K Number
K081259

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-06-15

(409 days)

Product Code
Regulation Number
870.5310
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DDU-2300 Semiautomatic External Defibrillator (AED) is indicated for use on victims of sudden cardiac arrest (SCA) who are:

  • · Unconscious and unresponsive
  • · Not breathing

For patients under 8 years old, or less than 55 pounds (25 kg), use child/infant electrode pads. Do not delay therapy to determine exact age or weight.

The DDU-2300 AED must be used by or on the order of a physician.

Device Description

The DDU-2300 is a portable, Automated External Defibrillator (AED) intended for use on victims of sudden cardiac arrest (SCA). It is powered by a user-replaceable nonrechargeable battery and supports both adult and pediatric user-replaceable single-use defibrillation/monitoring pads.

The DDU-2300 employs a Patient Analysis System that ensures proper pad/patient connection and analyzes the patient's ECG rhythm to determine whether a shock is required. If needed the DDU-2300 provides a 150 J (50J pediatric) impedance compensated, biphasic truncated exponential defibrillation shock to the patient. The cardiac rhythm analysis algorithm and defibrillation energy and waveform utilized are the same as previous Defibtech AEDs.

The DDU-2300 has a compact design and offers an improved user interface with an LCD display. Voice prompts and a graphical user interface provide simple instructions for the operator. The DDU-2300 AED is capable of recording event information including ECG, audio data and SHOCK/NO SHOCK recommendations.

AI/ML Overview

The provided text describes a 510(k) summary for the Defibtech DDU-2300 Semiautomatic External Defibrillator (AED). However, it does not contain the detailed information necessary to complete most of the requested fields regarding acceptance criteria and the study that proves the device meets them.

The document states that "Testing demonstrates that the DDU-2300 meets functional and performance specifications" and refers to "Safety testing assur[ing] compliance with applicable industry safety standards" and "performance evaluations demonstrate that the Defibtech DDU-2300 AED is substantially equivalent to the predicate device." However, it does not provide specific acceptance criteria values, reported performance metrics, or details about the studies conducted (sample size, data provenance, ground truth, experts, adjudication, or standalone performance).

The primary focus of this 510(k) summary is to demonstrate substantial equivalence to a predicate device based on similar technology and intended use, rather than presenting a detailed performance study with specific acceptance criteria.

Therefore, for most of your requested points, the answer is "Information not provided in the document."

Here's a breakdown of what can and cannot be answered based on the provided text:


1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Not provided. The document states, "Testing demonstrates that the DDU-2300 meets functional and performance specifications." However, it does not specify what those specifications or acceptance criteria are, nor does it provide quantitative performance results from such testing.Not provided. The document only states that the device meets unspecified "functional and performance specifications."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Information not provided.
  • Data Provenance: Information not provided.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Information not provided. The document does not describe the specific ground truth establishment process or the involvement of experts for any test sets.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Information not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Information not provided. The document highlights the automated nature of the device's analysis system (Patient Analysis System that ensures proper pad/patient connection and analyzes the patient's ECG rhythm to determine whether a shock is required) but does not mention any MRMC studies involving human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • The device is an "Automated External Defibrillator (AED)" and employs a "Patient Analysis System" and "cardiac rhythm analysis algorithm." While the document states that the "cardiac rhythm analysis algorithm and defibrillation energy and waveform utilized are the same as previous Defibtech AEDs" and that the DDU-2300 "replaces the LED/text display with an LCD display," it does not provide details or results of a standalone performance study. It only asserts that the DDU-2300 "uses similar technologies to provide functionally equivalent performance characteristics as the predicate device" and "Testing demonstrates that the DDU-2300 meets functional and performance specifications."

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Information not provided.

8. The sample size for the training set

  • Information not provided.

9. How the ground truth for the training set was established

  • Information not provided.

{0}------------------------------------------------

K081259

Section 5

10202 rev B

510(k) Summary

510(k) Summary

Date Summary Prepared: April 30, 2008

JUN 1 5 2009

510(k) Owner Information: Defibtech, LLC 741 Boston Post Road Guilford, CT 06437

Contact Information:

Mr. Ed Horton 203-453-6651, extension 138 (phone) 203-453-6657 (fax) ehorton@defibtech.com (e-mail)

Trade (Proprietary) Name: DDU-2300 Semiautomatic External Defibrillator Common Name: Semiautomatic External Defibrillator Classification Name: Automated External Defibrillator (21 CFR 870.5310, Product Code MKJ)

Substantial Equivalence Model

The design and intended use of the DDU-2300 AED is substantially equivalent in performance and safety to the Sentry (DDU-100) AED with Adult and Attenuated Pediatric Pads cleared under the following Proprietary Names:

Proprietary NameManufacturer510(k) Number
Sentry Semiautomatic ExternalDefibrillator, DDP-100Defibrillation PadsDefibtech, LLCK013896
Defibtech AED with AttenuatedDefibrillation/Monitoring PadsDefibtech, LLCK033896

Defibtech, LLC DDU-2300 510(k) Summary Page I of 3

{1}------------------------------------------------

Section 5

10202 rev B 510(k) Summary

Device Description

The DDU-2300 is a portable, Automated External Defibrillator (AED) intended for use on victims of sudden cardiac arrest (SCA). It is powered by a user-replaceable nonrechargeable battery and supports both adult and pediatric user-replaceable single-use defibrillation/monitoring pads.

The DDU-2300 employs a Patient Analysis System that ensures proper pad/patient connection and analyzes the patient's ECG rhythm to determine whether a shock is required. If needed the DDU-2300 provides a 150 J (50J pediatric) impedance compensated, biphasic truncated exponential defibrillation shock to the patient. The cardiac rhythm analysis algorithm and defibrillation energy and waveform utilized are the same as previous Defibtech AEDs.

The DDU-2300 has a compact design and offers an improved user interface with an LCD display. Voice prompts and a graphical user interface provide simple instructions for the operator. The DDU-2300 AED is capable of recording event information including ECG, audio data and SHOCK/NO SHOCK recommendations.

Intended Use

The DDU-2300 Semiautomatic External Defibrillator (AED) is indicated for use on victims of sudden cardiac arrest (SCA) who are:

  • · Unconscious and unresponsive
  • · Not breathing

For patients under 8 years old, or less than 55 pounds (25 kg), use child/infant electrode pads. Do not delay therapy to determine exact age or weight.

The DDU-2300 AED must be used by or on the order of a physician.

Comparison of Technology Characteristics

The DDU-2300 AED design characteristics are the same as those of the predicate device. Both devices employ the same underlying scientific technology for patient analysis and defibrillation therapy. Both user interface designs guide the user with voice prompts and visual guidance. The DDU-2300 replaces the LED/text display with an LCD display.

Defibtech, LLC DDU-2300 510(k) Summary

{2}------------------------------------------------

10202 rev B 510(k) Summary

202 REV D 510(K) Summary

K081259

The DDU-2300 AED uses similar technologies to provide functionally equivalent performance characteristics as the predicate device. Testing demonstrates that the DDU-2300 meets functional and performance specifications. Safety testing assures complance with applicable industry safety standards.

Conclusion Summary of Safety and Effectiveness

Testing and performance evaluations demonstrate that the Defibtech DDU-2300 AED is substantially equivalent to the predicate device. The introduction of the DDU-2300 AED does not present new issues of safety or effectiveness.

Defibtech, LLC DDU-2300 510(k) Summary

Page 3 of 3

{3}------------------------------------------------

ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 1 5 2009

DefibTech, LLC c/o Mr. Ed Horton VP, Quality Assurance & Regulatory Affairs 741 Boston Post Road, Suite 201. Guilford, CT 06437

Re: K081259

Trade/Device Name: DDU-2300 Semiautomatic Defibrillator and Accessories Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III (three) Product Code: MKJ Dated: June 8, 2009 Received: June 9, 2009

Dear Mr. Horton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. .

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Ed Horton

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Bram D. Zuckerman, M.D.

Director Division of Cardiovascular Device Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number: Ju81259

Device Name: DDU-2300 Semiautomatic External Defibrillator and Accessories

Indications for Use:

The DDU-2300 Semiautomatic External Defibrillator (AED) is indicated for use on victims of sudden cardiac arrest (SCA) who are:

  • · Unconscious and unresponsive
  • · Not breathing

For patients under 8 years old, or less than 55 pounds (25 kg), use child/infant electrode pads. Do not delay therapy to determine exact age or weight.

The DDU-2300 AED must be used by or on the order of a physician.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Defibtech LLC
DDU 2300 510(k) Indications for Use

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Confidential Information
Division of Cardiovascular Devices
510(k) NumberK081259

Page 1

§ 870.5310 Automated external defibrillator system.

(a)
Identification. An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.(b)
Classification. Class III (premarket approval)(c)
Date PMA or notice of completion of PDP is required. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED that was in commercial distribution before May 28, 1976. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976. Any other AED and AED accessory described in paragraph (a), shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.