(244 days)
CADstream is intended to be used in the visualization, analysis, and reporting of magnetic resonance imaging (MRI) studies. CADstream supports evaluation of dynamic MR data acquired during contrast administration. CADstream performs other user selected processing functions (such as image registration, subtractions, measurements, 3D renderings, and reformats).
CADstream also includes user-configurable features for reporting on findings in breast or general MRI studies. Additionally, CADstream assists users in planning MRI guided interventional procedures.
When interpreted by a skilled physician, this device provides information that may be used for screening, diagnosis, and interventional planning. Patient management decisions should not be made based solely on the results of CADstream.
CADstream may also be used as an image viewer of multi-modality, digital images, including ultrasound and mammography. CADstream is not intended for primary interpretation of digital mammography images.
CADstream is an image processing system designed to assist in the visualization, analysis, and reporting of magnetic resonance imaging (MRI) studies. CADstream also is intended to provide workflow efficiency and interventional planning tools.
CADstream receives DICOM magnetic resonance images from a PACS or directly from the MRI scanner. As they are received, CADstream processes and displays the results on the CADstream server or a client personal computer.
Available features support:
- Visualization (standard image viewing tools, MIPs, and reformats)
- Analysis (registration, subtractions, coil inhomogeneity correction, kinetic curves, parametric image maps, and 3D volume rendering)
- Reporting of user-selected findings and assessment
- Interventional planning
- Workflow efficiency
- Communication and storage (DICOM import/export, query/retrieve, and study storage)
The CADstream system consists of proprietary software developed by Confirma installed on an off-the-shelf computer.
Here's a breakdown of the acceptance criteria and study information for the CADstream Version 5 device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The provided text does not explicitly state specific numerical acceptance criteria for performance metrics (e.g., sensitivity, specificity, accuracy, or quantitative measures of visualization/analysis improvement). Instead, the performance testing is described in a more general sense, focusing on verification and validation against requirements and substantial equivalence to predicate devices.
Therefore, the table will reflect the general statements made about performance rather than specific, measurable criteria.
| Acceptance Criteria (Implied from text) | Reported Device Performance |
|---|---|
| Device meets input requirements (Verification Testing) | CADstream version 5 has successfully undergone extensive verification testing to verify the device meets input requirements. |
| Product modifications do not raise new safety or effectiveness concerns | Bench validation testing was also performed to demonstrate the product modifications from version 4 resulted in a substantially equivalent product and did not raise any new safety or effectiveness concerns. |
| Device conforms to user needs and intended uses (Clinical Validation) | Additionally, the completed clinical validation testing demonstrates the device conforms to user needs and intended uses per 21CFR820.30(g). |
| Device is safe and effective | "The successful completion of verification and validation testing has demonstrated conformance to design controls, user needs, and intended use, and that the device is safe and effective." "Based on the information supplied in this 510(k), we conclude that the subject device is safe, effective, and substantially equivalent to the predicate devices." |
| Substantially equivalent to predicate devices | Bench validation testing demonstrated substantial equivalence to version 4 and other predicate devices in terms of features (as shown in Table 1 and Table 2 comparison) and overall safety and effectiveness. The 510(k) summary explicitly concludes substantial equivalence to predicate devices based on the submitted information. |
| Provides features for visualization, analysis, and reporting of MR images | Device description and feature comparison tables illustrate the capability to perform: - Visualization (standard image viewing tools, MIPs, reformats, 3D volume rendering) - Analysis (registration, subtractions, coil inhomogeneity correction, kinetic curves, parametric image maps) - Reporting of user-selected findings and assessment - Interventional planning - Workflow efficiency - Communication and storage (DICOM import/export, query/retrieve, and study storage) |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size for Test Set: The document does not specify the sample size (number of patients or cases) used for the "clinical validation testing."
- Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Number of Experts: The document does not specify the number of experts used to establish ground truth for the test set.
- Qualifications of Experts: The document does not specify the qualifications of any experts involved in establishing ground truth. The "Indications for Use" section mentions interpretation "by a skilled physician," but this is a general statement about clinical use, not a description of ground truth establishment during testing.
4. Adjudication Method for the Test Set:
- The document does not mention any specific adjudication method (e.g., 2+1, 3+1, none) used for the test set.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
- The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it discuss the effect size of human readers' improvement with or without AI assistance. The focus is on device features and general validation, not comparative reader performance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:
- The document describes "Performance Testing" including "extensive verification testing," "bench validation testing," and "clinical validation testing." While these tests would involve evaluating the algorithm's functionality, the document does not explicitly state that a standalone performance study comparing the algorithm's diagnostic accuracy against a ground truth was conducted without a human in the loop. The device is described as an "image processing system designed to assist," implying a human-in-the-loop context for its intended use and evaluation.
7. The Type of Ground Truth Used:
- The document does not explicitly state the type of ground truth used for any testing (e.g., expert consensus, pathology, outcomes data). It broadly refers to "user needs and intended uses" for clinical validation.
8. The Sample Size for the Training Set:
- The document does not mention a training set or its sample size. This is typical for 510(k) submissions of this era for image processing systems, as they often leverage established algorithms or feature sets rather than reporting on deep learning model training.
9. How the Ground Truth for the Training Set Was Established:
- Since a training set is not mentioned, the document does not describe how ground truth for a training set was established.
{0}------------------------------------------------
K08 1556
FEB -2 2009
510(k) Summary 5
This 510(k) summary is submitted as part of the PreMarket Notification in accordance with the requirements of SMDA 1990, 21 CFR 807.87(h), and 21 CFR 807.92.
| 1. | Date Prepared | December 16, 2008 |
|---|---|---|
| 2. | 510(k) Submitter | Confirma, Inc.11040 Main Street, Suite 100Bellevue, WA 98004Phone: 425-691-1400Fax: 425-691-1599 |
| 3. | 510(k) Contact Person | Brent Lewis, Director RegulatoryAffairs and Quality AssuranceEmail: blewis@confirma.comPhone: 206-455-5398Fax: 425-691-1599 |
| 4. | Device Common Name | Image Processing System |
| 5. | Device Trade Name | CADstream Version 5 |
| 6. | Classification Regulation | 21 CFR 892.2050 |
| 7. | Class | 2 |
| 8. | Panel | Radiology Panel |
| 9. | Product Code | LLZ |
10. Comparison with Legally Marketed Devices
As with the predicate devices listed in Table 1, CADstream version 5 is indicated for use in the visualization, analysis, reporting, interventional planning, and improved workflow efficiency of magnetic resonance imaging (MRI) studies and supports the evaluation of dynamic MR data acquired during contrast administration. All these devices are intended to provide information that may be useful in screening or diagnosis.
Table 1: Predicate Devices
| Manufacturer | Product | Cleared 510(k) | Product Code | Classification |
|---|---|---|---|---|
| Confirma | CADstream 5 | Pending | LLZ | |
| CADstream 4 | K043216 | LLZ | ||
| CADstream 2 | K031779 | LLZ | ||
| Invivo | DynaCAD 1.0 | K041286 | LZ | |
| CADsciences | 3TP | K031350. K050862 | LIZ | |
| GE Medical | AW & Functool | K960265 | 17 | |
| Clario | Z3D Contrast Acuity | K080196 | LL7 |
{1}------------------------------------------------
Table 2 presents a detailed comparison of features in CADstream version 5 to the predicate devices.
| CADstream 5 Features | Predicate Devices | ||||
|---|---|---|---|---|---|
| CADstream2 & 4 | DynaCAD1.0 | AW &Functool | 3TP | Z3D | |
| General Body MR Functions | |||||
| Standard image viewing tools | X | X | X | X | X |
| View/play clips from ultrasoundimages | X | 0 | 0 | 0 | 0 |
| MIPs | X | X | 0 | X | X |
| Reformats | X | X | 0 | 0 | X |
| Registration | X | X | 0 | X | X |
| Subtraction series | X | X | 0 | X | X |
| Coil inhomogeneity correction | X | 0 | 0 | 0 | |
| View 3D volume rendering | X | X | X | 0 | X |
| Kinetic curves | X | X | X | X | X |
| Parametric image maps | X | X | X | X | X |
| Portfolio of user-selected findings | X | X | X | X | X |
| DICOM import/export andquery/retrieve | X | X | X | 0 | 0 |
| User-Customizable Reporting | X | X | 0 | X | X |
| Breast Specific: Module Features | |||||
| Reporting for Breast MR Studies | X | X | 0 | X | 0 |
| View/invert mammography images | X | X | 0 | 0 | 0 |
| View/edit finding volume | X | X | 0 | X | 0 |
| View/edit finding location | X | X | 0 | X | 0 |
| View/edit finding size | X | X | 0 | X | 0 |
| View/edit kinetic curve with highestuptake | X | X | 0 | X | 0 |
| View images and findings from twostudies from same patient | X | X | 0 | X | 0 |
| Interventional planning | X | X | 0 | 0 | 0 |
Table 2: Predicate Device Feature Comparison
Key
Substantially equivalent feature cleared in predicate device. X
0 Not known from literature.
11. Device Description
CADstream is an image processing system designed to assist in the visualization, analysis, and reporting of magnetic resonance imaging
{2}------------------------------------------------
(MRI) studies. CADstream also is intended to provide workflow efficiency and interventional planning tools.
CADstream receives DICOM magnetic resonance images from a PACS or directly from the MRI scanner. As they are received, CADstream processes and displays the results on the CADstream server or a client personal computer.
Available features support:
- Visualization (standard image viewing tools, MIPs, and 미 reformats)
- Analysis (registration, subtractions, coil inhomogeneity B correction, kinetic curves, parametric image maps, and 3D volume rendering)
- Reporting of user-selected findings and assessment ■
- Interventional planning 트
- Workflow efficiency l
- Communication and storage (DICOM import/export, query/retrieve, and study storage)
The CADstream system consists of proprietary software developed by Confirma installed on an off-the-shelf computer.
12. Indications for Use
CADstream is intended to be used in the visualization, analysis, and reporting of magnetic resonance imaging (MRI) studies. CADstream supports evaluation of dynamic MR data acquired during contrast administration. CADstream performs other user selected processing functions (such as image registration, subtractions, measurements, 3D renderings, and reformats).
CADstream also includes user-configurable features for reporting on findings in breast or general MRI studies. Additionally, CADstream assists users in planning MRI guided interventional procedures.
When interpreted by a skilled physician, this device provides information that may be used for screening, diagnosis, and interventional planning. Patient management decisions should not be made based solely on the results of CADstream.
{3}------------------------------------------------
CADstream may also be used as an image viewer of multi-modality, digital images, including ultrasound and mammography. CADstream is not intended for primary interpretation of digital mammography images.
13. Performance Testing
CADstream version 5 has successfully undergone extensive verification testing to verify the device meets input requirements. Bench validation testing was also performed to demonstrate the product modifications from version 4 resulted in a substantially equivalent product and did not raise any new safety or effectiveness concerns. Additionally, the completed clinical validation testing demonstrates the device conforms to user needs and intended uses per 21CFR820.30(g).
14. Conclusion
CADstream Version 5 provides features to integrate radiology department workflow by facilitating the visualization, analysis, and reporting of MR images. The potential hazards have been studied and controlled as part of the product development process, including risk analysis and design considerations. The successful completion of verification and validation testing has demonstrated conformance to design controls, user needs, and intended use, and that the device is safe and effective.
Based on the information supplied in this 510(k), we conclude that the subject device is safe, effective, and substantially equivalent to the predicate devices.
{4}------------------------------------------------
Truthful and Accurate Statement
PREMARKET NOTIFICATION
TRUTHFUL AND ACCURATE STATEMENT
[As required by 21 CFR 807.87(k)]
I certify that, in my capacity as Acting Director of Regulatory Affairs at Confirma, Inc., I believe to the best of my knowledge, that all data and information submitted in this 510(k) PreMarket Notification are truthful and accurate and that no material fact has been omitted.
ZZZ.
Brent Lewis Acting Director, Regulatory Affairs Confirma, Inc.
02 Jun 2008
Date
02Jun2008
6
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB - 2 2009
Mr. Brent Lewis Director, Regulatory Affairs and Quality Assurance Confirma, Incorporated 11040 Main Street. Suite 100 BELLEVUE WA 98004
Re: K081556
Trade/Device Name: CADstream@ Version 5 Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications systems Regulatory Class: II Product Code: LLZ
Dated: December 23, 2008 Received: December 23, 2008
Dear Mr. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|---|---|---|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry.suppot/index.html.
Sincerely yours,
Janine M. Morris
cting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indication(s) for Use Statement
510(k) Number: K081556
Device Name: CADstream@ Version 5
Indications for Use:
CADstream is intended to be used in the visualization, analysis, and reporting of magnetic resonance imaging (MRI) studies. CADstream supports evaluation of dynamic MR data acquired during contrast administration. CADstream performs other user selected processing functions (such as image registration, subtractions, measurements, 3D renderings, and reformats).
CADstream also includes user-configurable features for reporting on findings in breast or general MRI studies. Additionally, CADstream assists users in planning MRI guided interventional procedures.
When interpreted by a skilled physician, this device provides information that may be used for screening, diagnosis, and interventional planning. Patient management decisions should not be made based solely on the results of CADstream.
CADstream may also be used as an image viewer of multi-modality, digital images, including ultrasound and mammography. CADstream is not intended for primary interpretation of digital mammography images.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---|---|
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal and | |
| Radiological Devices | |
| 510(k) Number | K081656 |
4