K Number
K081992

Validate with FDA (Live)

Date Cleared
2008-08-05

(22 days)

Product Code
Regulation Number
862.1235
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CSAE method is an in vitro diagnostic test for the quantitative measurement of cyclosporine A (CSA) in human whole blood on the Dimension Vista® system. Measurements of CSA are used as an aid in the management of heart, liver and kidney transplant patients.

Device Description

The Dimension Vista® CSAE Flex® reagent cartridge is a pre-packaged in-vitro diagnostic test method (assay) that is specifically designed to be used on the Dimension Vista® Integrated system, a floor model, fully automated microprocessor-controlled, integrated instrument system. The Dimension Vista® system was previously cleared with seven associated test methods (K051087). This Special 510(k) is submitted for a packaging modification to the Dimension® Cyclosporine Extended Range (CSAE) Flex® reagent cartridge (K052017), an in-vitro diagnostic device that has been cleared under the 510(k) process for use on Dimension® clinical chemistry systems. The packaging change is to allow use on the Dimension Vista® system. The modifications also include a change in method parameters (sample size and reagent volume) but the final concentration of sample/reagent ratio in test milieu remains the same.

The reagents contained in the Dimension Vista® CSAE Flex® reagent cartridges are the same as those contained in the Dimension® CSAE Flex® reagent cartridges manyfactured for the Dimension® clinical chemistry systems, another family of Siemens analyzers. The packaging modification does not affect the intended use of the devices, nor does it alter the fundamental scientific technology of the device.

The CSAE method uses an immunoassay technique in which free and CSA bound antibody-enzyme species are separated using magnetic particles. The CSAE Flex® reagent cartridge contains a pretreatment reagent, B-galactosidase-CSA antibody conjugate, CSA immobilized on chromium dioxide particles, chlorophenol red B-dgalactopyranoside (CPRG) substrate, and diluent to hydrate the tablets. To perform the CSAE assay, a sample cup containing the whole blood sample to be analyzed and a CSAE Flex® reagent cartridge are placed appropriately on the Dimension Vista® system. The Dimension Vista® system mixes and lyses the whole blood sample. The lysed sample is then mixed with the antibody conjugate reagent. The CSA present in the sample is bound by the CSA antibody conjugate reagent. Magnetic particles coated with cyclosporine A are added to bind free (unbound) antibody-enzyme conjugate. The reaction mixture is then separated magnetically. Following separation, the supernatant containing the CSA antibody-enzyme complex is transferred to a cuvette and mixed with the substrate. B-galactosidase catalyzes the hydrolysis of CPRG (chlorophenol red B-dgalactopyranoside) to produce CPR (chlorophenol red) that absorbs light maximally at 577mm. The change in absorbance at 577nm due to the formation of CPR is directly proportional to the amount of CSA in the patient's sample and is measured using a bichromatic (577, 700nm) rate technique.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Dimension Vista® CSAE Flex® reagent cartridge, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document defines acceptance criteria implicitly through the demonstration of "substantial equivalence" to a predicate device. The primary performance metric presented is the method comparison study results.

Acceptance Criteria CategorySpecific Acceptance Criterion (Implicit)Reported Device Performance (Dimension Vista® CSAE Flex®)
Method ComparisonSlope of linear regression (vs. predicate device) close to 1.00.92
Intercept of linear regression (vs. predicate device) close to 0 ng/mL37.9 ng/mL
Correlation coefficient (r) close to 1.00.990
Intended UseMust match predicate device's intended useMatches predicate: Quantitative measurement of CSA in human whole blood on the Dimension Vista® system; aid in management of heart, liver, and kidney transplant patients.
Technological CharacteristicsMust demonstrate substantially equivalent characteristics to predicate (e.g., reagent form, sample type, range)Various characteristics shown to be substantially equivalent or improved (e.g., number of tests per cartridge).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 116 transplant samples.
  • Data Provenance: Not explicitly stated, but the samples were human "transplant samples." No information on country of origin or whether they were retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Not Applicable. This study is a method comparison between two in-vitro diagnostic devices, not a study evaluating human interpretation against a ground truth established by experts. The "ground truth" for comparison is the measurement obtained from the predicate device.

4. Adjudication Method for the Test Set

  • None. This was a quantitative method comparison study between two instruments. Adjudication is typically used when human interpretation or subjective assessments are involved.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No. This study is a technical comparison of an in-vitro diagnostic assay on a new instrument platform against an established assay on an older platform. It does not involve human readers or assess the effectiveness of AI assistance.

6. Standalone (Algorithm Only) Performance Study

  • Yes, in essence. The entire evaluation is a standalone performance assessment of the Dimension Vista® CSAE Flex® reagent cartridge's ability to quantitatively measure Cyclosporine A. There is no human-in-the-loop component for the measurement process itself, as it's an automated in-vitro diagnostic method. The algorithm here refers to the instrument's measurement and calculation process.

7. Type of Ground Truth Used

  • Comparison to a Legally Marketed Predicate Device. The "ground truth" for this study is the analytical measurement obtained from the Dimension® CSAE Flex® reagent cartridge (K052017), which is the legally marketed predicate device. The study aims to show that the new device's measurements are substantially equivalent to the predicate.

8. Sample Size for the Training Set

  • Not applicable / Not explicitly stated. This document describes a Special 510(k) for a packaging and platform modification, not the development of a novel algorithm that requires a "training set" in the machine learning sense. The assay chemistry and fundamental technology are consistent with the predicate device. If there was any internal development or optimization, the details of a "training set" are not provided in this regulatory summary.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As a Special 510(k) for a modified existing device, there isn't a "training set" in the context of a new algorithm development. The core methodology and reagents are the same as the predicate device, which would have undergone its own qualification and validation studies when it was initially cleared.

{0}------------------------------------------------

Dimension Vista® CSAE Flex® reagent cartridge Special 510(k) Premarket Notification

510(k) Summary of Safety and Effectiveness Dimension Vista® CSAE Flex® reagent cartridge

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

AUG - 5 2008

The assigned 510(k) number is: KO81992

1. Submitter's Contact Information and Date of Preparation

Siemens Healthcare Diagnostics Inc. Submitter's Contact Information: P.O. Box 6101 Newark, DE 19714-6601 Attn: Yuk-Ting Lewis Tel: 302-631-7626

Date of Preparation:

July 7, 2008

  • Proprietary Device Name / FDA Classification Name 2. Dimension Vista® CSAE Flex® reagent cartridge / Cyclosporine test system

Identification of the Predicate Device 3. Dimension® CSAE Flex® reagent cartridge, K052017

4. Device Description

The Dimension Vista® CSAE Flex® reagent cartridge is a pre-packaged in-vitro diagnostic test method (assay) that is specifically designed to be used on the Dimension Vista® Integrated system, a floor model, fully automated microprocessor-controlled, integrated instrument system. The Dimension Vista® system was previously cleared with seven associated test methods (K051087). This Special 510(k) is submitted for a packaging modification to the Dimension® Cyclosporine Extended Range (CSAE) Flex® reagent cartridge (K052017), an in-vitro diagnostic device that has been cleared under the 510(k) process for use on Dimension® clinical chemistry systems. The packaging change is to allow use on the Dimension Vista® system. The modifications also include a change in method parameters (sample size and reagent volume) but the final concentration of sample/reagent ratio in test milieu remains the same.

The reagents contained in the Dimension Vista® CSAE Flex® reagent cartridges are the same as those contained in the Dimension® CSAE Flex® reagent cartridges manyfactured for the Dimension® clinical chemistry systems, another family of Siemens analyzers. The packaging modification does not affect the intended use of the devices, nor does it alter the fundamental scientific technology of the device.

{1}------------------------------------------------

The CSAE method uses an immunoassay technique in which free and CSA bound antibody-enzyme species are separated using magnetic particles. The CSAE Flex® reagent cartridge contains a pretreatment reagent, B-galactosidase-CSA antibody conjugate, CSA immobilized on chromium dioxide particles, chlorophenol red B-dgalactopyranoside (CPRG) substrate, and diluent to hydrate the tablets. To perform the CSAE assay, a sample cup containing the whole blood sample to be analyzed and a CSAE Flex® reagent cartridge are placed appropriately on the Dimension Vista® system. The Dimension Vista® system mixes and lyses the whole blood sample. The lysed sample is then mixed with the antibody conjugate reagent. The CSA present in the sample is bound by the CSA antibody conjugate reagent. Magnetic particles coated with cyclosporine A are added to bind free (unbound) antibody-enzyme conjugate. The reaction mixture is then separated magnetically. Following separation, the supernatant containing the CSA antibody-enzyme complex is transferred to a cuvette and mixed with the substrate. B-galactosidase catalyzes the hydrolysis of CPRG (chlorophenol red B-dgalactopyranoside) to produce CPR (chlorophenol red) that absorbs light maximally at 577mm. The change in absorbance at 577nm due to the formation of CPR is directly proportional to the amount of CSA in the patient's sample and is measured using a bichromatic (577, 700nm) rate technique.

5. Device Intended Use

The CSAE method is an in vitro diagnostic test for the quantitative measurement of cyclosporine A (CSA) in human whole blood on the Dimension Vista® system. Measurements of CSA are used as an aid in the management of heart, liver and kidney transplant patients.

Summary of the devices technological characteristics 6.

The Dimension Vista® CSAE Flex is substantially equivalent to other cyclosporine test systems, such as the Dimension® CSAE Flex® reagent cartridge (K052017). A comparison of features is provided.

FeatureModified Device:Dimension Vista® CSAEFlex® reagent cartridgePredicate Device:Dimension® CSAE Flex®reagent cartridge
ReagentsPrepackaged, 12 well plastic Flex®reagent cartridgesPrepackaged, 8 well plastic Flex®reagent cartridges
Intended UseThe CSAE method is an in vitrodiagnostic test for the quantitativemeasurement of cyclosporine A(CSA) in human whole blood onthe Dimension Vista® system.Measurements of CSA are used asan aid in the management of heart,liver and kidney transplant patients.The CSAE method is an in vitrodiagnostic test for the quantitativemeasurement of cyclosporine A(CSA) in human whole blood onthe Dimension® clinicalchemistry system. Measurementsof CSA are used as an aid in themanagement of heart, liver andkidney transplant patients.

{2}------------------------------------------------

Dimension Vista® CSAE Flex® reagent cartridge Special 510(k) Premarket Notification

FeatureModified Device:Dimension Vista® CSAEFlex® reagent cartridgePredicate Device:Dimension® CSAE Flex®reagent cartridge
Final concentrationof sample/reagentratio in test milieuSame concentration as thepredicate deviceAs described in K052017
Reagent formSame liquids and tabletsLiquids and tablets
# of chrome tablets23
Total # of tests incach Flex®cartridge3020
Calibration30 days30 days
Sample TypeWhole bloodWhole blood
AnalyticalMeasurementRange350 – 2000 ng/mL350 - 2000 ng/mL
Sample Size1.12 µL3 µL
MeasurementBichromatic endpoint@ 577 and 700 nmBichromatic endpoint@ 577 and 700 nm

7. Method comparison to the predicate

A split sample method comparison study was conducted on the Dimension Vista® CSAE Flex vs, the Dimension® CSAE Flex using one hundred and sixteen (116) transplant samples. The data was analyzed using least squares linear regression and produced the following regression statistics:

Slope0.92
Intercept37.9 ng/mL
r0.990
n116

8. Conclusion

The proposed Dimension Vista® CSAE method is substantially equivalent to the legally marketed device, the Dimension® CSAE method in design, principle, and performance. They have the same intended use and indications for use. Comparative testing also demonstrates substantially equivalent performance.

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services, USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" around the top half of the circle. In the center of the seal is a stylized image of an eagle.

SEP 2-2008

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Siemens Healthcare Diagnostics Inc. c/o Ms. Yuk-Ting Lewis Regulatory Affairs and Compliance Manager P.O. Box 6101, Mailbox 514 Newark, DE 19714-6101

Re: K081992

Trade/Device Name: Dimension Vista CSAE Flex Regent Cartridge, Model k4108 Regulation Number: 21 CFR 862.1235 Regulation Name: Cyclosporine Test System. Regulatory Class: Class II Product Code: MKW Dated: July 11, 2008 Received: July 14, 2008

Dear Ms. Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR). Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2 --

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jean M. Cooper, M.S., D.V.M.

Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indication for Use

510(k) Number (if known):

Device Name: Dimension Vista® CSAE Flex® reagent cartridge

2081992

Indications For Use:

The CSAE method is an in vitro diagnostic test for the quantitative measurement of cyclosporine A (CSA) in human whole blood on the Dimension Vista® system. Measurements of CSA are used as an aid in the management of heart, liver and kidney transplant patients.

Prescription Use ______x (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Off

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K081992

§ 862.1235 Cyclosporine test system.

(a)
Identification. A cyclosporine test system is a device intended to quantitatively determine cyclosporine concentrations as an aid in the management of transplant patients receiving therapy with this drug. This generic type of device includes immunoassays and chromatographic assays for cyclosporine.(b)
Classification. Class II (special controls). The special control is “Class II Special Controls Guidance Document: Cyclosporine and Tacrolimus Assays; Guidance for Industry and FDA.” See § 862.1(d) for the availability of this guidance document.