K Number
K080334

Validate with FDA (Live)

Device Name
INSTARAD
Date Cleared
2008-02-21

(14 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

InstaRad™ is a software device (DICOM Gateway Application, InstaRad Enterprise Server, and Workstation) used for viewing and manipulating medical images. Digital images (including mammography and data from various sources (including CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or imaging sources) can be displayed, processed, stored and communicated across computer networks using this software. InstaRad™ can be integrated with an institution's existing Hospital Information System (HIS) or Radiology Information System (RIS) based on the study of the System, providing seamless access to reports for fully-integrated electronic patient records.

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

Device Description

InstaRad™ is an application used for viewing and manipulating medical images. Digital images and data from various sources (including CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or imaging sources) can be displayed, processed, stored and communicated across computer networks using this software. When viewing images, users can perform adjustments of window width and level, annotation, and various image manipulations. In addition, InstaRad™ can be integrated with an institution's existing Hospital Information System or Radiology Information System (based on the study of the System), providing seamless access to reports for fully-integrated electronic patient records.

InstaRad™ allows multiple centers or hospitals to send their images to a central server where Radiologists can view images over Web. It consists of following components:

  • DICOM Gateway Application is deployed at all remote centre or . hospitals. It receives images from modalities over LAN and uploads them to central server.
  • Central Server runs the Web Server and Image Server and provide . study and image data to doctors.
  • . WorkStation - Doctor at client side access the study list in the browser. They can select the patient and download the images.
AI/ML Overview

The provided text describes InstaRad™, a Picture Archiving Communications System (PACS), and its 510(k) summary for regulatory approval. However, the document does not contain any information about acceptance criteria, device performance studies, sample sizes, expert involvement, or ground truth establishment related to the device itself as a medical image processing system.

The text focuses on establishing substantial equivalence to a predicate device (VISAGE PACS/CS, RELEASE VERSION 4.1, K062490) based on its technological characteristics, intended use, and general safety and effectiveness.

Therefore, I cannot provide a table of acceptance criteria and reported device performance from the given information because that data is absent from the document. The subsequent points regarding studies are also not addressable for the same reason.

Here's a breakdown of what the document does provide and why the requested information is missing:

  • Device Description: InstaRad™ is an application for viewing, manipulating, processing, storing, and communicating digital medical images from various sources (CT, MR, US, RF, etc.) across computer networks. It can be integrated with HIS/RIS.
  • Intended Use: Viewing and manipulating medical images, including mammography (with specific caveats regarding lossy compression and monitor requirements).
  • Technological Characteristics: It's a software product handling digital medical images, not contacting the patient, and not controlling life-sustaining devices. Human intervention is expected for interpretation.
  • Predicate Device: VISAGE PACS/CS, RELEASE VERSION 4.1 (K062490).
  • Conclusion: The submission contains adequate information to determine substantial equivalence to the predicate device.

Missing Information:

  1. Acceptance criteria and reported device performance: Not specified. This type of detail is often found in performance testing reports, which are not included in this 510(k) summary.
  2. Sample sizes used for the test set and data provenance: No performance test data is mentioned.
  3. Number of experts used to establish ground truth for the test set and their qualifications: Not mentioned.
  4. Adjudication method: Not mentioned.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not mentioned.
  6. Standalone (algorithm-only) performance: Not mentioned, as this device functions as a PACS, not an AI algorithm performing a diagnostic task.
  7. Type of ground truth used: Not mentioned.
  8. Sample size for the training set: Not applicable for this type of PACS device, as it doesn't appear to use machine learning for diagnostic tasks that would require a "training set."
  9. How the ground truth for the training set was established: Not applicable.

The 510(k) process for a PACS system like InstaRad™ typically focuses on functional equivalence, compliance with DICOM standards, security, data integrity, and display capabilities compared to an existing predicate. It's not usually about diagnostic accuracy studies with ground truth in the same way an AI diagnostic algorithm would be evaluated.

{0}------------------------------------------------

K080334

FEB 2 1 2008

510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

Date Prepared:

January 2, 2008

Submitter's Information: 21 CFR 807.92(a)(1)

Mr. Nishant Mitta! Co-Founder and Director, R&D MedSphere Technologies Pvt. Ltd. GF, B-18, 2nd Main, KEB Layout, BTM 1st Stage Bangalore, Karnataka, India - 560 076 Tel: +91-80-64514222126684959 Fax: +91-80-26684959

Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)

Trade Name:InstaRad™
Common Name:Picture Archiving Communications System
Device Classification:892.2050 System, Image Processing
Product Code:LLZ

Predicate Device: 21 CFR 807. 92(a)(3)

InstaRad™ is substantially equivalent to:

Device Classification Namesystem, image processing, radiological
510(k) NumberK062490
Device NameVISAGE PACS/CS, RELEASE VERSION 4.1
ApplicantMERCURY COMPUTER SYSTEMS, INC199 Riverneck Road,Chelmsford, MA 01824
Regulation Number892.2050
Classification Product CodeLLZ
Date Received08/31/2006
Decision Date10/27/2006
Decisionsubstantially equivalent (SE)
Classification Advisory CommitteeRadiology
Review Advisory CommitteeRadiology

Device Description: 21 CFR 807 92(a)(4)

InstaRad™ is an application used for viewing and manipulating medical images. Digital images and data from various sources (including CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or imaging sources) can be displayed, processed, stored and communicated across computer networks using this software. When viewing

{1}------------------------------------------------

510(k) Summary of Safety and Effectiveness

images, users can perform adjustments of window width and level, annotation, and various image manipulations. In addition, InstaRad™ can be integrated with an institution's existing Hospital Information System or Radiology Information System (based on the study of the System), providing seamless access to reports for fully-integrated electronic patient records.

InstaRad™ allows multiple centers or hospitals to send their images to a central server where Radiologists can view images over Web. It consists of following components:

  • DICOM Gateway Application is deployed at all remote centre or . hospitals. It receives images from modalities over LAN and uploads them to central server.
  • Central Server runs the Web Server and Image Server and provide . study and image data to doctors.
  • . WorkStation - Doctor at client side access the study list in the browser. They can select the patient and download the images.

Indications for Use: 21 CFR 807 92(a)(5)

InstaRad™ is a software device (DICOM Gateway Application, InstaRad™ Enterprise Server, and Client Workstation) used for viewing and manipulating medical images. Digital images (including mammography and data from various sources (including CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or imaging sources) can be displayed, processed, stored and communicated across computer networks using this software. InstaRad™ can be integrated with an institution's existing Hospital Information System (HIS) or Radiology Information System (RIS) based on the study of the System, providing seamless access to reports for fully-integrated electronic patient records. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

Technological Characteristics: 21 CFR 807 92(a)(6)

InstaRad™ device is a software product that handles digital medical images. The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.

Conclusion: 21 CFR 807 92(b)(1)

The 510 (k) Pre-Market Notification for InstaRad™ contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device.

InstaRad™has been and will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The submission contains the results of a hazard analysis and the "Level of Concern for potential hazards has been classified as "Minor".

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines extending from its back, representing the department's mission to protect the health of all Americans. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

FEB 21 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Medsphere Technologies Pvt. Ltd. c/o Mr. Ned Devine Sr. Staff Engineer and Reviewer Underwriters Laboratories. Inc. 333 Pfingsten Road NORTHBROOK IL 60062-2096

K080334 Re:

Trade/Device Name: InstaRad" Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 30, 2008 Received: February 7, 2008

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA' may publish further announcements concerning your device in the Rederal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

K080334 510(k) Number (if known):

Device Name: InstaRad™

Indications for Use:

InstaRad™ is a software device (DICOM Gateway Application, InstaRad Enterprise Server, and Workstation) used for viewing and manipulating medical images. Digital images (including mammography and data from various sources (including CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or imaging sources) can be displayed, processed, stored and communicated across computer networks using this software. InstaRad™ can be integrated with an institution's existing Hospital Information System (HIS) or Radiology Information System (RIS) based on the study of the System, providing seamless access to reports for fully-integrated electronic patient records.

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Division of Reproductive, Abdominal and . Radiological Devices 510(k) Number

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).