K Number
K080800

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-04-07

(17 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACCUVIX V20 system and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: General, abdomen, obstetrics, gynecology, vascular, extremity, pediatric, cardiac, breast, urology, and etc.

Device Description

The ACCUVIX V20 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler mode, Power Doppler mode, PW Spectral Doppler, CW Spectral Doppler mode, and Tissue Doppler Image mode on the LCD display. It also provides the 3D/4D imaging mode using the 3D/4D probe in the Mechanical scan mode. The ACCUVIX V20 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed. The system also provides for the measurement of anatomical structures and for analysis packages that provide information used for clinical diagnostic purposes by competent health care professionals.

AI/ML Overview

The provided text is a 510(k) Summary of Safety and Effectiveness for the ACCUVIX V20 Diagnostic Ultrasound System. It primarily focuses on demonstrating substantial equivalence to predicate devices and outlining the intended uses of the system and its various transducers. Crucially, this document does not contain information about acceptance criteria for specific device performance metrics, nor does it describe any study (clinical or otherwise) that proves the device meets such criteria.

The document details the device's technological characteristics, intended uses, and lists the predicate devices. It also lists the specific transducers and their cleared indications for use, indicating which applications were "previously cleared" (P) under a predicate device, or "new indications" (N) for a particular transducer.

Therefore, it is not possible to provide the requested information based solely on the text provided, as it does not include:

  1. A table of acceptance criteria and reported device performance: This document only outlines the intended uses and general technological characteristics, not specific performance metrics or acceptance thresholds.
  2. Sample size used for the test set and data provenance: No performance studies are described.
  3. Number of experts used to establish ground truth and their qualifications: No ground truth establishment process is mentioned.
  4. Adjudication method for the test set: Not applicable as no test set or study is described.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not mentioned in the document. The device is a diagnostic ultrasound system, not an AI-assisted diagnostic tool in the sense of image analysis software.
  6. Standalone performance (algorithm only): Not applicable for a general-purpose ultrasound system.
  7. Type of ground truth used: Not applicable as no performance studies are described.
  8. Sample size for the training set: Not applicable as this is a hardware device, not an AI model requiring a training set in the conventional sense.
  9. How ground truth for the training set was established: Not applicable.

The document's purpose is to satisfy regulatory requirements for a 510(k) submission, confirming that the new device is "substantially equivalent" to existing legally marketed devices, rather than presenting detailed performance data from a clinical trial or algorithm validation study. It relies on the equivalence of the device's technical specifications and intended uses to existing cleared devices.

{0}------------------------------------------------

Koso800

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of safety and effectiveness is provided as part of this Premarket APR – 7 2008
Notified in annualiance with 21 GED. Box 807, Subman Explons 807 02 Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.

1. Submitter's Information: 21 CFR 807.92(a)(1)

MEDISON CO., LTD. 1003, Daechi-dong, Gangnam-gu, Seoul 135-280, Korea

Contact Person:

Mr. Kyung-Am, Shim Regulatory Affairs Manager

Telephone:82.2.2194.1381
Facsimile:82.2.2194.1399
Email:kashim@medison.com

Data Prepared: February 29, 2008

2. Name of the device:

Common/Usual Name:

Diagnostic Ultrasound System and Accessories

Proprietary Name:

ACCUVIX V20 Diagnostic Ultrasound System

Classification Names:FR NumberProduct Code
Ultrasonic Pulsed Doppler Imaging System892.155090-IYN
Ultrasound Pulsed Echo Imaging System892.156090-IYO
Diagnostic Ultrasound Transducer892.157090-ITX

3. Identification of the predicate or legally marketed device:

K070813, 04/31/2007, ACCUVIX X10 Diagnostic Ultrasound System K063580, 12/14/2006, SONOACE X8 Diagnostic Ultrasound System

4. Device Description:

The ACCUVIX V20 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler mode, Power Doppler mode, PW Spectral Doppler, CW Spectral Doppler mode, and Tissue Doppler Image mode on the LCD display. It also provides the 3D/4D imaging mode using the 3D/4D probe in the Mechanical scan mode.

The ACCUVIX V20 has real time acoustic output display with two basic indices, a

510(k) Summary of Safety and Effectiveness

ATTACHMENT 1(b)

{1}------------------------------------------------

mechanical index and a thermal index, which are both automatically displayed. The system also provides for the measurement of anatomical structures and for analysis packages that provide information used for clinical diagnostic purposes by competent health care professionals.

The ACCUVIX V20 has been designed to meet the following product safety standards:

  • UL 60601-1, Safety requirements for Medical Equipment
  • CSA C22.2 No. 601.1, Safety requirements for Medical Equipment
  • IEC60601-2-37, Diagnostic Ultrasound Safety Standards
  • EN/IEC60601-1, Safety requirements for Medical Equipment
  • EN/IEC60601-1-2, EMC requirements for Medical Equipment
  • NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
  • NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
  • IEC 61157, Declaration of the acoustic output
  • ISO10993-1, Biocompatibility

5. Intended Uses:

The ACCUVIX V20 system and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include:

General, abdomen, obstetrics, gynecology, vascular, extremity, pediatric, cardiac, breast, urology, and etc.

6. Technological Characteristics:

The ACCUVIX V20 is substantially equivalent to the ACCUVIX V10 Diagnostic Ultrasound System, cleared via K070813, and the SONOACE X8 Diagnostic Ultrasound System, cleared via K063580. All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All system allow for specialized measurements of structures and flow, and calculations.

END of 510(K) Summary

510(k) Summary of Safety and Effectiveness

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular fashion around the eagle. The eagle is depicted in black, and the text is also in black.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR - 7 2008

MEDISON CO., LTD. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K080800

Trade/Device Name: ACCUVIX V20 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: March 20, 2008 Received: March 21, 2008

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the ACCUVIX V20 Diagnostic Ultrasound System, as described in your premarket notification:

Transducer Model Number

3D2-6ETC3-7IML7-16IS
3D4-8ETEC4-9ISL8-15IS
3D4-9ESL4-7ELP2-4AC
3D5-9EKL5-12/50EPP3-5AC
C2-5ELL5-13ISCW 2.0
C2-6ICL6-12ISCW 4.0

{3}------------------------------------------------

If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

{4}------------------------------------------------

Page 3 -- Mr. Mark Job

If you have any questions regarding the content of this letter, please contact Ms. Lauren Hefner at (240) 276-3666.

Sincerely yours,

Argen M. White

So Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{5}------------------------------------------------

Section 4.3 INDICATIONS FOR USE

DIAGNOSTIC ULTRASOUND INDICATIONS STATEMENT

510/k) No .

Device Name:ACCUVIX V20 Diagnostic Ultrasound System
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)NNNNNNote 1Notes 2, 7, 8
AbdominalNNNNNNote 1Notes 2, 7, 8
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
& OtherPediatricNNNNNNote 1Note 2,4,5,6,7,8,9
Small Organ (See Note 5)NNNNNNote 1Note 2,5,6,8,9
Neonatal CephalicNNNNNNote 1
Adult CephalicNNNNNNote 1Note 4, 7
Trans-rectalNNNNNNote 1Note 2, 3, 7, 8
Trans-vaginalNNNNNNote 1Note 2, 3, 7, 8
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)NNNNNNote 1Note 2,5,6,8,9
Musculo-skel. (Superfic.)NNNNNNote 1Note 2,5,6,8,9
Intra-luminal
Other (spec.)
Cardiac AdultNNNNNNote 1Note 4, 7
CardiacCardiac PediatricNNNNNNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselNNNNNNote 1Note 5, 6, 9
VesselOther (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/C/VD (Doppler, B/PWD/Color Doppler, B/Color Doppler/M, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Amm Mhz

and Radiological Devic 510(k) Number

{6}------------------------------------------------

080800

510(k) No .: Device Name:

3D2-6ET for use with ACCUVIX V20

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Clinical Application Mode of Operation (includes simultaneous B-mode) General Specific B PWD M CWD Color Combined Other (Track I only) (Tracks I & III) Doppler* (Spec.) (Spec.) Ophthalmic Ophthalmic Fetal (See Note 3) P p p P Note 1 Notes 2, 7, 8 Abdominal P P Note 1 Notes 2, 7, 8 P P Intra-operative (See Note 6) Intra-operative (Neuro.) Fetal Imaging Laparoscopic Pediatric & Other P P P P Note 1 Notes 2, 7, 8 Small Organ (See Note 5) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Cardiac) Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Intra-luminal Other (spec.) Cardiac Adult Cardiac Cardiac Pediatric Trans-esophageal (Cardiac) Other (spec.) Peripheral Peripheral vessel Vessel Other (spec.)

N= new indication; P= previously cleared under K070813; E=added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: BM, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler/M, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Arne Thitz

Division of Reproductive, Abdominal, and Radiological Device

510(k) Number __

{7}------------------------------------------------

510(k) No.:
Device Name:

3D4-8ET for use with ACCUVIX V20

K080800

Intended Use:

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track 1 only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
Ophthalmic
Fetal Imaging& OtherFetal (See Note 3)PPPPNote INotes 2, 7, 8
AbdominalPPPPNote INotes 2, 7, 8
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
PediatricPPPPNote 1Notes 2, 7, 8
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared under K070813; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

tom In the

and Radio 510(k) Number

{8}------------------------------------------------

...

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:K080800
Device Name:3D4-9ES for use with ACCUVIX V20
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote 1Notes 2, 7, 8
Trans-vaginalPPPPNote 1Notes 2, 7, 8
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared under K070813; E=added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Arne Th. Witzke

roductive, Abdominal, and Radiolo 510(k) Number

{9}------------------------------------------------

510(k) No.:
Device Name:
Intended Use:

3D5-9EK for use with ACCUVIX V20

080800

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use: Clinical Application Mode of Operation (* includes simultaneous B-mode) General Specific B M PWD CWD Color Combined* Other (Tracks 1 & III) (Track I only) Doppler (Spec.) (Spec.) Ophthalmic Ophthalmic Fetal (See Note 3) Abdominal Intra-operative (See Note 6) Intra-operative (Neuro.) Fetal Imaging Laparoscopic Pediatric & Other Small Organ (See Note 5) Neonatal Cephalic Adult Cephalic Trans-rectal b p P P Notes 2, 7, 8 Note 1 Trans-vaginal P p P P Note 1 Notes 2, 7. 8 Trans-urethral Trans-esoph. (non-Cardiac) Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Intra-luminal Other (spec.) Cardiac Adult Cardiac Cardiac Pediatric Trans-esophageal (Cardiac) Other (spec.) Peripheral Peripheral vessel Vessel Other (spec.)

N= new indication; P= previously cleared under K063580; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

fory 722

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K080800

{10}------------------------------------------------

K080800

510(k) No .: Device Name: Intended Use:

C2-5EL for use with ACCUVIX V20

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Clinical Application Mode of Operation (* includes simultaneous B-mode) General Specific B M PWD CWD Color Combined* Other (Track I only) (Tracks I & III) Doppler* (Spec.) (Spec.) Ophthalmic Ophthalmic Fetal (See Note 3) P D P P Note I Notes 2, 7, 8 Abdominal P P P P Note I Notes 2, 7, 8 Intra-operative (See Note 6) Intra-operative (Neuro.) Fetal Imaging Laparoscopic & Other Pediatric P P P P Note | Notes 2, 7, 8 Small Organ (See Note 5) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph, (non-Cardiac) Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Intra-luminal Other (spec.) Cardiac Adult Cardiac Cardiac Pediatric Trans-esophageal (Cardiac) Other (spec.) Peripheral Peripheral vessel Vessel Other (spec.)

N= new indication; P= previously cleared under K070813; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Aron M. Wht

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K080800

{11}------------------------------------------------

080800

510(k) No .: Device Name

C2-6IC for use with ACCUVIX V20

Device Name:C2-6IC for use with ACCUVIX V20
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)PPPPNote 1Notes 2, 7, 8
AbdominalPPPPNote 1Notes 2, 7, 8
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Notes 2, 7, 8
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared under K070813; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler, B/Color Doppler/M, B/Color Doppler/CW/D

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Arne Na-Na

(Division Sion-Off Division of Reproductive. Abdominal and Radiolog 510(k) Number

{12}------------------------------------------------

1080800

510(k) No .: Device Name:

3-7IM for use with ACCUVIX V20

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal (See Note 3)PPPPNote 1Notes 2, 7, 8
AbdominalPPPPNote 1Notes 2, 7, 8
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
PediatricPPPPNote 1Notes 2, 7, 8
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

indication; P= previously cleared under K070813; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatic and neonatal patients

Note 6: Abdominal organs and peripheral vessei

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Division Sign Off

uctive, Abdominal and Radiological 510(k) Number

{13}------------------------------------------------

510(k) No .: Device Name: X 080 800

EC4-9IS for use with ACCUVIX V20

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote 1Note 2, 7, 8
Trans-vaginalPPPPNote 1Note 2, 7, 8
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared under K070813; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler/M, B/Color Doppler/M, B/Color Doppler/CWD Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Lorry Tatz

(Dision Sign-Off Division of Reproductive, Abdominal and Radiological Device 510(k) Number

{14}------------------------------------------------

510(k) No .: Device Name:

K080800

L4-7EL for use with ACCUVIX V20

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use: Clinical Application Mode of Operation (includes simultaneous B-mode) Specific B General M PWD CWD Color Combined Other (Track I only) (Tracks I & III) Doppler* (Spec.) (Spec.) Ophthalmic Ophthalmic Fetal (See Note 3) Abdominal Intra-operative (See Note 6) Intra-operative (Neuro.) Fetal Imaging Laparoscopic & Other Pediatric P P P P Note 2, 5, 6, 9 Note 1 Small Organ (See Note 5) P p P P Note l Note 2, 5, 6, 9 Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Cardiac) Musculo-skel. (Convent.) P P P P Note 2, 5, 6, 9 Note l Musculo-skel. (Superfic.) P P P Note 2, 5, 6, 9 P Note l Intra-luminal Other (spec.) Cardiac Adult Cardiac Pediatric Cardiac Trans-esophageal (Cardiac) Other (spec.) Peripheral Peripheral vessel P P p P Nate. I Note 5, 6, 9 Vessel Other (spec.)

N= new indication; P= previously cleared under K060087; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Hore 2n 2
(Division Sign Off)

Reproductive, Abdomin and Radiological Dev 510(k) Number

{15}------------------------------------------------

K080800

510(k) No.: Device Name: Intended Use:

L5-12/50EP for use with ACCUVIX V20

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Note 2, 5, 6, 7, 9
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 9
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 9
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 9
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 9
Other (spec.)

N= new indication; P= previously cleared under K070813; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler/M, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Vorge Th-Wh

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Dev

510(k) Number K080800

{16}------------------------------------------------

Kogogoo

510(k) No .: Device Name:

L5-13IS for use with ACCUVIX V20

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNote 1Note 2, 5, 6, 7, 9
Small Organ (See Note 5)NNNNNote 1Note 2, 5, 6, 7, 9
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)NNNNNote 1Note 2, 5, 6, 7, 9
Musculo-skel. (Superfic.)NNNNNote 1Note 2, 5, 6, 7, 9
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselNNNNNote 1Note 2, 5, 6, 7, 9
Other (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: BM, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler/M, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Tony Trinh

(Division Sign-Off) Division of Reproductive, Abdominal. and Radiological Devices

510(k) Number K080800

{17}------------------------------------------------

510(k) No.:
Device Name:
Concession Company Page of Children And

X080800 L6-12IS for use with ACCUVIX V20

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use: Mode of Operation (*includes simultaneous B-mode) Clinical Application

General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote 1Note 2, 5, 6, 7, 9
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 9
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 9
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 9
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 9
VesselOther (spec.)

N= new indication; P= previously cleared under K070813; E=added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Color Dopper natuador oner ( B/Color Doppler, B/PWD/Color Doppler, B/Color Doppler/M, B/Color Doppler/CWD Note 2: Includes imaging for guidance of biopsy

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatic and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Hingga Th Wh

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _

{18}------------------------------------------------

K080800

510(k) No.: Device Name: Intended Use:

L7-16IS for use with ACCUVIX V20

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNote 1Note 2, 5, 6, 9
Small Organ (See Note 5)NNNNNote 1Note 2, 5, 6, 9
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)NNNNNote 1Note 2, 5, 6, 9
Musculo-skel. (Superfic.)NNNNNote 1Note 2, 5, 6, 9
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselNNNNNote 1Note 2, 5, 6, 9
Other (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler/M, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

hoel 72

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K080800

{19}------------------------------------------------

Ko80800

510(k) No .: Device Name:

L8-15IS for use with ACCUVIX V20

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use: Mode of Operation (includes simultaneous B-mode) Clinical Application PWD CWD Color Combined Other Specific B M General (Tracks I & III) Doppler* (Spec.) (Spec.) (Track I only) Ophthalmic Ophthalmic Fetal (See Note 3) Abdominal Intra-operative (See Note 6) Intra-operative (Neuro.) Laparoscopic Fetal Imaging Note 2, 5, 6, 9 & Other Pediatric P P P Note I P Small Organ (See Note 5) P P P P Note I Note 2, 5, 6, 9 Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Cardiac) Musculo-skel. (Convent.) P P Note l Note 2, 5, 6, 9 P P Note 2, 5, 6, 9 Musculo-skel. (Superfic.) P Note 1 P P P Intra-luminal Other (spec.) Cardiac Adult Cardiac Pediatric Cardiac Trans-esophageal (Cardiac) Other (spec.) Peripheral vessel P Note 2, 5, 6, 9 Peripheral P P D Note, 1 Vessel Other (spec.)

N= new indication; P= previously cleared under K070813; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Norma 7272

Division of Reproductive, Abdominal. and Radiological Devices 510(k) Number

{20}------------------------------------------------

Logogoo

510(k) No .:

P22-4AC for use with ACCUVIX V20

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
AbdominalPPPPPNote 1Note 4, 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicPPPPPNote 1Note 4, 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: BM, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Torna M. Wh

(Division S gn-Off) Division of Reproductive, Abdomina ad Radiological Dev

i0(k) Number K980600

{21}------------------------------------------------

Kosobou

510(k) No .: Device Name: Intended Use:

P3-5AC for use with ACCUVIX V20

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
AbdominalPPPPPNote 1Note 4, 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicPPPPPNote 1Note 4, 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac AdultPPPPPNote 1Note 4 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by K070813; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: BM, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler/M, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

signature

(Division Sign-Off) (Lavision Orgil One)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

{22}------------------------------------------------

K080800
---------

510(k) No .: Device Name:

CW 2.0 for use with ACCUVIX V20

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use: Clinical Application Mode of Operation (* includes simultaneous B-mode) PWD General Specific B M CWD Color Combined* Other (Track I only) (Tracks I & III) Doppler* (Spec.) (Spec.) Ophthalmic Ophthalmic Fetal (See Note 3) Abdominal Intra-operative (See Note 6) Intra-operative (Neuro.) Fetal Imaging Laparoscopic & Other Pediatric Small Organ (See Note 5) Neonatal Cephalic Adult Cephalic p Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Cardiac) Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Intra-luminal Other (spec.) Cardiac Adult P Cardiac Cardiac Pediatric P Trans-esophageal (Cardiac) Other (spec.) Peripheral vessel Peripheral P Vessel Other (spec.)

N= new indication; P= previously cleared under K070813; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppier, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neoratal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

home In Wlz

(Division Sign-Off Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

{23}------------------------------------------------

K080800

510(k) No.: Device Name: Intended Use:

CW 4.0 for use with ACCUVIX V20

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
PediatricP
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
CardiacCardiac AdultP
Cardiac PediatricP
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselP
Other (spec.)

N= new indication; P= previously cleared under K070813; E=added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/CWD, B/Color Doppler, B/PWD/Color Doppler/M, B/Color Doppler/M, B/Color Doppler/CWD

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Vorzu Th Wh

(Division Sign-Off)
Division of Reproductive, Abdominal,
... Radiological Devices
0(k) Number K080800

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.