K Number
K081820

Validate with FDA (Live)

Date Cleared
2008-10-06

(101 days)

Product Code
Regulation Number
870.4210
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Avalon Elite™ Bi-Caval Dual Lumen Catheter is intended for use as a single catheter for both venous drainage and reinfusion of blood via the internal jugular vein during extracorporeal life support procedures.

Device Description

The Avalon Elite™ Bi-Caval Dual Lumen Catheter is a single catheter with two lumens within the cannula body to collect, separate, and return the blood. The product is offered in a range of sizes to address varying patient size requirements. Inlet ports/openings at the distal tip, and proximally, collect blood from the inferior and superior vena cavae and deliver it to the extracorporeal components consisting of at least, a pump and gas exchange device. Blood returns to the patient through the infusion lumen of the cannula where it exits at a port located in the right atrium. The lumens within the cannula body are completely separate throughout their length. This separation is achieved via a thin septum that divides the otherwise circular internal cross section of the main cannula body. The lumens become separate, or split out, at the proximal end of the cannula as is typical of many types of dual lumen cannulae/catheters so that they may be attached by conventional tube fittings. The ratio of the drainage lumen to reinfusion lumen cross sectional area is approximately 2 to 1 deriving from typical pre and post pump pressure/flow characteristics. The catheter is supplied with a dilator/introducer to facilitate placement into the vasculature by normal access techniques. The introducer/dilator is intended to follow a pre positioned standard guidewire (not supplied.) The device is supplied sterile, non-pryogenic and is intended for single use via prescription.

AI/ML Overview

This submission pertains to the Avalon Elite™ Bi-Caval Dual Lumen Catheter. The provided information is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a comprehensive clinical study to establish novel performance criteria. Therefore, the acceptance criteria and study details discussed below will reflect this regulatory pathway.

Here's a breakdown of the requested information based on the provided text:


1. Table of Acceptance Criteria and Reported Device Performance

For a 510(k) submission, "acceptance criteria" are generally demonstrated through substantial equivalence to a legally marketed predicate device across various characteristics and performance tests. The specific numerical acceptance criteria are not explicitly stated as they would be for a direct performance claim; instead, the "performance" here refers to the device meeting the same functional and safety requirements as its predicate.

Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance (Avalon Elite™ Bi-Caval Dual Lumen Catheter)
Same Intended Use as predicate device (Origen Dual Lumen Catheter, K003288).The Avalon Elite™ Bi-Caval Dual Lumen Catheter is intended for use as a single catheter for both venous drainage and reinfusion of blood via the internal jugular vein during extracorporeal life support procedures, which is identical to the predicate device.
Similar Design to predicate device.Shares similar design, bifurcations, and radiopaque tips with the predicate device.
Similar Size Ranges as predicate device.Offers similar size ranges to the predicate device.
Similar Materials of Construction as predicate device.Uses similar generic materials of construction as the predicate device.
Equivalent Performance in key areas affecting safety and effectiveness compared to the predicate device. This is assessed through various tests:Demonstrated equivalence to the predicate device through performance testing.
- Pressure/Burst strengthTesting conducted.
- Simulated UseTesting conducted.
- Kink ResistanceTesting conducted.
- Tensile StrengthTesting conducted.
- Flow CharacteristicsTesting conducted.
- HemolysisTesting conducted.
- BiocompatibilityTesting conducted.
Sterility and Non-PyrogenicitySupplied sterile and non-pyrogenic.
Single UseIntended for single use.

2. Sample Size Used for the Test Set and the Data Provenance

The document does not specify a "test set" in the context of a clinical trial or a dataset for an AI algorithm. Instead, it refers to performance testing conducted on the device itself (e.g., Pressure/Burst, Simulated Use, Kink Resistance, etc.).

  • Sample Size: The exact number of catheters or test specimens used for each performance test (Pressure/Burst, Simulated Use, Kink Resistance, Tensile Strength, Flow Characteristics, Hemolysis, and Biocompatibility) is not provided in this summary.
  • Data Provenance: The provenance of the data (e.g., country of origin, retrospective/prospective) is not specified. These tests would typically be conducted in a laboratory setting by the manufacturer (Avalon Laboratories, LLC).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable to this 510(k) submission. The "ground truth" concept is typically associated with studies involving subjective interpretation (e.g., medical imaging, pathology) where expert consensus is needed to define the correct outcome. For a physical medical device like a catheter, "ground truth" is established by objective engineering and safety standards, and validated through the performance tests mentioned above, not by expert human interpretation of a dataset.


4. Adjudication Method for the Test Set

This information is not applicable for the reasons stated in point 3. Adjudication methods (e.g., 2+1, 3+1) are relevant for reconciling differing expert opinions on a "ground truth" in subjective data analysis, not for direct physical performance testing of a device.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. An MRMC study is relevant for evaluating the impact of an AI system on human performance in diagnostic tasks, typically involving interpretation of medical images or data by multiple readers. This 510(k) summary is for a physical medical device (catheter), not an AI diagnostic tool. No AI component is mentioned.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This information is not applicable. No AI algorithm is described or evaluated in this submission.


7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

As discussed, the concept of "ground truth" in the context of expert interpretation is not applicable here. For this physical device, the "ground truth" for performance is established by objective engineering specifications and recognized safety standards validated through laboratory performance tests (e.g., burst pressure exceeding a certain threshold, flow rates matching specifications, non-hemolytic properties, biocompatibility according to ISO standards). It's based on empirical measurements against predefined criteria.


8. The Sample Size for the Training Set

This information is not applicable. There is no mention of a "training set" as this is not an AI/machine learning device submission.


9. How the Ground Truth for the Training Set Was Established

This information is not applicable. As there is no training set, there is no ground truth for a training set to be established.

{0}------------------------------------------------

OCT 0 6 2008

510(k) Summary Pursuant to 21 CFR 807,92

1.Submitted By:Avalon Laboratories, LLC2610 E. Homestead PlaceRancho Dominguez, CA 90220
2.Contact:David C. FurrFDC Services, LLC5200 Crystal Water DriveAustin, Texas 78735260-402-1598
3.Product:Avalon Elite™ Bi-Caval Dual Lumen CatheterCFR Section 870.4210Cardiopulmonary bypass vascular catheterClass IIProduct Code: DWF
  1. Common/Trade Name:

Avalon Elite™ Bi-Caval Dual Lumen Catheter

Description:

The Avalon Elite™ Bi-Caval Dual Lumen Catheter is a single catheter with two lumens within the cannula body to collect, separate, and return the blood. The product is offered in a range of sizes to address varying patient size requirements. Inlet ports/openings at the distal tip, and proximally, collect blood from the inferior and superior vena cavae and deliver it to the extracorporeal components consisting of at least, a pump and gas exchange device. Blood returns to the patient through the infusion lumen of the cannula where it exits at a port located in the right atrium. The lumens within the cannula body are completely separate throughout their length. This separation is achieved via a thin septum that divides the otherwise circular internal cross section of the main cannula body. The lumens

{1}------------------------------------------------

510(k) Premarket Notification Avalon Elite™ Bi-Caval Dual Lumen Catheter

become separate, or split out, at the proximal end of the cannula as is typical of many types of dual lumen cannulae/catheters so that they may be attached by conventional tube fittings. The ratio of the drainage lumen to reinfusion lumen cross sectional area is approximately 2 to 1 deriving from typical pre and post pump pressure/flow characteristics. The catheter is supplied with a dilator/introducer to facilitate placement into the vasculature by normal access techniques. The introducer/dilator is intended to follow a pre positioned standard guidewire (not supplied.)

The device is supplied sterile, non-pryogenic and is intended for single use via prescription.

Intended Use:

The Avalon Elite™ Bi-Caval Dual Lumen Catheter is intended for use as a single catheter for both venous drainage and reinfusion of blood via the internal jugular vein during extracorporeal life support procedures.

Technological Characteristics and Substantial Equivalence:

The Avalon Elite™ Bi-Caval Dual Lumen Catheter is substantially equivalent to the Origen Dual Lumen Catheter (K003288). Device comparisons show both products have the same intended use, and similar size ranges, materials, bifurcations, radiopaque tips, and packaging. Both devices are supplied as single use sterile products.

Performance Testing:

The Avalon Elite™ Bi-Caval Dual Lumen Catheter was subiected to numerous tests and comparisons to the predicate device. Testing included Pressure/Burst, Simulated Use, Kink Resistance, Tensile Strength, Flow Characteristics, Hemolysis, and Biocompatibility.

{2}------------------------------------------------

Conclusions:

The studies conducted on Avalon Elite™ Bi-Caval Dual Lumen Catheters demonstrate that the device is substantially equivalent to the predicate device currently in commercial distribution.

The predicate dual lumen catheters and the Avalon dual lumen catheters share similar design, size, and generic materials of construction. Comparisons show that the Avalon products are equivalent to the predicate products in all key areas of features and performance that could affect safety and effectiveness. The intended use is identical to the intended use of the predicate.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized lines forming its body and wings.

OCT 0 6 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Avalon Laboratories, LLC c/o Mr. David Furr Application Correspondent 5200 Crystal Water Dr. Austin, TX 78735

Re: K081820

Trade/Device Name: Avalon Elite™ Bi-Caval Dual Lumen Catheter Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary bypass vascular catheter, cannula, or tubing Regulatory Class: Class II Product Code: DWF Dated: September 15, 2008 Received: September 18, 2008

Dear Mr. Furr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or

{4}------------------------------------------------

Page 2 - Mr. David Furr

any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to promarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

D. Lachner

Bra
Dir
Dir

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

_ K08 1820 ___________________________________________________________________________________________________________________________________________________________________ 510(k) Number:

Device Name:

Avalon Elite™ Bi-Caval Dual Lumen Catheter

Indications for Use:

The Avalon Elite™ Bi-Caval Dual Lumen Catheter is indicated for use as a single catheter for both venous drainage and reinfusion of blood via the internal jugular vein during extracorporeal life support procedures.

Over-the-counter use _________________________________________________________________________________________________________________________________________________________ Prescription Use_ X or (per CFR 801.109)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Dana R. Vachner

(Division Sign-Off) (Division Sign-Sil)
Division of Cardiovascular Devices

510(k) Number_KO81820

§ 870.4210 Cardiopulmonary bypass vascular catheter, cannula, or tubing.

(a)
Identification. A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.(b)
Classification. Class II (performance standards).