Intended Use

The OSV II is an implantable system used in the treatment of patients with hydrocephalus, to shunt CSF from the ventricles to the peritoneal cavity or other appropriate drainage site such as the heart's right atrium.

Device Description

The OSV II is designed to shunt cerebrospinal fluid (CSF) from the ventricles of the brain to an appropriate drainage site, such as the atrium of the heart or the peritoneal cavity. Unlike conventional valves, it is a variable resistance valve which maintains a drainage rate constant within the physiological range (for the specified populations and disorders) of intracranial pressure.

The principle of operation consists of a flow restricting diaphragm, a seat, and a notched pin. The diaphragm, held between two polysulfone parts of the casing, reacts to variations in pressure. A synthetic ruby seat is inserted into the center of the diaphragm. As pressure varies, the clearance between the seat and the pin increases or decreases depending on seat movement along the pin.

AI/ML Overview

The provided text describes a 510(k) summary for the OSVII Hydrocephalus Valve. It does not contain information related to a study with acceptance criteria and device performance in the format requested, particularly for AI/ML-based device evaluation. The document focuses on demonstrating substantial equivalence to a previously cleared device (OSV II, K911799) based on design, principle of operation, intended use, materials, and manufacturing process.

The "Safety and Effectiveness" section does not report numerical performance metrics or acceptance criteria in the way that would typically be seen for a new performance claim. Instead, it makes qualitative statements about safety and effectiveness and focuses on the impact of a connector modification.

Below is an attempt to structure the available information, noting where the requested details are not present in the provided text.


Acceptance Criteria and Study for OSVII Hydrocephalus Valve

Summary: The OSVII Hydrocephalus Valve 510(k) submission primarily relies on demonstrating substantial equivalence to a previously cleared predicate device rather than presenting a performance study against specific acceptance criteria with numerical outcomes. The "Safety and Effectiveness" section addresses product attributes and the impact of a minor modification but does not provide quantitative performance metrics or acceptance criteria as typically found in studies for AI/ML devices or new performance claims.


1. Table of Acceptance Criteria and Reported Device Performance

As the provided text focuses on substantial equivalence rather than a detailed performance study with defined acceptance criteria and numerical results for the modified device, a table in the requested format cannot be fully generated. The "Safety and Effectiveness" section provides qualitative statements.

Acceptance Criterion (Implicit)Reported Device Performance (Qualitative)
Sterility and non-pyrogenicity"The OSV II Hydrocephalus Valve is provided sterile and non-pyrogenic."
Correlation of modified and unmodified test methods"The correlation established between the unmodified and modified test methods demonstrated that the modification did not adversely affect the safety or effectiveness of the product."
Leakage after connector modification"The connector modification to the OSV II had been tested for leakage..."
Resistance of connections after connector modification"...and for resistance of the connections at time of manufacturing, after shelf life simulation and after transport and environmental conditions."
Operational Stages (Flow Rate vs. Differential Pressure)The device is designed to operate in three stages: - Stage I (Low DP): 5-18 ml/hr flow rate (DP 30-120 mm H2O) - Stage II (Flow Regulation): 18-30 ml/hr flow rate (DP 120-300 mm H2O) - Stage III (Pressure Relief): Rapid flow for IVP normalization. (Note: These are design specifications/operational principles, not reported performance against acceptance criteria from a study for the modified device.)

2. Sample size used for the test set and the data provenance: Not explicitly stated. The document refers to "test methods" and "testing" related to the connector modification, implying physical and functional tests were conducted, but does not specify sample sizes or data provenance (e.g., country of origin, retrospective/prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The submission focuses on device characteristics and substantial equivalence, not on expert-adjudicated ground truth for a diagnostic task.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a medical device (hydrocephalus valve), not an AI/ML diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable in the context of diagnostic "ground truth." The "ground truth" for this device would relate to its physical and functional performance specifications (e.g., flow rate, pressure regulation, material integrity, sterility) typically established through engineering and biological testing.

8. The sample size for the training set: Not applicable. This is not an AI/ML device.

9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Ko8173

OSVII Hydrocephalus Valve 510(k) SUMMARY

Submitter's name and address:

Integra NeuroSciences Implants SA 2905 Route des Dolines 06921 Sophia Antipolis Cedex, France

NOV - 6 2008

Contact person and telephone number:

Anne Bigeard Regulatory Affairs Manager Telephone: +33 (0)4 93 95 5667 Facsimile: +33 (0)4 93 65 4030

Date summary was prepared:

June 10, 2008

Name of the device:

Proprietary Name:OSVII Hydrocephalus Valve
Common Name:Hydrocephalus Shunt Systems and Components
Classification Name:Central Nervous System Shunt and Components JXG

Substantial Equivalence:

The modified OSV II is substantially equivalent to the OSV II cleared by the FDA under 510(k) (K911799). The design, principle of operation, intended use, materials of composition, manufacturing process is equivalent to the previously cleared OSV II.

Intended use:

The OSVII is an implantable system used in the treatment of patients with hydrocephalus, to shunt CSF from the ventricles to the peritoneal cavity or other appropriate drainage site such as the heart's right atrium.

Device Description:

The OSV II is designed to shunt cerebrospinal fluid (CSF) from the ventricles of the brain to an appropriate drainage site, such as the atrium of the heart or the peritoneal cavity. Unlike conventional valves, it is a variable resistance valve which maintains a drainage rate constant within the physiological range (for the specified populations and disorders) of intracranial pressure.

The principle of operation consists of a flow restricting diaphragm, a seat, and a notched pin. The diaphragm, held between two polysulfone parts of the casing, reacts to variations in pressure. A synthetic ruby seat is inserted into the center of the diaphragm. As pressure varies, the clearance between the seat and the pin increases or decreases depending on seat movement along the pin.

{1}------------------------------------------------

Three Stages of Operation are defined:

Stage I - Low Differential Pressure

This stage begins when the flow rate through the valve reaches 5ml/hr (Differential Pressure (DP) will be between 30 and 80 mm H20). The valve remains in Stage I with CSF flow rates up to 18 ml/hr (DP will be between 40 and 120 mm H20).

Stage II - Flow Regulation

When the DP increases, the valve operates as a variable resistance flow regulator. At DP ranges between 120 and 300 mm H20, the valve restricts flow between 18 and 30 ml/hr.

Stage III - Pressure Relief Mode

Should the intraventricular pressure (IVP) elevate abruptly, the shunt operates in a rapid flow rate mode to facilitate IVP normalization. The valve then reverts to Stage II or I, depending upon conditions.

Safety and Effectiveness:

  • . The OSV II Hydrocephalus Valve is provided sterile and non-pyrogenic.
  • The correlation established between the unmodified and modified test methods . demonstrated that the modification did not adversely affect the safety or effectiveness of the product.
  • . The connector modification to the OSV II had been tested for leakage and for resistance of the connections at time of manufacturing, after shelf life simulation and after transport and environmental conditions.

Conclusion:

The modified OSV II is considered substantially equivalent to the unmodified OSV II cleared by the FDA under 510(k) (K911799). The modifications do not affect the intended use, the fundamental scientific technology of the device, and do not raise new issues of safety and effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings forming three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV - 6 2008

Integra LifeSciences Corporation % Mr. Jon Caparotta, RAC Director, Regulatory Affairs 311 Enterprise Drive Plainsboro, New Jersey 08536

Re: K081773

Trade/Device Name: OSV II Regulation Number: 21 CFR 882.5550 Regulation Name: Central nervous system fluid shunt and components Regulatory Class: II Product Code: JXG Dated: October 2, 2008 Received: October 7, 2008

Dear Mr. Caparotta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 -- Mr. Jon Caparotta, RAC

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Milliman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K081773

Device Name: OSV II

Indications For Use:

The OSV II is an implantable system used in the treatment of patients with hydrocephalus, to shunt CSF from the ventricles to the peritoneal cavity or other appropriate drainage site such as the heart's right atrium.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil Reath forman

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number K081773

Page 1 of 1

§ 882.5550 Central nervous system fluid shunt and components.

(a)
Identification. A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.(b)
Classification. Class II (performance standards).