K Number
K081655

Validate with FDA (Live)

Date Cleared
2008-07-21

(39 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

powder free vinyl patient examination gloves, (Light Yellow) that meets all of the requirements of ASTM standard D 5250-06.

AI/ML Overview

The provided text describes the 510(k) premarket notification for "Powder Free Vinyl Patient Examination Gloves (Light Yellow)." As such, the information contained does not pertain to the development or validation of a medical AI device. Instead, it details the safety and effectiveness information for a physical medical device (examination gloves) compared to a predicate device.

Therefore, many of the requested sections about AI device validation criteria, such as sample sizes for test and training sets, number of experts for ground truth, adjudication methods, multi-reader multi-case studies, and standalone performance, are not applicable to this document.

Here's an analysis of the available information based on the provided text, focusing on the acceptance criteria and the study (or tests) proving the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsStandard / Acceptance CriteriaReported Device Performance
DimensionASTM standard D 5250-06Meets
Physical PropertiesASTM standard D 5250-06Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06 and D6124-01<2mg/glove
Biocompatibility (Primary Skin Irritation)ISO10993-10Passes (in rabbits)
Biocompatibility (Dermal Sensitization)ISO10993-10Passes (in guinea pig), Not a Dermal sensitization

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify sample sizes for each test listed (e.g., how many gloves were tested for pinholes, or how many animals were used for biocompatibility). The studies are non-clinical, involving material performance and animal testing.

  • Provenance: Non-clinical tests, likely conducted in a laboratory setting by the manufacturer or contracted labs in China (submitter's country). These are a series of tests against established international and US standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This section is not applicable. For physical device performance and biocompatibility, ground truth is established through adherence to standardized test methods and criteria defined by regulatory bodies (e.g., ASTM, FDA, ISO). No human expert consensus for a "ground truth" in the way it's used for AI models is described or relevant here.

4. Adjudication Method for the Test Set

Not applicable. The tests are laboratory-based and follow predefined protocols for pass/fail criteria, rather than requiring adjudication by multiple human observers.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

Not applicable. This is not an AI device, and therefore no MRMC study or assessment of human reader improvement with AI assistance was conducted.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This is not an AI algorithm. The performance described is the standalone performance of the physical glove itself based on material science and manufacturing quality tests.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is based on:

  • Standardized Test Methods and Specifications: Adherence to defined parameters and limits set by ASTM (American Society for Testing and Materials), FDA regulations (e.g., 21 CFR 800.20 for freedom from pinholes), and ISO standards (e.g., ISO10993-10 for biocompatibility). These standards provide the objective criteria against which the device's performance is measured.
  • Predicate Device Comparison: The device is deemed substantially equivalent to a legally marketed predicate device (K070076) that also meets these standards.

8. The Sample Size for the Training Set

Not applicable. This is not an AI device, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable. There is no training set for a physical product like examination gloves. The "ground truth" for ensuring the product meets its specifications is established through quality control processes and adherence to manufacturing standards designed to consistently produce gloves meeting the performance criteria outlined in the standards (ASTM, FDA, ISO).

{0}------------------------------------------------

Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: KO81655 " (applicant leave blank)

21 2008

Premarket Notification [510(k)] Summary

[{a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

Submitter's name :JIANGSU CUREGUARD GLOVE CO., LTD.
Submitter's address :NO. 205 JIANLING ROAD, THE ECONOMICDEVELOPMENT ZONE OF SUQIAN CITY, JIANGSU223800 CHINA
Phone number :(86) 0527-4568102
Fax number :(86) 0527-4568100
Name of contact person:Mr. Zhang Bohoo
Date the summary was prepared:May.26.2008

[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Powder Free Vinyl Patient Examination Gloves ( Light Yellow)
Proprietary/Trade name:Rubber-Like
Common Name:Patient examination glove
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LYZ

[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* powder free vinyl patient examination gloves (Light Yellow) that meets all of the requirements of ASTM standard D 5250-06.

Predicate device : Powder-Free Vinyl Patient Examination Gloves, JIANGSU CUREGUARD GLOVE CO., LTD.. K070076

[(a)(4)} A description of the device

{1}------------------------------------------------

Device Description : powder free vinyl patient examination gloves, (Light Yellow) that meets all of the requirements of ASTM standard D 5250-06.

[(a)(5)] The summary describes the intended use of the device

Device Intended Use: powder free vinyl patient examination gloves, (Light Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 5250-06Meets
Physical PropertiesASTM standard D 5250-06Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06Meets
Biocompatabilityand D6124-01<2mg/glove
Primary Skin Irritation in rabbitsPasses
Dermal sensitization in the guinea pigPassesNot a Dermal sensitization

[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence ,

Powder free vinyl patient examination gloves , (Light Yellow) meet requirements per ASTM D5250-06, per ASTM D6124-01, per 21 CFR 800,20 and ISO10993-10.

[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

[(b)(3)] The conclusions drawn from the nonclinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL, meet labeling claims .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Jiangsu Cureguard Glove Company, Limited C/O Mr. Chu Xiaoan Beijing Easy-Link Company, Limited Room 1606 Building, 1 Jian Xiang Yuan No. 209 Bei Si Huan Zhong Road Hai Dian District, Beijing CHINA 100083

JUL 2 1 2008

Re: K081655

Trade/Device Name: Powder Free Vinyl Patent Examination Gloves (Light Yellow) Regulation Number. 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LYZ Dated: May 26, 2008 Received: June 12, 2008

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Qur

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

Applicant:_JIANGSU CUREGUARD GLOVE CO., LTD.

510(k) Number (if known): * KO8 | 655

Device Name:_Powder Free Vinyl Patient Examination Gloves (Light Yellow)

Indications For Use:

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Shula M. Murphy, MD
(Division Sign-Off)

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K081153

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.