K Number
K081091

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-06-11

(56 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Precision 500D R&F X-ray System is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations.

The Precision 500D R&F X-ray Systems are intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.

Device Description

The Precision 500D R&F X-ray System consists of an X-ray generator, angulating table with x-ray tube, collimator and image intensifier, wall stand, overhead tube suspension operator console and digital archive system

AI/ML Overview

The provided text describes a 510(k) Summary for the GE Healthcare Precision 500D R&F X-ray System. This document focuses on demonstrating substantial equivalence to a predicate device and does not involve AI/ML performance metrics, clinical studies with human readers, or detailed ground truth establishment as typically associated with AI-powered diagnostics.

However, based on the non-clinical testing section, we can extract information regarding the device's performance related to dose reduction and image quality (SNR).

Here's the information extracted from the provided text, adapted to your requested format as much as possible, with notes explaining the absence of certain categories:


Acceptance Criteria and Device Performance

This 510(k) Premarket Notification is for an improved version of an existing X-ray system, focusing on dose reduction while maintaining image quality. The "acceptance criteria" presented here refer to the performance targets for dose reduction and Signal-to-Noise Ratio (SNR) in pediatric modes compared to adult modes.

Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Target Performance)Reported Device Performance
Dose Reduction (Pediatric Grid-in Mode @ 7.5 fps vs. Adult Grid-in Mode @ 30 fps)Minimum 35% less dose when measured with a 32cm Image Intensifier (II) or 40cm II system.Achieves "a minimum of 35% less dose" when measured with a 32cm II or 40cm II system in Pediatric Grid-in mode at 7.5 frames per second (fps) as compared to Adult Grid-in Mode at 30fps. Dose savings represent a minimum specification, averaged over all magnification modes.
SNR Difference (Pediatric Grid-in Mode @ 7.5 fps vs. Adult Grid-in Mode @ 30 fps)Maximum difference in Signal to Noise Ratio (SNR) of 15%.Quantitatively maintains image quality with "a maximum difference in Signal to Noise Ratio (SNR) of 15%" when comparing Pediatric Grid-in mode at 7.5fps to Adult Grid-in Mode at 30fps. SNR calculations are performed with "Iodine" selected as contrast agent.
Dose Reduction (Pediatric Grid-in Mode @ 30 fps vs. Adult Grid-in Mode @ 30 fps)Minimum 35% less dose when measured with a 32cm II or 40cm II system.Achieves "a minimum of 35% less dose" when measured with a 32cm II or 40cm II system in Pediatric Grid-in mode at 30fps as compared to Adult Grid-in Mode at 30fps. Dose savings represent a minimum specification, averaged over all magnification modes.
SNR Difference (Pediatric Grid-in Mode @ 30 fps vs. Adult Grid-in Mode @ 30 fps)Maximum difference in Signal to Noise Ratio (SNR) of 15%.Quantitatively maintains image quality with "a maximum difference in SNR of 15%" when comparing Pediatric Grid-in mode at 30fps to Adult Grid-in Mode at 30fps. SNR calculations are performed with "Iodine" selected as contrast agent.

Additional Information (as per requested categories):

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Test Set Sample Size: Not applicable. The testing described is non-clinical bench testing, not a study involving patient data. It refers to measurements performed on the device itself.
    • Data Provenance: Not applicable. The data is from internal bench testing of the device, comparing performance characteristics under controlled conditions.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Number of Experts: Not applicable. Ground truth, in the context of clinical interpretation, is not established for this type of non-clinical, performance-based test. The "ground truth" for dose and SNR measurements is derived from the physical capabilities of the x-ray system and measurement instruments.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Adjudication Method: Not applicable. This was non-clinical bench testing, not an assessment requiring expert adjudication of clinical findings.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No. This submission pertains to an X-ray imaging system, not an AI/ML-powered diagnostic aid involving human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Standalone Performance: Not applicable. This is not an algorithm-only device. It is a an X-ray imaging system with improved performance characteristics.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Ground Truth Type: Not applicable in the clinical sense. For the non-clinical testing, the "ground truth" for dose measurements would be the actual dose output as measured by calibrated dosimeters, and for SNR, it would be the calculated signal-to-noise ratio based on image measurements under specified phantom conditions.
  7. The sample size for the training set:

    • Training Set Sample Size: Not applicable. This device does not use an AI/ML model that requires a training set.
  8. How the ground truth for the training set was established:

    • Training Set Ground Truth: Not applicable. This device does not use an AI/ML model that requires a training set.

{0}------------------------------------------------

Image /page/0/Picture/21 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, and there are three short, curved lines extending outward from the circle at different points.

K081091/

GE Healthcare

510(k) Summary

JUN 1 1 2008

The following 510(k) Summary was prepared in accordance with the requirements of 21 CFR §807.87(h) and §807.92.

  • (1) Identification of Submitter:
    Submitter: Address: Contact Person: Telephone #: Summary Prepared:

GE Healthcare 3000 N. Grandview Blvd W-709 Michael Petrowski (262)-548-2673 02-01-2008

Alternate Contact:

  • John L. Schmidt GE Healthcare Safety & Requlatory Enqineer T 262 548 4964 E John.L.Schmidt@med.ge.com 3000 N. Grandview Blvd. W709 Waukesha, WI, 53188-1678, USA
  • (2) Device Information:

GE Model Name: Classification Name: Class: Product Code:

Precision 500D R&F X-ray System Image-intensified fluoroscopic x-ray system Class II JAA

  • (3) Predicate Device:
    510(k) Number: GE Model Name:

Product Code:

Classification Name:

K011624

Precision 500D R&F X-ray System (Previous Name: Expedio 500D R&F X-ray System) Image-intensified fluoroscopic x-ray system Class II JAA -

  • (4) Device Description:
    Class:

The Precision 500D R&F X-ray System consists of an X-ray generator, angulating table with x-ray tube, collimator and image intensifier, wall stand, overhead tube suspension operator console and digital archive system

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the text "GE Healthcare". The text is written in a simple, sans-serif font. The words are arranged horizontally, with "GE" on the left and "Healthcare" on the right. The text is black and the background is white.

Image /page/1/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circle. The letters and the circle are black, while the background is white.

(5) Statement of Intended Use:

The Precision 500D R&F X-ray System is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations.

The Precision 500D R&F X-ray Systems are intended for use by a auglified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position

  • (6) Technological Characteristic Comparison:

Design:

Precision 500D R&F X-ray System is identical in design and technology to the previously cleared Precision 500D R&F X-ray System K011624. There are no changes to the design and/or technology.

Material:

Precision 500D R&F X-ray System uses the exact construction materials as used in the previously cleared Precision 500D R&F X-ray System K011624. There are no changes to the materials.

Chemical Composition:

Chemical composition of materials is not critical for this type of device, because no material is implanted or inserted into the patient body or orifice to contact body tissues or fluids. Patient contact is not required for proper functionina of the device. See Section 15 for a statement regarding patient support tabletop material.

Energy Source:

The Precision 500D R&F X-Ray System uses the same energy source as was used in the previously cleared Precision 500D R&F X-ray System K011624: High Voltage Generator 3 Phase 80kW, Model Jedi 80TF2T, 2290800

Non-clinical Testing:

Bench testing verification was performed and a complete testing summary can be found in Section 18. Verification supports a minimum of 35% less dose when measured with a 32cm Image Intensifier (II) or 40cm II system in Pediatric Grid-in mode at 7.5 frames per second (fps) as compared to Adult Grid-in Mode at 30fps while comparatively quantifying image quality with a maximum difference in Signal to Noise Ratio (SNR) of 15%. Additionally verification supports a minimum of 35% less dose when measured with a

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circular border. There are three small, stylized flourishes or dots placed around the circle, adding a decorative element to the design.

GE Healthcare

32cm II or 40cm II system in Pediatric Grid-in mode at 30fps as compared to Adult Grid-in Mode at 30fps while comparatively quantifying image quality with a maximum difference in SNR of 15%. Dose savings represent a minimum specification, averaged over all magnification modes, and SNR calculations are performed with "Iodine" selected as contrast agent.

Conclusion:

GE Healthcare considers the Precision 500D R&F X-ray System to be substantially equivalent to the identified predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The bird is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 11 2008

Mr. Michael Petrowski Safety & Regulatory Engineer GE Medical Systems 3000 N. Grandview Blvd. WAUKESHA WI 53188

Re: K081091

Trade/Device Name: Precision 500D R&F X-ray System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Dated: February 1, 2008 Received: April 16, 2008

Dear Mr. Petrowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device i can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.

Sincerely vours.

Nancy C. Burgdon

Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

GE Healthcare

Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular frame. The letters and the frame are solid black, creating a strong contrast against the white background.

Indication for Use

510(k) Number (if known): K081091

Device Name: Precision 500D R&F X-ray System

Indication For Use:

The Precision 500D R&F X-ray System is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations.

The Precision 500D R&F X-ray Systems are intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.

Prescription Use (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510k)

(Divisio h Sian-Off)

Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

4 - 1

N/A