K Number
K080936

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-05-12

(40 days)

Product Code
Regulation Number
888.3560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ConforMIS Total Knee Repair System ("iTotal") is a minimally invasive, bone preserving primary total knee system intended 'for use in patients with severe knee joint pain and disability. The indications for use include restoring joint function and relief of pain due to:
Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis of the knee
Post traumatic loss of joint function
Mild to moderate valgus or varus deformity of the knee
The ConforMIS Total Knee Repair System ("iTotal") is intended only for use with bone cement.

Device Description

The additional 6.0 mm thickness option for the ConforMIS Total Knee Repair System tibial component is intended to provide the surgeon with an additional option for sizing the tibial component. The X/Y dimensions of the component (i.e. the two-dimensional shape or "footprint") are designed to conform to the patient's anatomy as closely as possible based on images (MRI or CT scan) of the patient's knee. It is available as an UHMEPE component with minimal thickness of 6.0 mm.

AI/ML Overview

Acceptance Criteria and Device Performance for ConforMIS Total Knee Repair System (K080936)

This submission describes the ConforMIS Total Knee Repair System (iTotal): Tibial Component Extension of Size Range. The primary goal of the submission is to demonstrate substantial equivalence to previously cleared predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

The core of this 510(k) submission relies on demonstrating substantial equivalence to existing predicate devices. Therefore, the "acceptance criteria" are implicitly defined by the characteristics of these predicates, and the "reported device performance" is a direct comparison to those characteristics.

Acceptance Criteria CategorySpecific Criteria (Based on Predicate Devices)Reported Device Performance (ConforMIS Total Knee Repair System Tibial Component 6.0mm)
Design CharacteristicsEquivalent design and production process to ConforMIS Total Knee Repair System (K052687) in using imaging data for patient-matched geometry.Equivalent in terms of design and production process. Uses imaging data to design patient-matched implant geometry.
Material CompositionIdentical materials to ConforMIS Total Knee Repair System (K052687).Uses identical materials.
Intended Use & IndicationsShares the same intended use and indications for use as the ConforMIS Total Knee Repair System (K052687) and other cited predicates.Shares the same intended use and indications for use: "minimally invasive, bone preserving primary total knee system intended for use in patients with severe knee joint pain and disability due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, post traumatic loss of joint function, and mild to moderate valgus or varus deformity of the knee." Explicitly for cemented use only.
Mechanical Safety/FunctionMechanical safety and function equivalent to cited predicate devices (e.g., ConforMIS Uni-compartmental "iUni", Bi-compartmental "iDuo").Testing completed as part of design verification found the device to be as safe and effective as predicate devices. The additional 6.0 mm thickness option provides an alternative sizing option.
Minimum ThicknessAbsolute minimum thickness equivalent to tibial components cleared for use with ConforMIS uni-compartmental ("iUni") and bi-compartmental ("iDuo") devices."Substantially equivalent to the tibial components cleared for use with the ConforMIS uni-compartmental ('iUni') and bi-compartmental ('iDuo') Device in terms of absolute minimum thickness." Minimal thickness of 6.0 mm.
New Safety/Effectiveness IssuesNo new safety or effectiveness issues raised compared to predicate devices."raises no new safety or effectiveness issues."

2. Sample Size Used for the Test Set and Data Provenance

The submission does not refer to a "test set" or mention any clinical studies involving human subjects for this particular 510(k) submission (K080936). Instead, it relies on non-clinical performance testing for design verification and comparison to predicates.

  • Sample Size for Test Set: Not applicable, as no clinical test set was used.
  • Data Provenance: Not applicable for clinical data. Non-clinical performance data would be generated in a lab setting, likely within the US, by the manufacturer.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not applicable, as no clinical test set requiring expert ground truth establishment was conducted for this 510(k). The evaluation is based on mechanical testing and comparison to predicate device specifications and performance.

4. Adjudication Method for the Test Set

Not applicable, as no clinical test set requiring an adjudication method was conducted for this 510(k).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This 510(k) focuses on a mechanical device (knee implant) and its design/material properties, not an imaging or diagnostic AI device that would typically involve MRMC studies.

6. Standalone (Algorithm Only Without Human-in-the-loop) Performance Study

No, a standalone performance study was not done. This device is a component of a knee replacement system; its "performance" is inherently linked to its surgical implantation and function within the human body, and its evaluation in this submission is through non-clinical testing and comparison to predicates.

7. Type of Ground Truth Used for the Test Set

Not applicable, as no clinical test set requiring ground truth was conducted for this 510(k). The "ground truth" for the non-clinical testing would be the engineering specifications and performance standards established for such devices, against which the new device's test results are compared.

8. Sample Size for the Training Set

Not applicable. This device is a manufactured medical implant, not an AI/machine learning algorithm that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

K080936

MAY 12 2008

SECTION 2

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of the 510(k) pre-market notification for the ConforMIS Total Knee Repair System ("iTotal"): Tibial Component: Extension of Size Range is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Summary of Safety and Effectiveness

Submitted By:ConforMIS, Inc.
2 Fourth Avenue
Burlington, MA 01803

S. Michael Sharp, PhD Contact Person: Sr. Vice President, Regulatory/Clinical & Quality

  • March 24, 2008 Date:
  • Trade/Proprietary Name ConforMIS Total Knee Repair System (iTotal): Tibial Component Extended Thickness Range
  • Knee joint patellofemorotibial Common Name polymer/metal/polymer semi-constrained cemented prosthesis.

Reference/Classification Name: 21 CFR 888.3560 – Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis.

Predicate Devices

TechnologicalCharacteristicsDesign & InsertThicknessIndications for Use
ConforMIS Total KneeRepair System(K052687)ConforMIS Total KneeRepair System(K052687)ConforMIS Total KneeRepair System(K052687)
ConforMISBiCompartmentalKnee Repair System(K053488)ConforMISBiCompartmentalKnee Repair SystemPolyethylene Tibial Insert(K072368)ConforMISBiCompartmentalKnee Repair System(K053488)

{1}------------------------------------------------

Intended Use:

The ConforMIS Total Knee Repair System ("iTotal") is a minimally invasive, bone preserving primary total knee system intended 'for use in patients with severe knee joint pain and disability. The indications for use include restoring joint function and relief of pain due to:

  • Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid .
  • arthritis of the knee .
  • Post traumatic loss of joint function .
  • Mild to moderate valqus or varus deformity of the knee .

The ConforMIS Total Knee Repair System (iTotal): Tibial Component: is intended only for use with bone cement.

Device Description: The additional 6.0 mm thickness option for the ConforMIS Total Knee Repair System tibial component is intended to provide the surgeon with an additional option for sizing the tibial component. The X/Y dimensions of the component (i.e. the two-dimensional shape or "footprint") are designed to conform to the patient's anatomy as closely as possible based on images (MRI or CT scan) of the patient's knee. It is available as an UHMEPE component with minimal thickness of 6.0 mm.

Comparison to Predicates: The ConforMIS Total Knee Repair System tibial component in the additional size Component is substantially equivalent to the tibial components cleared for use with the ConforMIS uni-compartmental ("iUni") ลทd bi-compartmental ("iDuo") Device in terms of absolute minimum thickness. It is equivalent to the ConforMIS Total Knee Repair System in the use of imaging data to design patient-matched implant geometry; in terms of design and production process; in the use of identical materials and indications. It is substantially equivalent to the cited predicate devices in terms of design, materials, mechanical safety and intended use. All are intended for cemented use only.

Performance Data

Non-clinical Performance and Conclusions:

Testing completed as part of the design verification procedure for the ConforMIS Total Knee Repair System Tibial component found this device to be as safe and effective as the predicate devices, further confirming substantial equivalence.

{2}------------------------------------------------

Clinical Performance: Clinical data and conclusions are not necessary to demonstrate substantial equivalence.

SUMMARY

Based on the similarities in design, materials, function and intended use the ConforMIS Total Knee Repair System (iTotal): Tibial Component 6.0mm is substantially equivalent to the devices currently marketed under the Federal Food, Drug and Cosmetic Act. In addition, the ConforMIS Total Knee Repair System ("iTotal") 6 mm Tibial Component raises no new safety or effectiveness issues.

{3}------------------------------------------------

Public Health Service

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized caduceus symbol, which is a staff with two snakes coiled around it.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

ConforMIS, Inc. % S. Michael Sharp, PhD Senior Vice President Regulatory and Clinical Affairs 2 Fourth Avenue Burlington, Massachusetts 01803

MAY 12 2002

Re: K080936

Trade/Device Name: ConforMIS Total Knee Repair System Tibial Component: 6.0 mm Thickness Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semiconstrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: March 31, 2008 Received: April 2, 2008

Dear Dr. Sharp:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can he found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - S. Michael Sharp, PhD

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark M. Millman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

STATEMENT OF INDICATIONS FOR USE

510(k) Number (if known): #K0080936- KO80336

Device Name: ConforMIS Total Knee Repair System

Indications for Use:

The ConforMIS Total Knee Repair System ("iTotal") is a minimally invasive, bone preserving primary total knee system intended 'for use in patients with severe knee joint pain and disability. The indications for use include restoring joint function and relief of pain due to:

Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis of the knee

Post traumatic loss of joint function

Mild to moderate valgus or varus deformity of the knee

The ConforMIS Total Knee Repair System ("iTotal") is intended only for use with bone cement.

Prescription Use Over-The-Counter Use x AND/OR (Part 21 CFR 801 Subpart C)
Part 2 (Part 21 CFR 801 Subpart C)
CFUS LEVE CONTRACTURE ON ANOTEL (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Neil R. Ogle - from xm

(Division Sign-Off Division of General, Restorative, and Neurological Devices

510(k) Number K080936

§ 888.3560 Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.