K Number
K081726

Validate with FDA (Live)

Date Cleared
2008-08-26

(69 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TysonBio Link Health Management System is an optional software accessory for use with the following models with data management capabilities: a) EZ Smart-608 series meter, b) DIACHEX Series Blood Glucose Meter, and c) DIACHEX+ Series Blood Glucose Meter. When used with one of these meters, TysonBio Link Health Management System transfers data from the device's memory into a computer for enhanced data management.

TysonBio Link Health Management System is intended for use in home and clinical settings to aid people with diabetes and their health care professionals in the review, analysis and evaluation blood glucose test results to support effective diabetes management. It should be noted that the software does not recommend any medical treatment or medication dosage level.

Device Description

The TysonBio Health Management System software is an optional software accessory for use with TysonBio blood glucose monitors with data management capabilities. When use with one of these meters, TysonBio Link Health Management System software transfers data from the devices records into a computer for enhanced data management. It should be noted that the software does not recommend any medical treatment or medication dosage level.

AI/ML Overview

The TysonBio Link Health Management System is a data management software for blood glucose monitors. The provided 510(k) summary does not contain specific acceptance criteria or a detailed study report with quantitative performance data in the typical sense of a diagnostic device. Instead, the submission relies on demonstrating substantial equivalence to a predicate device through functional comparisons and general software validation.

Here's an analysis of the provided information based on your request, highlighting what is (and isn't) present:

1. Table of Acceptance Criteria and Reported Device Performance

As noted, the submission primarily compares features and functionalities rather than setting quantitative performance thresholds. The acceptance criteria essentially appear to be "functional equivalence" to the predicate device and meeting general software validation principles.

Feature/CriterionPredicate Device (IN TOUCH) PerformanceTysonBio Link HMS PerformanceAcceptance Status
Intended UseAid in review, analysis, evaluation of blood glucose results for diabetes management.Aid in review, analysis, evaluation of blood glucose results for diabetes management. (Does not recommend treatment/dosage)Met (Identical)
Data ImmutabilityData transferred cannot be changed or modified.Data transferred cannot be changed or modified.Met (Identical)
Optional AccessorySoftware accessory for blood glucose monitors with data management.Software accessory for blood glucose monitors with data management.Met (Identical)
No Control over Meter MeasurementsDoes not control or affect blood glucose monitor's measurements.Does not control or affect blood glucose monitor's measurements.Met (Identical)
Cable AvailabilitySerial cable, available separately.Serial cable, available separately.Met (Identical)
Uploading Data from Various DevicesSoftware driver must be uploaded on device or installed on PC.Software driver must be installed on PC.Met (Functionally Equivalent)
Minimum RAMMinimum of 128 MB.Minimum of 128 MB or better.Met (Equivalent or better)
Minimum Free Hard Disk Space100-200 MB during, 100 MB after installation.Minimum 100 MB or more.Met (Equivalent or better)
Port Required9-pin or 25 pin COM or USB port.USB port.Met (Functionally Equivalent, subset of predicate)
Screen Resolution800x600 pixels and 256 colors.Same as predicate.Met (Identical)
Software Use SettingsHome or clinic settings.Single or multiple user settings.Met (Functionally Equivalent)
Deleting Patients/RecordsYes.Yes.Met (Identical)
Uninstall ProgramYes.Yes.Met (Identical)
Personalize Target RangesYes.Yes.Met (Identical)
Unit of Measure DisplayChoice of mmol/L or mg/dL.mmol/L or mg/dL.Met (Identical)
Enter Hypoglycemic RangeYes.Yes.Met (Identical)
Report TypesData list, statistics, average reading, pie chart, line graph, histogram.Data list, statistics, average reading, pie chart, line graph, histogram.Met (Identical)
Auto-detect COM portYes.Yes.Met (Identical)
Downloaded Results Cannot Be Edited/DeletedYes.Yes.Met (Identical)
Modify Meter Average ResultsYes (7,14,30,60,90 days).Yes.Met (Functionally Equivalent)
Clear Meter Results in MemoryYes.Yes.Met (Identical)
Required Info on Use EntryYes.Yes (user ID, name, photo).Met (Functionally Equivalent or better)
Operating SystemWindows 98 SE, 2000, XP.Windows 2000, XP.Not Identical (More restrictive, but acceptable for clearance)
Manual EntryAbility to add records manually.No.Not identical, but acceptable difference as device is for transferring data.
Copy Database to Separate FileYes.No.Not identical, but acceptable difference.
Deleting ResultsYes.No.Not identical, but acceptable difference (emphasizes data immutability for transferred data).
Language CapabilitiesEnglish, Spanish.English.Not identical, but acceptable difference.
Viewing Owner's ManualLink provided via icon.No.Not identical, but acceptable difference.
Default Glucose Target RangesYes.No.Not identical, but acceptable difference (user can personalize).
Search Patient CapabilityYes.No.Not identical, but acceptable difference.
Result Type Display (Whole blood/Plasma)Choice of whole blood or plasma.No.Not identical, but acceptable difference.
Set Default Favorite ReportYes.No.Not identical, but acceptable difference.
Synchronize Meter Clock to PCYes.No.Not identical, but acceptable difference.
Display 12 or 24 hour formatYes.No.Not identical, but acceptable difference.
Change Date FormatYes.No.Not identical, but acceptable difference.
Link to Older Database VersionsYes.No old versions available at this time.Not identical, but acceptable difference (due to product lifecycle).
Diabetes Educator InfoOne diabetes educator may be entered.No.Not identical, but acceptable difference.
Diabetes Control (Insulin, medication, diet/exercise)Insulin list, medication list, diet/exercise options.No.Not identical, but acceptable difference.
Specifying/Entering Medication/InsulinYes, up to 3 types.No.Not identical, but acceptable difference.
Enter Insulin RegimentYes.No.Not identical, but acceptable difference.
Report Types (beyond basic)Standard Day, Average Reading, Insulin, Exception, Data List, Histogram, Health checks.No Insulin, Exception, Standard Day report.Not identical, but acceptable difference.
Set Meter Clock to Specific Date/TimeYes.No.Not identical, but acceptable difference.
Email Report from PC DirectlyYes.No.Not identical, but acceptable difference.
Input Additional Info on Manual ResultYes.No.Not identical, but acceptable difference.

2. Sample Size Used for the Test Set and Data Provenance

The 510(k) summary does not specify a sample size for a test set in the context of clinical performance or data accuracy. This is a software accessory for data management, not a diagnostic test device itself. The "testing" referred to is software verification and validation.

The "data provenance" (country of origin, retrospective/prospective) for user data or clinical samples is not applicable as no such data is being analyzed for clinical accuracy by the software. The software's function is to transfer and display data from a blood glucose meter.

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not provided and is not applicable to this type of software submission. There is no mention of expert review for establishing ground truth, as the software's role is data handling, not interpretation requiring expert consensus.

4. Adjudication Method

Not applicable. No clinical data requiring adjudication by experts is described as part of the submission.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not performed or mentioned. This type of study is typically used for diagnostic imaging or interpretation devices where human performance with and without AI assistance is being evaluated. This software is for data management, not interpretation.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

While there isn't a "standalone study" in the sense of a diagnostic algorithm evaluating patient data, the software verification and validation testing described is essentially evaluating the algorithm's (software's) performance in isolation against its functional requirements. The statement "Software verification and validation testing demonstrated that the TysonBio Health Management System meets the performance requirements for the intended use of the optional accessory device" indicates this. The performance requirements, in this context, are for accurate data transfer, display, reporting, and adherence to specified functionalities.

7. Type of Ground Truth Used

The "ground truth" for this device would be its functional specifications and software requirements. The testing would verify that the software accurately transfers numeric data (blood glucose readings) from the meter, stores it, and generates reports based on that data, without alteration. The ground truth for individual blood glucose values would originate from the blood glucose meters themselves, which are separate regulated devices. The software's ground truth is about its operation, not the medical accuracy of the data it handles.

8. Sample Size for the Training Set

Not applicable. This device is not an AI/ML model that requires a training set of data. It is a traditional software application designed for data transfer and management based on explicit programming rules, not learned patterns from data.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of software.

{0}------------------------------------------------

510(k) SUMMARY

1. DATE PREPARED

June16, 2008

K08 1726

2. SPONSOR INFORMATION

Address

TYSON BIORESEARCH, INC. 5 F., # 22, KE TUNG RD., SCIENCE BASED INDUSTRIAL PARK CHUN-NAN, MIAO-LI COUNTY, CHINA (TAIWAN) 350

Contact Person: WEN-HAI TSAI

PHONE: 886-37-585988 FACSIMILE: 886-37-585996

3. NAME OF DEVICE:

1. Trade Name:TysonBio Link Health Management System
2. Common Names:Data Management Software
3. Classification Names:Calculator/data processing module for clinical use, 21CFR862.2100Glucose Test System, 21 CFR862.1345
4. Product code:NBW - System, test, blood glucose, over the counter,JQP - Calculator/data processing module for clinical use

4. DEVICE DESCRIPTION:

The TysonBio Health Management System software is an optional software accessory for use with TysonBio blood glucose monitors with data management capabilities. When use with one of these meters, TysonBio Link Health Management System software transfers data from the devices records into a computer for enhanced data management. It should be noted that the software does not recommend any medical treatment or medication dosage level.

5. Principles of Operation

The TysonBio Link Health Management System software provides users the ability to export data from compatible TysonBio meters to a computer via a cable connecting to the computer's required USB port. It is a Microsoft Windows based software application for diabetes data management. TysonBio Link HMS software is designed to operate an Intel compatible PC with Microsoft Windows 2000/XP or later operating system. The export of the data to be viewed only in the TysonBio Link HMS software. Aside from the ability to export the data stored in the meter's memory, the software also allows the ability to clear the meter's memory. Data transferred from the device cannot be changed or modified.

AUG 2 6 2008

{1}------------------------------------------------

6. INTENDED USE:

The TysonBio Link Health Management System is an optional software accessory for use with the following models with data management capabilities: a) EZ Smart-608 series meter, b) DIACHEX series Blood Glucose Meter, and c) DIACHEX+ series Blood Glucose Meter. When used with one of these meters, TysonBio Link Health Management System transfers data from the device's memory into a computer for enhanced data management.

TysonBio Link Health Management System is intended for use in home and clinical settings to aid people with diabetes and their health care professionals in the review, analysis and evaluation blood glucose test results to support effective diabetes management. It should be noted that the software does not recommend any medical treatment or medication dosage level.

7. PREDICATE DEVICE:

Predicate device name(s): LifeScan IN TOUCH Diabetes Management System Predicate 510(k) number(s): K984527

8. Comparison with predicate:

The TysonBio Link Health Management System software is substantially equivalent to other products in commercial distribution intended for similar use. Most notably it is substantially equivalent to the currently marketed Lifescan IN TOUCH Diabetes Management software (K984527). Both management software programs can be described as follows:

  • have the same intended use.
  • -Data transferred from the device cannot be changed or modified in any way
  • -An optional software accessory for use with blood glucose monitors with data management capabilities
  • Do not in any way control or affect the blood glucose monitor's measurements -
ItemsIN TOUCH DIABETESManagement SoftwareTysonBio Link Health ManagementSystem
Intended UseThe IN TOUCH DiabetesManagement Software is intended foruse in home and clinical settings toaid people with diabetes and theirhealth care professionals in thereview, analysis and evaluation bloodglucose test results to supporteffective diabetes management.TysonBio Link Health ManagementSystem is intended for use in home andclinical settings to aid people with diabetesand their health care professionals in thereview, analysis and evaluation bloodglucose test results to support effectivediabetes management. It should be notedthat the software does not recommendany medical treatment or medicationdosage level
ItemsIN TOUCH DIABETESManagement SoftwareTysonBio Link HealthManagement System
Cable availabilitySerial cable, cable availableseparatelySerial cable, cable available separately
Capable for uploadingdata from variousdevicesSoftware driver must be uploaded onthe device or installed on PCSoftware driver must be installed on PC
Computer systemrequirementsMinimum of 128 MB of RAMMinimum of 128 MB of RAM or better
100-200 MB minimum free hard diskspace during, 100 MB afterinstallationMinimum 100 MB of free hard disk spaceor more
9-pin or 25 pin COM or USB portUSB port
Video monitor and adapter with atleast 800x600 pixels screenresolution and 256 colorssame as
Software useindicationsHome or clinic settingsSingle or multiple user settings
Deleting patients andall accompanyingrecordsYesYes
Ability to uninstallprogramYesYes
Ability to personalizetarget rangesYesYes
Unit of measuredisplayChoice of mmol/L or mg/dLmmol/L or mg/dL
Ability to enterhypoglycemic rangeYesYes
Report TypesData list, data statistics, averagereading, pie chart, Line graph,histogram,Data list, data statistics, average reading,pie chart, Line graph, histogram,
Auto-detect COM portYesYes
Downloaded resultscannot be edited ordeletedYesYes
Ability to modify meteraverage resultsYes,7,14,30,60,90 daysYes
Ability to clear meterresults in memoryYesYes
Required informationon use entryYesYes, user ID, name, photo
ltemsIN TOUCH DIABETESManagement SoftwareTysonBio Link HealthManagement System
Computer systemrequirementsWindows operating system 98SE, Windows 2000, WindowsXP home and Professionalonly Windows 2000,Windows XP home andProfessional
Manual EntryAbility to add records manuallyNo
Copy database toseparate fileYesNo
Deleting resultsYesNo
Language capabilitiesEnglish, SpanishEnglish
Viewing the Owner'smanualLink provided via iconNo
Default glucose targetranges availableYesNo
Search patient capabilityYesNo
Result type displayChoice of whole blood orplasma result typesNo
Ability to set defaultfavorite reportYesNo
Ability to synchronizemeter clock to PC upondownloadYesNo
Ability to display 12 or 24hour clock formatYesNo
Ability to change dateformatYesNo
Ability to link to olderdatabase versionsYesNo old versions available atthis time
Diabetes EducatorinformationOne diabetes educator may beenteredNo
Diabetes controlInsulin list, medication list.diet/exercise optionsNo
Specifying/Enteringmedication/insulinYes, up to 3 differentmedications/insulin typesNo
Ability to enter insulinregimentYesNo
ItemsIN TOUCH DIABETESManagement SoftwareTysonBio Link HealthManagement System
Report typesStandard Day, AverageReading, Insulin, Exception,Data List Histogram, HealthchecksNo Insulin, Exception,standard day report.
Ability to set meter clockto a specific date andtimeYesNo
Ability to email reportfrom PC directly fromprogramYesNo
Ability to input additionalinformation on manualresultYesno

Similarities

{2}------------------------------------------------

:

{3}------------------------------------------------

Differences

:

{4}------------------------------------------------

9. Performance Data

The TysonBio Health Management System software has been developed in accordance with the FDA's Guidance for Content of Premarket Submission for Software Contained in Medical Devices(May 11,2005) and general Principles of Software Validation 1/11/2002 : Final Guidance for Industry and FDA Staff, where applicable and appropriate. Software verification and validation testing demonstrated that the TysonBio Health Management System meets the performance requirements for the intended use of the optional accessory device.

10. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision. Software verification and validation testing demonstrated that the TysonBio Health Management System meets the performance requirements for the intended use of the optional accessory device.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

AUG 2 6 2008

Tyson Bioresearch, Inc. c/o Mr. Wen-Hai Tsai 5 F., # 22, Ke Tung Rd. Science Based Industrial Park Chun-Nan, Miao-Li County China (Taiwan) 350

Re: K081726 Trade/Device Name: Tysonbio Link Health Management System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW, JOP Dated: June 16, 2008 Received: June 18, 2008

Dear Mr. Tsai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{6}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Jean M. Cooper, M.S., D.V.M.

Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indication for Use

510(k) Number (if known): K081726

TysonBio Link Health Management System Device Name:

Indication For Use:

The TysonBio Link Health Management System is an optional software accessory for use with the following models with data management capabilities: a) EZ Smart-608 series meter, b) DIACHEX Series Blood Glucose Meter, and c) DIACHEX+ Series Blood Glucose Meter. When used with one of these meters, TysonBio Link Health Management System transfers data from the device's memory into a computer for enhanced data management.

TysonBio Link Health Management System is intended for use in home and clinical settings to aid people with diabetes and their health care professionals in the review, analysis and evaluation blood glucose test results to support effective diabetes management. It should be noted that the software does not recommend any medical treatment or medication dosage level.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (21 CFR Part 801 Subpart D) And/Or Over the Counter Use _ X (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Division Sign-Off

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K081726

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.