K Number
K080198

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-05-30

(126 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TSO3 Ozone Sterilization Wrap is a single-use, non-sterile device. It is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical device, and also to maintain sterility of the enclosed device until used. This wrap is intended for use with TSO3 Ozone Sterilization process.

Device Description

The TSO3 Ozone Sterilization Wrap is a single-use, non-sterile device. It is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical device, and also to maintain sterility of the enclosed device until used. This wrap is intended for use in the TSO3 Ozone Sterilization process.

The TSO3 Ozone Sterilization Wrap is a 2 layer laminate consisting of an ePTFE membrane and a 100% Polyethylene/Polyester (PE/PET) bi-component nonwoven backer.

AI/ML Overview

The TSO3 Ozone Sterilization Wrap is a single-use, non-sterile device intended to enclose a medical device for sterilization and maintain its sterility until use, specifically with the TSO3 Ozone Sterilization process.

Here's a breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

CriteriaAcceptance CriteriaReported Device Performance
Effectiveness (Sterilization Performance)Not explicitly stated, but implied to meet industry standards for sterilization wraps."Sterilization performance studies... were conducted and all acceptance criteria were met."
Effectiveness (Shelf-life Sterility)Not explicitly stated, but implied to meet industry standards for maintaining sterility over time."...and shelf-life sterility tests were conducted and all acceptance criteria were met."
Biocompatibility (Skin Irritation)No induction of skin irritation."A skin irritation test was performed by an independent laboratory to demonstrate that the final product does not induce any biocompatibility hazard." "All acceptance criteria established in the applicable portions of the standards were met."
Biocompatibility (Cytotoxicity)No induction of cytotoxicity."These materials [ePTFF/PE/PET] were evaluated for skin irritation, cytotoxicity testing, and sensitization... All acceptance criteria established in the applicable portions of the standards were met."
Biocompatibility (Sensitization)No induction of sensitization."These materials [ePTFF/PE/PET] were evaluated for skin irritation, cytotoxicity testing, and sensitization... All acceptance criteria established in the applicable portions of the standards were met."

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for the test set for the sterilization performance and shelf-life sterility studies. It only mentions that "sterilization performance studies and shelf-life sterility tests were conducted."

For biocompatibility testing, it states that "A skin irritation test was performed by an independent laboratory to demonstrate that the final product does not induce any biocompatibility hazard." The specific sample size for this test is not provided in the summary.

The provenance of the data (country of origin, retrospective/prospective) is not specified in the provided text.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the given text.

4. Adjudication Method for the Test Set

This information is not provided in the given text.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A multi-reader multi-case (MRMC) comparative effectiveness study was not done as this device is a sterilization wrap, not an AI-assisted diagnostic tool where human readers would be involved in interpreting results. Therefore, there is no effect size of human readers improving with AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This is not applicable as the device is a physical sterilization wrap, not an algorithm.

7. Type of Ground Truth Used

  • For Effectiveness (Sterilization Performance and Shelf-life Sterility): The ground truth would likely be established by microbiological testing, confirming the absence of viable microorganisms after sterilization and during the shelf-life. The document states "all acceptance criteria were met," implying these tests confirmed the desired outcome.
  • For Biocompatibility (Skin Irritation, Cytotoxicity, Sensitization): The ground truth was established by laboratory testing according to ISO 10993 standard, part 1. This involves observing biological responses (e.g., cell viability, skin reactions) to the device materials.

8. Sample Size for the Training Set

This information is not applicable as the device is a physical product, not an AI model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the letters "TSO" in a simple, sans-serif font. The letters are arranged horizontally, with "TS" appearing to be connected. There is a faint, dotted line that forms a partial circle above and to the right of the "O", suggesting a possible incomplete or erased element in the image.

510(k) Summary of Safety and Effectiveness

Applicant's Name and Address

TSO3 Inc. 2505, avenue Dalton Québec (Quebec) Canada G1P 3S5

Contact Person, Telephone, FAX

Marc Chaunet, Quality Assurance and Regulatory Affairs Tel : (418) 651-0003 ext.242 FAX : (418) 653-5726 E-mail: mchaunet@tso3.com

U.S. Agent

Charles O. Hancock Associates, Inc. Contact person: Charles O. Hancock, RAC Tel : (585) 223-1850 FAX : (585) 223-6855 E-mail : chancock@rochester.rr.com

Submission Date

January 23, 2008

Trade Namc TSO3 Ozone Sterilization Wrap

Common Name CSR Wrap or Sterilization Wrap

Classification Name Wrap, Sterilization Class II (as per 21CFR, part 880.6850 cquivalent device)

Legally Marketed Equivalent Device Name(s) Suprashield Express® Sterilization Wrap (K990300) MAY 3 0 2008

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the letters "TS" followed by a set of parentheses. The letters are in a simple, sans-serif font and appear to be bolded. The parentheses are slightly smaller in size compared to the letters. The image has a grainy texture, particularly around the edges of the letters and parentheses.

K080198
p2of2

Description of Device

The TSO3 Ozone Sterilization Wrap is a single-use, non-sterile device. It is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical device, and also to maintain sterility of the enclosed device until used. This wrap is intended for use in the TSO3 Ozone Sterilization process.

The TSO3 Ozone Sterilization Wrap is a 2 layer laminate consisting of an ePTFE membrane and a 100% Polyethylene/Polyester (PE/PET) bi-component nonwoven backer.

Effectiveness

Sterilization performance studies and shelf-life sterility tests were conducted and all acceptance criteria were met.

Safety

The material used in the composition of the TSO3 Ozone Sterilization Wrap (ePTFF/PE/PET) was evaluated and tested as required in ISO 10993 standard, part 1. These materials were evaluated for skin irritation, cytotoxicity testing, and sensitization. Results were submitted as part of the 510(k) submission for the TSO3 125L Ozone Sterilizer (K020875). All acceptance criteria established in the applicable portions of the standards were met.

A skin irritation test was performed by an independent laboratory to demonstrate that the final product does not induce any biocompatibility hazard. Refer to Annex V for test protocol and results (test report).

Substantial Equivalence

The TSO3 Sterilization Wrap is substantially equivalent to the Suprashield Express® Sterilization Wrap (K990300) in that the:

  • Intended use is the same
  • Performance attributes are the same

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a high-resolution image.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 3 0 2008

TSO3 Incorporated C/O Mr. Charles O. Hancock Regulatory Affairs Consultant Charles O. Hancock Associates, Incorporated 33 Black Watch Trail Fairport, New York 14450

Re: K080198

Trade/Device Name: TSO3 Ozone Sterilization Wrap Regulation Number: 880,6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: May 13, 2008 Received: May 13, 2008

Dear Mr. Hancock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Hancock

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Ching-Lin, Ph.D.

Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K080198

Device Name: TSO3 Ozone Sterilization Wrap

Indications For Use:

The TSO3 Ozone Sterilization Wrap is a single-use, non-sterile device. It is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed medical device, and also to maintain sterility of the enclosed device until used. This wrap is intended for use with TSO3 Ozone Sterilization process.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Office of Device Evaluation (ODE) Concurrence of CDRH.

Shala A. Murphy, MD

(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

Page 1 of __

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).