Intended Use

The VERTEBRON PSS™ Pedicle Screw System is intended for non-cervical, nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients. The VERTEBRON PSS™ Pedicle Screw System is intended for non cervical pedicle fixation for the following indications: spondylolisthesis (Grade 3 and 4 at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment); trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

Device Description

The modified VERTEBRON PSS™ Pedicle Screw System is comprised of non-sterile, single use, titanium alloy components. The VERTEBRON PSS™ Pedicle Screw System attaches to the vertebral body by means of screws to the non-cervical spine and allows a surgeon to build a spinal implant construct. This system's design is intended to stabilize the spinal operative site during the fusion process of a bone graft in the disc space. The VERTEBRON PSS™ Pedicle Screw System is made up of rods, multi-axial and standard screws, hooks and locking caps. This submission adds Curved Rods and Screw Spacers.

AI/ML Overview

The information primarily concerns the regulatory approval of a medical device, specifically the VERTEBRON PSS Pedicle Screw System, based on substantial equivalence to predicate devices, rather than a study evaluating its performance against pre-defined acceptance criteria using AI or image analysis. Therefore, most of the requested fields are not applicable to the provided text.

Here's a breakdown of the available and unavailable information based on your request:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Not specified in the document. The approval is based on substantial equivalence, not performance against specific criteria for a novel AI/image analysis device.Not specified in the document. The document states that "Testing in accordance with ASTM F1717 was performed and demonstrated that the modified VERTEBRON PSS Pedicle Screw System is substantially equivalent to the currently marketed VERTEBRON PSS Pedicle Screw System". This indicates mechanical testing for substantial equivalence, but not specific performance metrics like accuracy, sensitivity, or specificity that would be set for an AI/image analysis device.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: Not applicable. The document describes a medical device (pedicle screw system), not an AI/image analysis software that would have a test set of data. The testing mentioned (ASTM F1717) refers to mechanical testing of the physical device.
  • Data Provenance: Not applicable.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

  • Number of Experts: Not applicable. This information would be relevant for establishing ground truth for an AI/image analysis device.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method:

  • Adjudication Method: Not applicable. This is typically used in studies where multiple experts interpret data to establish a ground truth, which is not described here.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • MRMC Study Done: No. The document describes a 510(k) submission for a physical medical device (Pedicle Screw System), not an AI/image interpretation device that would involve human readers.
  • Effect Size of Human Readers' Improvement with AI: Not applicable.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

  • Standalone Study Done: Not applicable. This device is a physical implant, not an algorithm.

7. Type of Ground Truth Used:

  • Type of Ground Truth: Not applicable. For the mechanical testing (ASTM F1717), the "ground truth" would be the engineering specifications and performance characteristics of the predicate devices. This is not the same as ground truth for an AI/image analysis system (e.g., pathology, expert consensus).

8. Sample Size for the Training Set:

  • Sample Size for Training Set: Not applicable. This device is a physical implant, not an AI/machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

  • Ground Truth for Training Set Establishment: Not applicable.

Summary based on the provided text:

The provided 510(k) summary for the VERTEBRON PSS Pedicle Screw System describes a submission for a physical medical device, not an AI or image analysis software. The device's approval is based on demonstrating substantial equivalence to existing predicate devices (VERTEBRON PSS Pedicle Screw System K071376, K051716, K043152 & K033352) through mechanical testing in accordance with ASTM F1717.

Therefore, the concepts of "acceptance criteria" and "study proving the device meets acceptance criteria" as defined by your request (referring to AI/image analysis performance metrics, ground truth, expert review, sample sizes for training/test sets, etc.) are not applicable to this document. The "study" mentioned is the mechanical testing to show substantial equivalence, not a clinical performance study with human readers or AI algorithms.

{0}------------------------------------------------

K081597

5. 510(k) Summary or 510(k) Statement

SUBMITTER:VERTEBRON Inc.80 Hathaway DriveStratford, CT 06615(203) 380-9340
CONTACT PERSON:Luis NespridoSenior Manager, Regulatory and Quality Affairs
DATE PREPARED:May 23, 2008
CLASSIFICATION NAME:21 CFR §888.3050 Spinal Interlaminal Fixation Orthosis21 CFR §888.3060 Spinal Intervertebral Fixation Orthosis21 CFR §888.3070 Pedicle Screw Spinal System
COMMON NAME:Pedicle Screw Spinal System
PROPRIETARY NAME:VERTEBRON PSS Pedicle Screw System
PREDICATE DEVICES:VERTEBRON PSS Pedicle Screw System (K071376,K051716, K043152 & K033352)
DEVICE DESCRIPTION:The modified VERTEBRON PSS™ Pedicle Screw Systemis comprised of non-sterile, single use, titanium alloycomponents. The VERTEBRON PSS™ Pedicle ScrewSystem attaches to the vertebral body by means of screwsto the non-cervical spine and allows a surgeon to build aspinal implant construct. This system's design is intendedto stabilize the spinal operative site during the fusionprocess of a bone graft in the disc space. TheVERTEBRON PSS™ Pedicle Screw System is made up ofrods, multi-axial and standard screws, hooks and lockingcaps. This submission adds Curved Rods and ScrewSpacers.
INTENDED USE:The VERTEBRON PSS™ Pedicle Screw System isintended for non-cervical, non-pedicle fixation for thefollowing indications: degenerative disc disease (definedas back pain of discogenic origin with degeneration of thedisc confirmed by history and radiographic studies);spondylolisthesis; trauma {i.e., fracture or dislocation);spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/orlordosis); tumor; pseudoarthrosis; and failed previousfusion in skeletally mature patients. The VERTEBRONPSS™ Pedicle Screw System is intended for non cervicalpedicle fixation for the following indications:spondylolisthesis (Grade 3 and 4 at L5-S1 or degenerative

0011

{1}------------------------------------------------

ko 81597

spondylolisthesis with objective evidence of neurologic impairment); trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

MATERIALS:

The material used is titanium Alloy material that conforms to ASTM F136.

SUBSTANTIAL EQUIVALENCE:

Testing in accordance with ASTM F1717 was performed and demonstrated that the modified VERTEBRON PSS Pedicle Screw System is substantially equivalent to the currently marketed VERTEBRON PSS Pedicle Screw System (K071376, K051716, K043152 & K033352).

1 of L

Page 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL - 2 2008

Vertebron, Inc. % Mr. Luis Nesprido 400 Long Beach Boulevard Stratford, CT 06615

K081597 Re: Trade/Device Name: VERTEBRON PSS Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNI, MNH, KWP, KWQ Dated: June 3, 2008 Received: June 6, 2008

Dear Mr. Nesprido:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can he found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Luis Nesprido

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark M. Miller

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

4. Indications for Use Statement

510(k) Number (if known):

VERTEBRON PSS Pedicle Screw System

Indications For Use:

Device Name:

The VERTEBRON PSS™ Pedicle Screw System is intended for non-cervical, nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients. The VERTEBRON PSS™ Pedicle Screw System is intended for non cervical pedicle fixation for the following indications: spondylolisthesis (Grade 3 and 4 at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment); trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

Prescription Use: × AND / OR (Per 21 CFR 801 Subpart D)

Over-The-Counter Use: (21 CFR 801 Subpart C)

(Please do not write below this line - continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of General. Restorative. and Neurological De