CYTO-CHEX BCT

K080552 · Streck · GIM · Jul 31, 2008 · Clinical Chemistry

Device Facts

Record IDK080552
Device NameCYTO-CHEX BCT
ApplicantStreck
Product CodeGIM · Clinical Chemistry
Decision DateJul 31, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1675
Device ClassClass 2

Intended Use

Cyto-Chex BCT is intended for the collection and storage of blood specimens for immunophenotyping of WBC by flow-cytometry. Recovery of lymphocyte subset cell markers of the HIV panel can be accomplished over a 14-day period following collection.

Device Story

Cyto-Chex BCT is a 13x75mm glass blood collection tube containing 75.8ul of sterile K3EDTA anticoagulant and WBC preservative; vacuum-sealed for 5ml venipuncture draw. Device preserves lymphocyte subset cell markers (CD3, CD4, CD8) for up to 14 days post-collection. Used by clinicians/technicians in clinical settings to facilitate transport of blood specimens from remote or resource-poor locations to laboratories for flow cytometric analysis. By stabilizing markers, the device enables accurate immunophenotyping beyond the standard 6-hour window, supporting clinical decision-making for HIV patient management.

Clinical Evidence

Clinical study compared peripheral blood samples from healthy donors and HIV-positive patients collected in Cyto-Chex BCT versus K3EDTA tubes. Samples analyzed over 14 days using Becton-Dickinson FACSCalibur and Beckman Coulter EPICS XL flow cytometers. Reference values established from K3EDTA samples at 6 hours. Results showed stability of CD3, CD4, and CD8 markers with correlation statistics (trendline slope >0.90, R-values >0.85).

Technological Characteristics

13x75mm glass vacuum tube; contains 75.8ul K3EDTA anticoagulant and WBC preservative. Principle: chemical stabilization of cell-surface markers for flow cytometry. No electronic components, software, or energy source.

Indications for Use

Indicated for collection and storage of blood specimens for immunophenotyping of WBC by flow cytometry in patients, including those with HIV, requiring lymphocyte subset cell marker analysis.

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Konoss2 1 2008 # 510(k) Summary of Safety and Effectiveness | 510(k) Submitter: | Streck<br>7002 South 109th Street<br>Omaha, NE 68128 | |-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------| | Official Correspondent: | Carol Thompson, Quality Assurance Manager<br>(402) 537-5313 | | Date Prepared: | July 14, 2008 | | Names of Device:<br>Trade Name:<br>Common Name:<br>Classification Name: | Cyto-Chex® BCT<br>Blood specimen collection device<br>Blood specimen collection device, 21CFR862.1675 | | Predicate Device: | Cyto-Chex® BCT, K040107 | # Description: Cyto-Chex BCT consists of a standard 13 x 75mm glass blood collection tube containing 75.8ul of sterile KJEDTA anti-coagulant and WBC preservative. It is manufactured with a vacuum to draw 5ml of blood by venipuncture. #### Intended Use: Cyto-Chex BCT is intended for the collection and storage of blood specimens for immunophenotyping of WBC by flow-cytometry. Recovery of lymphocyte subset cell markers of the HIV panel can be accomplished over a 14-day period following collection. #### Comparison with Predicate Device: The New Cyto-Chex BCT is identical with Cyto-Chex BCT with one difference. The volume of reagent in each tube is 75.8ul instead of 57ul. The concentration of the active agent in the collected blood sample is the same in both tubes. It is used in exactly the same manner as the original Cyto-Chex BCT. Preservation of HIV markers has been established to 14 days from collection instead of 7 days in the original Cyto-Chex BCT. # Testing Performed: Flow cytometric data for lymphocyte subset cell-surface markers was obtained by analysis of peripheral blood samples collected from multiple healthy donors. A separate clinical study was set up in which samples from HIV positive patients were collected and analyzed. Samples were collected in both KyEDTA blood collection tubes and Cyto-Chex BCT tubes. Testing was performed over a period of 14 days using both Becton-Dickinson FACSCalibur and Beckman Coulter EPICS XL flow cytometers. Results were compared to those from fresh samples (6 hours after draw in KzEDTA). Testing was also performed to establish stability of Cyto-Chex BCT reagent and to verify that under filling the tube would not compromise results. # Conclusions Drawn from the Tests: Clinically important HIV markers (CD3, CD4 and CD8) are shown to be stable in samples collected in Cyto-Chex BCT. The results obtained in the EDTA tube within 6 hours represent the reference values for all comparisons. The agreement of these values with those recovered from the Cyto-Chex BCT stabilized samples is confirmed through the calculated correlation statistics. Acceptable stability is marked by a trendline with a slope better than 0.90, preferably 0.95. Re values of 0.85 or better confirm consistency of results. In both HIV positive and healthy donors, markers CD3, CD4 and CD8 are recovered well within the acceptance criteria for absolute cell count. CD19 shows some deterioration at 14 days but additional patient sampling would probably improve the results. HIV positive blood specimens collected in Cyto-Chex BCT can be stored/transported for analysis of HIV markers CD3, CD4 and CD8 as long as 14 days after collection. This will provide a benefit where specimens need to be transported over long distances for analysis such as in resource-poor countries. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three horizontal lines above it, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Streck c/o Kerrie Oetter Quality Assurance Coordinator 7002 South 109th Street Omaha, NE 68128 Re: k080552 Trade/Device Name: Cyto-Check BCT Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code: GIM Dated: July 14, 2008 Received: July 16, 2008 Dear Ms. Oetter: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in intestate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, youg, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listo recei devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. JUL 3 1 2008 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements aft the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing pratis f requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket {2}------------------------------------------------ Page 2 - Streck notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-041), preso, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRice of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may blain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 633-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html . Sincerely yours. Mama In Chan Maria M. Chan, Ph.D. Acting Division Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indication for Use 510(k) Number (if known): 长050552 Ko80552 Device Name: Cyto-Chex® BCTTM Indication For Use: Cyto-Chex® BCT™ is intended for collection and storage of blood specimens for immunophenotyping of WBC by flowcytometry. Recovery of lymphocyte subset cell markers of the HIV panel can be accomplished over a 14-day period following collection. Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use __ (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD) Josephine Bautista Division Sign Off Division Sig Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K080552
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%