K Number
K081388

Validate with FDA (Live)

Date Cleared
2008-08-21

(94 days)

Product Code
Regulation Number
868.1890
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GlucoCare IGC System is a software program indicated for the management of patient blood glucose levels in the hospital setting. The GlucoCare IGC System software, using an algorithm that is based on an established clinical protocol, calculates the dosage of intravenously administered insulin required to maintain the blood glucose level within a target range set by the protocol. No medical decisions are made by the GlucoCare IGC System software. The information provided by the software is intended to be used as a tool by qualified and trained medical personnel. The GlucoCare IGC System software is intended for use in patients over the age of 18 with no known insulin allergies. The device is not intended for use in patients with diabetic ketoacidosis.

Device Description

The GlucoCare IGC System is a software program that implements an insulin infusion protocol intended for the treatment of hyperglycemic adult patients in a healthcare facility. The software directs insulin infusion rates based on the patient's blood glucose level history and target blood glucose level. Audible and visual alerts remind the medical staff when it is time to take blood glucose readings or change insulin drip rates. The system provides a complete time/date/username-stamped history of all patient activity in text and graphical formats, both on-screen and in a printable form. The GlucoCare IGC System software resides on the GlucoCare application server. Local client computers interact with the GlucoCare IGC System software using HTTPS.

AI/ML Overview

The provided text does not contain specific acceptance criteria or a detailed study proving the device meets those criteria. The information focuses on establishing substantial equivalence to predicate devices rather than presenting performance metrics against predefined acceptance thresholds.

However, based on the limited information available, here's a breakdown of what can be inferred or explicitly stated about performance testing, and the missing elements:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Inferred from general software testing)Reported Device Performance
Functional Correctness: The software accurately calculates insulin dosages based on blood glucose levels and protocols."The results of the testing confirmed that the GlucoCare IGC System performed as intended." (General statement, no specific metrics)
Reliability/Stability: The software operates consistently without crashes or errors."The results of the testing confirmed that the GlucoCare IGC System performed as intended." (General statement)
Usability/User Interface: The system provides clear alerts and historical data."Audible and visual alerts remind the medical staff when it is time to take blood glucose readings or change insulin drip rates. The system provides a complete time/date/username-stamped history of all patient activity in text and graphical formats, both on-screen and in a printable form." (Feature description, not performance metric)
Security: Data integrity and security for patient information.Mention of "HTTPS" for client-server interaction implies some security consideration, but no specific performance or acceptance criteria are given.
System Testing: Overall system functions as expected."Software verification and validation testing activities were conducted to establish the performance, functionality, and reliability characteristics of the GlucoCare IGC System. Testing included unit testing, integration testing, system testing, and functional testing." (Methodology description, not specific results)
Accuracy of Insulin Dosage Calculation (Clinical Performance): (Expected for this type of device, but not detailed)Not provided. The text states "No medical decisions are made by the GlucoCare IGC System software. The information provided by the software is intended to be used as a tool by qualified, trained medical personnel." This indicates a lack of direct clinical outcome measurement for the device's performance, as the final decision rests with medical personnel.
Time within Target Blood Glucose Range: (Relevant clinical outcome, but not detailed for this software)Not provided.
Incidence of Hypoglycemia/Hyperglycemia: (Relevant safety outcome, but not detailed for this software)Not provided.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: Not specified. The document only mentions "Software verification and validation testing activities." There is no mention of a clinical test set with patient data.
  • Data Provenance: Not applicable, as there's no mention of using real patient data for "testing" in the sense of a clinical or retrospective patient study. The testing described appears to be software testing (unit, integration, system, functional).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • Not applicable. The document describes software verification and validation, not a clinical study where expert-established ground truth would be relevant for a test set. The algorithm is based on "an established clinical protocol," implying that domain experts (likely endocrinologists or critical care physicians) defined the protocol that the software implements, but this is distinct from establishing ground truth for a test set of the software's performance.

4. Adjudication Method for the Test Set:

  • Not applicable. There is no mention of a test set requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, a MRMC comparative effectiveness study was not done. The text does not describe any study comparing human readers (medical personnel) with and without AI (the GlucoCare IGC System) assistance. The device is described as a "tool" to assist medical personnel, implying it's always used with human-in-the-loop, rather than in comparison to unassisted human performance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

  • Yes, in the context of software performance, a standalone evaluation was implicitly done. The "Software verification and validation testing activities" including unit, integration, system, and functional testing are evaluations of the algorithm's correctness and functionality on its own, independent of live human interaction. However, this is not a standalone clinical performance study regarding patient outcomes. The document explicitly states: "No medical decisions are made by the GlucoCare IGC System software. The information provided by the software is intended to be used as a tool by qualified, trained medical personnel." This indicates the device is not intended for standalone clinical decision-making.

7. Type of Ground Truth Used:

  • For the software testing described, the "ground truth" would be the expected output or behavior of the software based on its design specifications and the underlying established clinical protocol. This is internal software validation, not a clinical ground truth like pathology or outcomes data.

8. Sample Size for the Training Set:

  • Not applicable. This device is described as implementing "an established clinical protocol" using an algorithm. It is not an AI/ML model that would typically require a "training set" in the machine learning sense to learn patterns. The algorithm's rules are pre-defined based on clinical knowledge.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable, as there is no training set in the typical AI/ML sense. The "ground truth" for the algorithm's logic would be rooted in established medical consensus and clinical guidelines that informed the "established clinical protocol" it implements.

{0}------------------------------------------------

K081388

510(k) Summary for the Pronia Medical Systems, LLC GlucoCare IGC System

1. SUBMITTER/510(K) HOLDER

AUG 21 2008

Pronia Medical Systems, LLC 7527 Beechspring Farm Boulevard Louisville, KY 40241

Contact Person: Brian J. Besterman, M.D. Telephone: (914) 261-6622

May 16, 2008 Date Prepared:

2. DEVICE NAME

Proprietary Name:GlucoCare IGC System
Common/Usual Name:Dose Calculation Software
Classification Name:Drug Dose Calculator
Classification:Class II (per 21 CFR 868.1890)
Product Code:NDC

3. PREDICATE DEVICES

  • EndoTool™ Glucose Management System, K053137 .
  • . MDN-CGS™ Insulin Dosing Calculator, K071713

4. DEVICE DESCRIPTION

The GlucoCare IGC System is a software program that implements an insulin infusion protocol intended for the treatment of hyperglycemic adult patients in a healthcare facility. The software directs insulin infusion rates based on the patient's blood glucose level history and target blood glucose level.

Audible and visual alerts remind the medical staff when it is time to take blood glucose readings or change insulin drip rates. The system provides a complete time/date/username-stamped history of all patient activity in text and graphical formats, both on-screen and in a printable form.

{1}------------------------------------------------

The GlucoCare IGC System software resides on the GlucoCare application server. Local client computers interact with the GlucoCare IGC System software using HTTPS,

5. INTENDED USE

The GlucoCare IGC System is a software program indicated for the management of patient blood glucose levels a healthcare facility setting. The GlucoCare IGC System software, using an algorithm that is based on an established clinical protocol, calculates the dosage of insulin required to maintain the blood glucose level within a target range set by the protocol. No medical decisions are made by the GlucoCare IGC System software. The information provided by the software is intended to be used as a tool by qualified, trained medical personnel. The GlucoCare IGC System software is intended for use in adults over the age of 18 with no known insulin allergies. The device is not intended for use in patients with diabetic ketoacidosis.

TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6.

The Pronia Medical Systems GlucoCare IGC System and the predicate devices are all software programs that dictate insulin dosing instructions based on the patient's blood glucose levels over time. The proposed and predicate software all issue alerts and provide instructions for the timing and frequency of blood glucose testing and the need to infuse glucose to prevent hypoglycemia.

The proposed and predicate devices have the same general intended use and similar indications, technological characteristics, and principles of operation. Any minor technological differences between the GlucoCare IGC System and its predicates do not present any new issues of safety or effectiveness. Thus, the GlucoCare IGC System is substantially equivalent to the EndoTool™ Glucose Management System (K053137) and the MDN-CGS™ Insulin Dosing Calculator (K071713). A comparison of the intended use and technological characteristics of the proposed and predicate devices is provided in Table 5-1.

7. PERFORMANCE TESTING

Software verification and validation testing activities were conducted to establish the performance, functionality, and reliability characteristics of the GlucoCare IGC System. Testing included unit testing, integration testing, system testing, and functional testing. The results of the testing confirmed that the GlucoCare IGC System performed as intended.

{2}------------------------------------------------

CategoryGlucoCare IGC SystemEndoTool™ GlucoseManagement System,K053137MDN-CGSTM InsulinDosing Calculator,K071713
Intended UseIntended for use by trained medical personnel for the management of patient bloodglucose levels in the hospital setting
PatientPopulationVariety of different patientpopulationsCritically ill patients oncontinuous feedingVariety of different patientpopulations
Principle ofOperationUsing blood glucose measurements, software algorithms determine appropriate insulindosing in accordance with established treatment protocols
HardwareWeb based network withPC running web browserWeb based network withWindows PCStandalone Windows PC
FeaturesProvides alerts forsubsequent blood glucosetesting and monitoringProvides alerts forsubsequent blood glucosetesting and monitoringProvides alerts forsubsequent blood glucosetesting and monitoring

Table 5-1. Comparison Table for Determination of Substantial Equivalence

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an emblem featuring a stylized eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Pronia Medical System, LLC C/O Cynthia J. M. Nolte, Ph.D., RAC Senior Regulatory Consultant Medical Device Consultants, Incorporated 49 Plain Street North Attleboro, Massachusetts 02760

AUG 2 1 2008

Re: K081388

Trade/Device Name: Pronia Medical Systems, LLC, GlucoCare Intensive Glycemic Control System Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive Pulmonary-Function Value Calculator Regulatory Class: II Product Code: NDC Dated: August 6, 2008 Received: August 7, 2008

Dear Dr. Nolte:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 – Dr. Nolte

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number: K081388

Pronia Medical Systems, LLC GlucoCare IGC System Device Name:

Indications for Use:

The GlucoCare IGC System is a software program indicated for the management of patient blood glucose levels in the hospital setting. The GlucoCare IGC System software, using an algorithm that is based on an established clinical protocol, calculates the dosage of intravenously administered insulin required to maintain the blood glucose level within a target range set by the protocol. No medical decisions are made by the GlucoCare IGC System software. The information provided by the software is intended to be used as a tool by qualified and trained medical personnel. The GlucoCare IGC System software is intended for use in patients over the age of 18 with no known insulin allergies. The device is not intended for use in patients with diabetic ketoacidosis.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Centum Vm
(Division Sign-Off)

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K
$$\phi$$
81288

§ 868.1890 Predictive pulmonary-function value calculator.

(a)
Identification. A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.(b)
Classification. Class II (performance standards).