K Number
K030596

Validate with FDA (Live)

Manufacturer
Date Cleared
2003-08-08

(164 days)

Product Code
Regulation Number
862.1675
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CellSave tubes are evacuated blood collection tubes that are designed to be used with standard phlebotomy supplies for venous blood collections. The tube contains 300µl of a solution that contains Na2EDTA and a cell preservative preservative preserves morphology and cell surface antigen expression of the epithelial cells and leukocytes. This tube may be used for monitoring of circulating epithelial cells (tumor cells) which may aid in the management of cancer patients. This tube may be used for monitoring CD3+/CD4+ T lymphocyte subsets which may aid in the management of patients with HIV/AIDS.

Device Description

The CellSave™ tube is an evacuated closed glass tube for collecting, transporting and processing blood. The assembly consists of a rubber stopper, a glass tube and anticoagulant. The anticoagulant contains EDTA (disodium EDTA), polyethylene glycol and a preservative reagent.

AI/ML Overview

Here's an analysis of the acceptance criteria and study detailed in the provided text, formatted as requested:

Acceptance Criteria and Device Performance

Device Name: The Immunicon CellSave™ Preservation Tube
Intended Use: Collection and preservation of circulating epithelial cells (tumor cells) and mature human T lymphocytes (CD3+) and helper/inducer (CD3+/CD4+) T lymphocyte subsets in whole blood, for enumeration and phenotyping.

Table of Acceptance Criteria and Reported Device Performance

Acceptance CriterionReported Device PerformanceStudy Type
Circulating Epithelial Cell (Tumor Cell) Recovery and Preservation
Maintains stability of circulating epithelial cells for 72 hours.Clinical Study: Linear correlation (R²) = 1.00, regression equation y=1.1x-7.1 for CTC counts at 24 hrs vs. 96 hrs. Wilcoxon sign-rank test: p > 0.90 (no significant difference).Clinical
T Lymphocyte (CD3+/CD4+) Subsets Preservation
Maintains stability of CD3/CD4 immunophenotyping for 72 hours.Clinical Study: R-values between 0.97 and 0.98, and slopes of 0.91 to 0.97 between Day 1, Day 2, and Day 3. No change in number of CD3 and CD4 positive lymphocytes over 72 hours.Clinical
Recovery of Spiked Tumor Cells (Low Spike)Regression equation y=0.8x+4.7 and R² = 0.98. Mean % recovery: 81.4% to 89.3% across different spike levels (50, 100, 200 cells/7.5 ml).Nonclinical (Spiking)
Recovery of Spiked Tumor Cells (High Spike)Regression equation y=0.9x+6.2 and R² = 0.99. Mean % recovery: 85.7% to 91.3% across different spike levels (100, 1,000, 10,000 cells/7.5 ml).Nonclinical (Spiking)
Non-interference of Hemolysis with Tumor Cell RecoveryMean % recovery: 74% (SD 6%) in hemolyzed samples. Conclusion: Hemolysis does not interfere significantly.Nonclinical (Interference)
Non-interference of Lipemia with Tumor Cell RecoveryMean % recovery: 90% (SD 10%) in lipemic samples. Conclusion: Lipemia does not interfere significantly.Nonclinical (Interference)
Non-interference of Icteris with Tumor Cell RecoveryMean % recovery: 85% (SD 9%) in icteric samples. Conclusion: Icteris does not interfere significantly.Nonclinical (Interference)
Compliance with ISO 6710Met applicable requirements.Nonclinical (Standard Compliance)
Compliance with NCCLS Standard H1-A4Met applicable requirements.Nonclinical (Standard Compliance)

Detailed Study Information:

  1. Sample Size Used for the Test Set and Data Provenance:

    • Circulating Epithelial Cell Preservation Study:
      • Test Set Size: 102 metastatic cancer patients providing 107 blood specimens. 70 specimens had sufficient volume for 24-hour and 96-hour testing, with 21 of these having ≥3 CTCs at both time points.
      • Data Provenance: Blood samples obtained from five geographically dispersed sites in the US. Retrospective or prospective is not explicitly stated but implies prospective collection for the study.
    • CD3/CD4 Immunophenotyping Study:
      • Test Set Size: 50 healthy volunteers and 12 patients with confirmed HIV.
      • Data Provenance: Not explicitly stated, but likely US given the context of other studies. Retrospective or prospective is not explicitly stated but implies prospective collection for the study.
    • Recovery and Interfering Substances Studies (Nonclinical):
      • Test Set Size: Blood from 5 normal donors for recovery; blood from 5 normal donors for interfering substances.
      • Data Provenance: Not explicitly stated, but likely originating from the US where Immunicon Corporation is based. These are in vitro spiking studies using donated blood.
  2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts:

    • No external human experts were explicitly stated as establishing ground truth. The method relies on flow cytometry measurements (using FACSCalibur flow cytometer) for enumeration and phenotyping, which is an objective measurement rather than an expert interpretation. The "ground truth" for recovery studies was the known number of spiked cells or the initial count at 24 hours (for preservation studies).
  3. Adjudication Method for the Test Set:

    • Not applicable as the ground truth relies on objective technical measurements (flow cytometry) rather than human interpretation requiring adjudication.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a MRMC comparative effectiveness study was not done. This device is a blood collection tube intended for sample preservation, not an imaging or diagnostic algorithm that human readers would interpret. Therefore, the effect size of human readers improving with AI vs. without AI assistance is not relevant to this device.
  5. Standalone (Algorithm-Only) Performance:

    • Yes, the performance presented for recovery, interference, and preservation is standalone (i.e., the performance of the tube in maintaining cell viability and antigen expression for subsequent automated analysis). There is no "algorithm" in the traditional sense, but the studies assess the device's ability to preserve the sample for downstream laboratory processing and machine-based analysis.
  6. Type of Ground Truth Used:

    • For Recovery Studies: The ground truth was the known number of spiked cells (SKBR-3 breast cancer cell line).
    • For Preservation Studies (Epithelial Cells): The ground truth was the initial count of circulating epithelial cells at 24 hours using flow cytometry, which was then compared to subsequent counts at 96 hours.
    • For Preservation Studies (T Lymphocyte Subsets): The ground truth was the initial counts of CD3/CD4 positive lymphocytes on Day 1 using flow cytometry, which was then compared to subsequent counts on Day 2 and Day 3.
    • For Interfering Substances Studies: The ground truth was the known number of spiked cells which was used to calculate percentage recovery.
    • In all cases, the primary measurement method for "ground truth" was flow cytometry.
  7. Sample Size for the Training Set:

    • Not applicable. This device is a physical blood collection tube, not an AI/ML algorithm that requires a training set. The studies described are validation studies for the device's performance characteristics.
  8. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

KC030596

510(k) Summary

Date prepared August 5, 2003.

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92 (c).

The assigned 510(k) number is K030596.

The submitter of this premarket notification is Immunicon Corporation, 3401 Masons Mill Road. Suite 100. Huntingdon Valley, PA 19006. The official correspondent is Peter J Scott. Vice President of Quality Assurance and Regulatory Affairs (215-830-0777 ext 235, fax 215-830-0751).

The subject of this summary of Safety and Effectiveness is the Immunicon CellSave™ Preservative Tube. The predicate device is the Becton Dickinson Vacutainer Brand Blood Collection Tube with EDTA Anticoagulant. The Immunicon CellSave Preservative Tube is intended for the collection and preservation of circulating epithelial cells (tumor cells) and mature human T lymphocytes (CD3+) and helper/inducer (CD3+/CD4+) T lymphocyte subsets in whole blood, to be used for enumeration and phenotyping. The device is for in vitro diagnostic use.

The CellSave™ tube is an evacuated closed glass tube for collecting, transporting and processing blood. The assembly consists of a rubber stopper, a glass tube and anticoagulant. The anticoagulant contains EDTA (disodium EDTA), polyethylene glycol and a preservative reagent.

NONCLINICAL TESTING

The CellSave™ Sample Tube was tested and met the applicable requirements of ISO 6710 Single Use containers for venous blood specimen collection and NCCLS Standard H1-A4 Evacuated Tubes and Additives for Blood Specimen Collection-Fourth Edition; Approved Standard.

Studies indicated that within a couple of hours epithelial cell counts decreased significantly when preserved in EDTA alone. It was also observed that separation of leukocyte sub-populations identified by cell surface antigens as measured by flow cytometry decreased over time. The use of the preservative contained within the CellSave™ tube preserved circulating epithelial cells for 72 hours and maintained the separation of the leukocyte sub-populations for 24 hours.

{1}------------------------------------------------

PERFORMANCE

Recoverv

Recovery was evaluated by spiking samples with low tumor cell numbers (0, 50, 100, and 200 cells/7.5 ml) and high tumor cell numbers (0, 100, 1000, and 10,000 cells/7.5 ml). Blood from 5 normal donors was collected into CellSave tubes and spiked with SKBR-3 cells (a breast cancer cell line). Samples were processed and stained with a nucleic acid dye, anti-CD45-APC and anti-CK-PE using the CellPrep™ Semi-Automated Processing System and analyzed using a FACSCalibur flow cytometer with beads to enable calculation of absolute counts of cells. For the low spike experiment, the regression equation was y=0.8x+4.7 and the R2 =0.98. For the high spike experiment, the regression equation was y=0.9x+6.2 and the correlation was R =0.99.

Low spikeHigh Spike
Donor05010020001001,00010,000
A231891642848768,259
B244971414747758,185
C551921751758809,342
D1468115321188468,030
E4528218121069599,014
Mean345881632918678,566
% Recovery89.3%88.2%81.4%91.3%86.7%85.7%

Interfering substances

Blood from 5 normal donors was collected into EDTA and CellSave™ tubes and spiked with approximately 800 SKBR-3 cells. CellSave™ tubes were spiked with potential interfering substances (hemolysis 5+, lipemia 1.94-2.04% emulsified fat, icteris 7.0 mg/dl) to determine the effect on recovery and enumeration of tumor cells. Duplicate samples were processed using CellPrep™ Semi-Automated Sample Processing System and analyzed using the FACSCalibur flow cytometer.

DonorEDTA ControlCellSave™ Control
# Cells Recovered# Cells Spiked% Recovery# Cells Recovered# Cells Spiked% Recovery
A145282855%38869656%
A244582854%48669670%
B1802749107%68969699%
B271174995%69069699%
C158077175%28971640%
C245177158%27271638%
D157177174%55271677%
D264277183%63671689%
E161077179%52671673%
E254177170%53571675%
Mean58175%50672%
SD11717%15022%

{2}------------------------------------------------

CellSave TM, HemolysisCellSave TM, LipemiaCellSave TM, Icteris
Donor# CellsRecovered# CellsSpiked%Recovery# CellsRecovered# CellsSpiked%Recovery# CellsRecovered# CellsSpiked%Recovery
A148269669%66469695%63869692%
A250269672%69172895%61272884%
B151469674%748696107%67869697%
B257169682%712696102%67969698%
C149971670%56871679%56171678%
C247071666%59971684%51471672%
D158271681%62871688%65171691%
D255171677%54971677%58971682%
E157171680%62071687%55471677%
E249971670%62071687%58471682%
Mean52474%64090%60685%
SD416%6310%559%

Hemolysis, lipemic and icteric whole blood samples collected into the CellSave™ tube do not interfere with the recovery and enumeration of tumor cells.

CLINICAL TESTING

A Clinical study was performed at Immunicon using blood samples obtained from five geographically dispersed sites from 102 metastatic cancer patients providing 107 blood specimens for analysis. Seventy of these specimens had sufficient blood volume for testing at approximately 24 hours and again at approximately 96 hours. Initial circulating epithelial cell (tumor cell, or CTC) counts ranged from 0 to 640 CTC per sample. Twenty-one of these specimens had an average of greater than or equal to 3 CTC at the two testing time points. The linear correlation for CTC recovery over this time period comparing 24 hours to 96 hours was an R2 equal to 1.00 with a regression equation of y=1.1x-7.1. A Wilcoxon signrank test indicated that the CTC counts obtained at 24 hours and those obtained at 96 hours were not significantly different (p > 0.90). These data demonstrate that the recovery of circulating epithelial cells from whole blood remains stable over a 72 hour time period using the CellSave™ blood collection tube.

A second clinical study was performed to compare CD3/CD4 immunophenotyping over a 72 hour time period using both EDTA and CellSave™ tubes. Whole blood samples were obtained from fifty healthy volunteers and twelve patients with confirmed HIV. Results of CD3/CD4 testing on days 1, 2, and 3 resulted in Revalues during the three days of between 0.97 and 0.98 and slopes of 0.91 to 0.97.

Together, these studies demonstrate that the CellSave™ blood collection tube is effective in preserving T lymphocytes and epithelial cells for phenotyping and enumeration over a 72 hour time period. The numbers of CD3 and CD4 positive lymphocytes and epithelial cells are unchanged over a 72-hour period. Preservation of T lymphocytes and their antigens are also effective using different instruments for enumeration of labeled cells, which demonstrates that the CellSave™ tube is useful across multiple instruments.

These studies justify the use of the CellSave tube for drawing, shipping and storing venous blood up to 72 hours for the counting and immunophenotyping of epithelial cells and leukocytes.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a caduceus, a traditional symbol of medicine.

AUG - 8 2003

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Peter J. Scott Vice President of Quality Assurance and Regulatory Affairs Immunicon® Corporation 3401 Masons Mill Road - Suite 100 Huntingdon Valley, PA 19006

Re: K030596

Trade/Device Name: The Immunicon CellSave™ Preservation Tube Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code: JKA Dated: June 3, 2003 Received: June 4, 2003

Dear Mr. Scott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page of

510(k) Number (if known):K030596
Device Name:The Immunicon CellSave TM Preservation Tube
Indications For Use:

CellSave tubes are evacuated blood collection tubes that are designed to be used with standard phlebotomy supplies for venous blood collections. The tube contains 300µl of a solution that contains Na2EDTA and a cell preservative preservative preserves morphology and cell surface antigen expression of the epithelial cells and leukocytes. This tube may be used for monitoring of circulating epithelial cells (tumor cells) which may aid in the management of cancer patients. This tube may be used for monitoring CD3+/CD4+ T lymphocyte subsets which may aid in the management of patients with HIV/AIDS.

  • prescription use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Josephine Bautista

Division Sign-Off

(Optional Format 3-10-98)

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K030596

§ 862.1675 Blood specimen collection device.

(a)
Identification. A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.(b)
Classification. Class II.