K Number
K081928

Validate with FDA (Live)

Date Cleared
2008-12-23

(169 days)

Product Code
Regulation Number
866.3480
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diagnostic Hybrids, Inc. device, D3 Duet DFA RSV/Respiratory Virus Screening Kit, is intended for the qualitative detection and identification of respiratory syncytial virus, while screening for influenza A virus, influenza B virus, adenovirus, and parainfluenza virus types 1, 2 and 3 viral antigens, in nasal and nasopharyngeal swabs and aspirates or in cell culture. The assay detects viral antigens by immunofluorescence using monoclonal antibodies (MAbs), from patients with signs and symptoms of respiratory infection.

It is recommended that specimens found to be negative after examination of the direct specimen result be confirmed by cell culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions.

Performance characteristics for influenza A virus detection and identification were established when influenza A (H3N2) and influenza A (H1N1) were the predominant influenza A strains circulating in the United States. Performance characteristics for influenza A virus detection and identification were established when influenza A H3N2 and influenza A H1N1 were the predominant influenza A strains circulating in the United States. When other Influenza A viruses are emerging, performance characteristics may vary. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to a state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Description

The Diagnostic Hybrids, Inc. device, D3 Duet DFA RSV/Respiratory Virus Screening Kit, uses a blend of viral antigen-specific murine MAbs. MAbs for RSV are directly labeled with R-phycoerythrin (R-PE) for the rapid detection and identification of RSV. MAbs for influenza A virus, influenza B virus, adenovirus, and parainfluenza virus types 1, 2, and 3 are directly labeled with fluorescein isothiocyanate (FITC), for rapid detection of these agents.

Kit components:

  • D3 Duet DFA RSV/Respiratory Virus Screening Reagent
  • Normal Mouse Gamma Globulin DFA Reagent
  • Respiratory Virus Antigen Control Slides
  • Wash Solution Concentrate
  • Mounting Fluid

The cells to be tested, derived from a clinical specimen or cell culture, are placed onto a glass slide and allowed to air dry. The cells are fixed in acetone. The D3 Duet DFA RSV/Respiratory Virus Screening Reagent is added to the cells which are then incubated for 15 to 30 minutes at 35° to 37°C in a humidified chamber or humidified incubator. The stained cells are then washed with the diluted wash solution, a drop of the supplied Mounting Fluid is added and a coverslip is placed on the prepared cells. The cells are examined using a fluorescence microscope. The respiratory syncytial virus infected cells will fluoresce golden-yellow, while cells infected with any of the other six viruses will fluoresce apple-green. Uninfected cells will contain no fluorescence but will be stained red by the Evans Blue counter-stain. If only golden-yellow fluorescent cells are present the specimen can be reported as positive for respiratory syncytial virus antigen. If only apple-green fluorescent cells are present, the particular virus is identified using the individual reagents from the D3 Ultra DFA Respiratory Virus Screening & ID Kit (D3 Ultra) on new, separate cell preparations. If both golden-yellow and apple-green are present, the additional virus may be identified using the individual reagents from the D3 Ultra on new, separate cell preparations.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the D' DUET DFA RSV/RESPIRATORY VIRUS SCREENING KIT:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as numerical targets for sensitivity and specificity in the provided text. Instead, the study uses the legally marketed predicate device (D3 Ultra DFA Respiratory Virus Screening & ID Kit) as the comparator, and "100% agreement" is repeatedly cited as the performance goal versus this comparator. The device met this implicit acceptance criterion by demonstrating 100% agreement in both positive and negative percentage agreements across various virus types and specimen sources.

Virus Target(s)Specimen TypePerformance MetricAcceptance Criteria (Implicit)Reported Device Performance95% Confidence Interval
RSVDirect (Nasal/NP)PPA (Positive Percent Agreement)100%100% (300/300)97.8% - 100%
Direct (Nasal/NP)NPA (Negative Percent Agreement)100%100% (887/887)99.6% - 100%
IAFVBAPF123*Direct (Nasal/NP)PPA (Positive Percent Agreement)100%100% (186/186)98.0% - 100%
IAFVBAPF123*Direct (Nasal/NP)NPA (Negative Percent Agreement)100%100% (1001/1001)99.6% - 100%
RSVCultured SpecimensPPA (Positive Percent Agreement)100%100% (33/33)89.5% - 100%
Cultured SpecimensNPA (Negative Percent Agreement)100%100% (265/265)98.6% - 100%
IAFVBAPF123*Cultured SpecimensPPA (Positive Percent Agreement)100%100% (104/104)96.4% - 100%
IAFVBAPF123*Cultured SpecimensNPA (Negative Percent Agreement)100%100% (194/194)98.1% - 100%

*IAFVBAPF123 refers to Influenza A virus, Influenza B virus, Adenovirus, and Parainfluenza virus types 1, 2, and 3.

2. Sample Sizes and Data Provenance

  • Direct Fresh Specimens (Test Set):
    • Total Initial Specimens: 1203
    • Excluded Specimens: 17 (due to site deviations, duplicate specimen, insufficient cell numbers, or high background)
    • Analyzed Specimens: 1187
    • Data Provenance: Prospective, collected at three study sites in the United States.
  • Cultured Specimens (Test Set):
    • Total Specimens: 298 frozen specimens
    • Data Provenance: Retrospective (frozen specimens), from a fourth study site. All specimens were derived from nasopharyngeal specimens. The age distribution of individuals is provided on page 13.

3. Number of Experts and Qualifications

The document does not explicitly state the number of experts used to establish ground truth for the test set or their specific qualifications (e.g., "Radiologist with 10 years of experience"). However, it mentions that the D3 Duet device was compared against a "cleared DSFA device" (D3 Ultra DFA Respiratory Virus Screening & ID Kit) as the comparator. This implies that the ground truth for the clinical studies was established by laboratory personnel using the predicate device, which is a standard immunofluorescence assay.

4. Adjudication Method

The document does not specify an adjudication method like 2+1 or 3+1. The results are presented as direct comparisons between the D3 Duet and the "D3 Ultra Final Identification," implying that the D3 Ultra's results served as the reference standard without further expert adjudication beyond its own established interpretation protocol.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No multi-reader multi-case (MRMC) comparative effectiveness study was done. This device is a diagnostic kit read by a single trained technician, not an AI system designed to assist human readers. Therefore, the effect size of human readers improving with AI vs. without AI assistance is not applicable.

6. Standalone Performance

Yes, a standalone performance study (algorithm only without human-in-the-loop performance) was done for the D3 Duet. The entire clinical performance section (pages 11-14) reports the performance of the D3 Duet device independently against the predicate device (D3 Ultra) as the gold standard.

7. Type of Ground Truth Used

The ground truth for the clinical studies was established using a cleared comparator device, specifically the D3 Ultra DFA Respiratory Virus Screening & ID Kit. This is a form of reference standard testing using an established laboratory method (immunofluorescence assay) for viral antigen detection. For the cultured specimens, cell culture methods were also employed.

8. Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of an AI/algorithm. Instead, the device is a diagnostic kit based on monoclonal antibodies. Therefore, there is no sample size for an algorithm training set. The various analytical and non-clinical performance studies (precision, detection limit, analytical reactivity, analytical specificity) act as development and validation steps for the kit itself, rather than training for an AI.

9. How Ground Truth for the Training Set Was Established

As there is no "training set" for an AI/algorithm, this point is not applicable. The development and validation of the D3 Duet kit relied on established laboratory techniques for virus isolation, identification using cell cultures, reference viral strains, and bacterial/fungal cultures. These methods served as the "ground truth" during the development and analytical validation of the antibody blend and assay performance.

{0}------------------------------------------------

D' DUET DFA RSV/RESPIRATORY VIRUS SCREENING KIT

SECTION 05, 510(K) SUMMARY

Applicant:

DIAGNOSTIC HYBRIDS, INC. 1055 East State Street Suite 100 Athens, OHIO 45701

Contact Information:

Gail R. Goodrum Vice President, Regulatory Affairs E-mail: goodrum(@dhiusa.com Telephone: 740-589-3300 Desk Extension: 740-589-3380 FAX: 740-593-8437

Date of preparation of 510(k) summary:

July 01, 2008

Device Name:

Trade name - D Duet DFA RSV/Respiratory Virus Screening Kit Common name - Fluorescent antibody test for detecting respiratory syncytial virus. and screening for other respiratory viruses Classification name - Respiratory Syncytial Virus, Antibody, Ifa Product Code - GNW Regulation - 21 CFR 866.3480, Class I, respiratory syncytial virus serological reagents; Panel Microbiology (83)

Legally marketed device to which equivalence is claimed:

K061101, D3 Ultra DFA Respiratory Virus Screening & ID Kit

Intended Use: The Diagnostic Hybrids, Inc. device, D3 Duet DFA RSV/Respiratory Virus Screening Kit, is intended for the qualitative detection and identification of respiratory syncytial virus, while screening for influenza A virus, influenza B virus, adenovirus, and parainfluenza virus types 1, 2 and 3 viral antigens, in nasal and nasopharyngeal swabs and aspirates or in cell culture. The assay detects viral antigens by immunofluorescence using monoclonal antibodies (MAbs), from patients with signs and symptoms of respiratory infection

Image /page/0/Picture/13 description: The image shows the number K081928 at the top. Below that is the logo for Diagnostic Hybrids. The logo includes the company name in bold text, followed by the tagline "Integrating Science and Humanity". To the right of the text is an image of a person with their arms raised, next to a DNA strand.

DEC 2 3 2008

{1}------------------------------------------------

It is recommended that specimens found to be negative after examination of the direct specimen result be confirmed by cell culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions.

Performance characteristics for influenza A virus detection and identification were established when influenza A (H3N2) and influenza A (H1N1) were the predominant influenza A strains circulating in the United States. When other Influenza A viruses are emerging, performance characteristics may vary. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to a state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens

Device Description:

The MAbs furnished with the D' Duet DFA RSV/Respiratory Virus Screening Kit (D3 Duet) are the same MAbs that are contained in the Diagnostic Hybrids, Inc. (DHI) device D' Ultra DFA Respiratory Virus Screening & ID Kit (D' Ultra) [510(k) number K061101. November 20, 2006].

The Diagnostic Hybrids, Inc. device, D3 Duet DFA RSV/Respiratory Virus Screening Kit, uses a blend of viral antigen-specific murine MAbs. MAbs for RSV are directly labeled with R-phycoerythrin (R-PE) for the rapid detection and identification of RSV. MAbs for influenza A virus, influenza B virus, adenovirus, and parainfluenza virus types 1, 2, and 3 are directly labeled with fluorescein isothiocyanate (FITC), for rapid detection of these agents.

Kit components:

  • . D3 Duet DFA RSV/Respiratory Virus Screening Reagent R-phycoerythrin-labeled murine MAbs directed against influenza A virus and a mixture of fluoresceinlabeled murine MAbs directed against influenza A. influenza B. adenovirus, and parainfluenza virus types 1, 2, and 3. The buffered, stabilized, aqueous solution also contains Evans Blue as a counter-stain and 0.1% sodium azide as preservative.
  • · Normal Mouse Gamma Globulin DFA Reagent a mixture of fluorescein labeled murine gamma globulin that has been shown to be non-reactive with any of the listed respiratory viruses. The buffered, stabilized, aqueous solution contains Evans Blue as a counter-stain and 0.1% sodium azide as preservative.
  • · Respiratory Virus Antigen Control Slides five individually packaged control slides containing wells with cell culture-derived positive and negative control cells. Each positive well is identified with the virus infected cells present, i.e., influenza A virus, influenza B virus, respiratory syncytial virus, adenovirus, and parainfluenza

{2}------------------------------------------------

virus types 1. 2 and 3. The negative well contains uninfected cultured cells. Each slide is intended to be stained only one time.

  • · Wash Solution Concentrate a 40X concentrate consisting of Tween 20 and 4% sodium azide (0.1% sodium azide after dilution in de-mineralized water) in a 40X phosphate buffered saline solution.
  • · Mounting Fluid an aqueous, buffered, stabilized solution of glycerol and 0.1% sodium azide.

The cells to be tested, derived from a clinical specimen or cell culture, are placed onto a glass slide and allowed to air dry. The cells are fixed in acetone. The D' Duet DFA RSV/Respiratory Virus Screening Reagent is added to the cells which are then incubated for 15 to 30 minutes at 35° to 37°C in a humidified chamber or humidified incubator. The stained cells are then washed with the diluted wash solution, a drop of the supplied Mounting Fluid is added and a coverslip is placed on the prepared cells. The cells are examined using a fluorescence microscope. The respiratory syncytial virus infected cells will fluoresce golden-yellow, while cells infected with any of the other six viruses will fluoresce apple-green. Uninfected cells will contain no fluorescence but will be stained red by the Evans Blue counter-stain. If only goldenyellow fluorescent cells are present the specimen can be reported as positive for respiratory syncytial virus antigen. If only apple-green fluorescent cells are present, the particular virus is identified using the individual reagents from the Dr Ultra™ DFA Respiratory Virus Screening & ID Kit (D- Ultra) on new, separate cell preparations. If both golden-yellow and apple-green are present, the additional virus may be identified using the individual reagents from the Dr Ultra on new, separate cell preparations.

Technological Characteristics:

The DHI device, D' Duet, has been compared directly to the DHI device, D' Ultra, as the legally marketed device. The technology used in both devices is based on a standard immunofluorescence assay technique utilizing either phycoerythrin- or fluorescein-labeled MAbs. A summary is provided in Table 5.1 below:

TABLE 12.1: Technological Characteristics Comparison
CharacteristicD³ Duet DFA RSV/ RespiratoryVirus Screening KitD³ Ultra DFA Respiratory VirusScreening & ID Kit
Monoclonal antibodies (MAbs)The RSV/Respiratory Virus DFAScreening Reagentcontains 12 MAbs to 6 differentrespiratory viruses (influenza A virus,influenza B virus, adenovirus,parainfluenza virus type 1,parainfluenza virus type 2,parainfluenza virus type 3), plus 2MAbs to respiratory syncytial virus.The Respiratory Virus DFA ScreeningReagentcontains 12 MAbs to 6 differentrespiratory viruses (influenza A virus,influenza B virus, adenovirus,parainfluenza virus type 1,parainfluenza virus type 2,parainfluenza virus type 3), plus 2MAbs to respiratory syncytial virus.
TABLE 12.1: Technological Characteristics ComparisonD³ Duet DFA RSV/ Respiratory Virus Screening KitD³ Ultra DFA Respiratory Virus Screening & ID Kit
CharacteristicDirect labeling,- using R-phycoerythrin (R-PE) to label the MAbs to respiratory syncytial virus antigens- using fluorescein isothiocyanate (FITC) to label all other MAbs with fluorescein moietyDirect labeling,- using fluorescein isothiocyanate (FITC) to label all MAbs with fluorescein moiety
Labeling methodInfluenza A virus, influenza B virus, adenovirus, parainfluenza virus type 1, parainfluenza virus type 2, parainfluenza virus type 3Influenza A virus, influenza B virus, respiratory syncytial virus, adenovirus, parainfluenza virus type 1, parainfluenza virus type 2, parainfluenza virus type 3
Fluorescein-labeled MAbsRespiratory syncytial virus (Phycoerythrin-labeled influenza A virus MAbs stain with golden-yellow fluorescence)None (Fluorescein-labeled respiratory syncytial virus MAbs stain with apple-green fluorescence)
Phycoerythrin-labeled MAbsCell Fixative is the same for both devices: Acetone
Cell Fixative
Performance characteristics
Staining patternsStaining patterns are the same for both devices:Influenza A and B: The fluorescence is cytoplasmic, nuclear or both. Cytoplasmic staining is often punctate with large inclusions while nuclear staining is uniformly bright.Respiratory Syncytial Virus: The fluorescence is cytoplasmic and punctate with small inclusions in the syncytia.Parainfluenza 1, 2, 3: The fluorescence is cytoplasmic and punctate with irregular inclusions. Types 2 and 3 cause the formation of syncytia.Adenovirus: The fluorescence is cytoplasmic and punctate or bright nuclear on both.
Analytical sensitivity, according to 96-well cell culture plates infected with Flu A diluted to give a TCID50 of 1 per 0.2-mL inoculum (reported as average of 4 runs)There is no significant difference between the two devices for analytical sensitivity.
27.7 ± 1.7 culture positives out of 9628.7 ± 1.3 culture positives out of 96
Analytical specificity (for influenza A virus strains; MAbs are reactive with all listed strains)Mabs to respiratory syncytial virus were shown to be reactive with these virus strains:
3 Respiratory syncytial virus strains: Long, VR-26 Group A, Wash, VR-1401 Group B, 9320, VR-955 Group B3 Respiratory syncytial virus strains: Long, VR-26 Group A, Wash, VR-1401 Group B, 9320, VR-955 Group B
Device Screening Reagent is not reactive with these numbers of microorganisms.*
Analytical specificity (cross reactivity studies; various strains of microorganisms and cell lines)Viruses3231
Bacteria2518
Chlamydia spp.31
Yeast10
Protozoan10
Cell lines1717
* Lists of these microorganisms are provided in Section 18 of this 510(k) submission.

{3}------------------------------------------------

{4}------------------------------------------------

Non-Clinical Performance:

Staining patterns of the phycoerythrin-labeled respiratory syncytial virus MAbs on respiratory syncytial virus infected cells were similar to those of the Predicate device.

Precision/Reproducibility:

Assay precision, intra-assay variability and inter assay variability were assessed with a panel of proficiency-level antigen control slides. The panel consisted of slides spotted with cell preparations of the following:

    1. Low level RSV (Washington strain)
    1. Mid level RSV (Washington strain)
    1. Low level influenza A (Victoria strain) mixed with Mid level RSV (Washington strain)
    1. Mid level influenza A (Victoria strain) mixed with Low level RSV (Washington strain)
    1. Low level respiratory virus (either influenza virus B {Taiwan strain}, adenovirus type 1, Parainfluenza virus types 1, 2, or 3 (strains C35, Grecr, C243 respectively). This panel member was rotated during the 5-days of testing so that each virus is tested twice.
    1. Negative no infected cells present

The low level is estimated to contain between 4 to 10% infected cells per cell spot. The mid level is estimated to contain between 20 to 25% infected cells per cell spot. Both levels were below the level used in quality control slides. Each pancl member was re-coded daily to prevent its identification. Each panel was stained twice per day for 5-days by three different laboratories.

The following results were recorded for both the control slide and the panel slide:

    1. Presence or absence of Yellow-gold fluorescence.
    1. Percent of cells exhibiting Yellow-gold fluorescence
    1. Presence or absence of Green fluorescence
  1. Percent of cells exhibiting Green fluorescence

The combined data for negative specimens - no infected cells present - from the three sites demonstrates that the R-PE labeled and FITC labeled MAbs reproducibly do not stain uninfected cells. No fluorescent cells were seen in 100% (60/60) of the wells lacking infected cells.

The combined data from the three sites demonstrates reproducible detection of respiratory syncytial virus by the R-PF labeled MAbs. The presence of respiratory syncytial virus infected cells was reported in 98% (147/150) of the wells in which the infected cells were expected:

Respiratory syncytial virus detection Summary
PositiveControl SlideLow LevelSlideMid-LevelSlideLow Level withMid-LevelInfluenza AMid-Level withLow LevelInfluenza A
100% (30/30)100% (30/30)100% (30/30)90% (27/30)100% (30/30)

{5}------------------------------------------------

The combined data demonstrates the reproducibility of the detection of Influenza A virus by the FITC labeled MAbs. The presence of Influenza A virus infected cells was reported in 96.7% (87/90) of the wells in which the infected cells were expected:

Influenza A virus detection Summary
Positive Control SlideLow Level Influenza Awith Mid-Level RSVMid-Level Influenza Awith Low Level RSV
100% (30/30)90% (27/30)100% (30/30)

The combined data demonstrates that the presence of R-PE fluorescent cells does not interfere with the detection of influenza A virus by the FITC labeled MAbs in a reproducible manner. The presence of influenza A virus infected cells was reported in 94.7% (54/57) of the wells in which the R-PE stained infected cells were present:

Influenza A virus detection in the presence of R-PE positive cells summary
Low Level R-PE stained cells with Mid-Level influenza A virusMid-Level R-PE stained cells with LowLevel influenza A virus
100% (27/27)90% (27/30)

The combined data from all three sites demonstrates that the presence of R-PE in the stain reproducibly does not interfere with the FITC staining of other viruses. The presence of influenza B virus infected cells was reported in 100% (36/36) of the wells in which the infected cells were expected. The presence of adenovirus infected cells was reported in 100% (36/36) of the wells in which the infected cells were expected. The presence of parainfluenza virus type 1 virus infected cells was reported in 100% (36/36) of the wells in which the infected cells were expected. The presence of parainfluenza virus type 2 virus infected cells was reported in 100% (36/36) of the wells in which the infected cells were expected. The presence of parainfluenza virus type 3 virus infected cells was reported in 100% (36/36) of the wells in which the infected cells were expected.

AdenovirusControl SlideLow LevelAdenovirusInfluenza BVirus ControlSlideLow LevelInfluenza BVirusParainfluenzatype 1Control SlideLow LevelParainfluenzatype 1
100% (30/30)100% (6/6)100% (30/30)100% (6/6)100% (30/30)100% (6/6)
Parainfluenzatype 2Control SlideLow LevelParainfluenzatype 2Parainfluenzatype 3Control SlideLow LevelParainfluenzatype 3
100% (30/30)100% (6/6)100% (30/30)100% (6/6)

miratory virus detection in the precence of D. DE. Summary

{6}------------------------------------------------

The reproducibility study data demonstrates that the presence of R-PE in the stain reproducibly does not interfere with the detection of the 5 respiratory viruses by their respective FITC labeled MAbs.

Detection limit:

Results for analytical detection limit for the seven viruses detected by the De Duet were reported in numbers of fluorescent cells per cell monolayer. Each master stock virus preparation was diluted in a ten-fold manner. Four wells of a 96-well cell culture plate were inoculated with each dilution. The plates were centrifuged at 700 xg for 60 minutes, and then incubated at 35° to 37°C for 24-hours. Four wells from each dilution were stained with the D3 Duet. Each well was then examined at 200x magnification and the number of fluorescent cells counted. The table below lists the virus identity and strain along with the fluorescent cell count.

Analytical Sensitivity of D3 Duet compared with that of
D3 Ultra MAbs
(values are numbers of fluorescent cells per cell monolayer)
Virus strainVirus Dilutions from master stockFluorescent staining cells/well
D3 DuetD3 Ultra
Influenza A virus(PR, VR-95 IIINI)1x10-54, 1, 5, 41, 3, 0, 5
1x10-61, 2, 0, 30, 0, 1, 0
1x10-70, 0, 0, 00, 0, 0, 0
Influenza B virus(Hong Kong, VR-823)1x10-43, 3, 4, 20, 4, 3, 5
1x10-51, 0, 1, 10, 0, 2, 2
1x10-60, 0, 0, 00, 0, 0, 0
Adenovirus (Type 8, VR-8)1x10-61, 1, 3, 31, 3, 2, 4
1x10-70, 0, 0, 00, 0, 0, 0
RSV (Washington, VR-1401)1x10-21, 1, 3, 32, 3, 2, 0
1x10-32, 0, 0, 12, 1, 0, 0
1x10-40, 0, 0, 00, 0, 0, 0
Parainfluenza 1 (C-35, VR-94)1x10-46, 5, 8, 69, 8, 4, 6
1x10-50, 2, 4, 21, 0, 2, 1
1x10-60, 0, 0, 00, 0, 0, 0
Parainfluenza 2(Greer, VR-92)1x10-65, 4, 2, 14, 3, 1, 2
1x10-70, 0, 1, 00, 1, 1, 1
1x10-80, 0, 0, 00, 0, 0, 0
Parainfluenza 3 (C 243, VR-93)1x10-61, 2, 0, 31, 1, 3, 5
1x10-71, 0, 1, 01, 1, 1, 0

{7}------------------------------------------------

Analytical Sensitivity of D. Duet compared with that ofDº Ultra MAbs(values are numbers of fluorescent cells per cellmonolayer)
VirusDilutions frommaster stockFluorescent staining cells/well
Virus strainD3 DuetD³ Ultra
1×10-80, 0, 0, 00, 0, 0, 0

Analytical reactivity (inclusivity); of the D3 Duet was evaluated using 10 influenza A virus and 4 influenza B virus strains. Four wells of a 96-well cell culture plate were inoculated with each viral strain (diluted to less than 20-TCID50 per 0.2-mL inoculum). The plates were centrifuged at 700xg for 60 minutes, and then incubated at 35° to 37°C for 24-hours. Four wells from each strain were stained with the D Duet, and each well was then examined at 200x magnification and the number of fluorescent cells counted. The table below lists the virus identity and strain along with the fluorescent cell count.

Analytical Reactivity (inclusivity) of D³ Dueton various influenza A virus and influenzaB virus strains (values are numbers offluorescent cells per cell monolayer)
Influenza strainFluorescent staining cells/cellmonolayer
Influenza A WS, VR-1520 (H1N1)10, 8, 7, 7
Influenza A HongKong, VR-544(H3N2)12, 11, 11, 12
Influenza A NewJersey, VR-897(H1N1)8, 11, 10, 14
Influenza A Victoria,VR-822 (H3N2)7, 9, 10, 11
Influenza A PR, VR-95 (H1N1)6, 9, 8, 11
Influenza A PortChalmers, VR-810(H3N2)8, 11, 15, 9
Influenza A Aichi,VR-547 (H3N2)16, 15, 14, 13
Influenza A Denver,VR-546 (H1N1)6, 9, 9, 8
Influenza A Mal, VR-98 (H1N1)16, 13, 11, 15
Influenza AA/NWS/33, VR-21912, 17, 15, 10

{8}------------------------------------------------

Analytical Reactivity (inclusivity) of D Duet on various influenza A virus and influenza B virus strains (values are numbers of fluorescent cells per cell monolayer)

Influenza strainFluorescent staining cells/cell monolayer
(H1N1)
Influenza BRussia/69, VR-79013, 14, 12, 15
Influenza BMass/3/66, VR-52312, 19, 14, 13
Influenza BHong Kong/5/72,VR-7918, 8, 9, 11
Influenza BMaryland/1/59,VR-29616, 12, 13, 12

Based on the data presented above, the assay can reliably detect influenza A virus and influenza B virus strains exhibiting both temporal and geographical diversity at viral levels near the limit of detection in cell culture.

Analytical specificity:

Cross-Reactivity Testing

The D3 Duet RSV/Respiratory Virus Screening Kit was tested for cross-reactivity against a variety of cells and microorganisms. Stringent conditions for crossreactivity testing were achieved by using a high concentration of the D3 Duet DFA Influenza RSV/Respiratory Virus Screening Reagent and relatively high titers of microorganisms. The D3 Duet DFA RSV/Respiratory Virus Screening Reagent was prepared at 1.5X the concentration that is provided in the kit. No cross-reactivity was observed for 32 virus strains or for 17 host culture cell types. Twenty-five bacterial strains, one yeast, three Chlamydia sp. and one protozoan were evaluated for cross-reactivity, including Staphylococcus aureus, a protein-Aproducing bacterium. Staining of S. aureus appeared as small points of fluorescence.

Thirty-two virus strains were tested for cross reactivity. Depending on the particular virus, 71 to 1,400 TCID50 were inoculated into shell vial culture and incubated for 24 to 48 hours, to yield a 1+ to 3+ infection, processed and stained with the 1.5X DFA Reagent according to the procedure as detailed in this product insert. No cross reactivity was observed for the viruses listed below:

Virus Strains Tested for Cross Reactivity with D3 Duet DFA RSV/Respiratory Virus Screening Reagent
OrganismStrain or TypeInoculum (TCID50)OrganismStrain or TypeInoculum (TCID50)
Parainfluenza 4aM-25, VR-13781,400CMVTowne, VR-977430
Parainfluenza 4bCH19503, VR-3771,400CMVDavis, VR-807430

{9}------------------------------------------------

Virus Strains Tested for Cross Reactivity with D³ Duet DFA RSV/Respiratory Virus Screening Reagent
OrganismStrain or TypeInoculum(TCID50)OrganismStrain or TypeInoculum(TCID50)
MetapneumovirusSubgroup A11,400CMVAD169, VR-538430
MetapneumovirusSubgroup A21,400Varicella-zosterWebster, VR-916430
MetapneumovirusSubgroup B11,400Varicella-zosterEllen, VR-1367430
MetapneumovirusSubgroup B21,400Rhinovirus 39209 Picornavirus,VR-3401,400
CoronavirusOC43, VR-15581,400RubeolaCommerciallyavailable slidesstained.*
Coronavirus229E, VR-7401,400MumpsCommerciallyavailable slidesstained.*
HSV-11F, VR-73371EchovirusTypes 4, 6, 9, 11,30, 34Commerciallyavailable slidesstained.*
HSV-1MacIntyre, VR-53971CoxsackievirusTypes B1, B2, B3,B4, B5, B6Commerciallyavailable slidesstained.*
HSV-2MS, VR-54071
HSV-2Strain G, VR-73471

Seventeen host culture cell types were tested for cross-reactivity. Cell cultures were prepared in shell vial format. Confluent monolayers were stained with the 1.5X preparation of the D' Duet DFA RSV/Respiratory Virus Screening Reagent according to the procedure as detailed in the product insert, and then examined for cross-reactivity. No cross-reactivity was observed for the following:

Cell lines Tested for Cross Reactivity with D³ Duet DFA InfluenzaA/Respiratory Virus Screening Reagent
A549monolayerpCMKcell spot
BGMKmonolayerpRhMKcell spot
HEp-2monolayerRDmonolayer
LLC-MK2monolayerRhMK IIcell spot
MDCKmonolayerpRKmonolayer
MRC-5monolayerR-Mixmonolayer
MRHFmonolayerVerocell spot
MvlLumonolayerW1-38cell spot
NCI-H292monolayer

Thirty microorganisms, including 25 bacterial and one yeast cultures, three Chlamydia sp. and one protozoan commercially available slides, were stained with the 1.5X DFA Reagent according to the procedure as detailed in the product insert, then examined for cross reactivity. Except for Staphylococcus aureus, which was cross reactive with the D Duet DFA Influenza A/Respiratory Virus Screening Reagent, all other microorganisms tested negative. Reactivity with Staphylococcus aureus is more than likely due to binding the protein A produced by Staphylococcus aureus. Concentrations for each bacterial organism cultured by DHI for cross reactivity testing were determined from suspensions of the bacteria in PBS by spectrophotometer according to McFarland standards of levels 1.0 and 2.0 (equaling approximately 3.0 x 106 and 6.0 x 100 CFU per mL). Slides

{10}------------------------------------------------

were prepared with spots of 0.01-mL of the suspensions to give either 3.0 x 104 or 6.0 x 104 per spot. At the same time, 1-mL of each suspension was plated on an appropriate agar dish for colony confirmation. According to the confirmation agar cultures, initial concentrations of the bacterial organisms in the study ranged from 6.4 x 104 to 2.9 x 10' CFU. Microorganisms tested are listed below.

Microorganisms Tested for Cross Reactivity with D³ Duet DFA Influenza
A/Respiratory Virus Screening Reagent
BACTERIACFU TESTED
Acholeplasma laidlawii~6 x 107
Acinetobacter calcoaceticus9.7 x 105
Bordetella bronchiseptica1.7 x 105
Bordetella pertussis4.6 x 106
Corynebacterium diphtheriac2.5 x 106
Escherichia coli2.6 x 105
Gardnerella vaginalis5.0 x 105
Haemophilis influenzae type A9.3 x 105
Klebsiella pneumoniae6.4 x 106
Legionella pneumophila6.5 x 104
Moraxella cartarrhalis6.4 x 104
Mycoplasma hominis~6 x 104
Mycoplasma orale~6 x 104
Mycoplasma pneumoniae~6 x 104
Mycoplasma salivarium~6 x 107
Neisseria gonorrhoeae1.3 x 106
Proteus mirabilis2.1 x 106
Pseudomonas aeruginosa1.0 x 107
Salmonella enteriditis2.5 x 106
Salmonella typhimurium1.8 x 106
Staphylococcus aureus*1.0 x 107
Streptococcus agalactiae9.6 x 106
Streptococcus pneumoniae8.0 x 106
Streptococcus pyogenes2.9 x 107
Ureaplasma uralyticum~6 x 104
Chlamydophila pneumoniaeCommercially available slides stained.
Chlamydophila psittaciCommercially available slides stained.
Chlamydia trachomatisCommercially available slides stained.
YEAST
Candida glabrata8.7 x 106
PROTOZOAN
Trichomonas vaginalisCommercially available slides stained.

*Reactivity with Staphylococcus aureus is more than likely due to binding the protein A produced by Staphylococcus aureus.

Clinical Performance:

Direct fresh specimens:

A study was performed prospectively at three sites with 1203 fresh specimens that were received for respiratory virus testing. Each specimen was evaluated by the D Duet DFA Influenza A/Respiratory Virus Screening Kit and a cleared DSFA device for the presence of respiratory syncytial virus, influenza A, influenza B, adenovirus, parainfluenza 1, parainfluenza 2 and parainfluenza 3 in cells derived from clinical specimens. Seventeen specimens were excluded from analysis due to a variety of reasons (site deviations, duplicate specimen, insufficient cell

{11}------------------------------------------------

numbers, or high background). These exclusions left 1187 specimen results for analysis.

and. John cleared DSFA comparator assay, combined for study sites 1, 2, and 3:

D3 Duet R-PE identification of respiratory syncytial virus positive specimens
Direct Specimen (1187 Specimens)D' Ultra Final Identification(respiratory syncytial virus)
PosNeg
D3 Duet R-PEPos3000
(respiratory syncytial virus)Neg0887
Positive Percent Agreement (PPA)100%(300/300)
95% CI-PPA97.8, 100%
Negative Percent Agreement (NPA)100% (887/887)
95% CI- NPA99.6. 100%

D3 Duet FITC detection of influenza A virus, influenza B virus, adenovirus, and parainfluenza virus types 1, 2, and 3

D3 Ultra Final Identification
Direct Specimen (1187 Specimens)D3 Duet FITC ScreenPosNeg
Pos1870
Neg01000*
Positive Percent Agreement (PPA)100% (186/186)
95% CI- PPA98.0,100%
Negative Percent Agreement (NPA)100% (1001/1001)
95% CI- NPA99.6,100%

Virus Follow-up Identification of 187 D3 Duet FITC Positive Specimens for influenza A virus, influenza B virus, adenovirus, and parainfluenza virus types 1, 2, and 3 viruses, using D3 Ultra Identification Reagents

Sensitivity95%CISpecificity95% CI
VirusTP /(TP+FN)percentforSensitivityTN/(TN+FP)percentforSpecificity
Influenza Avirus100/100100%96.3, 1001087/1087100%99.7, 100
Influenza Bvirus11/11100%74.1, 1001176/1176100%99.7, 100

{12}------------------------------------------------

Diagnostic Hybrids, Inc.

D3 DUET DFA RSV/RESPIRATORY VIRUS SCREENING KIT Section 5 - Page 13 of 15

Adenovirus52/52100%93.1, 1001135/1135100%99.7, 100
Parainfluenza type 14/4100%51.0, 1001183/1183100%99.7, 100
Parainfluenza type 21/1100%20.7, 1001186/1186100%99.7, 100
Parainfluenza type 319/19100%83.2, 1001168/1168100%99.7, 100

The D3 Duel's ability to identify respiratory syncytial virus using phycoerythrin in direct specimens was compared to the D3 Ultra's ability using fluorescein. The positive percent agreement was 100% (95% CI range of 98.7% to 100%). The negative percent agreement was 100% (95% CI range of 99.6% to 100%). When the ability of the D3 Duet to detect the six other respiratory viruses using fluorescein in direct specimens was compared to the D3 Ultra's ability using fluorescein, the positive percent agreement was 100% (95% CI range of 97.8% to 100%). The negative percent agreement was 100% (95% CI range of 99.6% to 100%).

Specimen type distribution:

Tables below show the study results by the claimed specimen type. Results from sites 1, 2, and 3 have been combined.

Respiratory syncytial virus by specimen type Study Sites 1, 2, and 3 Combined
SpecimentypePPA95%CI forNPA95% CI
TP /(TP+FN)percentPPATN/ (TN+FP)percentfor NPA
NPA155/155100%97.6, 100435/435100%99.1, 100
NPS132/132100%97.2, 100410/410100%99.1, 100

D' Duet FITC detection of influenza A virus, influenza B virus, adenovirus, and parainfluenza virus types 1, 2, and 3 viruses by specimen type Study Sites 1, 2, and 3 Combined

SpecimentypePPANPA
TP /(TP+FN)percent95% CIfor PPATN/(TN+FP)percent95% CIfor NPA
NPA103/103100%96.4, 100484/484100%99.2, 100
NPS79/79100%95.4, 100460/460100%99.2, 100

Cultured specimens:

To evaluate the performance of this device using cultured clinical specimens, a fourth study was performed with 298 frozen specimens to compare performance of the D3 Duet DFA RSV/Respiratory Virus Screening Kit with that of the predicate for the presence of respiratory syncytial virus, influenza A virus, influenza B virus, adenovirus, parainfluenza virus types 1, 2 and 3 from cultured

{13}------------------------------------------------

clinical specimens. At Study Site 4, 298 frozen specimens were processed for cell culture testing in accordance with the procedure in the Comparator product insert (same procedure for both Subject and Comparator devices) using R-Mix TooTM FreshCells™ in 48/24-fill multi-well plates. All specimens at study site 4 were derived from nasopharyngeal specimens. The results of this study are presented below. The table below shows the age distribution for individuals studied at site 4:

Site 4 (culture) – Age Distribution
0 - 1 month5
>1 month - 2 years130
>2 - 12 years44
>12 - 21 years28
22 - 30 years19
31 - 40 years20
41 - 50 years10
51 - 60 years9
61 - 70 years8
71 - 80 years6
81 - 90 years8
>90 years5
Unknown age6
Total298

The following tables detail the results of the cell culture study's comparison of D3 Duet's phycoerythrin-labeled MAbs identification of respiratory syncytial virus specimens positive specimens, and D3 Duet's fluorescein-labeled MAbs detection of influenza A virus, influenza B virus, adenovirus, and parainfluenza virus types 1, 2, and 3 positive specimens.

Cell Culture (298 Specimens)D³ Ultra Final Identification(respiratory syncytial virus)
PosNeg
D³ Duet R-PE(respiratory syncytial virus)Pos330
Neg0265
Positive Percent Agreement (PPA)100% (33/33)
95% CI- PPA89.5, 100%
Negative Percent Agreement (NPA)100% (265/265)
95% CI- NPA98.6, 100%
Study Site 4(culture) - D3 Duet FITC detection of influenza A virus, influenza B virus,adenovirus, and parainfluenza virus types 1, 2, and 3
Cell Culture (298 Specimens)D3 Ultra Final Identification
PosNeg

{14}------------------------------------------------

Diagnostic Hybrids, Inc.

D3 DUET DFA RSV/RESPIRATORY VIRUS SCREENING KIT Section 5 - Page 15 of 15

Pos1040
D3 Duet FITC ScreenNeg0194
Positive Percent Agreement (PPA)100%(104/104)
95% CI- PPA96.4,100%
Negative Percent Agreement (NPA)100% (194/194)
95% CI- NPA98.1,100%

A variety of viral respiratory pathogens were isolated. Virus identification of D3 Duet FITC Positive Specimens using D3 Ultra Identification Reagents yielded the following isolates: respiratory syncytial virus [prevalence 11.1% (33/298)], influenza A virus [prevalence 22.5% (67/298)], influenza B virus [prevalence 6.7% (20/298)], adenovirus [prevalence 3.4% (10/298)], parainfluenza type I virus [prevalence 1.7% (5/298)], parainfluenza type 2 virus [prevalence 1.0% (3/298)], and parainfluenza type 3 virus [prevalence 3.0% (9/298)]. There were sixteen co-infections as follows: three influenza A virus + parainfluenza type 3 virus, one influenza A virus + parainfluenza type 1 virus, one influenza A virus + parainfluenza type 2 virus, two influenza A virus + respiratory syncytial virus, one influenza A virus + adenovirus, one influenza B virus + parainfluenza type 2 virus, one influenza B virus + parainfluenza type 3 virus, one influenza B virus + respiratory syncytial virus, one respiratory syncytial virus + parainfluenza type 1 virus, two respiratory syncytial virus + parainfluenza type 3 virus, one adenovirus + parainfluenza type 1 virus and one adenovirus + parainfluenza type 3 virus.

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three bars forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Gail R. Goodrum Vice President of Regulatory Affairs Diagnostic Hybrids, Inc. 1055 East State Street Suite 100 Athens, Ohio 45701

Re: K081928

Trade/Device Name: D3 Duet DFA RSV/Respiratory Virus Screening Kit Regulation Number: 21 CFR 866.3480 Regulation Name: Respiratory syncytial virus serological reagents Regulatory Class: Class I Product Code: LKT Dated: November 26, 2008 Received: November 28, 2008

Dear Ms. Goodrum:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

DEC 2 3 2008

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements aft the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter

{16}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sally attaym

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{17}------------------------------------------------

SECTION 04, INDICATIONS FOR USE

510(k) Number (if known): K081928

Device Name: D3 Duet DFA RSV/Respiratory Virus Screening Kit

Indication for Use: The Diagnostic Hybrids, Inc. device, D3 Duet DFA RSV/Respiratory Virus Screening Kit, is intended for the qualitative detection and identification of respiratory syncytial virus, while screening for influenza A virus, influenza B virus, adenovirus, and parainfluenza virus types 1, 2 and 3 viral antigens, in nasal and nasopharyngeal swabs and aspirates or in cell culture. The assay detects viral antigens by immunofluorescence using monoclonal antibodies (MAbs), from patients with signs and symptoms of respiratory infection.

It is recommended that specimens found to be negative after examination of the direct specimen result be confirmed by cell culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for diagnosis, treatment or other management decisions.

Performance characteristics for influenza A virus detection and identification were established when influenza A (H3N2) and influenza A (H1N1) were the predominant influenza A strains circulating in the United States. Performance characteristics for influenza A virus detection and identification were established when influenza A H3N2 and influenza A H1N1 were the predominant influenza A strains circulating in the United States. When other Influenza A viruses are emerging, performance characteristics may vary. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to a state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

KO8 1928 SIG(K).

§ 866.3480 Respiratory syncytial virus serological reagents.

(a)
Identification. Respiratory syncytial virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to respiratory syncytial virus in serum. Additionally, some of these reagents consist of respiratory syncytial virus antisera conjugated with a fluorescent dye (immunofluorescent reagents) and used to identify respiratory syncytial viruses from clinical specimens or from tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of respiratory syncytial virus infections and provides epidemiological information on diseases caused by these viruses. Respiratory syncytial viruses cause a number of respiratory tract infections, including the common cold, pharyngitis, and infantile bronchopneumonia.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.