 - 491079501211.2%
50 - 5914210104128.9%
60 - 69523701528.9%
≥ 70342201235.3%

Expected results for the LIAISON® Anti-HAV assay from the Eastern U.S. (Low Prevalence n=501)

REPRODUCIBILITY: A 5 day reproducibility/precision study was conducted at three external laboratories. The CLSI document EP15-A2 was consulted in the preparation of the testing protocol.

A coded panel comprised of 12 frozen "engineered" serum samples was prepared by DiaSorin S.p.A. and provided to the sites. The coded panel samples were prepared by spiking positive samples into negative samples to achieve high negative, low positive and high positive results. The two negative panel samples were not spiked. The LIAISON® Control Anti-HAV set was also included in the 5 day study.

The coded panel was tested at all three sites, using four replicates per run in one run per day during five operating days. The mean Index value, standard deviation, and coefficient of variation (%CV) of the results were computed for each of the tested specimens for each of the sites.

Results .

The 5 day Index results are summarized in Table 6 (combined sites). The mean Index value, standard deviation, and coefficient of variation (%CV) of the results were computed for each of the tested specimens for each of the sites.

The result of the LIAISON® Anti-HAV Negative Control read above the reportable range of the assay >>2.7 therefore. SD and %CVs were not able to be calculated and are shown as (NA).

{7}------------------------------------------------

Site 1Site 2Site 3Overall (3 site)
TotalTotalTotalTotal
Sample IDMean Indexsd%CVMean Indexsd%CVMean Indexsd%CVMean Indexsd%CV
NC>>2.7NANA>>2.7NANA>>2.7NANA>>2.7NANA
PC0.420.037.20.360.037.30.200.0710.80.330.1030.0
HAVu-e11.130.108.90.920.088.90.990.043.91.010.1211.5
HAVu-e21.140.1311.20.930.088.40.980.033.31.010.1212.3
HAVu-e31.170.119.40.950.088.50.970.055.21.030.1312.5
HAVu-e41.210.1512.41.000.109.81.030.044.11.080.1413.2
HAVu-n11.470.1611.21.290.1411.21.200.064.91.330.1712.4
HAVu-n21.330.2116.11.070.087.61.160.064.91.190.1714.5
HAVu-p10.270.0312.40.230.028.60.290.027.40.260.0414.5
HAVu-p20.160.0211.60.100.0545.50.160.029.50.140.0428.1
HAVu-p30.840.1012.00.670.0812.20.670.034.30.730.1115.0
HAVu-p40.890.099.90.720.079.40.760.033.60.790.1012.5
HAVu-p51.000.088.30.800.1012.00.840.033.70.880.1112.8
HAVu-p60.840.0910.80.670.0811.70.720.034.40.740.1113.4

Table 6: Combined Sites

CROSS-REACTIVITY: The cross-reactivity study for the LIAISON® Anti-HAV assay was designed to evaluate potential interference from other viruses that may cause symptoms similar to HAV infection (EBV, CMV, Rubella, Measles, Mumps, HBV, HCV), other organisms that may cause infectious disease (VZV, HSV, HIV, Toxoplasma gondii) and from other conditions that may result from atypical immune system activity (i.e. rheumatoid factor, RF, antinuclear autoantibodies, ANA).

Organism/ConditionNComparatorAnti-HAVAssayLIAISON®Anti-HAVPositiveLIAISON®Anti-HAVNegativeLIAISON®Anti-HAVEquivocal
Anti-VZV (IgG)22Negative0220
Anti-VZV (IgM)3Negative030
Anti-EBV VCA IgG21Negative0210
Anti-EBV EA IgG14Negative0140
Anti-Toxoplasma IgG7Negative070
Anti-Toxoplasma IgM10Negative0100
Anti-HBs11Negative0110
Anti-HBc IgM13Negative0130
Anti-HBc25Negative0250
Anti-HBe3Negative030
HBeAg23Negative0230
HBsAg27Negative0270
Anti-CMV IgG14Negative0140
Anti-Rubella IgG/IgM29Negative0290
Anti-HSV 1 / 2 IgG25Negative0241
Anti-HSV 2 IgG19Negative0190

Section 5

Page 5 - 8

{8}------------------------------------------------

DiaSorin LIAISON® anti-HAV Premarket Notification

Organism/ConditionNComparatorAnti-HAVAssayLIAISON®Anti-HAVPositiveLIAISON®Anti-HAVNegativeLIAISON®Anti-HAVEquivocal
Anti-HIV 1 / 24Negative400
Anti-HCV5Negative050
Anti-Mumps IgG8Negative080
Anti-Measles IgG3Negative030
Anti-Measles IgM11Negative0110
RF+5Negative221
ANA24Negative2193
ENA3Negative021
Nucleotides1Negative010
y-globulin6Negative060
HAMA10Negative0100
Total34683326

WARNING: Assay interference due to circulating antibodies against HIV 1/2, Antinuclear autoantibodies and rheumatoid factor may occur.

POTENTIALLY INTERFERING SUBSTANCES: Twelve samples at different anti-HAV levels (2 positive, 4 borderline, 4 equivocal, 2 high negative) were prepared with and without the potentially interfering substances hemoglobin, triglycerides, billirubin and albumin. All samples were tested with the LIAISON® anti-HAV assay and results were compared using a paired t-test. The testing showed that the assay performance was not affected by hemolysis at ≤1000 mg/dL hemoglobin (2% change, 100% concordance of results with and without interferent), lipemia at ≤3000 mg/dL triglycerides (7% change, 92% (11/12) concordance of results with and without interferent), icterus at ≤5 mg/dL bilirubin (0% change, 100% concordance of results with and without interferent) and serum albumin at ≤5 g/dL (3% change, 92% (11/12) concordance of results with and without interferent).

{9}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 0 5 2008

Ms. Carol A. DePouw Regulatory Affairs Specialist DiaSorin Inc. 1951 Northwestern Avenue P.O. Box 285 Stillwater, MN 55082-0285

Re: K082049

Trade/Device Name: LIAISON® Anti-HAV and LIAISON® Control Anti-HAV Regulation Number: 21 CFR 866.3310 Regulation Name: Hepatitis A Virus Serological Reagents Regulatory Class: Class II Product Code: LOL, JJX Dated: October 24, 2008 Received: October 27, 2008

Dear Ms. DePouw:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{10}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)). please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sally attaym

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

510(k) Number (if known): K082049

Device Name:

Indication For Use:

LIAISON® Anti-HAV and LIAISON® Control Anti-HAV

The LIAISON® Anti-HAV assay is an in vitro chemiluminescent immunoassay intended for the qualitative detection of total antibodies to hepatitis A (anti-HAV) in human serum and sodium heparinized plasma samples using the automated LIAISON® Analyzer.

The assay is indicated as an aid in the laboratory diagnosis of current or previous HAV infections in conjunction with other serological and clinical information and to determine the presence of an antibody response to HAV in vaccine recipients.

This assay is not intended for screening blood or solid or soft tissue donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients. The user is responsible for establishing their own assay performance characteristics in these populations.

The LIAISON® Control Anti-HAV (negative and positive) is intended for use as assayed quality control samples to monitor the performance of the LIAISON® Anti-HAV assay.

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Uwe Schek

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K082049

§ 866.3310 Hepatitis A virus (HAV) serological assays.

(a)
Identification. HAV serological assays are devices that consist of antigens and antisera for the detection of hepatitis A virus-specific IgM, IgG, or total antibodies (IgM and IgG), in human serum or plasma. These devices are used for testing specimens from individuals who have signs and symptoms consistent with acute hepatitis to determine if an individual has been previously infected with HAV, or as an aid to identify HAV-susceptible individuals. The detection of these antibodies aids in the clinical laboratory diagnosis of an acute or past infection by HAV in conjunction with other clinical laboratory findings. These devices are not intended for screening blood or solid or soft tissue donors.(b)
Classification. Class II (special controls). The special control is “Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Hepatitis A Virus Serological Assays.” See § 866.1(e) for the availability of this guidance document.