K Number
K080178

Validate with FDA (Live)

Date Cleared
2008-07-28

(186 days)

Product Code
Regulation Number
870.4885
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Valvulotome by Koven Valvulotome is used for the treatment of vascular disorders and more particularly for excising or disrupting venous valves when performing in-situ vascular procedures, such as lower extremity arterial bypass surgery.

Device Description

The Valvulotome by Koven is a medical device that has a deployable assembly with multiple deployable cutting blades and guide fins which permit continual centralizing, and alignment of the cutting blades within the lumen of the vein as the valves are being cut. The device includes a handle operatively associated with this deployable assembly through a catheter operated by the handle in a one-handed operation for deploying the guide fins and cutting blades of the deployable assembly having respective outer diameters at each deployment position. The Valvulotome by Koven consists of three main parts; the handle, the catheter and the cutter head.

AI/ML Overview

The provided text describes a medical device, the "Valvulotome by Koven," and its 510(k) summary for FDA clearance. However, it does not contain information about acceptance criteria, device performance metrics, or details of a study that would typically quantify its performance against such criteria.

The 510(k) summary primarily focuses on:

  • Device Description: What the device is, its parts, and materials.
  • Intended Use: For excising or disrupting venous valves during in-situ vascular procedures.
  • Substantial Equivalence: Comparison to a predicate device (Lemaitre Valvulotome II, K946352).
  • Testing: It states that testing was done, including "laboratory bond strength testing of all joined elements" and "bench testing for the determination of adequacy and safety of the device for the disruption of valves in blood vessels of the leg." It concludes that the device "successfully completed all testing with positive end points achieved."

Missing Information:

The document explicitly lacks the following information required to answer your prompt adequately:

  • Specific Acceptance Criteria: No quantitative thresholds or qualitative standards are defined for "successful completion."
  • Reported Device Performance: No specific data (e.g., success rates, force measurements, blade sharpness metrics, time to complete a procedure, etc.) from the testing is provided.
  • Study Design Details:
    • No sample size for test sets (e.g., number of valves tested, number of devices tested).
    • No data provenance (country, retrospective/prospective).
    • No mention of experts, ground truth establishment, or adjudication for any performance testing.
    • No MRMC study details or AI components.
    • No standalone algorithm performance (as it's a mechanical device).
    • No details on training set size or ground truth for a training set.

Therefore, based solely on the provided text, I cannot complete the requested tables and details.

The document's statement: "The Valvulotome by Koven successfully completed all testing with positive end points achieved" is a summary of the outcome, but it does not provide the underlying data or the specific criteria for "successful completion" that would be needed to fill the table. It indicates that some form of "adequacy and safety" was determined through bench testing related to disrupting valves, but the metrics are not specified.

{0}------------------------------------------------

KC80178 page lot 3

510(k) (Traditional) Submission Section 5, Summary

Image /page/0/Figure/2 description: The image shows the logo for Koven Technology Incorporated. The words "Vascular Instruments" are at the top of the logo. The word "Koven" is in large font in the middle of the logo, and the words "Technology Incorporated" are at the bottom of the logo.

JUL 2 8 2008

Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92 for the Valvulotome by Koven.

1. Company making the submission:

Name:Koven Technology, Inc.
Address:12125 Woodcrest Executive Drive,
St. Louis, MO 63141
Telephone:314-542-2101 Voice
314-542-6020 Fax
Contact:Heather Bell
E-mail:hbell@koven.com
Submission Contact:J. Harvey Knauss, Delphi Consulting Group
Address:11874 South Evelyn Circle
Houston, Texas 77071-3404
Telephone:713-723-4080 Voice
832-615-3550 Fax
E-mail:Harvey@delphiconsulting.com

2. Device Name;

Trade/Proprietary Name: Valvulotome by Koven Common/Usual Name: Valvulotome - External Vein Stripper Classification Name: External Vein Stripper Regulation Number: 870.4885 Product Code: MGZ

3. Predicate Devices:

The Valvulotome by Koven is substantially equivalent to other Valvulotome devices in the market such as the Lemaitre Valvulotome II, K946352, manufactured by Vascutech, Inc., North Andover, MA 01854.

4. Indications for Use Statement

The Valvulotome by Koven Valvulotome is used for the treatment of vascular disorders and more particularly for excising or disrupting venous valves when performing in-situ vascular procedures, such as lower extremity arterial bypass surgery.

Summary Rey B don Houston, TX 7707

{1}------------------------------------------------

Korelitz page 2 of 3

510(k) (Traditional) Submission Section 5, Summary

5. Description of Device:

The Valvulotome by Koven is a medical device that has a deployable assembly with multiple deployable cutting blades and guide fins which permit continual centralizing, and alignment of the cutting blades within the lumen of the vein as the valves are being cut. The device includes a handle operatively associated with this deployable assembly through a catheter operated by the handle in a one-handed operation for deploying the guide fins and cutting blades of the deployable assembly having respective outer diameters at each deployment position.

The Valvulotome by Koven consists of three main parts; the handle, the catheter and the cutter head as shown.

    1. The handle is fabricated through injection molded plastics that are assembled using adhesive and mechanical fasteners. The material used is a Medical Grade Acrylic and a Medical Grade Thermoplastic Elastomer.
    1. The catheter is fabricated through an extrusion process and an adhesive bond. The materials used are a Medical Grade Natural Polyetheretherketone and

medical grade adhesive.

KOVEN

    1. The cutter head is fabricated through machining, etching, drawing and stamping and assembled using a welding process. All materials in the cutter head are made from 300 series Stainless Steel.
      Image /page/1/Figure/9 description: The image shows a diagram of a medical instrument with several measurements labeled. The total length of the instrument is 94 cm, and the handle is 16 cm long. The diameter of the shaft is 0.8 mm, and the diameter of the tip ranges from 2.6 to 5.5 mm.

Section 11 510(k) Submission Page #

Summary Rev R day Houston, TX 77071

{2}------------------------------------------------

The Valvulotome by Koven are packaged using a Medical Grade Polyethylene Terephthalate Glycol tray and lid. The tray with lid shall be placed inside a sealed Twek pouch, which will be placed, along with an "Instructions for Use" booklet, inside a shelf carton with two protective form pads. The device label will appear on both the pouch and the shelf carton. The device and its packaging will be sterilized using irradiation sterilization method.

6. Summary of the technological characteristics of the device compared to predicate device:

Differences in technology - None.

The basic method of construction and materials is very similar. The methods of operation and indications for use are the same for each device.

7. Testing:

Testing of the Valvulotome by Koven included laboratory bond strength testing of all joined elements, bench testing for the determination of adequacy and safety of the device for the disruption of valves in blood vessels of the leg.

The Valvulotome by Koven successfully completed all testing with positive end points achieved.

8. Rx or OTC:

The Valvulotome by Koven is a Rx prescription device per 21 CFR Subpart D. The device is labeled for clinical settings only. Federal Law restricts this device to sale by or on the order of a physician (or properly licensed practitioner).

9. Conclusions:

Based upon the testing and comparison to the predicts device the Valvulotome by Koven has the same intended use, with similar technological characteristics. Koven Technology, Inc., therefore posits that its device is substantiality equivalent to predicate devices.

Koven Technology. Inc.

D.R. Bell

President Koven Technology, Inc

Date: 7-17-08

Houston, TX 7707

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three bars representing the department's three main goals. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 2 8 2008

Koven Technology c/o Mr. Harvey Knauss Consultant Delphi Consulting Group 11874 South Evelyn Circle Houston, TX 77071

Rc: K080178

Trade/Device Namc: Valvulotome by Koven Regulation Number: 21 CFR 870.4885 Regulation Name: External Vein Stripper Regulatory Class: Class II Product Code: MGZ Dated: July 1, 2008 Received: July 8, 2008

Dear Mr. Knauss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

Page 2 Mr. Harvey Knauss

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labcling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) promarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

una R. V. Auner

Bram Zuckerman Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) (Traditional) Submission Section 3, 510(k) Cover Letter

Indications for Use Statement

510(k) Number K080178 Device Name: Valvulotome by Koven

The Valvulotome by Koven is used for the treatment of vascular disorders and more particularly for excising or disrupting venous valves when performing in-situ vascular procedures, such as lower extremity arterial bypass surgery.

Prescription Use Yes (Part 21 CFR 801 Subpart D) OR

Over-The-Counter Use (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anna P. Mc James

Division Sign-Off) on of Cardiovascula

510(k) Number K080178

§ 870.4885 External vein stripper.

(a)
Identification. An external vein stripper is an extravascular device used to remove a section of a vein.(b)
Classification. Class II (performance standards).