K Number
K081164

Validate with FDA (Live)

Date Cleared
2008-06-13

(50 days)

Product Code
Regulation Number
866.3175
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diagnostic Hybrids, Inc. device, D3 DFA Cytomegalovirus Immediate Early Antigen Identification Kit, is intended for use in the qualitative detection and identification of human Cytomegalovirus (CMV) immediate early antigen (IEA) in cell cultures by immunofluorescence using fluoresceinated monoclonal antibodies (MAbs).

This product is not intended for use in testing blood or plasma donors and is not intended for use in direct detection of cytomegalovirus in clinical specimens.

Device Description

Two murine derived monoclonal antibodies (MAbs) are used in the Diagnostic Hybrids, Inc. (DHI) device, D3 DFA Cytomegalovirus Immediate Early Antigen Identification Kit (CMV-IEA ID Kit), and are directed against CMV immediate early antigen (pp 72). The MAbs used in the Kit have been shown to be highly specific, with no cross-reactivity to other cultured viruses. The MAbs have been labeled by DHI using Fluorescein Isothiocyanate (FITC).

Kit Components:

  1. CMV-IEA DFA Reagent, 10-mL. One dropper bottle containing a mixture of two murine MAbs directed against CMV immediate early antigen (pp 72). The MAbs are both IgG1 (k) isotype. The buffered, stabilized, aqueous solution contains Evans Blue as a counter-stain and 0.1% sodium azide as preservative.
  2. CMV Antigen Control Slides, 5-slides. Individually packaged control slides containing wells with cell culture derived positive and negative control cells. Each slide contains one Negative well of uninfected cells and one Positive well of CMV infected cells. Each slide is intended to be stained only one time.
  3. Mounting Fluid, 7-mL. One dropper bottle of an aqueous, buffered, stabilized solution of glycerol (ph 8.2 ± 0.2) and 0.1% sodium azide.
  4. 40X PBS Concentrate, 25-mL. One bottle containing a 40X concentrate consisting of 4% sodium azide (0.1% sodium azide after dilution to 1X using de-mineralized water) in a phosphate buffered saline (PBS) solution.

Patient samples are inoculated onto susceptible cell monolayers and cultured. After a defined incubation period, the cells to be tested for the presence of CMV-IEA are fixed in acetone. The CMV-IEA DFA Reagent is added to the cells which are then incubated for 15 to 30 minutes at 35° to 37°C, the stained cells are washed with the supplied PBS Solution (diluted), and a drop of the supplied Mounting Fluid is placed on the prepared cells. The cells are examined using a fluorescence microscope. The cells infected with CMV and expressing the CMV-IEA will have apple-green fluorescent nuclei while uninfected cells will contain no fluorescence but will be stained red by the Evans Blue counter-stain.

If no fluorescent cells are found, report result as "No cytomegalovirus detected". If fluorescent cells are found in the CMV-IEA DFA Reagent stained monolayer, report result as "Cytomegalovirus isolated by cell culture".

AI/ML Overview

The provided 510(k) summary describes the D3 DFA Cytomegalovirus Immediate Early Antigen Identification Kit and its performance in comparison to predicate devices. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in terms of specific numerical thresholds for positive or negative percent agreement that the sponsor had to meet to claim substantial equivalence. Instead, the study aims to show that the D3 DFA CMV-IEA ID Kit performs comparably to legally marketed predicate devices. The performance is reported as percent agreement with the comparator devices, which implicitly serves as the standard for acceptance.

Performance MetricAcceptance Criteria (Implied by Comparison)Reported Device Performance
Analytical Sensitivity (D3 vs. Comparator)Not statistically differentAverage 35.3 ± 2.3 positive wells (D3) vs. 38.3 ± 2.1 positive wells (Comparator) - not statistically different.
Detection Limit (D3 vs. Comparator)Comparable detection limit0.7-TCID50 as the minimum viral dilution detected - not statistically different.
Analytical Specificity / Cross-ReactivityNo cross-reactivity (with exception of S. aureus)No cross-reactivity observed with viruses or bacteria (except Staphylococcus aureus due to Protein A binding, which is distinguishable by morphology).
Clinical Performance (Positive Percent Agreement)Comparable to predicate devicesStudy Site 1 (vs. Bartels Kit): 94.1% (95% CI: 73.0, 98.9)
Study Site 2 (vs. Light Diagnostics): 100% (95% CI: 94.7, 100)
Study Site 3 (vs. Light Diagnostics): 83.3% (95% CI: 43.6, 97.0)
Study Site 4 (vs. Light Diagnostics): 83.3% (95% CI: 55.2, 95.3)
Clinical Performance (Negative Percent Agreement)Comparable to predicate devicesStudy Site 1 (vs. Bartels Kit): 99.7% (95% CI: 98.1, 99.9)
Study Site 2 (vs. Light Diagnostics): 98.7% (95% CI: 96.1, 99.5)
Study Site 3 (vs. Light Diagnostics): 100% (95% CI: 96.9, 100)
Study Site 4 (vs. Light Diagnostics): 100% (95% CI: 98.6, 100)

2. Sample Size Used for the Test Set and Data Provenance

  • Total Clinical Specimens: 1060 specimens were cultured initially, with 1026 included in the final analysis after exclusions. This represents the sample size for the test set.
  • Data Provenance: The clinical study involved three external clinical laboratory sites and the DHI internal laboratory.
    • Study Site 1: Collected and cultured 314 fresh specimens during August 2006.
    • Study Site 2: Cultured 300 specimens (72 fresh and 228 archival). Archival specimens were collected from June 2005 through September 2006 and stored at -70°C.
    • Study Site 3: Tested 146 fresh specimens from February 2007 through May 2007.
    • Study Site 4: Cultured 300 frozen prospectively collected archival specimens from a clinical reference laboratory in the Eastern U.S. in February 2007. These were stored at -70°C.
  • Retrospective/Prospective: The study included both prospective (fresh specimens) and retrospective (archival/frozen specimens) data. Most sites included fresh specimens, while Study Site 2 and 4 specifically mention archival/frozen specimens. "Prospectively collected" for archival specimens refers to how they were collected for the archive, not necessarily how they were collected for this specific study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number or qualifications of experts used to establish the ground truth for the test set. The ground truth for clinical performance was established by comparison to legally marketed predicate devices (Bartels Cytomegalovirus Immediate Early Antigen Kit and Light Diagnostics CMV Direct Immunofluorescence Assay), meaning the results from these predicate devices were used as the reference standard. The interpretation of these predicate device results would have typically been done by trained laboratory personnel, but no specific expert qualifications are provided.

4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for resolving discrepancies or establishing the ground truth when comparing the D3 DFA Cytomegalovirus Immediate Early Antigen Identification Kit to the comparator devices. The comparison is mainly presented as percentage agreement.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No multi-reader multi-case (MRMC) comparative effectiveness study, evaluating the effect size of human readers improving with AI vs. without AI assistance, was explicitly reported. This device is a diagnostic kit (fluorescent antibody test), not an AI algorithm assisting human readers. The performance is assessed by comparing its direct output to predicate devices.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This device is a standalone diagnostic kit. Its performance, as described by the analytical and clinical studies, refers to the kit's ability to qualitatively detect and identify CMV IEA in cell cultures. The results are interpreted by trained laboratory personnel examining the stained cells under a fluorescence microscope, but the "performance" described is of the kit itself, rather than an AI or algorithm in a human-in-the-loop context.

7. Type of Ground Truth Used

The ground truth used for the clinical performance evaluation was the results obtained from legally marketed predicate devices (Bartels Cytomegalovirus Immediate Early Antigen Kit and Light Diagnostics CMV Direct Immunofluorescence Assay) when applied to cultured specimens. In the context of the analytical studies, ground truth was based on known concentrations of CMV (TCID50) in cell cultures.

8. Sample Size for the Training Set

The document does not describe a "training set" in the context of machine learning or AI models. This device is a traditional laboratory diagnostic kit utilizing monoclonal antibodies, not an AI-driven system. Therefore, there is no training set in the typical sense. The "training" for such a kit would involve optimization and validation during its development phase, but these details are not provided as a distinct "training set size."

9. How the Ground Truth for the Training Set Was Established

Since there is no training set mentioned in the context of an AI/ML device, the concept of establishing ground truth for a training set does not apply here. The "development" of the kit would be based on well-established immunological principles and laboratory validation procedures to ensure the monoclonal antibodies are specific and effective, rather than training an algorithm on a dataset with ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Diagnostic Hybrids. The logo features the company name in bold, sans-serif font, with the words "DIAGNOSTIC" stacked above "HYBRIDS." Below the name is the tagline "Integrating Science and Humanity" in a smaller, italicized font. To the right of the text is a stylized graphic resembling a DNA double helix intertwined with a human figure with arms raised.

510(K) SUMMARY - K081164

D3 DFA CYTOMEGALOVIRUS IMMEDIATE EARLY ANTIGEN IDENTIFICATION KIT

Applicant:

DIAGNOSTIC HYBRIDS, INC. 1055 East State Street Suite 100 Athens, OHIO 45701

JUN 1 3 2008

Contact Information:

Gail R. Goodrum Vice President, Regulatory Affairs E-mail: goodrum@dhiusa.com Telephone: 740-589-3300 Desk Extension: 740-589-3380 FAX: 740-593-8437

Date of preparation of 510(k) summary:

June 06, 2008

Device Name:

Trade name - D3 DFA CYTOMEGALOVIRUS IMMEDIATE EARLY ANTIGEN IDENTIFICATION KIT Common name - Fluorescent antibody test for identification of Cytomegalovirus (CMV) Classification name - Antisera, Conjugated Fluorescent, Cytomegalovirus Product Code - LIN Regulation - 21 CFR 866.3175, Cytomegalovirus serological reagents

Legally marketed devices to which equivalence is claimed:

    1. K951821, Light Diagnostics CMV Direct Immunofluorescence Assay (Millipore) Intended Use: Light Diagnostics Cytomegalovirus Direct Immunofluorescence Assay is intended for pre-cytopathic effect (CPE) detection and identification of immediate early antigen of human CMV in cell culture. This product is not FDA approved for use in testing blood or plasma donors and is not intended for use in direct detection of CMV in clinical specimens.
    1. K904036, Bartels Cytomegalovirus Immediate Early Antigen Indirect Fluorescent Antibody (Trinity Biotech PLC) Intended Use: Bartels Cytomegalovirus Immediate Early Antigen Indirect Fluorescent Antibody is intended for qualitative detection of CMV pre-cytopathic effect (CPE) immediate early antigen in centrifugation-enhanced inoculated cell cultures. This product is not FDA cleared (approved) for use in testing (i.e., screening) blood or plasma donors.

{1}------------------------------------------------

Device Description:

Two murine derived monoclonal antibodies (MAbs) are used in the Diagnostic Hybrids, Inc. (DHI) device, D3 DFA Cytomegalovirus Immediate Early Antigen Identification Kit (CMV-IEA ID Kit), and are directed against CMV immediate early antigen (pp 72). The MAbs used in the Kit have been shown to be highly specific, with no cross-reactivity to other cultured viruses. The MAbs have been labeled by DHI using Fluorescein Isothiocyanate (FITC).

Kit Components:

    1. CMV-IEA DFA Reagent, 10-mL. One dropper bottle containing a mixture of two murine MAbs directed against CMV immediate early antigen (pp 72). The MAbs are both IgG1 (k) isotype. The buffered, stabilized, aqueous solution contains Evans Blue as a counter-stain and 0.1% sodium azide as preservative.
    1. CMV Antigen Control Slides, 5-slides. Individually packaged control slides containing wells with cell culture derived positive and negative control cells. Each slide contains one Negative well of uninfected cells and one Positive well of CMV infected cells. Each slide is intended to be stained only one time.
    1. Mounting Fluid, 7-mL. One dropper bottle of an aqueous, buffered, stabilized solution of glycerol (ph 8.2 ± 0.2) and 0.1% sodium azide.
    1. 40X PBS Concentrate, 25-mL. One bottle containing a 40X concentrate consisting of 4% sodium azide (0.1% sodium azide after dilution to 1X using de-mineralized water) in a phosphate buffered saline (PBS) solution.

Patient samples are inoculated onto susceptible cell monolayers and cultured. After a defined incubation period, the cells to be tested for the presence of CMV-IEA are fixed in acetone. The CMV-IEA DFA Reagent is added to the cells which are then incubated for 15 to 30 minutes at 35° to 37°C, the stained cells are washed with the supplied PBS Solution (diluted), and a drop of the supplied Mounting Fluid is placed on the prepared cells. The cells are examined using a fluorescence microscope. The cells infected with CMV and expressing the CMV-IEA will have apple-green fluorescent nuclei while uninfected cells will contain no fluorescence but will be stained red by the Evans Blue counter-stain.

If no fluorescent cells are found, report result as "No cytomegalovirus detected". If fluorescent cells are found in the CMV-IEA DFA Reagent stained monolayer, report result as "Cytomegalovirus isolated by cell culture".

Intended Use:

The Diagnostic Hybrids, Inc. device, D3 DFA Cytomegalovirus Immediate Early Antigen Identification Kit, is intended for use in the qualitative detection and identification of human cytomegalovirus (CMV) immediate early antigen (IEA) in cell cultures by immunofluorescence using fluoresceinated monoclonal antibodies (MAbs).

This product is not intended for use in testing blood or plasma donors and is not intended for use in direct detection of cytomegalovirus in clinical specimens.

{2}------------------------------------------------

Technological Characteristics:

The Diagnostic Hybrids, Inc. device, D3 DFA Cytomegalovirus Immediate Early Antigen Identification Kit, has been compared directly to Bartels Cytomegalovirus Immediate Early Antigen Kit and the Light Diagnostics CMV Direct Immunofluorescence Assay as the legally marketed devices. The technology used in all devices is based on a standard immunofluorescence assay technique utilizing fluorescein-labeled MAbs and viral isolation in cell culture. A summary is provided in Table 5.1 below:

TABLE 5.1: Technological Characteristics Compared Among DHI Device and Predicate Devices
CharacteristicSubject DeviceDHIPredicateBartels(K904036)PredicateLight Diagnostics(K951821)
Immediate Early AntigenXXX
MAb Directly Labeled with FluoresceinX--X
MAb Indirect labeled with Fluorescein(requires labeled secondary anti-mouseantibody)--X--
Culture ConfirmationXXX

Non-Clinical Performance:

Staining patterns of the fluorescein-labeled MAbs on CMV infected cells were similar to those of the predicate devices.

Analytical sensitivity was studied for purposes of demonstrating the effectiveness of the D3 CMV-IEA DFA Reagent with that of a comparator device. This was done by first inoculating two 96-well cell culture plates (Hs27) with CMV diluted to a value of 1-TCID50 and incubating at 37°C for 48 hours; then, one plate was stained with the subject D3 CMV-IEA DFA Reagent and the other plate was stained using the CMV-IEA DFA Reagent from the comparator device. This assay was performed three times, with an average of 35.3 ±2.3 positive wells out of a total 96 wells detected with the subject, and an average of 38.3 ±2.1 positive wells out of a total 96 wells with the predicate. These results were not statistically different by a paired t-test".

Detection limit for the subject device CMV-IEA ID Kit was addressed under the conditions: CMV, at a starting concentration 350-TCID50 per mL, was serially diluted to a final value of 0.7-TCID50 per mL using 2-fold dilutions. Each dilution was inoculated into confluent monolayers of Hs27 cells contained in multi-well plates, centrifuged at 700 xg for 60 minutes and incubated at 35° to 37°C for 48 hours. The subject CMV-IEA ID Kit or the comparator device, was used to stain 3 monolayers of each viral dilution according to the respective device's product insert. The number of positive cells per well was counted. The experiment was performed three times. The results suggest that the detection limit of both fluorescent antibody stains are

DHI-CMV-IE Sec05 510k-Summary_rev_08-0609 (2).doc

a Microsoft Office Excel, Microsoft Corporation

{3}------------------------------------------------

D² DFA CYTOMEGALOVIRUS IMMEDIATE EARLY ANTIGEN IDENTIFICATION KIT

Section 05, 510(k) Summary

comparable, with 0.7-TCID50 as the minimum viral dilution detected, as indicated by at least one well having no detectable infection. These results were not statistically different by a paired t-test3.

Analytical specificity was evaluated according to cross reactivity studies against a number of strains of viruses, strains of bacteria, and different cell lines. No crossreactivity or non-specific staining with any of the other agents was observed, except for Staphylococcus aureus (Protein A producing bacteria will bind the Fc portion of some of the fluoresccin-labeled MAb), which was cross-reactive with the CMV-IEA DFA Reagent. Microorganisms and cell lines which were evaluated against the CMV-IEA DFA Reagent are listed in Table 5.2, below:

TABLE 5.2: Microorganisms tested for Cross-Reactivity with D3 CMV-IEA DFA Reagent
Virus (no reactivity with CMV-IEA DFA Reagent)Strain or TypeInoculum (TCID50)OrganismStrain or TypeInoculum (TCID50)
AdenovirusType 11,400RSVLong1,400
AdenovirusType 31,400RSVWash1,400
AdenovirusType 51,400RSV93201,400
AdenovirusType 61,400Parainfluenza 1C-351,400
AdenovirusType 71,400Parainfluenza 2Greer1,400
AdenovirusType 81,400Parainfluenza 3C 2431,400
AdenovirusType 101,400Parainfluenza 4aM-251,400
AdenovirusType 131,400Parainfluenza 4bCH195031,400
AdenovirusType 141,400HSV-1IF140
AdenovirusType 311,400HSV-1MacIntyre140
Influenza AAichi1,400HSV-2MS140
Influenza AMalaya1,400HSV-2Strain G140
Influenza AHong Kong1,400VZVWebster140
Influenza ADenver1,400VZVEllen140
Influenza APort Chalmers1,400VZVAV92-3140
Influenza AVictoria1,400Epstein-Barr
Influenza ANew Jersey1,400RubeolaCommercially available slides stained b
Influenza APR1,400Mumps
Influenza AWS1,400EchovirusTypes 4, 6, 9, 11, 30, 34Commercially available slides stained.
Influenza BHong Kong1,400
Influenza BMaryland1,400CoxsackievirusTypes B1, B2, B3, B4, B5, B6
Influenza BMass1,400
Influenza BTaiwan1,400PoliovirusTypes 1, 2, 3
Influenza BGL1,400
Influenza BRussia1,400
BACTERIAResultCFU TESTEDBACTERIAResultCFU TESTED
Acholeplasma laidlawiiNegative~6 x 107MycoplasmasalivariumNegative~6 x 107
AcinetobactercalcoaceticusNegative9.7 x 105NeisseriagonorrhoeaeNegative1.3 x 106
BordetellabronchisepticaNegative1.7 x 105Proteus mirabilisNegative2.1 x 106
Bordetella pertussisNegative4.6 x 106PseudomonasaeruginosaNegative1.0 x 107
CorynebacteriumdiphtheriaeNegative2.5 x 106Salmonella enteriditisNegative2.5 x 106
Escherichia coliNegative2.6 x 105SalmonellatyphimuriumNegative1.8 x 106

b Test material is from commercially available prepared slides. Each positive well contains 10 to 50% reactive cells.

{4}------------------------------------------------

Diagnostic Hybrids, Inc.

Gardnerella vaginalisNegative5.0 x 105StaphylococcusaureusPOSITIVE1.0 x 107
Haemophilis influenzaetype ANegative9.3 x 105StreptococcusagalactiaeNegative9.6 x 106
Klebsiella pneumoniacNegative6.4 x 106StreptococcuspneumoniaeNegative8.0 x 105
Legionella pneumophilaNegative6.5 x 104StreptococcuspyogenesNegative2.9 x 107
Moraxella cartarrhalisNegative6.4 x 104UreaplasmauralyticumNegative~6 x 104
Mycoplasma hominisNegative~6 x 104ChlamydiophilapneumoniaeNegativeCommercially availableslides stained.
Mycoplasma oraleNegative~6 x 104ChlamydiophilapsittaciNegativeCommercially availableslides stained.
MycoplasmapneumoniaeNegative~6 x 104ChlamydiatrachomatisNegativeCommercially availableslides stained.
YEASTPROTOZOAN
Candida glabrataNegative8.7 x 106TrichomonasvaginalisNegativeCommercially availableslides stained.
CELL LINES (no reactivity with CMV-IEA DFA Reagent)
A549MonolayerMRC-5MonolayerRhMK IICell Spot
BGMKMonolayerMRIIFMonolayerRK (passage 1)Monolayer
HEp-2MonolayerMv1LuMonolayerR-MixMonolayer
HFF / Hs27MonolayerNCI-H292MonolayerVeroCell spot
LLC-MK2MonolayerpCMKCell spotVero 76Cell spot
McCoyMonolayerpRhMKCell spotWI-38Cell spot
MDCKMonolayerRDMonolayer

Section 05, 510(k) Summary

SECTION 05- 5 of 6

Clinical Performance:

A total of 1060 specimens were cultured and stained with one of two comparative devices and the D DFA Cytomegalovirus Immediate Early Antigen Identification Kit at three external clinical laboratory sites and at the DHI internal laboratory. A total of 34 specimens were excluded from final analysis, resulting in a total of 1026 results reported. Reasons for exclusion were specimen toxicity to cell culture (29), bacterial contamination of cell culture (1), non-specific fluorescence seen prohibiting interpretation (2), and unacceptable specimens (2).

Study site 1 collected and cultured a total of 314 fresh specimens during August, 2006. There were no specimens excluded from final analysis. A wide variety of specimen sources were cultured from a diverse age population. The DS DFA Cytomegalovirus Immediate Early Antigen Identification Kit positive percent agreement compared with Bartels Cytomegalovirus Immediate Early Antigen kit for cultured specimens had a positive percent agreement of 94.1% (exact 95% C1 range of 73.0, 98.9) and a negative percent agreement of 99.7% (exact 95% CI range of 98.1, 99.9).

Study site 2 cultured a total of 300 specimens (72 fresh and 228 archival) specimens from August 31 through November 8, 2006. The archival specimens were collected from June, 2005 through September, 2006. They were stored at -70° C until re-

Staining of S. aureus appeared as small points of fluorescence while all other cultures were negative. This will be noted in labeling in the section "Limitations of the Assay": The Protein A produced by the bacterium, Staphylococcus aureus, will bind the Fc portion of the fluorescein-labeled monoclonal antibodies used in this kit. Such binding can be distinguished from viral antigen binding on the basis of morphology, i.e., S. aureusbound fluorescence appears as small (~1 micron diameter), bright dots. Results from cell cultures with hacterial contamination must, therefore, be interpreted with caution."

{5}------------------------------------------------

Diagnostic Hybrids, Inc.

D3 DFA Cytomegalovirus Immediate early antigen Identification Kit

Section 05, 510(k) Summary

cultured for this study. The specimens were not selected for this study based on previous culture results. Of the 228 archival specimens, seven were excluded from the final analysis due to toxicity in cell culture. The D DFA Cytomegalovirus Immediate Early Antigen Identification Kit combined archival and fresh positive percent agreement, compared with Light Diagnostics CMV Direct Immunofluorescence Assay, for cultured specimens had positive percent agreement of 100% (exact 95% CI range of 94.7, 100) and a negative percent agreement of 98.7% (exact 95% CI range of 96.1, 99.5).

Study site 3 tested specimens from February 2007 through May 2007. A total of 146 fresh specimens were cultured. Of these 146 specimens, 18 were excluded from the final analysis due to toxicity, contamination, and specimen acceptability. The D DFA Cytomegalovirus Immediate Early Antigen Identification Kit positive percent agreement, compared with Light Diagnostics Cytomegalovirus Direct Immunofluorescence Assay kit for culture confirmation specimens. had a positive percent agreement of 83.3% (exact 95% CI range of 43.6, 97.0) and a negative percent agreement of 100% (exact 95% CI range of 96.9, 100).

Study site 4 cultured 300 specimens that were collected at a clinical reference laboratory located in the Eastern U.S. in February, 2007. These frozen prospectively collected specimens were stored at -70° C until cultured for this study. The specimens were not selected for this study based on previous culture results. Of these 300 archival specimens, 9 were excluded from the final analysis due to toxicity and nonspecific fluorescence. The D DFA Cytomegalovirus Immediate Early Antigen Identification Kit positive percent agreement, compared with Light Diagnostics Cytomegalovirus Direct Immunofluorescence Assay kit for culture confirmation specimens, had a positive percent agreement of 83.3% (exact 95% CI range of 55.2, 95.3) and a negative percent agreement range of 100% (exact 95% CI 98.6. 100).

The overall study results indicate that the D DFA Cytomegalovirus Immediate Early Antigen Identification Kit performs comparably to the legally marketed CMV immunofluorescence assay kits for the qualitative detection and identification of CMV IEA in cell cultures.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUN 1 3 2008

Gail R. Goodrum Vice President of Regulatory Affairs Diagnostic Hybrids, Inc. 1055 East State Street Suite 100 Athens, OH 45701

Re: K081164

Trade/Device Name: D3 DFA Cytomegalovirus Immediate Early Antigen Identification Kit Regulation Number: 21 CFR 866.3175 Regulation Name: Cytomegalovirus Serological Reagents Regulatory Class: Class II Product Code: LIN Dated: April 21, 2008 Received: April 24, 2008

Dear Ms. Goodrum:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{7}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Sally attym

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use

510(k) Number (if known): K081164

Device Name: D3 DFA Cytomegalovirus Immediate Early Antigen Identification Kit

Indications for Use: The Diagnostic Hybrids, Inc. device, D3 DFA Cytomegalovirus Immediate Early Antigen Identification Kit, is intended for use in the qualitative detection and identification of human Cytomegalovirus (CMV) immediate early antigen (IEA) in cell cultures by immunofluorescence using fluoresceinated monoclonal antibodies (MAbs).

This product is not intended for use in testing blood or plasma donors and is not intended for use in direct detection of cytomegalovirus in clinical specimens.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of GDRH2 Office of Device Evaluation (ODE) D C 120 Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

K081164 510(k)_

§ 866.3175 Cytomegalovirus serological reagents.

(a)
Identification. Cytomegalovirus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to cytomegalovirus in serum. The identification aids in the diagnosis of diseases caused by cytomegaloviruses (principally cytomegalic inclusion disease) and provides epidemiological information on these diseases. Cytomegalic inclusion disease is a generalized infection of infants and is caused by intrauterine or early postnatal infection with the virus. The disease may cause severe congenital abnormalities, such as microcephaly (abnormal smallness of the head), motor disability, and mental retardation. Cytomegalovirus infection has also been associated with acquired hemolytic anemia, acute and chronic hepatitis, and an infectious mononucleosis-like syndrome.(b)
Classification. Class II (performance standards).