K Number
K080837

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-09-24

(183 days)

Product Code
Regulation Number
866.1640
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cepheid Xpert MRSA/SA Skin and Soft Tissues Infection Assay (Xpert MRSA/SA SSTI Assay) performed in the GeneXpert® Dx System is a qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) from skin and soft tissue infection swabs. The test utilizes automated real-time polymerase chain reaction (PCR) to detect MRSA/SA DNA. The Xpert MRSA/SA SSTI Assay is indicated for use in conjunction with other laboratory tests such as microbiology culture, and clinical data available to the clinician as an aid in the detection of MRSA/SA from skin and soft tissue infections. The Xpert MRSA/SA SSTI Assay is not intended to monitor treatment for MRSA/SA infections. Concomitant cultures for SA and MRSA are necessary to recover organisms for susceptibility testing or epidemiological typing. In a mixed culture containing MRSA/SA and other organisms (e.g. Gram negative bacilli, yeast), results can be false negative or variable depending on the concentration of MRSA/SA present, particularly if the concentration of MRSA/SA is close to the LoD of the assay.

Device Description

The Cepheid Xpert MRSA/SA Skin and Soft Tissue Infection Assay (Xpert MRSA/SA SSTI Assay) is a rapid, automated DNA test for simultaneously detecting MRSA and SA directly from skin and soft tissue specimen is collected on a double swab, which is placed in a tube containing elution reagent. Following brief vortexing, the eluted matcrial and two single-use reagents (Reagent 1 and Reagent 2) that are provided with the assay are transferred to different, uniquely-labeled chambers of the disposable fluidic cartridge (the Xpert MRSA/SA cartridge). The user initiates a test from the system user interface and places the cartridge into the GeneXpert® Dx System instrument platform, which performs hands-off real-time, multiplex polymerase chain reaction (PCR) for detection of DNA. In this platform, additional sample preparation, amplification, and real-time detection are all fully-automated and completely integrated. The GeneXpert® System consists of a GeneXpert instrument, personal computer, and the multi-chambered fluidic cartridges that are designed to complete sample preparation and real-time PCR for detection of MRSA and SA in approximately 50 minutes. Each system has 1 to 16 randomly accessible modules that are each capable of performing separate sample preparation and real-time PCR tests. Each module contains a syringe drive for dispensing fluids, an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and detection. The Xpert MRSA/SA Assay includes reagents for the simultaneous detection of the target organisms, SA and MRSA. The primers and probes in the Xpert MRSA/SA Assay detect nucleic acid sequences of the staphylococcal protein A (spa), the gene for MecA-Mediated Oxacillin resistance (mecA), and staphylococcal cassette chromosome (SCCmec) inserted in the SA chromosomal attB site. The test includes a sample processing control (SPC) to control for adequate processing of the target bacteria and to monitor the presence of inhibitor(s) in the PCR assay. The Probe Check Control (PCC) verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Cepheid Xpert MRSA/SA SSTI Assay, based on the provided text:


1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" numerical targets in a table format. Instead, it presents performance data (Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA)) in comparison to a reference culture method. These performance metrics implicitly serve as the acceptance criteria for regulatory approval.

Implicit Acceptance Criteria (Performance Targets) and Reported Device Performance

Metric (Relative to Reference Culture)Performance Target (Implicit)Reported Device Performance (No Antibiotic Use, n=441)Reported Device Performance (Unknown Antibiotic Use, n=200)Reported Device Performance (Known Antibiotic Use, n=207)
MRSA Detection:
Positive Percent Agreement (MRSA+)High agreement expected for detecting MRSA93.8% (95% CI: 88.6-97.1)94.0% (95% CI: 83.5-98.7)88.0% (95% CI: 75.7-95.5)
Negative Percent Agreement (MRSA+)High agreement expected for correctly identifying absence of MRSA97.3% (95% CI: 94.7-98.8)97.3% (95% CI: 93.3-99.3)92.4% (95% CI: 87.0-96.0)
SA Detection:
Positive Percent Agreement (SA+/MRSA+)High agreement expected for detecting SA (including MRSA, as MRSA is a type of SA)95.7% (95% CI: 92.2-97.9)96.9% (95% CI: 91.2-99.4)95.2% (95% CI: 88.3-98.7)
Negative Percent Agreement (SA+/MRSA+)High agreement expected for correctly identifying absence of SA (and MRSA)89.5% (95% CI: 84.6-93.3)88.3% (95% CI: 80.5-93.8)76.4% (95% CI: 67.9-83.6)
Analytical Performance:
Analytical Specificity (Cross-reactivity)100% (No detection of non-target organisms)100% (All 105 tested isolates reported MRSA negative and SA negative)N/AN/A
Limit of Detection (LoD) - SALowest CFU/swab at which 95% of replicates are positive (e.g., ≤ 150 CFU/swab)51 CFU/swab (N7129, USA900) to 123 CFU/swab (29213, unknown PFGE)N/AN/A
Limit of Detection (LoD) - MRSALowest CFU/swab at which 95% of replicates are positive (e.g., ≤ 350 CFU/swab)82 CFU/swab (MW2, USA400) to 242 CFU/swab (ST59-MRSA-V, USA1000)N/AN/A
Reproducibility (Total Agreement)High agreement across sites and days (e.g., ≥ 95%)94.2% (565/600, 1st study); 99.6% (239/240, 2nd study)N/AN/A
Carry-Over ContaminationNo evidence of carry-over contamination (0 false positives after high positives)0% (All 21 negative samples correctly reported negative after 21 high positives)N/AN/A

Note: The document states "The test results showed the Xpert MRSA/SA SSTI to be substantially equivalent to the current standard of care" and explicitly lists performance data, which are interpreted as meeting implicit acceptance criteria for regulatory clearance.


2. Sample Size Used for the Test Set and the Data Provenance

  • Clinical Test Set Sample Size: A total of 848 specimens were collected from patients.
    • No Antibiotic Use (within 3 weeks): 441 subjects
    • Unknown Antibiotic Use (within 3 weeks): 200 subjects
    • Known Antibiotic Use (within 3 weeks): 207 subjects
  • Data Provenance: The study was a multi-site prospective investigation conducted at four US institutions. This means the data was collected specifically for this study in a forward-looking manner, and from within the United States.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document states that the reference culture method involved "confirmation of presumptive positive colonies was performed with catalase, tube coagulase, and Gram stain. MecA-Mediated Oxacillin resistance was tested by disk diffusion test using a 30 µg cefoxitin disk and cutoff of 21/22 mm." This implies standard microbiology laboratory procedures were followed by trained laboratory personnel.

  • Number of Experts: Not explicitly stated as a specific number of individual "experts."
  • Qualifications of Experts: Implied to be trained microbiologists or clinical laboratory technologists experienced in performing and interpreting standard microbiology culture, identification (catalase, coagulase, Gram stain), and susceptibility testing (cefoxitin disk diffusion). The reference culture was performed at a "centralized laboratory," suggesting specialized expertise.

4. Adjudication Method for the Test Set

The document does not describe a formal adjudication method (like 2+1 or 3+1 consensus) for the ground truth of the clinical test set. The "reference culture" is treated as the definitive ground truth, performed at a centralized laboratory following established protocols. Implicitly, any discrepancies would be resolved by the standard operating procedures of that centralized laboratory, but a multi-reader adjudication process is not detailed.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This device is an automated in vitro diagnostic test (Nucleic Acid Amplification Test) performed on a GeneXpert Dx System, designed to provide results directly without requiring human interpretation of complex images or data that AI might otherwise augment for human readers. Its comparison is against a reference culture method.


6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Yes, a standalone performance study was done. The Xpert MRSA/SA SSTI Assay is an automated system (algorithm only without human interpretation in the loop beyond sample preparation and loading). The "Performance Characteristics" and "Clinical Performance" sections detail the direct comparison of the Xpert MRSA/SA SSTI Assay's results against the reference culture method, making it a standalone evaluation.


7. The Type of Ground Truth Used

The primary ground truth used for the clinical performance study was reference microbiology culture and susceptibility testing. This involved:

  • Enrichment in trypticase soy broth.
  • Streaking onto plates with and without cefoxitin.
  • Sub-culturing presumptive colonies onto blood agar.
  • Confirmation of positive colonies with catalase, tube coagulase, and Gram stain for Staphylococcus aureus (SA) identification.
  • MecA-Mediated Oxacillin resistance tested by disk diffusion using a 30 µg cefoxitin disk and a cutoff of 21/22 mm for MRSA identification.

8. The Sample Size for the Training Set

The document does not explicitly state a sample size for a "training set" in the context of machine learning (AI). This device is a real-time PCR assay, which is a molecular diagnostic method based on known biological reactions, not typically developed using machine learning algorithms that require distinct training sets.

However, if "training set" is interpreted more broadly as data used for analytical development and optimization, the document mentions extensive analytical studies:

  • Analytical Inclusivity Study on CDC Staphylococcus aureus Specimens: 21 MRSA strains and 3 MSSA strains (total 24 strains).
  • Analytical Inclusivity Study on Expanded Panel of Staphylococcus aureus Specimens: 121 additional Staphylococcus aureus strains (78 MRSA, 43 SA).
  • Evaluation of Empty Cassette Variants: 22 isolates.
  • Limit of Detection Studies: Multiple individual isolates (6 MRSA, 3 SA) each tested with 20 replicates at various concentrations.
  • Linearity Studies: SA and MRSA isolates tested across ten-fold serial dilutions.
  • Cross-reactivity Study: 105 strains (various bacteria and yeast).
  • Evaluation of BORSA Strains: 7 BORSA strains.

These analytical tests provide the data and parameters for the assay's design and demonstrate its performance characteristics, which is analogous to the role of a training set for algorithm development, even though it's a traditional molecular test.


9. How the Ground Truth for the Training Set Was Established

For the analytical studies (which can be considered analogous to a training/development phase for a molecular assay), the ground truth was established through:

  • Confirmed Identification: Using CDC-reported strains, phylogenetically characterized strains, ATCC cultures, and NARSA strains.
  • Phenotypic Testing: Catalase, tube coagulase, Gram stain, and cefoxitin disk diffusion (for oxacillin resistance) were used to characterize discrepant results and confirm the identity and resistance profile of isolates.
  • Genetic Characterization: PFGE for USA types and SCCmec typing for MRSA strains provided detailed genetic ground truth.
  • Known Concentrations: For LoD and linearity studies, bacterial concentrations were carefully quantified (CFU/sample) to establish precise analytical ground truth.

{0}------------------------------------------------

K080837

5.0 510(k) Summary

SEP 2 4 2008

As required by 21 CFR Section 807.92(c).

Submitted by:Cepheid904 Caribbean DriveSunnyvale, CA 90489Phone number: (408) 400-8230Fax number: (408) 541-6439
Contact:Russel K. Enns, Ph.D.
Date of Preparation:September 19, 2008
Device:
Trade name:Xpert MRSA/SA SSTI Assay
Common name:Staphylococcus aureus (SA) and methicillin-resistantStaphylococcus aureus (MRSA) from skin and soft tissueinfections Assay.
Type of Test:Nucleic Acid Amplification Test, DNA, Staphylococcusaureus (SA) and Methicillin-resistant Staphylococcusaureus (MRSA), qualitative
Classification name:Antimicrobial susceptibility test powder
Regulation number:866.1640
Procode:NQX
ClassificationAdvisory Committee:Microbiology
Panel:83
Predicate Device:Cepheid Xpert MRSA Assay [510(k) no. K070462]

Device Description:

The Cepheid Xpert MRSA/SA Skin and Soft Tissue Infection Assay (Xpert MRSA/SA SSTI Assay) is a rapid, automated DNA test for simultaneously detecting MRSA and SA directly from skin and soft tissue specimen is collected on a double swab, which is placed in a tube containing elution reagent. Following brief vortexing, the eluted matcrial and two single-use reagents (Reagent 1 and Reagent 2) that are provided with the assay are transferred to different, uniquely-labeled chambers of the disposable fluidic cartridge (the Xpert MRSA/SA cartridge). The user initiates a test from the system user interface and places the cartridge into the GeneXpert® Dx System instrument platform, which performs hands-off real-time, multiplex polymerase chain reaction (PCR) for detection of DNA. In this platform, additional sample preparation, amplification, and real-time detection are all fully-automated and completely integrated.

{1}------------------------------------------------

The GeneXpert® System consists of a GeneXpert instrument, personal computer, and the multi-chambered fluidic cartridges that are designed to complete sample preparation and real-time PCR for detection of MRSA and SA in approximately 50 minutes. Each system has 1 to 16 randomly accessible modules that are each capable of performing separate sample preparation and real-time PCR tests. Each module contains a syringe drive for dispensing fluids, an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and detection.

The Xpert MRSA/SA Assay includes reagents for the simultaneous detection of the target organisms, SA and MRSA. The primers and probes in the Xpert MRSA/SA Assay detect nucleic acid sequences of the staphylococcal protein A (spa), the gene for MecA-Mediated Oxacillin resistance (mecA), and staphylococcal cassette chromosome (SCCmec) inserted in the SA chromosomal attB site.

The test includes a sample processing control (SPC) to control for adequate processing of the target bacteria and to monitor the presence of inhibitor(s) in the PCR assay. The Probe Check Control (PCC) verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.

Device Intended Use:

The Cepheid Xpert MRSA/SA Skin and Soft Tissues Infection Assay (Xpert MRSA/SA SSTI Assay) performed in the GeneXpert® Dx System is a qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) from skin and soft tissue infection swabs. The test utilizes automated real-time polymerase chain reaction (PCR) to detect MRSA/SA DNA. The Xpert MRSA/SA SSTI Assay is indicated for use in conjunction with other laboratory tests such as microbiology culture, and clinical data available to the clinician as an aid in the detection of MRSA/SA from skin and soft tissue infections. The Xpert MRSA/SA SSTI Assay is not intended to monitor treatment for MRSA/SA infections. Concomitant cultures for SA and MRSA are necessary to recover organisms for susceptibility testing or epidemiological typing.

In a mixed culture containing MRSA/SA and other organisms (e.g. Gram negative bacilli, ycast), results can be false negative or variable depending on the concentration of MRSA/SA present, particularly if the concentration of MRSA/SA is close to the LoD of the assay.

Substantial Equivalence:

The Xpert MRSA/SA SSTI Assay is substantially equivalent to two molecular-based MRSA and SA assays, Cepheid Xpert™ MRSA Assay (K070462) and the BD GeneOhm StaphSR Assay (K071026). All three assays detect MRSA and the Xpert MRSA/SA and the BD GeneOhm StaphSR Assay detect both SA and MRSA; all three assays determine the presence of the target organisms through real-time PCR amplification and fluorogenic target-specific hybridization detection. Both Cepheid assays utilize the same fullyautomated instrument system, the Cepheid GeneXpert Dx System.

{2}------------------------------------------------

The Xpert MRSA/SA SSTI Assay simultaneously detects SA and MRSA from skin and soft tissue specimens. The Xpert MRSA Assay detects MRSA from nasal swab specimens. The BD GeneOhm StaphSR Assay detects SA and MRSA from positive blood cultures. Table 5.1 shows the similarities and differences between the Xpert MRSA/SA SSTI Assay and the two molecular-based predicate devices.

The Xpert MRSA/SA SSTI is also substantially equivalent to conventional microbiologybased predicate devices that identify (ID) and/or test for antimicrobial susceptibility (AST) of gram positive organisms, including Staphylococcus species of human origin from pure culture isolates. These conventional microbiology-based predicates accommodate multiple specimen types, including swabbed specimens of skin and soft tissue infections. These conventional microbiology-based predicates are:

  • . Remel Staphaurex Latex Agglutination Test (K851949),
  • BBL (BD) Oxacillin Screen Agar (K863821), .
  • . BD BBL CHROMagar MRSA (K042812
  • BD Phoenix Automated Microbiology ID/AST System (K020322 and K023301). .

Table 5.2 compares the new device with the conventional microbiology-based assays for SA. Table 5.3 compares the new device with the conventional microbiology-based assays for MRSA.

A multi-center study was conducted on 848 patients to determine the performance characteristics of the device relative to the sensitivity and specificity of culture and susceptibility testing, the current standard of care. The test results showed the Xpert MRSA/SA SSTI to be substantially equivalent to the current standard of care, identification of Staphyloccoccus aureus from solid media by catalase, coagulase, and Gram stain, and susceptibility by cefoxitin disk diffusion test.

{3}------------------------------------------------

Table 5.1

Similarities and Differences Between the Xpert MRSA/SA SSTI Assay and the Molecular-based Predicate Devices ______________________________________________________________________________________________________________________________________________________________________________

Similarities
DevicePredicates
ItemXpert MRSA/SA SSTI AssayXpert MRSAAssay(K070462)BD GeneOhm™StaphSR Assay(K071026)
Intended UseRapid detection of MRSA andSAMRSA onlySame
Indication for UseIdentification of MRSA and SAMRSA onlySame
TechnologicalPrinciplesFully-automated nucleic acidamplification (DNA); real-timePCRSameSame
Test CartridgeDisposable single-use, multi-chambered fluidic cartridge.SameDisposablesingle-use PCRtube
Instrument SystemCepheid GeneXpert Dx SystemSameCepheidSmartCycler
FluidicsSelf-contained and automatedafter swab elution and twosingle-dose reagent additions.SameManual
ProbesTaqMan® ProbesSameMolecularBeacons
Internal ControlsSample processing control(SPC) and probe check control(PCC).SameOne internalreagent controland externalpositive andnegative controlsrequired per run
DNA TargetSequenceSequence incorporating theinsertion site (attBssc) ofStaphylococcal CassetteChromosome mec (SCCmec)for detection of MRSA.SameSame
Sequence specific tomethicillin/oxacillin resistance(mecA gene)N/AN/A
Rapid test resultsApproximately 50 minutes toresults.Approximately75 minutes toresults.Approximately60-75 minutes.
Similarities
DevicePredicates
ItemXpert MRSA/SA SSTI AssayXpert MRSAAssay(K070462)BD GeneOhm™StaphSR Assay(K071026)
UsersOperators with no clinical labexperience to experiencedclinical laboratorytechnologists.Same.Categorized asa CLIAmoderatecomplexityassay.CLIA HighComplexityLaboratory Users
Differences
ItemDevicePredicates
Xpert MRSA/SA SSTI AssayXpert MRSAAssay(K070462)BD GeneOhm™StaphSR Assay(K071026)
Intended UseSimultaneous rapid detection ofSA and MRSA.Only detectsMRSA.Same
Specimen TypeDirect from skin and soft tissueinfection swabs.Direct fromnasal swabs.Direct fromPositive BloodCultures
DNA TargetSequenceSequence specific toStaphylococcus aureus species( spa gene)N/ASequence specific toStaphylococcus aureus species( nuc gene)
Ability to identifycorrectly "EmptyCassette Variants"Yes, sequence specific toStaphylococcus aureus species( mecA gene)NoNo

{4}------------------------------------------------

Table 5.2

Similarities and Differences Between the Xpert MRSA/SA SSTI Assay and the Conventional Microbiology-based Predicate Devices for

Staphylococcus aureus (SA) only

*******----------------------------------------------------------------------------------------Allery Production of Children Company of Children--------------------------------------14- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -476 184 1800.
..1. WALLComment on Children and ACALL B.ALL AND AND AND ANPVICE------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Orar10.00LEASE IS AND AND A------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{5}------------------------------------------------

Xpert MRSA/SA SSTIAssayStaphaurexLatexAgglutinationTest for SAK851949BD PhoenixAutomatedMicrobiologySystem for SAK020322
Intended UseDetection of SASameSame
Single useYesSameSame
Assay ControlsPositive Control: SANegative Control: S.epidermidisSameSame

:

:

Differences
ItemDevicePredicates (SA only)
Xpert MRSA/SA SSTIAssayStaphaurexLatexAgglutinationTest for SAK851949BD PhoenixAutomatedMicrobiologySystem for SAK020322
Mode ofDetectionSequence specific toStaphylococcus aureusspecies (spa gene)Clumping factorand protein AMicrobialutilization anddegradation ofspecificsubstrates
Specimen TypeDirect from skin and softtissue infection swabs.StaphylococcusspeciesGram Positiveorganisms
Assay formatAmplification: PCRDetection: Fluorogenic target-specific hybridizationAgglutinationwith latexparticlessensitized withfibrinogen andIgGConventional,chromogenic andfluorogenicbiochemical testsfor identification(ID) andantimicrobialresistance test(AST)
Interpretation oftest resultsDiagnostic software of theGeneXpert Dx SystemVisualinterpretationAutomated

{6}------------------------------------------------

Table 5.3

Similarities and Differences Between the Xpert MRSA/SA SSTI Assay and the Conventional Microbiology-based Predicate Devices for

Similarities
ItemDevicePredicates (MRSA only)
XpertMRSA/SA SSTIAssayMuellerHinton Agarw/ 4% NaCland Oxacillin(OxacillinScreen AgarTest)K863821BBLCHROMagarMRSA(K042812)BD PhoenixAutomatedMicrobiologySystemK023301
IntendedUseDetection ofMRSASameSameSame
Single useYesSameSameSame
AssayControlsPositive Control:MRSANegative Control:SASameSameSame

Methicillin-Resistant Staphylococcus aureus (MRSA)

Differences
ItemDevicePredicates (MRSA only)
XpertMRSA/SA SSTIAssayMuellerHinton Agarw/ 4% NaCland Oxacillin(OxacillinScreen AgarTest)K863821BBLCHROMagarMRSA(K042812)BD PhoenixAutomatedMicrobiologySystemK023301
Differences
ItemDevicePredicates (MRSA only)
XpertMRSA/SA SSTIAssayMuellerHinton Agarw/ 4% NaCland Oxacillin(OxacillinScreen AgarTest)K863821BBLCHROMagarMRSA(K042812)BD PhoenixAutomatedMicrobiologySystemK023301
Mode ofDetectionformethicillinresistanceSCCmec genespecific forMRSAmecA genespecific formethicillin/oxacillinresistanceGrowth onMueller HintonAgar with 4%NaCl and 6 ug/mloxacillinUse of specificChromogenicsubstrates andcefoxitin todifferentiateMRSA fromother organismsUtilizes aredox indicatorfor detection oforganismgrowth in thepresence of anantimicrobialagent
SpecimenTypeDirect from skinand soft tissueinfection swabs.Pure cultureisolate ofStaphylococcusaureusDirect fromAnterior naresPure cultureisolate ofStaphylococcusaureus
AssayformatAmplification:PCRDetection:Fluorogenictarget-specifichybridizationPhenotypicdetection basedon a 24 hourgrowth of SAinoculated onmediaPhenotypicdetection baseon a 24-48 hourgrowth of SA(mauvecolonies)inoculated onmediaAST panelscontainingMIC tests forseveralantimicrobialagents
Interpretation of testresultsDiagnosticsoftware of theGeneXpert DxSystemManual: VisualinterpretationManual: VisualinterpretationAutomated

{7}------------------------------------------------

Non-Clinical Studies:

Analytical Inclusivity Study on CDC Staphylococcus aureus Specimens

The analytical inclusivity of the Xpert MRSA/SA Assay was determined using Staphylococcus aureus strains that were reported by the Center for Disease Control

{8}------------------------------------------------

(CDC) to be representative of MRSA and MSSA strains currently encountered in the healthcare community. All strains were tested in triplicate using 100 µL of stationary phase cell suspensions diluted 10 million-fold. The panel consisted of MRSA strains representing SCCmec types II. IV. IVa. IVb. and IVc in addition to several unknown types. Data supplied by the Center for Disease Control (CDC) indicated these strains, when characterized by pulsed-field gel electrophoresis (PFGE), represent numerous USA types including USA100, the most common hospital-acquired strain and USA300 and USA400, the most common community-acquired strains.1

All 21 MRSA strains were correctly reported MRSA positive; SA positive using the Xpert MRSA/SA Assay. Additionally, each MSSA strain (n=3) was correctly reported MRSA negative: SA positive. Culture using BBL CHROMagar MRSA confirmed all Xpert test results. Colony forming units per assay were determined by plate counts in duplicate.

Analytical Inclusivity Study on Expanded Panel of Staphylococcus aureus Specimens

One hundred twenty-one (121) additional Staphylococcus aureus strains were tested using the Xpert MRSA/SA Assay. Overnight cultures were grown in Brain Heart Infusion (BHI) broth and adjusted to 0.5 McFarland units. All strains were tested in triplicate using 100 µL of cultures further diluted 100 thousand to one million-fold.

MRSA (78) and SA (43) strains were selected to broadly represent the range of genetic diversity found in the species Staphylococcus aureus based on phylogenetic structure. Selections represent primary lineages with emphasis on specific clonal complexes within which MRSA is predominantly observed. Lineages that contain MRSA and SA, as well as those that contain SA exclusively were included.

The Xpert MRSA/SA Assay correctly identified 116 of 121 strains. The 5 discordants were charactcrized by catalase, tube coagulase, and Gram stain. MecA-Mediated Oxacillin resistance was assessed by disk diffusion using a 30 ug cefoxitin disk and a diameter cut-off of 21/22 mm.

Three (3) of 78 MRSA strains were reported MRSA negative; SA positive using the Xpert assay. Further characterization indicates these strains are not resistant and were correctly reported MRSA negative; SA positive.

Two (2) of 43 SA strains were reported MRSA positive; SA positive using the Xpert assav. Further characterization indicates these strains arc resistant and were correctly reported MRSA negative; SA positive.

Each of the 12 known USA300 isolates were correctly reported MRSA positive and SA positive as expected.

1 McDougal L. Steward C. Killgore G. Chaitram J. McAllister S. Tenover F. Pulsed-Field Gel Electrophoresis Typing of Oxacillin-Resistant Staphylococcus aureus Isolates from the United States: Establishing a national Database. J Clin Micro 2003;41(11):5113-20.

{9}------------------------------------------------

Evaluation of Empty Cassette Variants

Twenty-two (22) Staphylococcus aureus isolates identified as "empty cassette variants" were tested using the Xpert MRSA/SA Assay. Overnight cultures were adjusted to 0.5 McFarland units. All strains were tested from cultures further diluted 100-fold (high) and 100 thousand-fold (low).

The Xpert MRSA/SA Assay correctly identified all 22 isolates as MRSA negative and SA positive. At both cell concentrations tested, only Cts for the spa and SCCmec targets were reported. No mecA Cts were reported

Analytical Sensitivity

Limit of Detection Studies

Studies were performed to determine the 95% confidence intervals for the analytical limit of detection (LoD) of Staphylococcus aureus (SA) cells and methicillin-resistant Staphylococcus aureus (MRSA) cells diluted into a surrogate wound matrix of human origin. The surrogate wound matrix consisted of a white blood cell (WBC) concentrate prepared from whole blood by centrifugation. The matrix also contained red blood cells (RBC) and plasma, and a negligible amount of anticoagulant (CPD or CPDA-1). The limit of detection is defined as the lowest number of colony forming units (CFU) per sample that can be reproducibly distinguished from negative samples with 95% confidence or the lowest concentration at which 19 of 20 replicates were positive.

For MRSA. replicates of 20 were evaluated at each MRSA concentration tested (CFU/swab) for 6 individual isolates representing SCCmec types I, II, III, IVa, V, and VI. When characterized by pulsed-field gel electrophoresis (PFGE), USA100, the most common healthcare-acquired strain and USA400, one of the most common communityacquired strains were represented.

For SA. replicates of 20 were evaluated at each SA concentration (CFU/swab) for 3 individual SA isolates. USA types USA900 and USA1200 werc represented.

The estimate and confidence intervals were determined using logistic regression with data (number of positive results per number of replicates at each level) over the range of CFU/swab tested. The confidence intervals were determined using maximum likelihood estimates on the logistic model parameters using the large sample variance-covariance matrix. The LoD point estimates and 95% upper and lower confidence intervals for each SA and each MRSA SCCmec type tested are summarized in Tables 5.4 and 5.5.

Image /page/9/Figure/9 description: This image is a table titled "Table 5.4: LoD and Confidence Intervals - SA". The table contains data for SA Strain ID N7129. The PFGE is USA900, the LoD (CFU/swab) is 51, the lower 95% CI is 42, and the upper 95% CI is 69.

{10}------------------------------------------------

102-04USA12008776109
29213unknown12397188
MRSA Strain IDSCCmec TypePFGELoD(CFU/swab)Lower95% CIUpper95% CI
64/4176IUSA500221195271
N315IIUSA100122106152
11373IIIunknown124115155
MW2IVaUSA4008268113
ST59-MRSA-VVUSA1000242208305
HDE288VIUSA800183161223
64/4176IUSA500221195271

Table 5.5: LoD and Confidence Intervals - MRSA

The results of this study indicate that the Xpert MRSA/SA SSTI Assay will produce a positive SA result 95% of the time with 95% confidence for a wound swab containing 123 CFU and a positive MRSA result 95% of the time with 95% confidence for a wound swab containing 300 CFU.

Linearity

A study was conducted to define the reportable range of the Xpert MRSA/SA Assay and demonstrate a linear relationship between SA and MRSA input and assay output (Ct). Linearity was evaluated using ten-fold serial dilutions (1e8 CFU/sample - 10 CFU/sample) of SA and MRSA isolates.

The Xpert MRSA/SA Assay responds linearly (r2 = 0.998) with respect to spa detection as a function of SA cell input over 6 logs. The mean reportable Ct range = 13.4 to 33.1. PCR efficiency for the spa reaction is 100%.

The Xpert MRSA/SA Assay responds linearly (r2 = 0.999) with respect to spa detection as a function of MRSA cell input over 6 logs. The mean reportable Ct range = 14.3 to 35.0. PCR efficiency for the spa reaction is 95.4%.

The Xpert MRSA/SA Assay responds linearly (r2 = 0.999) with respect to mecA detection as a function of MRSA cell input over 6 logs. The mean reportable Ct range = 14.2 to 35.3. PCR efficiency for the mecA reaction is 93.3%.

The Xpert MRSA/SA Assay responds linearly (r2 = 0.999) with respect to SCCmec detection as a function of MRSA cell input over 5 logs. The mean reportable Ct range for SCCmec is 16.6 to 33.8. PCR efficiency for the mec reaction is 94.6%

Analytical Specificity

{11}------------------------------------------------

Cross-reactivity Study

One hundred five (105) strains were collected, quantitated and tested using the Xpert MRSA/SA Assay. The 98 cultures from the American Type Culture Collection (ATCC) and 7 strains from the Network on Antimicrobial Resistance in Staphylococcus aureus (NARSA) represent species phylogenetically related to Staphylococcus aureus or those potentially encountered in a hospital environment.

Of these, methicillin-sensitive coagulase negative staphylococci (29) and methicillinresistant coagulase negative staphylococci (9) were included. The organisms tested were identified as either Gram positive (74), Gram negative (28), or yeast (3). The organisms were further classified as either aerobic (95) or anaerobic (10).

Two (2) or more replicates of each isolate were tested at 1.7 - 3.2 McFarland units. Under the conditions of the study, all isolates were reported MRSA negative and SA negative; none of the isolates were detected by the Xpert MRSA/SA Assay. Positive and Negative controls were included in the study. The analytical specificity was 100%.

Evaluation of BORSA Strains

Seven (7) well characterized borderline oxacillin-resistant Staphylococcus aureus (BORSA) strains were tested, including one apparent "empty cassette" (see above). Methicillin-resistant Staphylococcus aureus is resistant to all ß-lactam drugs through the alternative penicillin-binding protein PBP2a encoded by meca15. BORSA strains are mecA negative, but exhibit an oxacillin minimum inhibitory concentration (MIC) ≥ 2 and < 8 ug/mL. It is especially valuable to distinguish MRSA from BORSA to prevent the unnecessary and inappropriate use of vancomycin and isolation precautions not warranted for patients infected with a ß-lactam susceptible strain16

Under the conditions of this study, all 7 BORSA isolates (including the apparent "empty cassette" isolate) were reported MRSA negative; SA positive at both high and low cell concentrations using the Xpert MRSA/SA Assay. No mecA signals were reported. These results demonstrate that a BORSA strain will be correctly identified as MRSA negative; SA positive and will not report a false positive MRSA test result using the Xpert MRSA/SA Assay.

Interfering Substances

In the investigational study for Xpert MRSA/SA SSTI Assay, 428 of the 848 specimens were observed to contain blood, and 404 were observed to contain other non-specific substances, which could potentially interfere with the assay (note that some specimens contained more than one type of potential contaminant). Fisher's exact tests conducted on the data generated from swabs with and without these potential interfering substances demonstrated that their presence did not affect the assay performance.

{12}------------------------------------------------

In a non-clinical study, potentially interfering substances that may be present in clinical skin and soft tissue infection specimens were evaluated directly relative to the performance of the Xpert MRSA/SA Assay. Potentially interfering substances in skin and soft tissue infections may include, but are not limited to: blood, pus, plasma, topical ointments (antibiotic/antiseptic/pain relieving), debriding agents, and tinctures. These substances are listed in Table 5.6a and Table 5.6b with the active ingredients and concentrations tested shown. Inhibition of the MRSA/SA assay has been observed with the following substances: StaphA + Septic (5% w/v), Hydrocortisone (5% w/v), and antibacterial hand sanitizer (5% w/v).

SubstanceActive Ingredient% Tested
TET Buffer (control)ControlControl
Buffy Coat (woundstimulant)WBC (1.5e9/mL)50% (v/v)
Whole Blood (MRSA/SAfree)N/A50% (v/v)
Plasma50% (v/v)
Neosporin400 units Bacitracin5,000 units Polymyxin B3.5 mg Neomycin1% and5% (w/v)
StaphA+Septic0.2% BenzethoniumChloride, 2.5% LidocaineHCl1% and5% (w/v)
Hyrdocortisone1% Hyrdocortisone1% and5% (w/v)
Biol-Ease20% Benzocaine1% and5% (w/v)
Iodine Tincture2% Iodine50% (v/v)

Table 5.6a: Potentially Interfering SSTI Substances Tested

Table 5.6b: Potential Interfering SSTI Substances Tested
--------------------------------------------------------------------
SubstanceActive Ingredient% Tested
TET Buffer(control)ControlControl
Mupirocin0.2% BenzethoniumChloride 2.5% LidocaineHCl5% (w/v)
Saline0.65% Sodium Chloride50% (v/v)
Antibacterial handsanitizer62% Ethyl alcohol1% (w/v)

{13}------------------------------------------------

Antibacterial handsanitizer62% Ethyl alcohol5% (w/v)
70% Isopropylalcohol70% Isopropyl alcohol50% (v/v)

Clinical Studies

Performance Characteristics

Clinical Performance

Performance characteristics of the Xpert MRSA/SA SSTI Assay were determined in a multi-site prospective investigation study at four US institutions by comparing the Xpert MRSA/SA SSTI Assav with reference culture. Subjects included individuals whose routine care called for collection of a swab from the patient's skin and soft tissue infection for culture.

Double swabs were collected from each subject. One swab was tested by the Xpert MRSA/SA SSTI Assay at the enrolling center and the other swab was tested by the site's standard method, and the remaining specimen was sent to the central laboratory for reference culture testing.

At the centralized laboratory, the specimen was enriched overnight in trypticase soy broth with 6.5% NaCl. The trypticase soy broth was then streaked onto plates with cefoxitin (for MRSA) and without cefoxitin (for SA). If either or both the SA or MRSA plates showed S. aureus presumptive colonies were sub-cultured onto a blood agar plate. Confirmation of presumptive positive colonies was performed with catalase, tube coagulase, and Gram stain. MecA-Mediated Oxacillin resistance was tested by disk diffusion test using a 30 µg cefoxitin disk and cutoff of 21/22 mm. If the cultures for both the SA and MRSA plates were determined to be negative, the archived trypticase sov broth with 6.5% NaCl was subcultured onto blood agar followed by workup for SA/MRSA as previously described.

Performance of the Xpert MRSA/SA SSTI Assay was calculated relative to the reference culture results.

{14}------------------------------------------------

Overall Results

A total of 848 specimens were tested for MRSA and SA by Xpert MRSA/SA SSTI Assay and culture.

Among the 848 cases in the eligible dataset, antibiotic use within the 3 weeks prior to sample collection was reported for 207 subjects, and no antibiotic use was confirmed for 441 subjects; for 200 cases, antibiotic status was unknown. A statistically significant decrease in the positivity rate of SA with respect to culture results was observed when antibiotics were used (p=0.007); this phenomenon has also been reported in the literature. 10-14 The MRSA positivity rate for culture was also decreased, although to a lesser extent (p=0.022). The decrease in positivity was not observed with the Xpert MRSA/SA Assay when antibiotics were used nor was any inhibition observed in the assay in the presence of topical antibiotics (see Interfering Substances). The decreased culture positivity rates for MRSA and SA in the presence of antibiotics caused the higher than expected false positive rates observed with the Xpert MRSA/SA SSTI Assay.

Five (5) of the 246 MRSA positive cultures had mixed infections of MRSA and SA. Xpert MRSA/SA SSTI identified 3 of the 5 mixed infections as MRSA positive and 2 of the five as SA positive/MRSA negative.

The performance of the Xpert MRSA/SA SSTI Culture Assay is summarized in Tables 5.7a-5.7c.

Table 5.7a: MRSA/SA Performance in Subjects with No Antibiotic Use (within 3 weeks of sample collection) vs. Reference Culture

Culture
XpertMRSA+SA+/MRSA-Neg/No GrowthTotal
MRSA+137a26145
SA+/MRSA-3b791698
SA-64188198
Total14685210441

41 of the 137 were mixed infections of MRSA and SA.

b2 of the 3 were mixed infections of MRSA and SA.

Positive Percent Agreement (MRSA+) = 93.8%; 95%Confidence Interval = 88.6-97.1 Negative Percent Agreement (MRSA+) = 97.3%; 95%Confidence Interval = 94.7-98.8 Positive Percent Agreement (SA+/MRSA+) = 95.7%; 95%Confidence Interval = 92.2-97.9

Negative Percent Agreement (SA+/MRSA+) = 89.5%; 95%Confidence Interval = 84.6-93.3

Among subjects with no antibiotic use within the 3 weeks prior to sample collection, the Xpert MRSA/SA SSTI Assay identified 93.8% of the specimens positive for MRSA and 97.3% of the specimens negative for MRSA relative to the reference culture method, and

{15}------------------------------------------------

95.7% of the specimens positive for SA and 89.5% of the specimens negative for SA relative to the reference culture method.

Among these subjects with no antibiotic use. 96.8% (427/441) were successful on the first attempt with the Xpert MRSA/SA Assay. The remaining 14 gave indeterminate results on the first attempt (6 "INVALID", 7 "ERROR" and 1 "NO RESULT"). Of the 14 indeterminate on the first attempt, all gave a result on the second attempt.

Table 5.7b: MRSA/SA Performance in Subjects with Unknown Antibiotic Use (within 3 weeks of sample collection) vs. Reference Culture

XpertMRSA+SA+/MRSA-Neg/No GrowthTotal
MRSA+47°0451
SA+/MRSA-245855
SA-129194
Total5047103200

Culture

62 of the 47 were mixed infections of MRSA and SA.

Positive Percent Agreement (MRSA+) = 94.0%; 95%Confidence Interval = 83.5-98.7 Negative Percent Agreement (MRSA+) = 97.3%; 95%Confidence Interval = 93.3-99.3 Positive Percent Agreement (SA+/MRSA+) = 96.9%; 95%Confidence Interval = 91.2-99.4

Negative Percent Agreement (SA+/MRSA+) = 88.3%; 95%Confidence Interval = 80.5-93.8

When it was unknown if subjects took antibiotics within the 3 weeks prior to sample collection, the Xpert MRSA/SA SSTI Assay identified 94.0% of the specimens positive for MRSA and 97.3% of the specimens negative for MRSA relative to the reference culture method, and 96.9% of the specimens positive for SA and 88.3% of the specimens negative for SA relative to the reference culture method.

Among these subjects with unknown antibiotic use, 97.0% (194/200) were successful on the first attempt with the Xpert MRSA/SA Assay. The remaining 6 gave indeterminate results on the first attempt (2 "INVALID", 3 "ERROR" and 1 "NO RESULT"). Of the 6 indeterminate on the first attempt, all gave a result on the second attempt.

Table 5.7c: MRSA/SA Performance in Subjects with Known Antibiotic Use (within 3 weeks of sample collection) vs. Reference Culture

Culture
MRSA+SA+/MRSA-Neg/No GrowthTotal
XpertMRSA+4421056
SA+/MRSA-3311953
SA-319498
Total5034123207

Positive Percent Agreement (MRSA+) = 88.0%; 95%Confidence Interval = 75.7-95.5

{16}------------------------------------------------

Negative Percent Agreement (MRSA+) = 92.4%; 95%Confidence Interval = 87.0-96.0 Positive Percent Agreement (SA+/MRSA+) = 95.2%; 95%Confidence Interval = 88.3-98.7

Negative Percent Agreement (SA+/MRSA+) = 76.4%; 95%Confidence Interval = 67.9-83.6

Among subjects with known antibiotic use within the 3 weeks prior to sample collection, the Xpert MRSA/SA SSTI Assay identified 88.0% of the specimens positive for MRSA and 92.4% of the specimens negative for MRSA relative to the reference culture method, and 95.2% of the specimens positive for SA and 76.4% of the specimens negative for SA relative to the reference culture method.

Among these subjects with antibiotic use, 96.1% (199/207) of these eligible specimens were successful on the first attempt with the Xpert MRSA/SA Assay. The remaining 8 gave indeterminate results on the first attempt (5 "INVALID" and 3 "ERROR"). Of the 8 indeterminate on the first attempt, all gave a result on the second attempt.

{17}------------------------------------------------

Empty Cassette Variants

For an isolate to be identified as MRSA positive with the Xpert MRSA/SA SSTI Assay, the test for spa must be positive as well as the test for mecA and SCCmec. An isolate that is positive for spa and SCCmec, but not mecA is reported SA because it is methicillinsensitive. This situation can occur when the portion of the SCCmec element carrying mecA is excised, but the ends of this mobile element remain in place, yielding a positive SCCmec signal. These isolates are sometimes referred to as "empty cassette variants" and are not uncommon in the clinical environment. The significance of these isolates is to potentially confound an assay for MRSA that does not detect the mecA gene directly. The Xpert MRSA/SA Assay was designed to correctly identify these variants as SA.

Among the eligible specimens included in the data analyses presented in this report, a total of 16 isolates fit the empty cassette profile resulting in positive spa and SCCmec test results, but no mecA detection (Ct = 0) as shown in Table 5.8. Fifteen of the 16 were verified MRSA true negative isolates relative to culture, and 14 of 16 were verified true positive SA isolates relative to culture. One isolate was identified as MRSA by culture and 2 isolates were both MRSA and SA negative by culture.

Subject #XpertResultspa(Ct)mecA(Ct)SCCmec(Ct)CultureXpert v. Culture
MRSASA
1SA23.6026.0SATNTP
2SA14.7016.5SATNTP
3SA20.5034.0SATNTP
4SA18.4021.0SATNTP
5SA15.6028.4MRSAFNTP
6SA17.2031.6SATNTP
7SA34.1035.6NegTNFP
8SA29.1033.0SATNTP
9SA12.7023.5SATNTP
10SA18.2027.6SATNTP
11SA18.4022.0SATNTP
12SA25.5027.7SATNTP
13SA20.0022.1NegTNFP
14SA26.0028.3SATNTP
15SA23.9025.7SATNTP
16SA19.9034.0SATNTP

Table 5.8 : MRSA/SA SSTI Performance vs. Reference Culture - Empty Cassette Variants

{18}------------------------------------------------

Carry-Over Contamination Study

A study was conducted to demonstrate that single-use, self-contained GeneXpert cartridges prevent carry-over contamination in negative samples run following very high positive samples in the same GeneXpert module. The study consisted of a negative sample processed in the same GeneXpert module immediately following a very high MRSA positive sample (roughly 10' CFU/test). This was repeated 20 times between 2 GeneXpert modules for a total of 42 runs. There was no evidence of any carry-over contamination. All 21 positive samples were correctly reported MRSA positive: SA positive. All 21 negative samples were correctly reported MRSA negative; SA negative.

Reproducibility Study

A panel of 10 specimens with varying concentrations of SA, MRSA and Staphylococcus epidermidis (negative) were tested in duplicate on 10 different days at each of the three sites (10 specimens x 2 times/ day x 10 days x 3 sites). One lot of Xpert MRSA/SA kit was used at each of the 3 testing sites. Xpert MRSA/SA assays were performed according to the Xpert MRSA/SA procedure.

Specimen IDSite 1Site 2Site 3% Total Agreement
Neg (MSSE)100%(20/20)100%(20/20)100%(20/20)100% (60/60)
SA High Neg100%(20/20)100%(20/20)90%(18/20)96.7% (58/60)
SA Low Pos190%(18/20)95%(19/20)75%(15/20)86.6% (52/60)
SA Moderate Pos2100%(20/20)100%(20/20)95%(19/20)98.3% (59/60)
MRSA1 High Neg100%(20/20)90%(18/20)100%(20/20)96.6% (58/60)
MRSA1 Low Pos185%(17/20)70%(14/20)65%(13/20)73.3% (44/60)
MRSA1 Moderate Pos2100%(20/20)100%(20/20)95%(18/20)96.6% (58/60)
MRSA2 High Neg100%(20/20)100%(20/20)100%(20/20)100% (60/60)
MRSA2 Low Pos1100%(20/20)100%(20/20)95%(19/20)98.3% (59/60)
MRSA2 Moderate Pos2100%(20/20)95%(19/20)95%(19/20)96.6% (58/60)
% Total Agreementby Site97.5%(195/200)95%(190/200)90%(180/200)94.2% (565/600)

Table 5.9 - Summary of Reproducibility Results

{19}------------------------------------------------

SPC
LevelMeanStd Dev%CV
MRSA1 High Neg34.520.822.36
MRSA2 High Neg34.460.852.46
Neg (MSSE)34.440.902.62
SA High Neg34.380.922.66

Table 5.10 - Summary of Ct Value Results by Sample Level and Probe

Spa
LevelMeanStd Dev%CV
MRSA1 Low Pos32.960.82.44
MRSA2 Low Pos31.050.692.21
SA Low Pos33.910.82.35
mecA
LevelMeanStd Dev%CV
MRSA1 Low Pos33.250.802.40
MRSA2 Low Pos31.500.682.16
SCCmec
LevelMeanStd Dev%CV
MRSA1 Low Pos34.190.902.63
MRSA2 Low Pos33.130.682.05

A second reproducibility study was performed using a panel of 4 specimens of (SA: 10X LoD, MRSA1: 10X LoD, MRSA2: 10X LoD, and Negative Control: Staphylococcus epidermidis). The panels were tested in duplicate on 10 different days at each of the three sites (4 specimens x 2 times/ day x 10 days x 3 sites). One lot of Xpert MRSA/SA SSTI Assay was used at each of the 3 testing sites. Xpert MRSA/SA assays were performed according to the Xpert MRSA/SA procedure. The correct results were obtained in 239 of 240 tests.

Table 5.11 - Summary of Second Reproducibility Results

Specimen IDSite 1Site 2Site 3TotalAgreement
Neg (MSSE)100%(20/20)100%(20/20)100%(20/20)100%(60/60)
SA Moderate Pos1100%100%100%100%

{20}------------------------------------------------

(20/20)(20/20)(20/20)(60/60)
MRSA1 Moderate Pos¹100%(20/20)100%(20/20)100%(20/20)100%(60/60)
MRSA2 Moderate Pos¹100%(20/20)100%(20/20)95%(19/20)98.3%(59/60)
% Total Agreement by Site100%(80/80)100%(80/80)98.8%(79/80)99.6%(239/240)

'10X LoD

Table 5.12 – Summary of Ct Value Results by Sample Level and Probe

SPC
LevelMeanStd Dev%CV
MRSA1 Moderate Pos35.721.875.24
MRSA2 Moderate Pos36.292.667.34
SA Moderate Pos34.551.193.44
NEG34.451.063.09
Spa
LevelMeanStd Dev%CV
MRSA1 Moderate Pos29.521.304.40
MRSA2 Moderate Pos28.911.033.57
SA Moderate Pos30.590.912.99
mecA
LevelMeanStd Dev%CV
MRSA1 Moderate Pos29.781.284.29
MRSA2 Moderate Pos29.321.244.22
SCCmec
LevelMeanStd Dev%CV
MRSA1 Moderate Pos31.491.263.99
MRSA2 Moderate Pos31.051.123.59

{21}------------------------------------------------

Conclusions

The results of the nonclinical and clinical performance studies summarized above demonstrate that the MRSA/ST SSTI Assay is as effective, and performs as well as or better than the predicate device.

1000

.

{22}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/22/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a stylized caduceus symbol. The caduceus is a symbol of medicine and healthcare, often depicted as a staff with two snakes coiled around it.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

SEP 2 4 2008

Russel K. Enns, Ph.D. Senior Vice President Regulatory & Clinical Affairs, Quality System and Reimbursement Cepheid, Inc. 904 Caribbean Drive Sunnyvale, CA 94089

Re: K080837

Trade/Device Name: Xpert™ MRSA/SA SSTI Assay Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susccptibility Test Powder Regulatory Class: Class II Product Code: NOX Dated: August 1, 2008 Received: August 4, 2008

Dear Dr. Enns:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{23}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sally attaym

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{24}------------------------------------------------

4.0 Indications for Use Statement

510(k) Number (if known): K080837

Device Name: Xpert MRSA/SA SSTI

Indications for Use:

The Cepheid Xpert MRSA/SA Skin and Soft Tissues Infection Assay (Xpert MRSA/SA SSTI Assay) performed in the GeneXpert® Dx System is a qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) from skin and soft tissue infection swabs. The test utilizes automated real-time polymerase chain reaction (PCR) to detect MRSA/SA DNA. The Xpert MRSA/SA SSTI Assay is indicated for use in conjunction with other laboratory tests such as microbiology culture, and clinical data available to the clinician as an aid in the detection of MRSA/SA from skin and soft tissue infections. The Xpert MRSA/SA SSTI Assay is not intended to monitor treatment for MRSA/SA infections. Concomitant cultures for SA and MRSA are necessary to recover organisms for susceptibility testing or epidemiological typing. In a mixed culture containing MRSA/SA and other organisms (e.g. Gram negative bacilli, yeast), results can be false negative or variable depending on the concentration of MRSA/SA present, particularly if the concentration of MRSA/SA is close to the LoD of the assay.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Freddie L. Poole

Vision Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K080837

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).