K Number
K080857

Validate with FDA (Live)

Date Cleared
2008-04-10

(14 days)

Product Code
Regulation Number
878.4580
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CHROMOPHARE® Exxx is intended to be used to provide of the device: visible illumination of the surgical field or the patient. The surgical lights BERCHTOLD CHROMOPHARE® E805, E800, E 655, E 650, E 550 and E 520 are intended to illuminate locally the operating site on the patient's body with a high intensity, shadow free, "cold" light.

Device Description

The new BERCHTOLD CHROMOPHARE® E805, E800, E655, E650, E550, and E520 (collectively known and referred to in this document as CHROMOPHARE® Exxx) surgical lights are suitable for all types of surgical procedures. With the use of gas-discharging lamps or Halogen bulbs in these lights, BERCHTOLD realizes high illumination intensity with a lower heat radiation and less pattern variation compared to previous generations of BERCHTOLD products. The light features easy-to-operate swivel arms, auto switching to the secondary lamp in case of failure of the primary lamp, and an easy-to-exchange lamp cartridge. Each light functions with an optional integrated CCD video camera and/or with upward "EndoLite" for endoscopic procedures. The lights could be combined among each other.

AI/ML Overview

The CHROMOPHARE® Exxx surgical lamps' acceptance criteria and the study that proves the device meets the acceptance criteria are described below.

Acceptance Criteria and Reported Device Performance

The acceptance criteria for the CHROMOPHARE® Exxx surgical lights are based on their similarity to predicate devices (CHROMOPHARE® X65 and CHROMOPHARE® D650) and compliance with relevant electrical and surgical lamp performance standards.

The table below summarizes the key performance characteristics of the new devices (CHROMOPHARE® E805 and CHROMOPHARE® E655 are shown as examples for the metal halide discharging lamp technology, and CHROMOPHARE® E800, E650, E550, and E520 for the Halogen lamp technology) and compares them to the predicate devices, effectively serving as the reported device performance against the implicit acceptance criteria set by the predicate devices and performance standards.

CharacteristicLegally marketed device (CHROMOPHARE® X65) Acceptance CriteriaNew devices (CHROMOPHARE® E805) Reported PerformanceNew devices (CHROMOPHARE® E655) Reported PerformanceLegally marketed device (CHROMOPHARE® D650) Acceptance CriteriaNew devices (CHROMOPHARE® E800) Reported PerformanceNew devices (CHROMOPHARE® E650) Reported PerformanceNew devices (CHROMOPHARE® E550) Reported PerformanceNew devices (CHROMOPHARE® E520) Reported Performance
Intended useIllumination of the operating site on a patient's bodysamesameIllumination of the operating site on a patient's bodysamesamesameSame
Input power120V, 1- phase lines, 60Hzsamesame120V, 1- phase lines, 60HzsamesamesameSame
Protection against electrical shockClass IsamesameClass IsamesamesameSame
Diameter of light body650mm799mm649mm650mm799mm649mm579mm579mm
Diameter of polygon reflectors580mm680mm520mm580mm680mm520mm460mm460mm
Number of mirrored reflector elements800720720800720720720720
Lamp technologyMetal halide dischargingsamesameHalogensamesamesameSame
Polygon ReflectorMulti-pieceSingle pieceSingle pieceMulti-pieceSingle pieceSingle PieceSingle pieceSingle piece
Power consumption of bulb72W150Wsame150Wsamesamesame150W
Light / heat filter technology incl. UV light filter mechanismThermoSorb®samesameThermoSorb®samesamesameSame
Color rendering index Ra94samesame9493939393
Color temperature4500°K4300°K4300°K4500°K4300°K4300°K4300°K4300°K
Central illuminance (at 1m)80000 - 160000luxsamesame70000 - 130000lux80000 - 160000lux80000 - 160000lux65000 - 130000lux80000lux
Light field diameter170 - 280mm200 - 330mmsame150 - 280mm200-300mm170 - 280mm150 - 250mm200mm
Depth of Illumination1250mm1200mm1300mm1200mmsame1300mm1250mm1250mm
Total irradiance Ee550W/m²560W/m²560W/m²485W/m²560W/m²507W/m²455W/m²280W/m²
UV- irradiance (≤400nm)2.0W/m²samesame2.6W/m²sameSamesameSame
Light focusing mechanismRotating of HandlesamesameRotating of handlesameSamesameNo
Life time of bulb5000hsamesame1000hsameSamesameSame
Bulb life time indicatorYessamesameNosameSamesameSame
Automatic switching to the reserve bulbYessamesameYessameSamesameSame
Bulb replacement indicatorYessamesameYessameSamesameSame
Reusable steam sterilizable lamp handleYessamesameYessameSamesameSame
Additional light controls in separate wall boxStandardsamesameOptional featureStandardStandardStandardNo
EndoLite in the light head (ambient illumination for minimally invasive surgeries)Optional featuresamesameOptional featuresameSamesameNo
CCD video camera located in sterilizable lamp handleOptional featuresamesameOptional featuresameSamesameNo
Automatic regulation of intensityNoAUTOluxNoNot specifiedNot specifiedNot specifiedNot specifiedNot specified

Study Proving Device Meets Acceptance Criteria

The study demonstrating that the CHROMOPHARE® Exxx devices meet the acceptance criteria is a performance summary and comparison to predicate devices, and compliance with recognized standards. This is a common approach for 510(k) submissions where substantial equivalence is being claimed.

  1. Sample size used for the test set and the data provenance:

    • The document** does not specify a separate "test set" in the context of a clinical study or a dataset for an algorithm.**
    • The evaluation is based on a comparison of the technical specifications and performance metrics of the new devices against established predicate devices (CHROMOPHARE® X65 and CHROMOPHARE® D650) and relevant standards.
    • Data provenance: The data appears to be derived from the product specifications and performance testing conducted by the manufacturer, BERCHTOLD Holding GmbH, presumably in Germany (country of origin of the submitter). The nature of the data is technical specifications and performance characteristics, not patient-derived data, and thus doesn't fit the retrospective/prospective dichotomy in the typical clinical study sense.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This question is not applicable as the submission does not describe a study involving expert assessment of a test set to establish "ground truth" for a diagnostic or AI algorithm. The device is a surgical lamp, and its performance is evaluated against engineering and performance standards, as well as comparison to existing legally marketed devices.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This question is not applicable for the same reasons as above. There was no test set requiring expert adjudication.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices (often AI-powered) where human reader performance is being evaluated, typically with and without AI assistance. The CHROMOPHARE® Exxx is a surgical lamp, not a diagnostic or AI-assisted device in that context.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, a standalone algorithm performance study was not done. The CHROMOPHARE® Exxx is a physical medical device (surgical lamp), not an algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" in this context refers to established engineering and performance standards for surgical lamps, specifically:
      • IEC 60601-2-41:2001 for performance of surgical lamps.
      • IEC 60601-1 for electrical safety.
      • IEC 60601-1-2 for electromagnetic compatibility (implied by electrical safety).
    • Additionally, the performance of the predicate devices (CHROMOPHARE® D650 and CHROMOPHARE® X65) serves as a benchmark for "ground truth" to establish substantial equivalence.
  7. The sample size for the training set:

    • This is not applicable. The device is not an AI/ML algorithm that requires a training set.
  8. How the ground truth for the training set was established:

    • This is not applicable for the same reasons as above.

{0}------------------------------------------------

510 (k) Summary

Submitter:

BERCHTOLD Holding GmbH Ludwigstaler Straße 125 D- 78532 Tuttlingen Germany

Contact Person:

Silke Goral Requiatory Affairs Phone: 0049 7461 181-155 Fax: 0049 7461 181-8155 Silke.Goral@BERCHTOLD.biz

..............................................................................................................................................................................

January 16, 2008 Preparation Date:

Trade Name:

CHROMOPHARE® E 805 CHROMOPHARE® E 800 CHROMOPHARE® E 655 CHROMOPHARE® E 650
CHROMOPHARE® E 550 CHROMOPHARE® E 520

Common Name: Surgical lamp

Classification Name: Light, Surgical, Ceiling mounted

BERCHTOLD CHROMOPHARE® D650 (K965130) Predicate Device: BERCHTOLD CHROMOPHARE® X65 (K024132)

Device Description: The new BERCHTOLD CHROMOPHARE® E805, E800, E655, E650, E550, and E520 (collectively known and referred to in this document as CHROMOPHARE® Exxx) surgical lights are suitable for all types of surgical procedures. With the use of gas-discharging lamps or Halogen bulbs in these lights, BERCHTOLD realizes high illumination intensity with a lower heat radiation and less pattern variation compared to previous generations of BERCHTOLD products. The light features easy-to-operate swivel arms, auto switching to the secondary lamp in case of failure of the primary lamp, and an easy-to-exchange lamp cartridge. Each light functions with an optional integrated CCD video camera and/or with upward "EndoLite" for endoscopic procedures. The lights could be combined among each other.

Intended Use The CHROMOPHARE® Exxx is intended to be used to provide of the device: visible illumination of the surgical field or the patient.

The surgical lights BERCHTOLD CHROMOPHARE® E805, Indications for Use: E800, E 655, E 650, E 550 and E 520 are intended to illuminate locally the operating site on the patient's body with a high intensity, shadow free, "cold" light.

APR 1 0 2008

4080857

{1}------------------------------------------------

Summary of technological: characteristics compared to the predicate device

. . .

The BERCHTOLD CHROMOPHARE® Exxx is substantially equivalent to the surgical lights BERCHTOLD CHROMOPHARE® D 650 (K96513 ) and CHROMOPHARE® X 65 (K024132) Similarities and differences are tabulated below. Any differences between the CHROMOPHARE Exxx and the predicate devices do not alter the safety or efficacy of the device.

Legally marketed deviceNew devices
CHROMOPHARE® X65CHROMOPHARE® E805CHROMOPHARE® E655
Intended useIllumination of the operatingsite on a patient's bodysamesame
Input power120V, 1- phase lines, 60Hzsamesame
Protection againstelectrical shockClass Isamesame
Diameter of light body650mm799mm649mm
Diameter of polygonreflectors580mm680mm520mm
Number of mirroredreflector elements800720720
Lamp technologyMetal halide dischargingsamesame
Polygon ReflectorMulti-pieceSingle pieceSingle piece
Power consumption ofbulb72W150Wsame
Light / heat filtertechnology incl. UV lightfilter mechanismThermoSorb®samesame
Color rendering index Ra94samesame
Color temperature4500°K4300°K4300°K
Central illuminance (at 1m)80000 - 160000luxsamesame
Light field diameter170 - 280mm200 - 330mmsame
Depth of Illumination1250mm1200mm1300mm
Total irradiance Ee550W/m²560W/m²560W/m²
UV- irradiance (≤400nm)2.0W/m²samesame
Light focusing mechanismRotating of Handlesamesame
Life time of bulb5000hsamesame
Bulb life time indicatorYessamesame
Automatic switching to thereserve bulbYessamesame
Bulb replacement indicatorYessamesame
Reusable steamsterilizable lamp handleYessamesame
Additional light controls inseparate wall boxStandardsamesame
EndoLite in the light head(ambient illumination forminimally invasivesurgeries)Optional featuresamesame
CCD video camera locatedin sterilizable lamp handleOptional featuresamesame
Automatic regulation ofintensityNoAUTOluxNo
Legally marketeddeviceCHROMOPHARE®D650New devices
CHROMOPHARE®E800CHROMOPHARE®E650CHROMOPHARE®E550CHROMOPHARE®E520
Intended useIllumination of theoperating site on apatient's bodysamesamesameSame
Input power120V, 1- phase lines,60HzsamesamesameSame
Protectionagainstelectrical shockClass IsamesamesameSame
Diameter oflight body650mm799mm649mm579mm579mm
Diameter ofpolygonreflectors580mm680mm520mm460mm460mm
Number ofmirroredreflectorelements800720720720720
LamptechnologyHalogensamesamesameSame
PolygonReflectorMulti-pieceSingle pieceSingle PieceSingle pieceSingle piece
Powerconsumption ofbulb150Wsamesamesame150W
Light / heatfiltertechnology incl.UV light filtermechanismThermoSorb®samesamesameSame
Color renderingindex Rₐ9493939393
Colortemperature4500°K4300°K4300°K4300°K4300°K
Centralilluminance (at1m)70000 - 130000lux80000 - 160000lux80000 - 160000lux65000 - 130000lux80000lux
Light fielddiameter150 - 280mm200-300mm170 - 280mm150 - 250mm200mm
Depth ofillumination1200mmsame1300mm1250mm1250mm
Total irradianceEₑ485W/m²560W/m²507W/m²455W/m²280W/m²
UV- irradiance(≤400nm)2.6W/m²sameSamesameSame
Light focusingmechanismRotating of handlesameSamesameNo
Life time ofbulb1000hsameSamesameSame
Bulb life timeindicatorNosameSamesameSame
Automaticswitching to thereserve bulbYessameSamesameSame
BulbreplacementindicatorYessameSamesameSame
Reusablesteamsterllizablelamp handleYessameSamesameSame
Additional lightcontrols inseparate wallboxOptional featureStandardStandardStandardNo
EndoLite in thelight head(ambientillumination forminimallyinvasivesurgeries)Optional featuresameSamesameNo
CCD videocamera locatedIn sterilizablelamp handleOptional featuresameSamesameNo

{2}------------------------------------------------

Performance Summary:

·

This device conforms to IEC 60601-2-41:2001 specifications for performance of surgical lamps. This device conforms to IEC 60601-1 and IEC 60601-1-2 for electrical safety.

.

.

·

{3}------------------------------------------------

Public Health Service

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the eagle. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 0 2008

Berchtold Holding GmbH % Underwriters Laboratories, Inc. Mr. Jeff D. Rongero Senior Project Engineer 12 Laboratory Drive Research Triangle Park, North Carolina 27709

Re: K080857

I Trade/Device Name: CHROMOPHARE® E 805, CHROMOPHARE® E 800, CHROMOPHARE® E 655, CHROMOPHARE® E 650, CHROMOPHARE® E 550, CHROMOPHARE® E 520 Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: II Product Code: FSY Dated: March 25, 2008 Received: March 27, 2008

Dear Mr. Rongero:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 – Mr. Jeff D. Rongero

CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, pcrmits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Mark N. Milburn

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K080857

Device Name:

CHROMOPHARE® E 805 CHROMOPHARE® E 800 CHROMOPHARE® E 655 CHROMOPHARE® E 650 CHROMOPHARE® E 550 CHROMOPHARE® E 520

Indications for Use:

The surgical lights BERCHTOLD CHROMOPHARE® E 805, E 800, E 655, E 650, E 550 and E 520 are intended to illuminate locally the operating site on the patient's body with a high intensity, shadow free,

"cold" light.

Prescription Use × (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division
Division of Gascal, Restorative,
and Neurological Dovices

Page 5 of 27
510(k) Number: K680857

§ 878.4580 Surgical lamp.

(a)
Identification. A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.(b)
Classification. Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.