K Number
K081485

Validate with FDA (Live)

Device Name
SOMNOWATCH
Manufacturer
Date Cleared
2008-09-18

(113 days)

Product Code
Regulation Number
868.2375
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SOMNOwatch is a non-life-supporting portable physiological signal recording device intended to be used for testing adult patients suspected of having movement-correlated sleep disturbances.

Device Description

The SOMNOwatch is a small, portable physiological signal recording system intended to be used to record, display, monitor, print and store biophysical events to aid in the diagnosis of sleep disorders. The device is intended to be prescribed for use by a physician in the office, sleep laboratory or patient's home.

The SOMNOwatch is a small, typically wrist-worn activity monitor. The device is intended to be used to analyze circadian rhythms, automatically collect and score data for sleep parameters. These parameters, representing the number and intensity of limb movements, are directly associated to movement-correlated sleep disturbances. The unit can also be used to assess activity in any instance where quantifiable analysis of physical motion is desired. For PLM-detection two identical SOMNOwatches may be affixed to the patients legs, one to each leg.

AI/ML Overview

The SOMNOwatch 510(k) Premarket Application (K081485) describes the device, its intended use, and a summary of nonclinical and clinical testing. However, it does not explicitly define acceptance criteria in terms of specific performance metrics (e.g., sensitivity, specificity, accuracy) or provide a detailed study that proves the device meets such criteria. Instead, the submission focuses on establishing substantial equivalence to predicate devices primarily through nonclinical performance testing and a "clinical comparison" study, without providing quantitative results against predefined thresholds.

Based on the provided text, here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

The provided document does not contain explicit numerical acceptance criteria or quantifiable performance metrics (like sensitivity, specificity, or accuracy for diagnosing sleep disturbances) for the SOMNOwatch itself. The "reported device performance" is broadly stated as:

Acceptance Criteria (Explicitly Stated)Reported Device Performance
Compliance to device specificationsAll functions were verified to operate as designed.
Compliance to international standards for electrical safety and electromagnetic compatibilityFound to be compliant with the requirements of these standards for its intended use.
Safety and effectiveness outcomes substantially equivalent to predicate devicesClinical comparison studies found the subject device can be expected to provide safety and effectiveness outcomes substantially equivalent to the predicates.

2. Sample sized used for the test set and the data provenance

  • Test Set Sample Size: Not specified. The document only mentions "clinical comparison studies" without detailing the number of participants or the data size.
  • Data Provenance: Not specified. It's unclear if the data was retrospective or prospective, or the country of origin.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not specified. The document does not describe how ground truth was established for any clinical comparison.

4. Adjudication method for the test set

Not specified. There is no mention of an adjudication method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was described. The SOMNOwatch is an activity recording device that provides raw data and analysis software (DOMINOlight) but the document does not discuss human reader performance improvement with or without AI assistance. The focus is on the device's ability to record and score data for sleep parameters related to movement-correlated sleep disturbances.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device is intended to "automatically collect and score data for sleep parameters," implying a standalone algorithmic analysis. The "DOMINOlight software retrieves the data from the SOMNOwatch, displays the data, and can store data for future reference and comparison. DOMINOlight also allows automatic analysis of all signals including the body position." However, no specific standalone performance metrics (e.g., for automated PLM detection accuracy) are provided in this summary. The evaluation focuses on equivalence to predicate devices rather than direct algorithmic performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Not specified. The document does not describe the specific type of ground truth used in its "clinical comparison studies." Given the device's function to "automatically collect and score data for sleep parameters" related to movement, it's possible that the comparison was against similar data from predicate devices or established polysomnography (PSG) techniques, but this is not explicitly stated.

8. The sample size for the training set

Not applicable/Not specified. The document does not mention a training set, as it does not describe the development or validation of a new AI/ML algorithm through traditional training/testing splits with a separate training set. The clinical evaluation mentioned is a "comparison study" to predicate devices for substantial equivalence.

9. How the ground truth for the training set was established

Not applicable/Not specified as no training set was described.

{0}------------------------------------------------

SOMNOwatch 510(k) Premarket Application

Section 5: 510(k) Summary

5681485 SEP 1 8 2008

510(k) Summary

Applicant:

Somnomedics GmbH Am Sonnenstuhl 63 Randersacker, Germany D-97236 Phone: +49 931 35 90 94 0 Facsimile: +49 931 35 90 94 49

US Contact:

M Squared Associates, Inc. Cherita James 901 King Street, Suite 200 Alexandria, Virginia 22314 Phone: 703-562-9800 ext. 257 Facsimile: 703-562-9797 E-mail: CJames(amsquaredassociates.com

May 28, 2008 Date submitted:

Proprietary Name:Somnomedics SOMNOwatch
Common Name:Activity Recording Device
Classification Status:Class II per regulations §882.1400, §868.2375
Product Codes:GWQ, MNR

Establishment Registration Number: Applied for, not yet assigned

Manufacturing/

Distribution Address:

Somnomedics GmbH Am Sonnenstuhl 63 Randersacker, Germany D-97236

Predicate Devices:

  • ActiTrac (K992410); Individual Monitoring Systems . (IM-systems), Inc.
  • SOMNOscreen (K060708), SOMNOmedics GmbH .

{1}------------------------------------------------

Device Description

The SOMNOwatch is a small, portable physiological signal recording system intended to be used to record, display, monitor, print and store biophysical events to aid in the diagnosis of sleep disorders. The device is intended to be prescribed for use by a physician in the office, sleep laboratory or patient's home.

This device is NOT designed to be used in a Life Support situation.

Indication for Use

The SOMNOwatch is a non-life-supporting portable physiological signal recording device intended to be used for testing adult patients suspected of having movement-correlated sleep disturbances.

Intended Use

The SOMNOwatch is a small, typically wrist-worn activity monitor. The device is intended to be used to analyze circadian rhythms, automatically collect and score data for sleep parameters. These parameters, representing the number and intensity of limb movements, are directly associated to movement-correlated sleep disturbances. The unit can also be used to assess activity in any instance where quantifiable analysis of physical motion is desired. For PLM-detection two identical SOMNOwatches may be affixed to the patients legs, one to each leg.

Technical Specifications of the SOMNOwatch

CHANNELS6 Internal Channels:(3 Activity Sensors for X-Y-Z Axis,Body Position, Ambient Light, Patient Marker)
DATA PROCESSING12 Bit ADCDifferent Sampling Rates Adjustable (1/120s - 128/s)Different Storage Rates Adjustable (1/120s - 128/s)

Summary of Technological Characteristics

The SOMNOwatch utilizes a motion sensor, an accelerometer ,to monitor the occurrence and degree of motion. This type of sensor provides an analog signal where the amplitude and speed of motion produces a signal whose magnitude and duration depend on the amount of motion. The activity signals are amplified and digitized by on-board circuits within the device. This information is stored in the internal memory of the device. The system provides

{2}------------------------------------------------

up to 6 channels for data acquisition; 3 Acceleration for X-Y-Z Axis, Ambient Light, Patient Marker and Body Position.

Information is stored on an internal 8 MB Flash and can be transferred to a PC via a USB docking station. The DOMINOlight software retrieves the data from the SOMNOwatch, displays the data, and can store data for future reference and comparison. DOMINOlight also allows automatic analysis of all signals including the body position.

Summary of Nonclinical Testing

Performance testing was conducted to confirm compliance to device specifications. All functions were verified to operate as designed. Testing to the international standards for electrical safety and electromagnetic compatibility were performed. The SOMNOwatch was found to be compliant with the requirements of these standards for it's intended use.

Clinical Testing

Comparison studies of the SOMNOwatch and the predicate devices found the subject device can be expected to provide safety and effectiveness outcomes substantially equivalent to the predicates.

Conclusion

SOMNOmedics SOMNOwatch has the same principles of operation and similar technological characteristics as the Actitrac. While the principle of operation and method of data collection differ from the SOMNOscreen, the differences do not present new issues of safety or effectiveness as demonstrated in the clinical comparison.

Based on performance testing, the SOMNOmedics SOMNOwatch is substantially equivalent to devices already on the market and presents no new concerns of safety and effectiveness. Additionally, the device has similar indications to the predicate devices and the labeling of the device is consistent both with FDA's guidance as well as current medical practice.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three wing-like shapes extending upwards. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

SEP 18 2008

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Somnomedics GmbH C/O Ms. Cherita James Regulatory Consultant M Squared Associates, Incorporated 901 King Street, Suite 200 Alexandria, Virginia 22314

Re: K081485

Trade/Device Name: SOMNOwatch Regulation Number: 21 CFR 868.2376 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: May 28, 2008 Received: May 28, 2008

Dear Ms. James:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include controls provide of annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 wrt), it may of election in the Code of Federal Regulations, Title 21, Parts 800 to 898. In your device our of found in firsher announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. James

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Dr. Samuels Ludewig foey

Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosures

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: SOMNOwatch

Indications For Use: The SOMNOwatch is a non-life-supporting portable physiological signal recording device intended to be used for testing adult patients suspected of having movement-correlated sleep disturbances.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M Th

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number. K081485

§ 868.2375 Breathing frequency monitor.

(a)
Identification. A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.(b)
Classification. Class II (performance standards).